The retrospective dataset was used to evaluate the sensitivity, specificity, and AUC of the cmTriage AI algorithm against established clinical benchmarks (BCSC) to support the device's performance claims.
Retrospective study; Mammography; AI algorithm validation; Clinical performance
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective, blinded, multi-center study of cmTriage; Retrospective, blinded, multi-center study; Follow-up/Duration: Two-year follow-up for normal cases
Screening mammography patients (enriched with biopsy-confirmed cancers); Sample Size: 1255 mammographic studies (400 cancer, 855 normal); Number of Sites: Multiple clinical sites
Not applicable for this study
Sensitivity, Specificity, AUC, and processing time
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Suspicious finding triage
—
Sensitivity > 80% (BCSC 80% CI lower bound), Specificity comparable to BCSC median specificity
Sensitivity 86.9% (95% CI 83.6% to 90.2%), Specificity 88.5% (95% CI 86.4% to 90.7%)
—
—
Retrospective, blinded, multi-center study: 1255 mammographic studies (400 biopsy-proven cancer studies and 855 normal studies).
—
Indications for Use
cmTriage is a passive notification-only, parallel-workflow software tool used by radiologists to prioritize specific patients within the standard-of-care image worklist for 2D FFDM screening mammograms. cmTriage uses an artificial intelligence algorithm to analyze 2D FFDM screening mammograms and flags those that are suggestive of the presence of at least one suspicious finding at the exam level. These flags are viewed by the radiologist via their Picture Archiving and Communication System (PACS) worklist. The decision to use cmTriage codes and how to use cmTriage codes is ultimately up to the radiologist. cmTriage does not send a proactive alert directly to the radiologist. Radiologists are responsible for reviewing each exam on a diagnostic viewer according to the current standard of care. cmTriage is limited to the categorization of exams, does not provide any diagnostic information beyond triage and prioritization, does not remove images from the radiologist's worklist, and should not be used in lieu of full patient evaluation, or relied upon to make or confirm diagnosis. cmTriage is for prescription use only.
Device Story
cmTriage is a cloud-based software tool for radiologists to prioritize 2D FFDM screening mammograms. It operates in parallel to standard clinical workflows. Input: 2D FFDM DICOM images captured by standard FFDM systems. Process: Local forwarding software de-identifies and transmits images to the CureMetrix cloud; AI algorithm analyzes images for suspicious findings; cloud generates a DICOM Structured Report (SR) with a 'Suspicious' or blank flag. Output: DICOM SR returned to the facility PACS, updating the worklist column for the exam. Usage: Radiologists view flags in the PACS worklist to prioritize reading; the device does not alter original images, provide diagnostic information, or send proactive alerts. Benefit: Enables efficient prioritization of potentially suspicious screening mammograms, potentially improving workflow efficiency. The device is not a diagnostic tool and does not replace standard-of-care review.
Clinical Evidence
Retrospective, blinded, multi-center study using 1,255 mammographic studies (400 biopsy-proven cancer, 855 normal). Primary endpoints: sensitivity and specificity. Results: Mean sensitivity 86.9% (95% CI: 83.6%–90.2%); mean specificity 88.5% (95% CI: 86.4%–90.7%). AUC 0.951 (95% CI: 0.937–0.964). Performance held across breast densities and lesion types (masses/calcifications). Timing performance averaged 3.35 minutes. Population-adjusted mark rate (6.37% at 84.4% sensitivity) compared favorably to radiologist recall rates.
Technological Characteristics
Software-only device. Components: (1) Image Forwarding/Result Receiving software (on-premise), (2) Cloud-based Image Processing/Analysis software, (3) DICOM SR Notification Result File. Connectivity: Networked via secure channel to cloud. Algorithm: AI-based analysis for suspicious regions of interest. Output: DICOM SR file for PACS integration. No hardware components.
Indications for Use
Indicated for use by radiologists to prioritize 2D FFDM screening mammograms within the standard-of-care worklist by flagging exams suggestive of at least one suspicious finding. Prescription use only.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
CureMetrix, Inc. % Mr. Kevin Harris CEO 9404 Genesee Avenue, Suite 330 LA JOLLA CA 92037
March 8, 2019
Re: K183285
Trade/Device Name: cmTriage Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: OFM Dated: February 4, 2019 Received: February 4, 2019
Dear Mr. Harris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Thalia Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183285
Device Name
cm Triage
#### Indications for Use (Describe)
cm Triage is a passive notification for prioritization-only, parallel-workflow software tool used by radiologists to prioritize specific patients within the standard-of-care image worklist for 2D FFDM screening mammograms, cmTriage uses an artificial intelligence algorithm to analyze 2D FFDM screening mammograms and flags those that are suggestive of the presence of at least one suspicious finding at the exam level. These flags are viewed by the radiologist via their Picture Archiving and Communication System (PACS) worklist. The decision to use cmTriage codes and how to use on Triage codes is ultimately up to the radiologist. cmTriage does not send a proactive alert directly to the radiologist.
Radiologists are responsible for reviewing each exam on a diagnostic viewer according to the current standard of care.
cm Triage is limited to the categorization of exams, does not provide any diagnostic information beyond triage and prioritization, does not remove images from the radiologist's worklist, and should not be used in lieu of full patient evaluation, or relied upon to make or confirm diagnosis.
cmTriage is for prescription use only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@ fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| Submitter | CureMetrix, Inc. |
|---------------------------|-----------------------------------------------------------------|
| Contact Person | Kevin Harris |
| | 9404 Genesee Ave |
| | Suite 330 |
| | La Jolla, CA 92037 |
| | 619-339-9889 |
| | Kevin@CureMetrix.com |
| Common Name | Radiological Computer-Assisted Triage and Notification Software |
| Trade Name | cmTriage™ |
| Classification | |
| Name: | Radiological Computer-Assisted Prioritization Software |
| Regulation: | 21 CFR 892.2080 |
| Code: | QFM |
| Class: | Class II |
| Predicate Device | |
| Name: | ContaCT |
| Company: | Viz.AI |
| Code: | QAS |
| 510(k): | DEN170073 |
| 510(k) Submission Number: | K183285 |
## Device Description
CureMetrix's cmTriage is a radiological computer-assisted triage and notification software device. Digital two-dimensional (2D) mammograms are captured by a Full-Field Digital Mammography (FFDM) system and deposited on the PACS. The CureMetrix image forwarding software, acting as a PACS listener, receives a copy of the mammography DICOM image(s), creates a local copy of the image(s), de-identifies the local copy, transmits the local copy, transmits the local copy to the CureMetrix cloud, and then deletes the local copy.
Within the CureMetrix cloud, the cmTriage service receives the DICOM image(s), groups them by study, analyzes the image(s) within the study, and produces a result for each study. The results are produced in the form of DICOM Structured Report (SR) file with a cmTriage result.
The result file is encrypted and transmitted from the CureMetrix cloud back to cmEdge where it is decrypted and reassociated with the original study. The DICOM SR is then routed to the PACS.
Once the PACS receives the DICOM SR, the file is opened, the cmTriage code ("Impression Description") is extracted for the exam, and the worklist column for the exam is updated. The cmTriage code will either indicate "Suspicious" or "" (blank).
Within the PACS worklist the cmTriage code can be displayed in a separate column. Each PACS may have different features and functionality depending on the manufacturer which are outside of the scope and control of CureMetrix and cmTriage. However, in general, at a minimum, the user is able to sort their worklist based on values in columns. This sorting functionality (if present) would allow the radiologist to group Suspicious exams together. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The standard of care for breast cancer screening in the US is quickly becoming one in which both FFDM and digital breast tomosynthesis (DBT) are acquired during the exam. However, cmTriage only operates on 2D images. If a site does not use FFDM and 2D but instead only uses 3D and DBT, they will not be able to use this current device.
In summary, the cmTriage device is intended to provide a passive notification through the PACS to the radiologist indicating the existence of a case that may potentially benefit from that radiologist's prioritization.
{5}------------------------------------------------
## Intended Use/Indications for Use
cmTriage is a passive notification-only, parallel-workflow software tool used by radiologists to prioritize specific patients within the standard-of-care image worklist for 2D FFDM screening mammograms. cmTrificial intelligence algorithm to analyze 2D FFDM screening mammograms and flags those that are suggestive of the presence of at least one suspicious finding at the exam level. These flags are viewed by the radiologist via their Picture Archiving and Communication System (PACS) worklist. The decision to use cmTriage codes and how to use cmTriage codes is ultimately up to the radiologist. cmTriage does not send a proactive alert directly to the radiologist.
Radiologists are responsible for reviewing each exam on a diagnostic viewer according to the current standard of care.
cmTriage is limited to the categorization of exams, does not provide any diagnostic information beyond triage and prioritization, does not remove images from the radiologist's worklist, and should not be used in lieu of full patient evaluation, or relied upon to make or confirm diagnosis.
cmTriage is for prescription use only.
## Technological Characteristics
cmTriage is a software only device that can be segmented into three components: (1) Image Forwarding and Result File Receiving Software, (2) Image Processing and Analysis Software, and (3) Notification Result File.
The Image Forwarding and Result File Receiving Software is configured by the hospital to interact with a healthcare facility's PACS or local DICOM router and is responsible for automatically transmitting a copy of DICOM files from the local router through a secured channel to the Image Processing and Analysis Software based on DICOM metadata.
The Image Processing and Analysis Software component of the device is hosted in the cloud and is reseiving, assembling, processing, and storing DICOM images. This software component includes the software algorithm that is responsible for identifying and quantifying image characteristics that are consistent with a suspicious region of interest in a screening mammogram. When the software algorithm detects imaging characteristics suggestive of a suspicious region of interest, the Image Processing and Analysis Software component sets a flag in the Notification Result File indicating that the exam is "Suspicious" and sends the file back to the Image Forwarding and Result File Receiving Software. While the software algorithm informs the notification process, no other diagnostic information is generated from the algorithm or available to the user beyond the notification.
The Notification Result File is a DICOM SR file that can be used by the PACS or viewing workstation to update the worklist with an indication of that reflects the cmTriage notification of either "Suspicious" or "" (blank).
| Characteristic | Predicate: Viz.AI ContaCT | cmTriage | Similarities or<br>Differences |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Clearance | DEN170073 | TBD | |
| Clearance Date | 2/13/18 | TBD | |
| Product Code | QAS | QFR | |
| Class | II | II | Same |
| Regulation | 892.2080 | 892.2080 | Same |
| OTC vs. PUO | PUO | PUO | Same |
| | | | |
| Intended<br>Use.<br>Indications<br>for<br>Use | ContaCT is a notification-only, parallel workflow<br>tool for use by hospital networks and radiologists<br>to identify and communicate images of specific<br>patients to a radiologist, independent of standard<br>of care workflow.<br><br>ContaCT uses an artificial intelligence algorithm to<br>analyze images for findings suggestive of a pre-<br>specified clinical condition and to notify an<br>appropriate medical radiologist of these findings<br>in parallel to standard of care image<br>interpretation. Identification of suspected findings<br>is not for diagnostic use beyond notification.<br><br>Specifically, the device analyzes CT angiogram<br>images of the brain acquired in the acute setting,<br>and sends notifications to neurovascular<br>radiologist that a suspected large vessel occlusion<br>has been identified and recommends review of<br>those images. Images can be previewed through a<br>mobile application.<br><br>Images that are previewed through the mobile<br>application are compressed and are for<br>informational purposes only and not intended for<br>diagnostic use beyond notification. Notified<br>radiologists are responsible for viewing non-<br>compressed images on a diagnostic viewer and<br>engaging in appropriate patient evaluation and<br>relevant discussion with a treating physician<br>before making care-related decisions or requests.<br>ContaCT is limited to analysis of imaging data and<br>should not be used in-lieu of full patient evaluation<br>or relied upon to make or confirm diagnosis. | cmTriage is a passive notification for<br>prioritization-only, parallel-workflow software<br>tool used by radiologists to prioritize specific<br>patients within the standard-of-care image<br>worklist for 2D FFDM screening mammograms.<br>cmTriage uses an artificial intelligence algorithm<br>to analyze 2D FFDM screening mammograms and<br>flags those that are suggestive of the presence of at<br>least one suspicious finding at the exam level.<br>These flags are viewed by the radiologist via their<br>Picture Archiving and Communication System<br>(PACS) worklist. The decision to use cmTriage<br>codes and how to use cmTriage codes is ultimately<br>up to the radiologist. cmTriage does not send a<br>proactive alert directly to the radiologist.<br><br>Radiologists are responsible for reviewing each<br>exam on a diagnostic viewer according to the<br>current standard of care.<br><br>cmTriage is limited to the categorization of exams,<br>does not provide any diagnostic information<br>beyond triage and prioritization, does not remove<br>images from the radiologist's worklist, and should<br>not be used in lieu of full patient evaluation, or<br>relied upon to make or confirm diagnosis.<br><br>cmTriage is for prescription use only. | Similarities<br>• Used by radiologists<br>• Identifies specific<br>patients<br>• Operates in parallel<br>to standard of care<br>workflow<br>• Uses artificial<br>intelligence to<br>analyze images<br>• Looking for findings<br>suggestive of a pre-<br>specified clinical<br>condition<br>• Not intended for<br>diagnostic use<br>• Radiologists are<br>responsible for<br>viewing images on a<br>diagnostic viewer<br>• Limited to analysis<br>of imaging data and<br>should not be used<br>in-lieu of full patient<br>evaluation or relied<br>upon to make or<br>confirm diagnosis<br><br>Noted difference:<br>cmTriage notification<br>is passive vs. proactive<br>and cmTriage codes<br>are accessed by<br>radiologist through<br>PACS worklist and<br>used for prioritization. |
| Technical<br>Method | The device provides triage or notification that is<br>informed by machine learning, artificial<br>intelligence or other image analysis algorithms | The device provides triage or notification that is<br>informed by machine learning, artificial<br>intelligence or other image analysis algorithms | Same |
| Target Area | The device operates on radiological images of the<br>human body. | The device operates on radiological images of the<br>human body. | Same |
| Anatomical Site | Head | Breast | Different anatomic site |
| Where Used | Hospital or Clinic | Hospital or Clinic | Same |
| User Population | Radiologist | Radiologist | Same |
| Software | Device is software only | Device is software only | Same |
| Software Level<br>of Concern | Moderate | Moderate | Same |
| Communication<br>with Patient | Communicates images of patients to a radiologist. | Does not communicate images of patients.<br>cmTriage passively notifies the radiologist via the<br>PACS worklist, whereas ContaCT notifies a pre-<br>identified specialist via direct message. | Noted difference: No<br>images are<br>communicated from<br>cmTriage to<br>radiologist. Only<br>cmTriage codes are<br>returned for viewing<br>within the PACS<br>worklist. |
| Notification/<br>Priority | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.