K182336 · Subtle Medical, Inc. · KPS · Nov 30, 2018 · Radiology
Device Facts
Record ID
K182336
Device Name
SubtlePET
Applicant
Subtle Medical, Inc.
Product Code
KPS · Radiology
Decision Date
Nov 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
PET image noise reduction
Convolutional neural network with residual learning
—
Significant average increase in quantitative metrics
—
—
Representative cases of human data gathered under IRB-approved clinical protocols
—
Indications for Use
SubtlePET is an image processing software intended for use by radiologists and nuclear medicine physicians for transfer, storage, and noise reduction of fluorodeoxyglucose (FDG) and amyloid PET images (including PET/CT and PET/MRI).
Device Story
SubtlePET is image processing software for PET/CT and PET/MRI scans. It functions as a DICOM node, receiving images from modalities or PACS, processing them, and forwarding enhanced images to a workstation. It uses a convolutional neural network (CNN) with residual learning to predict and separate noise from structural components, enhancing image quality. It operates on a virtual machine (VM) either onsite or in the cloud. Radiologists and nuclear medicine physicians view the output on existing PACS workstations. By reducing noise, the device aims to improve image quality for clinical interpretation. It integrates into existing radiology workflows without requiring changes to scanner acquisition protocols.
Clinical Evidence
Bench testing only. Performance evaluated using representative human PET data gathered under IRB-approved protocols. Testing confirmed design traceability, software requirement fulfillment, and end-to-end functionality of the DICOM dispatcher. Noise reduction bench tests demonstrated a significant average increase in quantitative metrics across all cases, confirming noise reduction performance.
Technological Characteristics
Software package operating on a virtual machine (VM) using CentOS 7 Linux. Employs a convolutional neural network (CNN) with residual learning for image enhancement. DICOM 3.0 compliant. Operates as a DICOM node for image transfer and processing. Deployment supports onsite or cloud-based VM configurations.
Indications for Use
Indicated for radiologists and nuclear medicine physicians to perform transfer, storage, and noise reduction of FDG and amyloid PET images (including PET/CT and PET/MRI).
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA) along with the Department of Health & Human Services. The FDA logo features a blue square with the letters 'FDA' in white, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue. To the left of the FDA logo is the Department of Health & Human Services seal, which includes an emblem with a caduceus-like symbol and the text 'DEPARTMENT OF HEALTH & HUMAN SERVICES-USA' arranged around it.
November 30, 2018
Subtle Medical, Inc. % Ms. Terese Bogucki Regulatory Consultant Decus Biomedical Inc. 2342 Shattuck Ave #333 BERKELEY CA 94704
Re: K182336
Trade/Device Name: SubtlePET Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: LLZ Dated: August 27, 2018 Received: August 28, 2018
Dear Ms. Bogucki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D. O'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K182336
Device Name SubtlePET
Indications for Use (Describe)
SubtlePET is an image processing software intended for use by radiologists and nuclear medicine physicians for transfer, storage, and noise reduction of fluorodeoxyglucose (FDG) and amyloid PET images (including PET/CT and PET/MRI).
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{4}------------------------------------------------
# 5 510(k) Summary
| Table 5-1. Subject Device Overview. | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name: | Subtle Medical, Inc. |
| Address: | 880 Santa Cruz Ave, Suite 200<br>Menlo Park, CA 94025 |
| Contact Person: | Terese Bogucki |
| Title: | Regulatory Consultant |
| Telephone Number: | 650-488-7799 |
| Fax Number: | 650-227-2264 |
| Email: | terri@decusbiomedical.com |
| Date Summary Prepared: | 5-NOV-2018 |
| Device Proprietary Name: | SubtlePET |
| Model Number: | V 1.0.0 |
| Common Name: | SubtlePET |
| Regulation Number: | 21 CFR 892.2050 |
| Regulation Name: | Picture archiving and communications system |
| Primary Product Code: | LLZ |
| Secondary Product Code: | KPS |
| Device Class: | Class II |
| Predicate Device: | Trade name: Sapheneia Clarity<br>Manufacturer: Sapheneia Commercial Products AB<br>Address: Teknikringen 8<br>SE-583 30 Linkoping, Sweden<br>Regulation Number: 21 CFR 892.2050<br>Regulation Name: Picture Archiving and<br>Communications System<br>Device Class: Class II<br>Product Code: LLZ<br>510(k) Number: K063391<br>510(k) Clearance Date: April 26, 2007 |
| Table 5-1. Subiect Device Overview | | |
|------------------------------------|--|--|
{5}------------------------------------------------
#### 5.1 Device Description
The SubtlePET image processing software reduces noise to increase image quality using a deep neural network-based algorithm.
The software employs a convolutional neural network-based method in a pixel's neighborhood to generate the value for each pixel. Using a residual learning approach, the software predicts the noise components and structural components. The software separates these components, which enhances the structure while simultaneously reducing the noise.
The workflow of the product can be easily adapted to existing radiology departmental workflow. The product acts as a DICOM node that receives DICOM 3.0 digital medical image data from the modality or another DICOM source, processes the data and then forwards the enhanced study to the selected destination. This destination can be any DICOM node, typically either the PACS system or a specific workstation.
## 5.2 Indications for Use
SubtlePET is an image processing software intended for use by radiologists and nuclear medicine physicians for transfer, storage, and noise reduction of fluorodeoxyglucose (FDG) and amyloid PET images (including PET/CT and PET/MRI).
#### 5.3 Summary of Technological Characteristics Comparison
Table 5-2 shows the similarities and differences between the technological characteristics of the two products. The key difference is the imaging modality type. Testing demonstrates that the differences do not raise new questions of safety or effectiveness.
| Topic | Predicate Device | Subject Device |
|------------------------------------------------|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Physical<br>Characteristics | Software package that operates on<br>off-the-shelf hardware | Software package that operates<br>on a virtual machine (VM) |
| Computer | PC Compatible | Virtual machine host-<br>compatible system |
| Image<br>Processing<br>Enhancement<br>Location | Onsite on the desktop computer<br>server | Onsite on the facility VM<br>and/or offsite on the cloud VM,<br>depending on the site's<br>configuration |
| DICOM<br>Standard<br>Compliance | The software processes DICOM<br>compliant image data | Same |
| Operating<br>System | Windows | CentOS 7 Linux |
Table 5-2. Summary of Technological Characteristics Comparison
{6}------------------------------------------------
| Topic | Predicate Device | Subject Device |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Modalities | Multi-modality | Multi-modality; specifically<br>processes PET, PET/CT and<br>PET/MR images |
| User Interface | The software is designed for use<br>on a radiology workstation. It is<br>unknown whether there is a user<br>interface. | None - enhanced images are<br>viewed on existing PACS<br>workstations |
| Protocols | Custom low dose protocols | Standard scanner protocols |
| Image<br>Enhancement<br>Algorithm<br>Description | Sapheneia Clarity™ employs a<br>sophisticated statistical analysis<br>of the image structure in the<br>neighborhood of each pixel.<br>Using robust estimation methods<br>the dominant structures are<br>separated from the embedding<br>noise. Once the structure has been<br>determined, it is possible to<br>strengthen the interesting parts<br>while simultaneously reducing the<br>noise. | The software employs a<br>convolutional neural network-<br>based method in a pixel's<br>neighborhood to generate the<br>value for each pixel.<br>Using a residual learning<br>approach, the software predicts<br>the noise components and<br>structural components. The<br>software separates these<br>components, which enhances<br>the structure while<br>simultaneously reducing the<br>noise. |
| Image<br>Acquisition | The acquisition remains the same,<br>i.e. the image processing can be<br>generated from multiple<br>modalities and with predefined or<br>specific acquisition protocol<br>settings. | The acquisition remains the<br>same. |
#### 5.4 Performance Data
Subtle Medical conducted the following non-clinical performance tests:
- . Design traceability confirming all requirement tracing is complete from design inputs and verification/validation and that all risk controls are implemented
- Design verification testing which included confirming all labeling complies . with 21CFR801 and all software requirements work as expected
- Design validation testing simulated intended use to confirm that the end-to-. end functionality of the SubtlePET DICOM Dispatcher in conjunction with the SubtlePET algorithm meets the design requirements
{7}------------------------------------------------
- . Noise reduction bench test utilizing representative cases of human data already gathered under the auspices of IRB-approved clinical protocols. The study showed a significant average increase in quantitative metrics for all cases demonstrating that the software reduced noise in PET scans.
Based upon the results of this testing, it was determined the SubtlePET performance was substantially equivalent to the predicate device.
### 5.5 Substantial Equivalence Conclusion
SubtlePET is an image enhancement software which has similar intended use and indications for use statement as the predicate device. The main difference in indications for use is that the predicate and subject devices apply enhancement to different multimodality image types. The two devices have similar technological characteristics: both algorithms are fixed image-domain nonlinear filtering that uses neighborhood information, and both methods have optimized parameters to ensure robustness and adaption to variable structures, tissues, noises and scales. This 510(k) submission includes information on the SubtlePET technological characteristics, as well as performance data and verification and validation activities demonstrating that SubtlePET is as safe and effective as the predicate, and does not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.