K173821 · Thirona Corporation · JAK · Jun 5, 2018 · Radiology
Device Facts
Record ID
K173821
Device Name
LungQ Software
Applicant
Thirona Corporation
Product Code
JAK · Radiology
Decision Date
Jun 5, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Lung and lobar volume measurement
—
Difference <= 10%
—
—
—
Equivalence study: 250 CT scans from the COPDGene study.
—
Lung and lobar density measurement
—
LAA-950HU: -1% to 1%; LAA-910HU: -10% to 10%; 15th Percentile: -10 HU to 10 HU
—
—
—
Equivalence study: 250 CT scans from the COPDGene study.
—
Fissure evaluation
—
Statistically higher Az value compared to previously published version
Average Az of 0.95 (vs 0.76 for previous version, p<0.001)
—
—
Fissure analysis study: 55 scans from a multicentre study (NCT01101958).
—
Indications for Use
The Thirona LungQ software provides CT values for pulmonary tissue which is essential for providing quantitative support for diagnosis and follow up examination. The Lung Q software can be used to support physician in the diagnosis and documentation of pulmonary tissues images (e.g., abnormalities) from CT thoracic datasets. Three-D segmentation and isolation of sub-compartments, volumetric analysis, density evaluation, fissure evaluation, and reporting tools are provided.
Device Story
LungQ is a standalone command-line software tool for processing thoracic CT scans. It ingests DICOM-formatted CT datasets to perform automated 3D segmentation of lungs, lobes, and airways. The software calculates volumetric measurements and density evaluations, including low attenuation areas (LAA-950HU, LAA-910HU) and 15th percentile density histograms, as well as fissure analysis. Operated by clinicians, the software provides quantitative reports to support the diagnosis and documentation of pulmonary abnormalities. By automating complex volumetric and density measurements, the device assists physicians in clinical decision-making for pulmonary disease management and follow-up.
Clinical Evidence
Equivalence study compared 250 CT scans from the COPDGene cohort analyzed by both LungQ and the predicate VIDA PW2. Metrics included lung/lobar volumes and density scores (LAA-950HU, LAA-910HU, 15th percentile). Results showed agreement within pre-defined limits (e.g., volume difference ≤10%). Fissure analysis performance was validated against a manual reference standard using 55 scans; LungQ 1.1.0 achieved an average Az of 0.95, significantly outperforming a previous version (Az 0.76, p<0.001).
Technological Characteristics
Standalone command-line software; processes DICOM CT thoracic datasets. Features 3D segmentation of lungs, lobes, and airways; volumetric and density quantification; fissure evaluation. Utilizes filtered back projection reconstruction. No graphical user interface. Software version 1.1.0.
Indications for Use
Indicated for use by physicians to provide quantitative support for diagnosis and follow-up examination of pulmonary tissue abnormalities using CT thoracic datasets. Applicable to patients undergoing CT scans for pulmonary assessment.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 5th , 2018
Thirona Corporation % Jean-Paul Charbonnier Managing Director Toernooiveld 300 6525 EC Nijmegen THE NETHERLANDS
Re: K173821
Trade/Device Name: LungQ Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: May 31, 2018 Received: June 1, 2018
Dear Jean-Paul Charbonnier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jeff Bargar
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K173821
Device Name LungQ
Indications for Use (Describe)
The Thirona LungQ software provides CT values for pulmonary tissue which is essential for providing quantitative support for diagnosis and follow up examination. The Lung Q software can be used to support physician in the diagnosis and documentation of pulmonary tissues images (e.g., abnormalities) from CT thoracic datasets. Three-D segmentation and isolation of sub-compartments, volumetric analysis, density evaluation, fissure evaluation, and reporting tools are provided.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# thir
| Submitter | Thirona Corporation |
|--------------------------|-----------------------------------------|
| Contact Person | Eva van Rikxoort |
| | Managing Director |
| | Thirona BV |
| | Toernooiveld 300 |
| | 6525 EC Nijmegen |
| | the Netherlands |
| | Phone: +31 (0)647142838 |
| | E-mail: evavanrikxoort@thirona.eu |
| Date Prepared | June X, 2018 |
| Trade Name | LungQ |
| Common Use/Usual<br>Name | Computer Tomography X-ray system |
| Product Code | JAK |
| Classification | Class II, 21 CFR 892.1750 |
| Device Panel | Radiology |
| Predicate Device | VIDA PW2 |
| Predicate Classification | Class II, 21 CFR 892.1750 |
| Reference Device | Imbio CT Lung Density Analysis Software |
| Reference Classification | Class II, 21 CFR 892.1750 |
Section 2.0 510(k) Summary
### Device Description
The LungQ software is designed to aid in the interpretation of Computed Tomography (CT) scans of the thorax that may contain pulmonary abnormalities. LungQ is standalone command-line software which must be run from a command-line interpreter and does not have a graphical user interface.
### Indications for Use
The Thirona LungQ software provides CT values for pulmonary tissue which is essential for providing quantitative support for diagnosis and follow up examination. The LungQ software can be used to support physician in the diagnosis and documentation of pulmonary tissues images (e.g., abnormalities) from CT thoracic datasets. Three-D segmentation and isolation of sub-compartments, volumetric analysis, density evaluations, fissure evaluation, and reporting tools are provided.
{5}------------------------------------------------
### Summary of Technical Comparisons
The table below compares the Thirona LungQ software to both the predicate and reference devices.
| Item | LungQ<br>Thirona<br>(Subject Device) | VIDA PW2<br>VIDA<br>Diagnostics<br>K083227<br>(Predicate Device) | Imbio CT<br>Lung Density<br>Analysis<br>Software<br>Imbio LLC<br>K141069<br>(Reference Device) |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Product Code | JAK | Identical | Identical |
| Regulation<br>Number | 21 CFR 892.1750 | Identical | Identical |
| Device<br>Classification | Class II | Identical | Identical |
| Common Name | Software Accessory to a<br>Computed tomography<br>x-ray system | Identical | Identical |
| Item | LungQ<br>Thirona<br>(Subject Device) | VIDA PW2<br>VIDA<br>Diagnostics<br>K083227<br>(Predicate<br>Device) | Imbio CT<br>Lung Density<br>Analysis<br>Software<br>Imbio LLC<br>K141069<br>(Reference<br>Device) |
| Intended Use | The Thirona LungQ<br>software provides CT<br>values for pulmonary<br>tissue which is essential<br>for providing<br>quantitative support for<br>diagnosis and follow up<br>examination. The<br>LungQ software can be<br>used to support<br>physician in the<br>diagnosis and<br>documentation of<br>pulmonary tissues<br>images (e.g.,<br>abnormalities) from CT<br>thoracic datasets. Three-<br>D segmentation and<br>isolation of sub-<br>compartments,<br>volumetric analysis,<br>density evaluations,<br>fissure evaluation, and<br>reporting tools are<br>provided. | Nearly Identical | Nearly Identical |
| Modality | CT | Identical | Identical |
| Data Loading | DICOM | Identical | Identical |
| Application | Command-line interface | Includes a<br>workstation | Identical |
| Segmentation | Provides 3D<br>segmentation | Identical | Identical |
| Item | LungQ<br>Thirona<br>(Subject Device) | VIDA PW2<br>VIDA<br>Diagnostics<br>K083227<br>(Predicate<br>Device) | Imbio CT<br>Lung Density<br>Analysis<br>Software<br>Imbio LLC<br>K141069<br>(Reference<br>Device) |
| | Provides Segmentation<br>of the:<br>• Left Lung<br>• Right Lung<br>• Left Upper Lobe<br>• Left Lower Lobe<br>• Right Upper<br>Lobe<br>• Right Middle<br>Lobe<br>• Right Lower<br>Lobe | Identical | Similar |
| | Provides Airways<br>Segmentation | Identical | Different |
| | User cannot manually<br>edit segmentation | User can manually<br>edit segmentation | Identical |
| Lung Volume<br>Analysis<br>Support | Ability to measure<br>volume for:<br>• Both Lungs<br>• Left Lung<br>• Right Lung<br>• Left Upper Lobe<br>• Left Lower Lobe<br>• Right Upper Lobe<br>• Right Middle Lob<br>• Right Lower Lobe | Identical | Similar |
| Item | LungQ<br>Thirona<br>(Subject Device) | VIDA PW2<br>VIDA<br>Diagnostics<br>K083227<br>(Predicate<br>Device) | Imbio CT<br>Lung Density<br>Analysis<br>Software<br>Imbio LLC<br>K141069<br>(Reference<br>Device) |
| Volume<br>Density<br>Analysis | Ability to measure<br>volume at multiple<br>density ranges for:<br>• Both Lungs<br>• Left Lung<br>• Right Lung<br>• Left Upper Lobe<br>• Left Lower Lobe<br>• Right Upper Lobe<br>• Right Middle Lob<br>• Right Lower Lobe | Identical | Similar |
| | Ability to measure the<br>15th percentile density<br>analysis | Identical | Different |
| | Does not perform low<br>density cluster analysis | Does perform low<br>density cluster<br>analysis | Identical |
| Fissure<br>Analysis | Ability to perform<br>fissure evaluations | Identical | Does not<br>perform fissure<br>evaluations |
| Analyzed Data<br>Output | Provides a report | Identical | Identical |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### Non-Clinical Testing
Software Verification testing was conducted to ensure that the Lung Q software met its requirements. The verification testing included white box testing to verify implantation and system integration testing. The LungQ software successful passed the verification testing.
{9}------------------------------------------------
### thi
Software Validation was conducted to ensure the software met the user needs (i.e. input requirements). This validation was based on user scenarios. The LungO software successfully passed the software validation.
An equivalence study comparing 250 CT scans from the COPDGene study (http://www.copdgene.org) analyzed by the predicate device, VIDA PW2, and the same 250 CT scans analyzed by Thirona LungQ. LungQ 1.1.0 was run allowing for interaction with the LungQ results in the Image Analysis Service as described in the Service Manual. The scans were randomly selected from the entire COPDGene cohort, leading to a population of subjects with and without COPD and with different stages of COPD. The distribution of subjects with respect to their disease state is provided below.
| Category | Disease state | # of subjects | % of data set |
|---------------------------------------------------------------------------|------------------|---------------|---------------|
| Control subjects: current or former<br>smokers without airflow limitation | GOLD stage 0 | 112 | 44.80% |
| Subjects with COPD but with minimal<br>airflow limitations | GOLD stage 1 | 26 | 10.40% |
| Subjects with COPD and with moderate to<br>severe airflow limitations | GOLD stage 2-4 | 76 | 30.40% |
| Subjects with preserved ratio but impaired<br>spirometry | GOLD stage PRISm | 33 | 13.20% |
| Control subjects: non-smokers | None | 3 | 1.20% |
The 250 scans were taken with a wide variety of scanners. The imaging parameters of the 250 scans are provided below.
The outputs between VIDA PW2 and Thirona LungQ 1.1.0 were compared. The following measurements were analyzed for equivalence:
- Lung and lobar volumes ●
- Lung and lobar density scores: .
- o Low attenuation areas below -950 HU (LAA-950HU)
- o Low attenuation areas below -910 HU (LAA-910HU)
- o 15th percentile of density histogram (perc15)
Equivalence was determined using the following criteria:
- Lung and lobar volume: Difference ≤ 10% ●
- . Lung and lobar density measurements:
- o LAA-950HU: Agreements limits -1% to 1%
- o LAA-910HU: Agreement limits -10% and 10%
- o 15th Percentile: Agreement limits -10 HU to 10 HU
{10}------------------------------------------------
## thir
The results showed that outputs from Thirona LungQ 1.1.0 are equivalent to the predicate device, VIDA PW2.
Additional performance testing of Thirona LungQ fissure analysis was conducted to compare a previously published version to LungQ 1.1.0. This study was performed on a set of 55 scans taken from a multicentre study conducted at five clinical study sites located in Germany, the Netherlands and Sweden (NCT01101958). The 55 scans were taken with a wide variety of scanners. The imaging parameters of the 55 scans are provided below.
Voxel-wise fissure completeness was compared to a manual reference standard in order to evaluate the performance of the two algorithms. The area under the receiver operating characteristic curve (Az) was used as a performance measure. A statistically higher Az value was set as the criteria to prove that LungQ 1.1.0 outperforms the previously published version.
For each individual scan. the Az value for LungO 1.1.0 was higher compared to the previously published version. The average Az of LungQ 1.1.0 was 0.95, compared to a statistically significantly lower average Az value of 0.76 for the previously published version (tested with a paired-sample t-test, p<0.001). This study showed that the fissure analysis of LungQ 1.1.0 significantly outperforms the previously published version.
| Imaging parameters | Equivalence study | Fissure analysis |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| # of scans | 250 | 55 |
| Voxel spacing | 0.50 - 0.97 mm | 0.44 - 0.81 mm |
| Slice thickness | 0.625 - 0.9 mm | 0.6 - 1.5 mm |
| Slice spacing | 0.45 - 0.625 mm | 0.45 - 1.2 mm |
| Peak Kilovoltage | 120 pKv | 120 - 130 pKv |
| Scanner manufacture | GE MEDICAL SYSTEMS; SIEMENS;<br>Philips | SIEMENS; Philips |
| Scanner types | LightSpeed16; LightSpeed VCT;<br>Sensation 64; Definition; Sensation 16;<br>Definition AS+; SOMATOM Definition<br>Flash; Brilliance 64; LightSpeed Pro 16;<br>Discovery CT750 HD; SOMATOM<br>Definition | Mx8000 IDT 16; iCT<br>128; Volume Zoom;<br>Emotion 16; Sensation<br>64; Definition;<br>SOMATOM Definition; |
| Reconstruction algorithms | Filtered back projection | Filtered back projection |
| Reconstruction kernels | STANDARD; B35f; B | B40f; B50s; B60f; B60s;<br>B80f |
| milliampere second | 200 mAs | NA |
{11}------------------------------------------------
#### Conclusions
Based on the comparison of intended use and key technological characteristics, Thirona believes that the Subject Device (LungQ) is substantially equivalent to the Predicate Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.