K083227 · Vida Diagnostics, Inc. · JAK · Nov 18, 2008 · Radiology
Device Facts
Record ID
K083227
Device Name
VIDA PULMONARY WORKSTATION 2 (PW2)
Applicant
Vida Diagnostics, Inc.
Product Code
JAK · Radiology
Decision Date
Nov 18, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
The VIDA Pulmonary Workstation 2 (PW2) software provides reproducible CT values for pulmonary tissue, which is essential for providing quantitative support for diagnosis and follow up examinations. The PW2 can be used to support the physician in the diagnosis and documentation of pulmonary tissue images (e.g., abnormalities) from CT thoracic datasets. Three-D segmentation and isolation of sub-compartments, volumetric analysis, density evaluations, low density cluster analysis and reporting tools are combined with a dedicated workflow. The VIDA Pulmonary Workstation 2 (PW2) software package is also intended to be a real-time interactive evaluation in space and time for CT volume data sets that provides the reconstruction of two-dimensional images into a three-dimensional image format.
Device Story
PW2 is a self-contained image analysis software package for pulmonary CT thoracic datasets. It imports 2D CT slice data to reconstruct 3D models. Physicians use the workstation to visualize and analyze lung parenchyma and bronchial airways. Core functions include 3D segmentation, isolation of sub-compartments, volumetric analysis, density evaluations (Hounsfield units), and low-density cluster analysis. The software automates tedious manual tracing tasks to provide quantitative measurements and aggregate statistics, assisting in therapy effectiveness assessment and clinical decision-making. It does not interface directly with CT scanners; it processes previously generated data files. The output is displayed to the physician for diagnostic support and documentation.
Clinical Evidence
No clinical data. Bench testing only, including software development, verification, and validation testing to ensure performance and safety.
Technological Characteristics
Software-based image analysis workstation. Processes CT voxel data using Hounsfield units (HU) for tissue characterization. Features 3D segmentation, volumetric analysis, and airway tree extraction. Operates on imported CT datasets. No direct hardware interface. Voluntary standards used for development and manufacturing processes.
Indications for Use
Indicated for use by physicians to support diagnosis and documentation of pulmonary tissue abnormalities in CT thoracic datasets. Provides quantitative support for diagnosis and follow-up examinations of chest diseases.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
γ0γ3227
Image /page/0/Picture/1 description: The image shows the logo for VIDA Diagnostics. The logo is in black and features the word "VIDA" in large, bold letters. Below the word "VIDA" is the word "Diagnostics" in a smaller font. There is a curved line that extends from the bottom of the "V" to the right of the "A".
NOV 1 8 2008
VIDA Pulmonary Workstation 2 (PW2)
# 5 510(k) Summary 510(k) Summary of Safety and Effectiveness [As required by 21 CFR 807.92(c)]
General information regarding the VIDA Diagnostics, Inc., Inc., VIDA Pulmonary Workstation 2 (PW2) is contained in the following table 1.
# Table 1: General Information
| Manufacturer | VIDA Diagnostics<br>100 Oakdale Campus, Suite 225 TIC<br>lowa City, IA 52242<br>USA<br>FDA Establishment Registration No. Pending |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name, Title<br>and Phone Contact | Jack Slovick, RA/QA, 763-639-0238 (phone)<br>RA/QA Affairs<br>Fax: 763-434-0966<br>VIDA Diagnostics<br>100 Oakdale Campus,<br>Suite 225 Technology Innovation Center<br>lowa City, IA 52242 |
| Trade Name | VIDA Pulmonary Workstation 2 (PW2) |
| Common Name | Computed tomography x-ray system |
| Classification | This device has been classified by the Reproductive,<br>Abdominal and Radiological Panel into Class II (21 CFR<br>892.1750), Product Code 90 JAK Class 11 |
| Date Prepared | October 13, 2008 |
| Intended Use | The VIDA Pulmonary Workstation 2 (PW2) software<br>provides reproducible CT values for pulmonary tissue,<br>which is essential for providing quantitative support for<br>diagnosis and follow up examinations. The PW2 can be<br>used to support the physiclan in the diagnosis and<br>documentation of pulmonary tissue images (e.g.,<br>abnormalities) from CT thoracic datasets. Three-D<br>segmentation and isolation of sub-compartments,<br>volumetric analysis, density evaluations, low density<br>cluster analysis and reporting tools are combined with a<br>dedicated workflow. The VIDA Pulmonary Workstation 2<br>(PW2) software package is also intended to be a real-time<br>interactive evaluation in space and time for CT volume |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for VIDA Diagnostics. The logo is in black and white and features the word "VIDA" in large, bold letters. Below the word "VIDA" is the word "Diagnostics" in a smaller font. There is a curved line that extends from the bottom of the "V" to the right of the "A", with a dot at the end.
# VIDA Pulmonary Workstation 2 (PW2)
| | data sets that provides the reconstruction of two- dimensional images into a three-dimensional image format. | |
|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|
| Identification of Equivalent<br>Devices | Manufacturer: | Siemens Medical Solutions |
| | Trade name: | InSpace 4D - Software Package with<br>Extended Functionality |
| | 510(k) Number: | K071513 |
| | Date Cleared: | June 26, 2007 |
| Compliance with Special<br>Controls or Performance<br>Standards | Special Controls: No special controls or mandatory<br>performance standards for this device have been<br>established. Voluntary standards were used throughout<br>the development; testing and manufacturing processes<br>(see section 9.0 for detailed list of voluntary standards that<br>were used) | |
| Product Code | 90 JAK, Class II | |
| Regulation Number | 21 CFR 892.1750 | |
| Reason for Premarket<br>Notification | VIDA Diagnostics, Inc. intends to introduce the VIDA<br>Pulmonary Workstation 2 (PW2) into interstate commerce<br>for commercial distribution. | |
## 1 General Safety and Effectiveness Concerns
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, VIDA Diagnostics adheres to recognized and established industry practice and standards.
## 2 Substantial Equivalence
The VIDA Pulmonary Workstation 2 (PW2), addressed in this pre-market notification, is substantially equivalent to the following commercially available software package.
## 3 Manufacturer Product 510(k) Clearance date
Siemens InSpace 4D K071513 June 26, 2007
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for VIDA Diagnostics. The logo consists of the word "VIDA" in large, bold, sans-serif font, with a swoosh-like line extending from the bottom of the "A" to the right, ending in a solid circle. Below the word "VIDA" is the word "Diagnostics" in a smaller, serif font.
#### VIDA Pulmonary Workstation 2 (PW2)
In summary, VIDA Diagnostics is of the opinion that VIDA Pulmonary Workstation 2 (PW2) does not introduce any new potential safety risks and is substantially equivalent to and performs as well as the predicate software components and the predicate device.
#### 3. Intended Use
The VIDA Pulmonary Workstation 2 (PW2) software provides reproducible CT values for pulmonary tissue, which is essential for providing quantitative support for diagnosis and follow up examinations. The PW2 can be used to support the physician in the diagnosis and documentation of pulmonary tissue images (e.g., abnormalities) from CT thoracic datasets. Three-D segmentation and isolation of sub-compartments, volumetric analysis, density evaluations, low density cluster analysis and reporting tools are combined with a dedicated workflow. The VIDA Pulmonary Workstation 2 (PW2) software package is also intended to be a real-time interactive evaluation in space and time for CT volume data sets that provides the reconstruction of two-dimensional images into a three-dimensional image format.
#### 4. Device Description
VIDA Pulmonary Workstation 2 (PW2) is a self-contained image analysis software package. This real-time interactive evaluation in space and time for CT volume data sets provides the reconstruction of two-dimensional images into a three-dimensional image format. VIDA Pulmonary Workstation 2 (PW2) can be used to support the physician in the diagnosis and documentation of chest diseases, e.g. when examining the pulmonary tissue (i.e. lung parenchyma) in CT thoracic datasets. Evaluation tools (3D seamentation & isolation of sub-compartments. volumetric analysis, density evaluations, and low density cluster analysis) and reporting tools are combined with a dedicated workflow.
The PW2 is designed to analyze pulmonary CT slice data and display analysis results. Each voxel of the scan is measured by Hounsfield units (HU), a measurement of x-ray attenuation that is applied to each volume element in threedimensional space ("voxel"). The HU are utilized to distinguish between air, water, tissue and bone, such distinction is common in the industry. Figure 1 shows the basic Input/ Output of the CT volume converting to a digitized, colorized picture.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for VIDA Diagnostics. The logo is in black and features the word "VIDA" in large, bold letters. A curved line extends from the bottom of the "V" to the right of the "A", with a dot at the end. Below the word "VIDA" is the word "Diagnostics" in a smaller font.
VIDA Pulmonary Workstation 2 (PW2)
Image /page/3/Figure/2 description: The image shows a diagram of an airway tree extraction process. On the left side of the image, the input is a CT volume with dimensions 512x512x550 voxels, a voxel size of 0.65x0.65x0.6 mm^3, a 16-bit gray-scale, and a size of 250-300 MByte. On the right side of the image, the output is a binary image of the airway tree.
Figure 1: Input/ Output
PW2 provides computed tomography (CT) viewing, airway analysis, and parenchymal density analysis in one application. PW2 provides imaging of bronchial airways that can be used to assess therapy effectiveness based on CT scan data. PW2 reconstructs multiple cross-section images from CT data into a computer model displaying complex bronchial branches.
PW2 does not interface directly with any CT or data collection equipment; instead PW2 imports data files previously generated by such equipment.
PW2 provides quantitative measurements and tabulates quantitative properties. PW2 focuses on what is visible to the eye and applies volumetric methods that might otherwise be too tedious to use. The software does not perform any function which cannot be accomplished by a trained user utilizing manual tracing methods; the intent of the software is to save time and automate potential errorprone manual tasks.
The software has functions for loading, analyzing, saving datasets and will generate screen displays, computations and aggregate statistics.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a bird or a person with outstretched arms. The symbol is composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# NOV 1 8 2008
VIDA Diagnostics, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K083227
Trade/Device Name: VIDA Pulmonary Workstation 2 (PW2) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: October 31, 2008 Received: November 3, 2008
#### Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/odrh/industry/support/index.html.
Sincerely yours.
forque M. Whang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# KO83227
Image /page/6/Picture/1 description: The image shows the logo for VIDA Diagnostics. The logo is in black and white and features the word "VIDA" in large, bold letters. Below the word "VIDA" is the word "Diagnostics" in a smaller font. There is a curved line that extends from the bottom of the "D" in VIDA to the right, ending in a small circle.
VIDA Pulmonary Workstation 2 (PW2)
# 4 Indications for Use Statement
The VIDA Pulmonary Workstation 2 (PW2) software provides reproducible CT values for pulmonary tissue, which is essential for providing quantitative support for diagnosis and follow up examinations. The PW2 can be used to support the physician in the diagnosis and documentation of pulmonary tissue images (e.g., abnormalities) from CT thoracic datasets. Three-D segmentation and isolation of sub-compartments, volumetric analysis, density evaluations, low density cluster analysis, and reporting tools are combined with a dedicated workflow. The VIDA Pulmonary Workstation 2 (PW2) software package is also intended to be a real-time interactive evaluation in space and time for CT volume data sets that provides the reconstruction of two-dimensional images into a three-dimensional image format.
Hills Penn
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Device 510(k) Number
Prescription Use __ (per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.