KWT · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented

Orthopedic · 21 CFR 888.3650 · Class 2

Overview

Product CodeKWT
Device NameProsthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented
Regulation21 CFR 888.3650
Device ClassClass 2
Review PanelOrthopedic
ImplantYes

Identification

A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Classification Rationale

Class II. The special controls for this device are:

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,” (vii) F 1378-97 “Specification for Shoulder Prosthesis,” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Recent Cleared Devices (20 of 72)

Showing 20 most recent of 72 cleared devices.

RecordDevice NameApplicantDecision DateDecision
K262173N22 Offset Humeral HeadShoulder Innovations, Inc.Sep 8, 2026SESE
K250713Equinoxe® Humeral Reconstruction Prosthesis; Equinoxe® Stemless Shoulder ImplantsExactech, Inc.Dec 1, 2025SESE
K251686Tornier Humeral Reconstruction System Max (Tornier HRS Max)Tornier, Inc.Oct 30, 2025SESE
K241944AGILON® XO Shoulder Replacement SystemImplantcast GmbHMar 10, 2025SESE
K241817InSet Total Shoulder SystemShoulder Innovations, Inc.Feb 11, 2025SESE
K231657AGILON® XO Shoulder Replacement SystemImplantcast GmbHFeb 16, 2024SESE
K223631Comprehensive Segmental Revision System(SRS)Biomet OrthopedicsJul 12, 2023SESE
K222482AGILON XO Shoulder SystemImplantcast GmbHSep 28, 2022SESE
K210899LINK Embrace Shoulder System- Anatomical ConfigurationWaldemar Link GmbH & Co. KGJun 21, 2021SESE
K191698TiN Coated Humeral HeadFx Shoulder USA, Inc.Jul 16, 2020SESE
K192365Shoulder Innovations Total Shoulder SystemShoulder Innovations, Inc.Nov 29, 2019SESE
K191811Catalyst OrthoScience CSR Shoulder SystemCatalyst Orthoscience, Inc.Sep 11, 2019SESE
K190595Signature ONE SystemOrthosoft Inc (D/B/A Zimmer Cas)Jun 5, 2019SESE
K182500Catalyst CSR Press-Fit Humeral ComponentsCatalyst Orthoscience, Inc.Jan 11, 2019SESE
K181287Catalyst CSR Shoulder SystemCatalyst Orthoscience, Inc.Jul 12, 2018SESE
K173812Catalyst CSR 3 Peg GlenoidsCatalyst Orthoscience, Inc.Mar 9, 2018SESE
K162903Exactech Equinoxe Humeral Reconstruction Prosthesis Extra-Small Proximal BodiesExactech, Inc.Feb 27, 2017SESE
K161476SMR Shoulder SystemLima Corporate S.P.A.Jun 24, 2016SESE
K152825Catalyst CSR Shoulder SystemCatalyst Orthoscience, LLCMay 11, 2016SESE
K150730Zimmer PSI Shoulder SystemZimmerJun 16, 2015SESE

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...