IMJ · Diathermy, Shortwave, For Use In Applying Therapeutic Deep Heat

Physical Medicine · 21 CFR 890.5290 · Class 2

Overview

Product CodeIMJ
Device NameDiathermy, Shortwave, For Use In Applying Therapeutic Deep Heat
Regulation21 CFR 890.5290
Device ClassClass 2
Review PanelPhysical Medicine
3rd-Party ReviewableYes

Identification

A nonthermal shortwave therapy is a prescription device that applies to the body pulsed electromagnetic energy in the RF bands of 13.56 MHz or 27.12 MHz and that is intended for adjunctive use in the palliative treatment of postoperative pain and edema of soft tissue by means other than the generation of deep heat within body tissues as described in paragraph (a) of this section.

Classification Rationale

Class II (special controls). The device is classified as class II.

Special Controls

*Classification: Class II (special controls).* The device is classified as class II. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. (ii) Appropriate analysis/testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment. (iii) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Non-clinical performance testing must characterize the output waveform of the device and demonstrate that the device meets appropriate output performance specifications. The output characteristics and the methods used to determine these characteristics, including the following, must be determined: (A) Peak output power; (B) Pulse width; (C) Pulse frequency; (D) Duty cycle; (E) Characteristics of other types of modulation that may be used; (F) Average measured output powered into the RF antenna/applicator; (G) Specific absorption rates in saline gel test load or other appropriate model; (H) Characterization of the electrical and magnetic fields in saline gel test load or other appropriate model for each RF antenna and prescribed RF antenna orientation/position; and (I) Characterization of the deposited energy density in saline gel test load or other appropriate model. (iv) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device in its intended use. (v) Labeling must include the following: (A) Output characteristics of the device; (B) Recommended treatment regimes, including duration of use; and (C) A detailed summary of the clinical testing pertinent to the use of the device and a summary of the adverse events and complications. (vi) Nonthermal shortwave therapy devices marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.

Recent Cleared Devices (20 of 45)

Showing 20 most recent of 45 cleared devices.

RecordDevice NameApplicantDecision DateDecision
K252963BTL-899TTBTL Industries, Inc.Jun 12, 2026SESE
K250376ViVYViatherm Therapeutics, LLCNov 5, 2025SESE
K243112Curapuls 670Enraf-Nonius, B.V.Jun 25, 2025SESE
K223620Reprieve by RegenesisTMRegenesis Biomedical, Inc.May 24, 2023SESE
K182363BTL-703BTL Industries, Inc.Jan 24, 2019SESE
K173300ViaTherm BOOSTViatherm Therapeutics, LLCMay 1, 2018SESE
K162240ProMedTek Model C1400 Shortwave Diathermy DevicePromedtek, Inc.Nov 17, 2016SESE
K161862ThermoProZimmer Medizinsysteme GmbHNov 14, 2016SESE
K153456Megapulse III Shortwave DiathermyAccelerated Care PlusJun 14, 2016SESE
K131926THERMOPULSEIbramed Equipamentos MedicosJul 24, 2014SESE
K131899SW-THERMShenzhen Dongdixin Technology Co., Ltd.Mar 14, 2014SESE
K121123BIOFUSIONARY BEBERocky Mountain Biosystems, Inc.Mar 10, 2014SESE
K120093BTL ELITEBTL Industries, Inc.Oct 5, 2012SESE
K083433INTELECT SWD 100, MODEL 1600; SENIOR SOLUTIONS SHORTWAVE DIATHERMY, MODEL 1601Chattanooga GroupMar 18, 2009SESE
K042554AUTO THERM 390, MODEL ME 390Mettler Electronics Corp.Oct 8, 2004SESE
K030382SELITHERM, MODEL C100Selicor, Inc.Apr 1, 2003SESE
K022846SW-180 SHORTWAVE THERAPY UNITIto Co., Ltd.Jan 28, 2003SESE
K022458AUTO*THERM 395, MODEL ME 395Mettler Electronics Corp.Sep 26, 2002SESE
K020227SELICOR, MODELS A100 & D100Selicor, Inc.Jul 30, 2002SESE
K973732MEGAPULSE IIPtiJan 23, 1998SESE

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...