DTE · Pulse-Generator, Pacemaker, External

Cardiovascular · 21 CFR 870.3600 · Class 2

Overview

Product CodeDTE
Device NamePulse-Generator, Pacemaker, External
Regulation21 CFR 870.3600
Device ClassClass 2
Review PanelCardiovascular

Identification

An external pacemaker pulse generator (EPPG) is a prescription device that has a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. This device, which is used outside the body, is used as a temporary substitute for the heart's intrinsic pacing system until a permanent pacemaker can be implanted, or to control irregular heartbeats in patients following cardiac surgery or a myocardial infarction. The device may have adjustments for impulse strength, duration, R-wave sensitivity, and other pacing variables.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must validate electromagnetic compatibility (EMC) within a hospital environment. (2) Electrical bench testing must demonstrate device safety during intended use. This must include testing with the specific power source ( *i.e.,* battery power, AC mains connections, or both).(3) Non-clinical performance testing data must demonstrate the performance characteristics of the device. Testing must include the following: (i) Testing must demonstrate the accuracy of monitoring functions, alarms, measurement features, therapeutic features, and all adjustable or programmable parameters as identified in labeling; (ii) Mechanical bench testing of material strength must demonstrate that the device and connection cables will withstand forces or conditions encountered during use; (iii) Simulated use analysis/testing must demonstrate adequate user interface for adjustable parameters, performance of alarms, display screens, interface with external devices ( *e.g.* data storage, printing), and indicator(s) functionality under intended use conditions; and(iv) Methods and instructions for cleaning the pulse generator and connection cables must be validated. (4) Appropriate software verification, validation, and hazard analysis must be performed. (5) Labeling must include the following: (i) The labeling must clearly state that these devices are intended for use in a hospital environment and under the supervision of a clinician trained in their use; (ii) Connector terminals should be clearly, unambiguously marked on the outside of the EPPG device. The markings should identify positive (+) and negative (−) polarities. Dual chamber devices should clearly identify atrial and ventricular terminals; (iii) The labeling must list all pacing modes available in the device; (iv) Labeling must include a detailed description of any special capabilities ( *e.g.,* overdrive pacing or automatic mode switching); and(v) Appropriate electromagnetic compatibility information must be included.

Recent Cleared Devices (20 of 65)

Showing 20 most recent of 65 cleared devices.

RecordDevice NameApplicantDecision DateDecision
K241781Solo Pace ControlSolo Pace, Inc.Jan 10, 2025SESE
K232721Lifetech Cardio Temporary PacemakerShenzhen Lifetech Cardio Medical Electronics Co., Ltd.Jan 7, 2024SESE
K201011Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.May 13, 2020SESE
K190825Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.Apr 30, 2019SESE
K182839Lifetech Cardio Model 8301 Temporary PacemakerShenzhen Lifetech Cardio Medical Electronics Co., Ltd.Feb 28, 2019SESE
K181973Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.Aug 21, 2018SESE
K180873Medtronic Model 53401 External Pulse Generator (EPG)Medtronic, Inc.Apr 19, 2018SESE
K162054Medtronic Temporary External Pacemaker 53401Medtronic, Inc.Oct 18, 2016SESE
K162550Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.Oct 12, 2016SESE
K150246Medtronic Model 5392 External Pulse Generator (EPG)Medtronic, Inc.Feb 18, 2015SESE
K132924MEDTRONIC EXTERNAL PULSE GENERATOR (EPG)Medtronic, Inc.Oct 31, 2013SESE
K123916PSA SERIESOsypka Medical, Inc.Apr 9, 2013SESK
K042708ERA 3000 PACING SYSTEM ANALYZERBiotronik, Inc.Oct 20, 2004SESE
K033130PACE 101/PACE 101H, MODEL 3077 SSI TEMPORARY PULSE GENERATOROsypka Medical, Inc.Oct 31, 2003SESE
K023080GALIX PACESTARGalix Biomedical Instrumentation, Inc.Jun 11, 2003SESE
K022939OSCOR PACE 101/101H AND ACCESSORIES, ST.JUDE MEDICAL MODEL 3077 AND ACCESSORIES,CARDIOTRONIC PACE 101 AND ACCESSORIESOsypka Medical, Inc.Oct 3, 2002SESE
K020896OSCOR PACE 203/PACE 203H DUAL-CHAMBER TEMPORARY PACEMAKER, ST. JUDE MEDICAL, MODEL 3085 DUAL-CHAMBER EXTERNAL PULSE GENEOsypka Medical, Inc.Jun 14, 2002SESE
K011980ADAPTER, PACING ANALYZER/UNIVERSAL PACING ADAPTER, MODELS 4820 AND 4825Pace MedicalSep 12, 2001SESE
K002204PACEMAKER EXTENSION CABLE MODELS 4265,4265A, 4265V, 5265, 5265A, 5265VPace MedicalOct 18, 2000SESE
K001180TEMPORARY PACEMAKER EXTENSION CABLE, MODELS 5235, 5235A, 5235V, TEMPORARY PACEMAKER SURGICAL EXTENSION CABLE, MODELS 528Pace MedicalJun 27, 2000SESE

Top Applicants

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...