GALIX PACESTAR
Device Facts
| Record ID | K023080 |
|---|---|
| Device Name | GALIX PACESTAR |
| Applicant | Galix Biomedical Instrumentation, Inc. |
| Product Code | DTE · Cardiovascular |
| Decision Date | Jun 11, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.3600 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The GALIX PaceStar External Paccmaker may be used in any clinical situation in which the use of a temporary pacemaker on a patient provides therapentic or diagnostic value. Specifically, indications for teaporary paccusions include, but are not limited to, the following: - temporary treatment of arrhythmias and heart block . - pro-, intra- and postoperative temporary stimulation of patients undergoing cardiac surgery .
Device Story
External pacemaker pulse generator; provides temporary cardiac stimulation. Used in clinical settings for patients requiring temporary pacing therapy or diagnostics. Device delivers electrical impulses to heart; manages arrhythmias and heart block; supports cardiac surgery patients. Operated by clinicians. Output consists of electrical pacing pulses; assists in maintaining cardiac rhythm; provides therapeutic benefit to patients requiring temporary hemodynamic support.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use.
Technological Characteristics
External pacemaker pulse generator; Class III device (21 CFR 870.3600). Electrical stimulation delivery system. Specific materials and dimensions not detailed in provided text.
Indications for Use
Indicated for patients requiring temporary cardiac pacing for treatment of arrhythmias, heart block, or pro-, intra-, and postoperative stimulation during cardiac surgery.
Regulatory Classification
Identification
An external pacemaker pulse generator (EPPG) is a prescription device that has a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. This device, which is used outside the body, is used as a temporary substitute for the heart's intrinsic pacing system until a permanent pacemaker can be implanted, or to control irregular heartbeats in patients following cardiac surgery or a myocardial infarction. The device may have adjustments for impulse strength, duration, R-wave sensitivity, and other pacing variables.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must validate electromagnetic compatibility (EMC) within a hospital environment. (2) Electrical bench testing must demonstrate device safety during intended use. This must include testing with the specific power source ( *i.e.,* battery power, AC mains connections, or both).(3) Non-clinical performance testing data must demonstrate the performance characteristics of the device. Testing must include the following: (i) Testing must demonstrate the accuracy of monitoring functions, alarms, measurement features, therapeutic features, and all adjustable or programmable parameters as identified in labeling; (ii) Mechanical bench testing of material strength must demonstrate that the device and connection cables will withstand forces or conditions encountered during use; (iii) Simulated use analysis/testing must demonstrate adequate user interface for adjustable parameters, performance of alarms, display screens, interface with external devices ( *e.g.* data storage, printing), and indicator(s) functionality under intended use conditions; and(iv) Methods and instructions for cleaning the pulse generator and connection cables must be validated. (4) Appropriate software verification, validation, and hazard analysis must be performed. (5) Labeling must include the following: (i) The labeling must clearly state that these devices are intended for use in a hospital environment and under the supervision of a clinician trained in their use; (ii) Connector terminals should be clearly, unambiguously marked on the outside of the EPPG device. The markings should identify positive (+) and negative (−) polarities. Dual chamber devices should clearly identify atrial and ventricular terminals; (iii) The labeling must list all pacing modes available in the device; (iv) Labeling must include a detailed description of any special capabilities ( *e.g.,* overdrive pacing or automatic mode switching); and(v) Appropriate electromagnetic compatibility information must be included.