UNIPURE C3F8 Ophthalmic Gas in the UNIFEYE Gas Delivery System, UNIPURE C3F8 Ophthalmic Gas in the UNIPEXY Gas Delivery
Applicant
Alcon Research, LLC
Product Code
LPO · Ophthalmic
Decision Date
Jul 2, 2024
Decision
APPR
Regulation
21 CFR 886.4270
Device Class
Class 3
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
P220030 · Jul 2, 2024
UNIPURE C3F8 Ophthalmic Gas in the UNIFEYE Gas Delivery System, UNIPURE C3F8 Ophthalmic Gas in the UNIPEXY Gas Delivery
Alcon Research, LLC
Published clinical literature (Medline, Embase, Cochrane, Clinicaltrials.gov, WHO International Clinical Trial Registry Platform)
A systematic literature review of 46 studies was conducted to evaluate the safety (rate of elevated IOP) and effectiveness (primary retinal reattachment rate) of C3F8 gas in routine clinical settings for the treatment of uncomplicated retinal detachments.
Systematic Clinical Literature Review; Systematic review of published clinical literature (retrospective and observational cohorts); Follow-up/Duration: At least 3 months postoperatively for effectiveness; 1 to 7 days for safety; Study Period: January 1, 1980 to May 16, 2022
Patients with uncomplicated retinal detachments treated with C3F8 gas; Sample Size: 2270 participants treated with C3F8 (out of 4612 total study participants); Number of Sites: Not specified (multi-center literature)
Not applicable for this study
Primary retinal reattachment rate (effectiveness); Rate of elevated IOP >25 mm Hg (safety)
Indications for Use
The UNIPURE C₃F₈ Ophthalmic Gas in the UNIFEYE Gas Delivery System is indicated for intraocular injection into the eye for the treatment of uncomplicated retinal detachments. Associated measures used include vitrectomy, fluid/air exchange, transconjunctival and transscleral cryotherapy, laser photocoagulation, and air/gas exchange. The UNIPURE C₃F₈ Ophthalmic Gas in the UNIPEXY Gas Delivery System is indicated for intraocular injection into the eye for the treatment of uncomplicated retinal detachments. Associated measures used include vitrectomy, fluid/air exchange, transconjunctival and transscleral cryotherapy, and laser photocoagulation.
Device Story
UNIPURE C₃F₈ Ophthalmic Gas is a non-toxic, inert, liquefied perfluoropropane gas supplied in pico-cylinders within the UNIFEYE or UNIPEXY delivery systems. UNIFEYE mixes gas with filtered air for injection; UNIPEXY injects undiluted gas. Used by ophthalmologists in clinical/surgical settings for intraocular endotamponade during retinal detachment repair. The gas forms a bubble in the vitreous cavity to provide internal pressure against the retina. Healthcare providers monitor the bubble via ophthalmoscopy; the bubble dissipates over time. Patients must maintain specific head positioning post-operatively to ensure the bubble contacts the retinal break. The device benefits patients by facilitating retinal reattachment and reducing the risk of detachment recurrence. Air travel is strictly prohibited until the bubble is fully absorbed to prevent dangerous intraocular pressure spikes.
Clinical Evidence
Systematic literature review of 46 studies (1980–2022) involving 4,612 participants. Primary effectiveness endpoint: primary retinal reattachment rate at ≥3 months (81.9%; 95% CI: 76.4%, 86.3%), exceeding the 72% threshold. Primary safety endpoint: rate of elevated IOP (>25 mmHg) at 1–7 days post-op (13%; 95% CI: 2.6%, 46.2%), below the 34% threshold. Evidence supports safety and effectiveness for uncomplicated retinal detachment.
Technological Characteristics
Perfluoropropane (C₃F₈) gas; molecular weight 188 g/mol; boiling point -36.7 °C. Delivery systems include pico-cylinders, filters (0.2-micron), and mechanical mixing/injection mechanisms. Biocompatibility per ISO 10993-1. Sterilization via ethylene oxide (EO) to 10^-6 SAL. Mechanical components tested for leak, force, and mix ratio accuracy (±1.5%).
Indications for Use
Indicated for intraocular injection for treatment of uncomplicated retinal detachments in patients requiring vitrectomy, fluid/air exchange, cryotherapy, or laser photocoagulation. Contraindicated in patients with PVR > Stage C, inability to maintain therapeutic positioning for 5 days, severe glaucoma with significant visual field loss, uveitis, severe peripheral retinal degeneration, congenital malformations (e.g., coloboma), or conditions facilitating gas migration; contraindicated for high-altitude or airline travel until gas bubble dissipation.
Regulatory Classification
Identification
An intraocular gas is a device consisting of a gaseous fluid intended to be introduced into the eye to place pressure on a detached retina.
{1}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
Device Generic Name: Intraocular Gas
Device Trade Name: UNIPURE C₃F₈ Ophthalmic Gas in the UNIFEYE Gas Delivery System
UNIPURE C₃F₈ Ophthalmic Gas in the UNIPEXY Gas Delivery System
Device Procode: LPO
Applicant’s Name and Address: Alcon Research, LLC
20511 Lake Forest Drive,
Lake Forest, CA 92630, USA
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P220030
Date of FDA Notice of Approval: 7/02/2024
II. INDICATIONS FOR USE
The UNIPURE C₃F₈ Ophthalmic Gas in the UNIFEYE Gas Delivery System is indicated for intraocular injection into the eye for the treatment of uncomplicated retinal detachments. Associated measures used include vitrectomy, fluid/air exchange, transconjunctival and transscleral cryotherapy, laser photocoagulation, and air/gas exchange.
The UNIPURE C₃F₈ Ophthalmic Gas in the UNIPEXY Gas Delivery System is indicated for intraocular injection into the eye for the treatment of uncomplicated retinal detachments. Associated measures used include vitrectomy, fluid/air exchange, transconjunctival and transscleral cryotherapy, and laser photocoagulation.
III. CONTRAINDICATIONS
Proliferative vitreoretinopathy (PVR) greater than Stage C, the mental or physical inability to maintain the therapeutic position for 5 postoperative days, severe glaucoma with more than a minimum of vision field loss and a cup to disc ratio equal to or greater than 0.6; uveitis; severe peripheral retinal degeneration; congenital malformations (such as coloboma), and any other condition that may facilitate the migration of the gas bubble out of the vitreous chamber; and high-altitude travel, including but not limited to airline travel.
PMA P220030: FDA Summary of Safety and Effectiveness Data
{2}
- Air travel is contraindicated until the gas/air bubble has completely dissipated. Normal cabin pressure changes will cause a severe enlargement of the gas/air bubble with a resultant potential blinding, due to an increase in intraocular pressure $(\mathrm{IOP})^{1-7}$ .
- Patients should not travel through high elevations and over mountain ranges until the mixed gas/air bubble has dissipated $^{8}$ .
# IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas in the UNIFEYE Gas Delivery System and in the UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas in the UNIPEXY Gas Delivery System labeling.
# V. DEVICE DESCRIPTION
The integrated UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas pico-cylinder in the UNIFEYE Gas Delivery System contains undiluted, non-sterile, liquefied perfluoropropane $(\mathrm{C_3F_8})$ gas under pressure. The gas is non-toxic, inert, non-flammable, odorless, and colorless. The UNIFEYE Gas Delivery System is used to mix filtered UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas with filtered air and inject the gas/air mix into the vitreous cavity of the eye. The prepared gas/air mixture for injection is sterile via the 0.2-micron filtration of the gas and air.

Figure 1. UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas in the UNIFEYE Gas Delivery System
The integrated UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas pico-cylinder in the UNIPEXY Gas Delivery System contains undiluted, non-sterile, liquefied perfluoropropane $(\mathrm{C_3F_8})$ gas under pressure. The gas is non-toxic, inert, non-flammable, odorless, and colorless. The UNIPEXY Gas Delivery System is used to inject filtered UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas directly into the vitreous cavity to form a gas/air bubble. The prepared gas for injection is sterile via the 0.2-micron filtration of the gas.
PMA P220030: FDA Summary of Safety and Effectiveness Data
{3}

Figure 2. UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas in the UNIPEXY Gas Delivery System
The physical properties for the UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas in the UNIFEYE Gas Delivery System and the UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas in the UNIPEXY Gas Delivery System are presented in Table 1.
Table 1: Summary of Physical Properties
| Property | Description |
| --- | --- |
| Physical Attributes | colorless, odorless, non-toxic, and non-flammable gas at room temperature and atmospheric pressure |
| Molecular Formula | C3F8 |
| Molecular Weight | 188 g/mol |
| CAS Registry Number | 76-19-7 |
| Boiling Point | -34.1 °F / -36.7 °C |
| Vapor Pressure | 114.8 psia / 792 kPa at 21.1 °C |
| Liquid Density | 1.352 kg/L at 20 °C |
| Gas Density | 8.003 kg/m3at 20 °C, 1 atm |
| Viscosity | 0.01454 Cp at 25 °C, 1 atm |
| Thermal Conductivity | 0.01381 W/mol °K at 25 °C, 1 atm |
| Freezing Point | -297.4 °F/ -183.0 °C |
| Critical Temperature | 161.4 °F/ 71.9 °C |
| Critical Pressure | 338.7 psia/ 2680 kPa |
| Critical Volume | 1.590 dm3/kg |
| Critical Density | 0.629 kg/dm3 |
| Critical Compressibility Factor | 0.279 |
PMA P220030: FDA Summary of Safety and Effectiveness Data
{4}
VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several other alternatives for the correction of uncomplicated retinal detachments.
The primary alternative practice to use of UNIPURE C₃F₈ Ophthalmic Gas is the use of other sources of the same type of gas, other types of gases or air to form an ocular endotamponade. Currently, both perfluoropropane (C₃F₈) and sulfur hexafluoride (SF₆) gases are approved, for commercial sale as gas ocular endotamponades in the U.S. No other gas ocular endotamponades have been approved for commercial sale in the U.S. These two ocular endotamponades are sold separately from accessory devices that would be needed to prepare them for use in intraocular surgery. Air-only ocular endotamponades are also used.
An alternative to using a gas ocular endotamponade with or without vitrectomy is vitrectomy with a silicone oil endotamponade. This is usually for patients who are not capable of or who wish to avoid postoperative positioning regimens, and for patients wishing to have uninterrupted vision while the endotamponade is in the eye.
Another alternative is a scleral buckling procedure, which is frequently chosen for younger, more active patients who wish to avoid postoperative positioning regimen, or to avoid the risk of cataract due to iatrogenic damage to the lens and who are open to spectacle use (due to distortion of the eye by the buckle). In many cases, scleral buckling procedures are combined with vitrectomy either with or without gas ocular endotamponades.
Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
VII. MARKETING HISTORY
The UNIPURE C₃F₈ Ophthalmic Gas in the UNIFEYE Gas Delivery System and the UNIPURE C₃F₈ Ophthalmic Gas in the UNIPEXY Gas Delivery System have not been marketed in the United States or any foreign country.
VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device.
Operative complications associated with the use of gas/air ocular endotamponades in eyes with or without vitrectomy may include:
- Central retinal artery occlusion
- Subconjunctival gas
PMA P220030: FDA Summary of Safety and Effectiveness Data
{5}
- Subretinal hemorrhage
- Small subretinal gas bubble
- Hypotony
- Choroidal hemorrhage
- Choroidal detachment
- Crystalline lens touch by needle
- Hyphema
- Escape of mixed gas/air through the surgical incisions
- Vitreous or iris incarceration at the wound
- Elevated IOP, which may require additional medical or surgical intervention to reduce pressure
Postoperative complications associated with surgical procedures using gas/air ocular endotamponades in eyes with or without vitrectomy may include:
- Elevated IOP, which may require additional medical or surgical intervention to reduce pressure
- Central retinal artery occlusion
- Malignant glaucoma
- Choroidal hemorrhage
- Choroidal detachment
- Changes to the crystalline lens
- Cataract
- Severe elevated IOP that has been known to result in vision decrease or blindness if N₂O is administered during a subsequent surgical or dental procedure with a gas bubble present in the eye
- Uveitis
- Hyphema
- Cystoid macular edema/macular edema
- New, missed, or recurrent retinal detachment or retinal breaks
- Subconjunctival gas
- Subconjunctival hemorrhage
- Vitreous hemorrhage
- Subretinal fluid
- Subretinal hemorrhage
- Subretinal gas
- Macular hole
- Macular pucker/epiretinal membrane
- Proliferative vitreoretinopathy
- Extrafoveal subretinal pigment migration
- Vitreal opacification (known as "floaters" or "tobacco dust")
- Endophthalmitis
- Refractive changes
- Escape of mixed gas/air through the surgical incisions
PMA P220030: FDA Summary of Safety and Effectiveness Data
5 of 27
{6}
- Vitreous or iris incarceration at the wound
For the specific adverse events that occurred in the clinical studies reported in literature, please see Section X below.
# IX. SUMMARY OF NON-CLINICAL STUDIES
# A. Laboratory Studies
Table 2: Physical and Chemical Characterization Summary
| Test | Acceptance Criteria | Results | Analysis Type |
| --- | --- | --- | --- |
| Characterization Study | Comparison of quantities of the targeted and unknown impurities (control [marketed] and test [proposed] gases). | N/A | There is no pass/fail criteria for this study. Test results from the characterization study demonstrated the impurity profile of the as-delivered UNIPURE C3F8 Ophthalmic Gas and the C3F8 control gas is equivalent. No targeted impurities were detected above specification limits and no unknown impurities were detected in the test UNIPURE C3F8 Ophthalmic Gas. |
| Extractable and Leachable | Performed in compliance with ISO 10993- 18:2020. | Pass | UNIPEXY Gas Delivery System components were successfully characterized for extractables per ISO 10993-18:2020. |
PMA P220030: FDA Summary of Safety and Effectiveness Data
{7}
A toxicological risk assessment was performed on the as-delivered UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas per ISO 10993-17.
# B. Animal Studies
# 1. Intraocular Implantation Test - UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas
The UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas is categorized per ISO 10993-1:2018, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, as an implant medical device with long-term contact duration (greater than 30 days) with tissue. Intraocular implantation in New Zealand White (NZW) rabbit eyes with histopathology was conducted according to Annex A of ISO 16672:2020 to determine the ocular toxicity of perfluoropropane $(\mathrm{C_3F_8})$ . The contralateral eyes served as control and received either the commercially available and unmodified ISPAN Perfluoropropane $(\mathrm{C_3F_8})$ gas (P900066) or air. The ophthalmic tissue responses in the eyes implanted with the UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas as delivered by the UNIFEYE and UNIPEXY Gas Delivery Systems were compared to tissue responses of the control eyes. The rabbit intraocular implantation study (Table 3) showed that the tissue responses in eyes implanted with UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas and the commercially available ISPAN $\mathrm{C_3F_8}$ gas are comparable, and the UNIPURE $\mathrm{C_3F_8}$ Ophthalmic Gas does not lead to toxicity and inflammation in the animal model. The animal implantation study was conducted in accordance with FDA Title 21, CFR Part 58: Good Laboratory Practice for Non-Clinical Laboratory Studies.
Table 3: UNIPURE ${\mathrm{C}}_{3}{\mathrm{\;F}}_{8}$ Ophthalmic Gas Animal Study Results
| Endpoint | Test | Animal model | Results |
| --- | --- | --- | --- |
| Implantation | Ocular Implantation (Eight-Weeks; Rabbit) in accordance with ISO 16672:2020 | NZW rabbit | Non-toxic; Non-Inflammatory |
PMA P220030: FDA Summary of Safety and Effectiveness Data
{8}
PMA P220030: FDA Summary of Safety and Effectiveness Data
8 of 27
2. Biocompatibility Test - Pico-cylinder, UNIFEYE and UNIPEXY Gas Delivery Systems
The UNIFEYE Gas Delivery System and pico-cylinder components have indirect tissue contact via the UNIPURE C₃F₈ Ophthalmic Gas.
The UNIPEXY Gas Delivery System is intended to provide the surgeon with a filled syringe of undiluted UNIPURE C₃F₈ Ophthalmic Gas for delivery to the eye via an included syringe needle. Except for the syringe needle and hub that have direct limited (≤ 24 h) tissue contact, all the components of the UNIPEXY Gas Delivery System have indirect tissue contact via the UNIPURE C₃F₈ Ophthalmic Gas.
Biocompatibility assessment of the pico-cylinder and the UNIFEYE and UNIPEXY Gas Delivery System was conducted in accordance with ISO 10993-1:2018 components. The biocompatibility tests were conducted in accordance with 21 CFR Part 58. The biocompatibility test results are summarized in Table 4.
Table 4: UNIFEYE, UNIPEXY, Pico-cylinder Biocompatibility Results
| Endpoint | Test | Results |
| --- | --- | --- |
| Cytotoxicity | MEM Elution Assay using L-929 Fibroblasts Cells (ISO 10993-5:2009) | No evidence of cell lysis or cell toxicity |
| Irritation | Intraocular Irritation (ISO 10993-10:2013) | Not irritating to intraocular tissues |
| Sensitization | Guinea Pig Maximization Test(ISO 10993-10:2013) | No evidence of delayed dermal contact sensitization |
C. Additional Studies
{9}
Table 5: Summary of Additional Studies
| Test | Acceptance Criteria | Results | Analysis Type |
| --- | --- | --- | --- |
| Sterilization | | | |
| Sterilization | Acceptance criteria in accordance with ISO 11135:2014 and EN 556-1:2001. | Pass | Sterility assurance level of 10^{-6} was demonstrated for EO sterilization process through microbiological testing of control microorganisms. |
| EO/ECH Residual | Acceptance criteria in accordance with ISO 10993-7:2008. | Pass | Results for EO and ECH residual levels met permanent intraocular device acceptance criteria per the limits in EN ISO 10993-7:2008. |
| Endotoxin | Acceptance criteria in accordance with ANSI/AAMI ST72 and 2015 FDA Guidance, Endotoxin Testing Recommendations for Single-Use Intraocular Ophthalmic Devices | Pass | Endotoxin test results conform to the intraocular device limit of ≤0.2 Endotoxin Units (EU)/device |
| Shelf Life | | | |
| Stability | Acceptance criteria in accordance with ASTM F1980-16 and ASTM D4169-16. | Pass | Results indicated that product is stable at elevated temperature and transportation conditioning at 22-month shelf-life. |
| Package Integrity | Acceptance criteria in accordance with ISO 11607-1:2019+LC2020, ISO 11607-2:2019, ASTM F1929-12, | Pass | Results of dye penetration, bubble leak test, and seal strength testing met respective ASTM test method acceptance |
PMA P220030: FDA Summary of Safety and Effectiveness Data
{10}
| | ASTM F2096-11, and ASTM F88/F88M-15 | | criteria, and conform to ISO 11607-1 and ISO 11607-2 requirements. |
| --- | --- | --- | --- |
| Functional and Performance | | | |
| Functional and Performance Tests
UNIFEYE Gas Delivery System
- Latch
Mechanical Hold
- Ratchet
Mechanism
Break Force
- Plunger Draw
Force
- Force to Expel
Mixed Gas
- User Selection
- Gas Release
- Automatic
Plunger
Movement
- Unmixed Gas
Volume
Indicators
- Air Draw
- 50 mL Stop
- Ratchet
Mechanism
Engagement
- Expelling Mixed Gas
- Lever Packaging
Securement
- Purge Air
- Visual Inspection
of Piston and O-ring Assembly
- Destructive Testing
- Helium Leak
Testing | Acceptance criteria in accordance with product specification. | Pass | The device met all functional and performance acceptance criteria per product specification throughout the product shelf life. |
PMA P220030: FDA Summary of Safety and Effectiveness Data
10 of 27
{11}
| - Gap and Height Inspection
- Pre- to Post-Weight Check
- Semi-Automatic Weight Assessment of Microcylinders (SWAMY) Leak Check
- Batch Testing for Purity | | | |
| --- | --- | --- | --- |
| Functional and Performance Tests
**UNIPEXY Gas Delivery System**
- Latch Break Test
- Package Protection
- Device Activation
- Plunger Stop
- Gas Vent
- Latch Open and Fill Volume
- Expel Gas
- Needle Pull Test
- Visual Inspection of Piston and O-ring Assembly
- Destructive Testing
- Helium Leak Testing
- Gap and Height Inspection
- Pre- to Post-Weight Check
- SWAMY Leak Check
- Batch Testing for Purity
- Pad Printing Inspection | Acceptance criteria in accordance with product specification. | Pass | The device met all functional and performance acceptance criteria per product specification throughout the product shelf life. |
PMA P220030: FDA Summary of Safety and Effectiveness Data
11 of 27
{12}
PMA P220030: FDA Summary of Safety and Effectiveness Data
12 of 27
| - Body Bleed Filter and Flow Restrictor Leak Testing
- Component Cleaning Water Conductivity Testing | | | |
| --- | --- | --- | --- |
| Mix Ratio Accuracy Test | Upper and lower tolerance limits within the ±1.5% per product specification. | Pass | Test results demonstrated that the UNIFEYE Gas Delivery System upper and lower tolerance limits are within the ±1.5% specification for each gas/air mix ratio throughout product shelf life. |
| Filter Integrity Tests | Acceptance criteria in accordance with ASTM F1929-12. | Pass | Gas filter integrity was maintained throughout the product shelf life and met acceptance criteria per ASTM F1929-12. |
| Particulate Test (Gas) | Acceptance criteria in accordance with USP <788>:2012 and USP <789>:2012. | Pass | Test results provided evidence that the test articles meet the acceptance criteria per USP <788>:2012 and <789>:2012, respectively. |
| Canister Burst Test | Withstand at least 1300 psi of internal pressure without catastrophic failure. | Pass | Test results demonstrated that the pico-cylinder could withstand at least 1300 psi of |
{13}
PMA P220030: FDA Summary of Safety and Effectiveness Data
13 of 27
| | | | internal pressure without catastrophic failure. |
| --- | --- | --- | --- |
| Gas Weight Verification | At least 200 mg through product shelf-life. | Pass | Test results for the pico-cylinder were within the acceptance criteria throughout product shelf-life. |
| **Usability and Human Factors** | | | |
| Summative Human Factors and DFU Comprehension Studies | All critical tasks were evaluated based on defined successful task performance | Two (2) user groups performed all critical tasks, use problems were captured and assessed for root cause analysis, residual risks and effectiveness of risk control measures were evaluated to demonstrate use-related risks were reduced to acceptable levels | The subject device and its accessories are safe and effective for the intended users, uses, and use environments in alignment with FDA's 2016 Guidance - Applying Human Factors and Usability Engineering to Medical Devices, and in adherence to IEC 62366-1:2015 and ANSI/AAMI HE75:2018. |
{14}
PMA P220030: FDA Summary of Safety and Effectiveness Data
14 of 27
# X. SUMMARY OF PRIMARY CLINICAL STUDY
The applicant performed a systematic clinical literature review to establish a reasonable assurance of safety and effectiveness of the UNIPURE C₃F₈ Ophthalmic Gas in the UNIFEYE and UNIPEXY Gas Delivery Systems for the treatment of uncomplicated retinal detachment. Data from this clinical literature review were the basis for the PMA approval decision. A summary of the clinical literature review is presented below.
## A. Study Design
Data was evaluated from published literature to generate clinical evidence that supports the safety and effectiveness of the UNIPURE C₃F₈ Ophthalmic Gas in the UNIFEYE and UNIPEXY Gas Delivery Systems. Online literature databases, including Medline, Embase, Cochrane, Clinicaltrials.gov, and WHO International Clinical Trial Registry Platform, were searched for a period spanning January 1, 1980 to May 16, 2022. The range of databases searched aimed to provide adequate coverage of treatments in use worldwide and identified relevant publications of user experience for the indication of treatment of uncomplicated retinal detachments. The output from the search results were then reviewed for inclusion in the clinical literature review using the PICO (patient characteristics, type of intervention, control, and outcome of interest) review method. The first review was performed by title/abstract and articles remaining were then reviewed as full text according to the following criterion.
### 1. Clinical Inclusion and Exclusion Criteria
For an article to be recommended for inclusion in the safety and effectiveness analysis, it must have met at least one of the following criteria:
a. Clinical studies and/or studies in animals evaluating the safety and/or effectiveness of the evaluated device or equivalent device; or
b. Isolated case reports that describe new risk(s).
Articles were excluded based on the following criteria:
a. Articles not related to the device of interest or equivalent device.
b. Articles that did not provide data on the safety and/or effectiveness of the device of interest or equivalent device in humans or in animals.
c. Articles with date of publication outside the reporting period.
d. Letters to the editor, opinions, editorials, surveys, manufacturer’s advertisements, and press releases.
e. Laboratory research, such as in vitro or ex vivo studies, post-mortem studies, biomechanical studies, simulation studies.
f. Non-peer reviewed articles.
g. Abstracts or conference proceedings.
h. Review articles.
i. Isolated case reports, unless new risks are described.
j. Duplicate articles or duplicate publications of the same study data.
k. Articles using the device of interest in ways unrelated to the indicated use.
{15}
1. Articles using the device of interest with associated procedures that are not included in the indicated use.
m. Articles from registries without final results.
2. **Clinical Endpoints**
With regards to safety, the primary safety endpoint was the Rate of Elevated Intraocular Pressure (IOP): Report of elevated IOP >25 mm Hg in literature.
Target Rate: Elevated IOP (defined as >25 mm Hg) reported in literature will be no more than 34% reported 1 to 7 days postoperatively.
With regards to effectiveness, the primary effectiveness endpoint was the Anatomical Success Rate: Primary retinal reattachment rate.
Target Rate: The primary operational success rate reported in literature will be at least 72% reported at least 3 months postoperatively.
3. **Statistical Methods**
Primary safety and effectiveness endpoint data were analyzed using a logistic regression model fit with a random effect to account for study heterogeneity. Data were reported with the 95% confidence intervals of the primary endpoints and forest plots and comparisons were made between the point estimate and the proposed primary endpoint target rates.
B. **Accountability of PMA Cohort**
N/A
C. **Study Population Demographics and Baseline Parameters**
The demographics of the literature review population are shown in Table 6 and are typical for uncomplicated retinal detachment treatments in the US. The mean age for C₃F₈-treated participants was 59.2 years with an age range of 10 to 93 years. Nearly 65% of the 2270 participants were male.
PMA P220030: FDA Summary of Safety and Effectiveness Data
15 of 27
{16}
Table 6 Demographics of Literature Review Population
| Total Study Participants* N=4612 | | | Participants Treated With C3F8 N=2270 | | |
| --- | --- | --- | --- | --- | --- |
| Mean Age, year | Age Range, year | Male, | Mean Age, year | Age Range, year | Male, |
| Weighted mean = 57.9 | Range = 8-94 | 66.2% | Weighted mean = 59.2 | Range = 10-93 | 64.9% |
*Treatments including ${\mathrm{C}}_{3}{\mathrm{\;F}}_{8}$ and other tamponade agents and procedures
# D. Safety and Effectiveness Results
The literature review included 46 studies that reported safety and effectiveness of the use of $\mathrm{C_3F_8}$ gas for the treatment of uncomplicated retinal detachments. Of these, 16 studies
reported rates of elevated IOP and 35 studies reported primary retinal reattachment rates. Studies with data eligible for quantitative analyses included 6 for safety outcomes and 21 for effectiveness outcomes.
# 1. Safety Results
The analysis of safety was based on the quantitative cohort of 153 eyes reported with rates of elevated IOP $>25\mathrm{mmHg}$ at 1- to-7-day postoperative visits. The overall rate of elevated IOP $>25\mathrm{mmHg}$ reported 1 to 7 days postoperatively in the analyzed studies was $13\%$ (20/153), $(95\% \mathrm{CI}: 2.6\%, 46.2\%)$ , which is less than the
prespecified threshold of $34\%$ . The key safety outcomes for this study are presented below in Table 7 and Figure 1. Adverse effects are reported in Tables 8 and 9.
Table 7 Safety - Elevated IOP Rate,
All Studies and Subgrouped by Study Type
| Study type | Number of studies (n) | Mean | 95% CI |
| --- | --- | --- | --- |
| All studies | 6 | 0.1307 | (0.0257, 0.4616) |
| B | 2 | 0.5667 | (0.3464, 0.7634) |
| C | 3 | 0.0316 | (0.0048, 0.1813) |
| E | 1 | 0.0000 | NA |
PMA P220030: FDA Summary of Safety and Effectiveness Data
{17}

Figure 1 Forest Plot: Safety - Elevated IOP Rate-All Studies
Adverse events (AEs) and complications were reported in the literature review in 37 of 46 included articles. All events were tabulated, including events that were not likely device-related. Twenty-nine articles reported AEs specifically in eyes treated with $\mathrm{C_3F_8}$ gas only, while eight reported on other tamponades or treatments (mixed). AEs reported without specification of the agent/treatment (mixed) were tabulated separately.
AEs were categorized by treatment (C₃F₈ or mixed treatment), time of occurrence and by anterior segment/posterior segment/surgical complication/vision/other AEs.
Tables 8 and 9 show the reported intraoperative adverse events for $\mathrm{C_3F_8}$ and mixed studies.
PMA P220030: FDA Summary of Safety and Effectiveness Data
{18}
Table 8 Intraoperative Adverse Events of C₃F₈ Reported in Included Studies
| Category | Intraoperative AEs | S #* | n | N | % |
| --- | --- | --- | --- | --- | --- |
| Posterior segment | New retinal breaks | 2 | 4 | 82 | 4.9 |
| | Retinal artery occlusion | 2 | 37 | 68 | 54.4 |
| | Retained Perfluoro-n-octane | 1 | 1 | 22 | 4.5 |
| | Tear enlargement | 1 | 2 | 22 | 9.1 |
| Surgical complications | Sclerotomy tear | 1 | 1 | 22 | 4.5 |
*Number of studies
Table 9 Intraoperative Adverse Events of C₃F₈ with Mixed Agent or Procedure Reported in Included Studies
| Category | Intraoperative AEs | S#* | n | N | % |
| --- | --- | --- | --- | --- | --- |
| Anterior Segment | Hyphema | 1 | 1 | 65 | 1.5 |
| | Subconjunctival gas | 1 | 3 | 100 | 3.0 |
| Posterior Segment | Vitreous hemorrhage | 2 | 2 | 120 | 1.7 |
| | Vitreous strand incarcerated in the paracentesis site | 1 | 1 | 100 | 1.0 |
| Surgical complications | Detached pars plana epithelium | 1 | 1 | 100 | 1.0 |
| | Sclerotomy leakage | 1 | 3 | 126 | 2.4 |
*Number of studies
Tables 10 and 11 show the reported postoperative adverse events for C₃F₈ and mixed studies.
PMA P220030: FDA Summary of Safety and Effectiveness Data
{19}
Table 10: Postoperative Adverse Events of ${\mathrm{C}}_{3}{\mathrm{\;F}}_{8}$ Reported in Included Studies
| Category | Reported AE | Timeframe Reported | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Day1 to ≤ Month 1 | | | >Month 1 to < Month 6 | | ≥ Month 6 | | | Unspecified | | |
| | | S#* | n / N | % | S#* | n / N | % | S#* | n / N | % | S#* | n / N |
| Anterior segment | Cataract | - | - | - | - | - | - | 1 | 5 / 27 | 18.5 | 4 | 89 / 306 |
| | Collapsed / shallow anterior chamber | 1 | 9 / 215 | 4.2 | - | - | - | - | - | - | 1 | 6 / 60 |
| | Glaucoma | - | - | - | 1 | 8 / 23 | 34.8 | - | - | - | 2 | 7 / 90 |
| | Hypotony | - | - | - | - | - | - | - | - | - | 1 | 3 / 60 |
| | Increased intraocular pressure (IOP) | 3 | 20 / 57 | 35.1 | - | - | - | - | - | - | 2 | 28 / 169 |
| | Subconjunctival gas | - | - | - | - | - | - | - | - | - | 1 | 1 / 28 |
| | Subconjunctival hemorrhage | - | - | - | - | - | - | - | - | - | 1 | 11 / 60 |
| | Transient lens opacification | 1 | 5 / 10 | 50.0 | - | - | - | - | - | - | 3 | 15 / 107 |
| Posterior segment | Cystoid macular edema (CME) | - | - | - | - | - | - | - | - | - | 2 | 5 / 169 |
| | Epiratinal membrane (ERM)/macular pucker | - | - | - | 2 | 16 / 54 | 29.6 | 1 | 27 / 58 | 46.6 | 5 | 19 / 256 |
| | New retinal breaks | 1 | 7/58 | 12.1 | - | - | - | - | - | - | 3 | 11 / 123 |
| | New retinal detachments | 1 | 5 / 58 | 8.6 | - | - | - | - | - | - | - | - |
| | Proliferative vitreoretinopathy (PVR) | - | - | - | - | - | - | - | - | - | 2 | 11 / 94 |
| | PVR progression | - | - | - | - | - | - | - | - | - | 1 | 7 / 54 |
| | Retained Perfluoro-n-octane | - | - | - | - | - | - | - | - | - | 3 | 10 / 173 |
| | Retinal re-detachment | - | - | - | 1 | 1 / 5 | 20.0 | - | - | - | 2 | 10 / 88 |
| | Subretinal gas | - | - | - | - | - | - | - | - | - | 1 | 1 / 63 |
| | Vitreous cell reaction | 1 | 10 / 17 | 58.8 | - | - | - | - | - | - | - | - |
PMA P220030: FDA Summary of Safety and Effectiveness Data
{20}
Table 11: Postoperative Adverse Events of ${\mathrm{C}}_{3}{\mathrm{\;F}}_{8}$ with Mixed Agent or Procedure Reported in Included Studies
| Category | Reported AE | Timeframe Reported | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Day 1 to ≤ Month 1 | | > Month 1 to < Month 6 | | | ≥ Month 6 | | | Unspecified | | |
| | | S#* | n / N | % | S#* | n / N | % | S#* | n / N | % | S#* | n / N |
| Anterior segment | Cataract | - | - | - | - | - | - | 1 | 1 / 43 | 2.3 | 2 | 52 / 154 |
| | Choroidal detachment | 1 | 1 / 43 | 2.3 | - | - | - | - | - | - | - | - |
| | Choroidal hemorrhage | - | - | - | - | - | - | - | - | 1 | 1 / 43 | 2.3 |
| | Glaucoma | - | - | - | - | - | - | - | - | 1 | 6 / 82 | 7.3 |
| | Hypotony | - | - | - | - | - | - | - | - | 2 | 2 / 208 | 1.0 |
| | Increased IOP | - | - | - | - | - | - | - | - | 2 | 38 / 158 | 24.1 |
| | Pupillary block | - | - | - | - | - | - | - | - | 1 | 1 / 126 | 0.8 |
PMA P220030: FDA Summary of Safety and Effectiveness Data
{21}
| | Subconjunctival gas | - | - | - | - | - | - | - | - | 1 | 2 / 20 | 10.0 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Subconjunctival hemorrhage | - | - | - | - | - | - | - | - | 1 | 12 / 20 | 60.0 |
| Transient lens opacification | - | - | - | - | - | 1 | 9 / 43 | 20.9 | - | - | - |
| Uveitis | - | - | - | - | - | - | - | - | 2 | 2 / 120 | 1.7 |
| Posterior segment | Central retinal artery occlusion | - | - | - | - | - | 1 | 1 / 43 | 2.3 | - | - | - |
| CME | - | - | - | - | - | - | - | - | 4 | 16 / 396 | 4.0 |
| ERM (macular pucker) | - | - | - | - | - | - | - | - | 4 | 10 / 420 | 2.4 |
| New retinal breaks | - | - | - | - | - | - | - | - | 4 | 20 / 323 | 6.2 |
| PVR | - | - | - | - | - | - | - | - | 3 | 8 / 303 | 2.6 |
| Retinal displacement | - | - | - | 1 | 12 / 21 | 57.1 | - | - | - | - | - |
| Retinal re-detachment | - | - | - | - | - | - | - | - | 2 | 5 / 138 | 3.6 |
| Subretinal fluid | 1 | 15 / 32 | 46.9 | - | - | - | - | - | 2 | 7 / 87 | 8.0 |
| Subretinal hemorrhage | - | - | - | - | - | - | - | - | 1 | 6 / 56 | 10.7 |
| Subretinal pigment | - | - | - | - | - | - | - | - | 2 | 2 / 120 | 1.7 |
| Subretinal residual perfluorocarbon | - | - | - | - | - | - | - | - | 1 | 2 / 147 | 1.4 |
| Tear enlargement | - | - | - | - | - | - | - | - | 1 | 1 / 56 | 1.8 |
| Vitreous floaters / opacities | 1 | 1 / 43 | 2.3 | - | - | - | - | - | 3 | 19 / 140 | 13.6 |
| Vitreous hemorrhage | -- | -- | -- | - | - | - | - | - | 1 | 2 / 56 | 3.6 |
| Secondary intervention | 1 | 1 / 126 | 0.8 | - | - | - | 1 | 4 / 88 | 4.5 | 6 | 50 / 846 | 5.9 |
| Vision | VA deterioration | - | - | - | - | - | - | - | - | 1 | 5 / 100 | 5.0 |
| Visual disturbance (metamorphopsia) | - | - | - | - | - | 1 | 59 / 60 | 98.3 | - | - | - |
PMA P220030: FDA Summary of Safety and Effectiveness Data
{22}
| Other ocular AE | Intravitreal dexamethasone injection | - | - | - | - | - | - | 1 | 27 / 88 | 30.7 | - | - | - |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Non-ocular AE | Headache | - | - | - | - | - | - | - | - | - | 1 | 1 / 20 | 5.0 |
| | Nausea | - | - | - | - | - | - | - | - | - | 1 | 1 / 20 | 5.0 |
| | Neck pain | - | - | - | - | - | - | - | - | - | 1 | 1 / 20 | 5.0 |
*Number of studies
PMA P220030: FDA Summary of Safety and Effectiveness Data
22 of 27
{23}
The most commonly reported possible device-related intraoperative AEs were retinal artery occlusion (54.4% of eyes in 2 studies) and subconjunctival gas (3% of eyes in 1 study). Other reported intraoperative AEs were likely associated with the surgical procedure.
The most commonly reported possible $\mathrm{C_3F_8}$ device-related postoperative AEs were cataract (28% of eyes in 5 studies), increased IOP (21.2% of eyes in 5 studies), secondary surgical intervention (18.4% of eyes in 21 studies), transient lens opacification (17.1% of eyes in 4 studies), glaucoma (13.3% of eyes in 3 studies), subconjunctival gas (3.6% of eyes in 1 study), and subretinal gas (1.6% of eyes in 1 study). Other postoperative AEs were likely associated with the surgical procedure or not related to either the device or procedure.
AEs reported in the mixed agent/procedure studies were similar in events and frequency.
# 2. Effectiveness Results
The analysis of effectiveness was based on the quantitative cohort of 1131 eyes reported with rates of primary retinal reattachment at $\geq 3$ -months postoperatively. The overall primary retinal reattachment rate reported $\geq 3$ -months postoperatively in the analyzed studies was $81.9\%$ (95% CI: $76.4\%$ , $86.3\%$ ), which exceeds the prespecified threshold of $72\%$ . The key effectiveness outcomes for this study are presented below in Table 12 and Figure 2.
Table 12 Effectiveness: Primary Retinal Reattachment Rate, All Studies and Subgrouped by Study Type
| Study Type | Number of Studies (n) | Mean | 95% CI |
| --- | --- | --- | --- |
| All studies | 21 | 0.8187 | (0.7641, 0.8630) |
| A | 2 | 0.7634 | (0.7460, 0.7800) |
| C | 7 | 0.8586 | (0.7287, 0.9321) |
| D | 1 | 0.6667 | (0.4733, 0.8166) |
| E | 11 | 0.8092 | (0.7496, 0.8574) |
PMA P220030: FDA Summary of Safety and Effectiveness Data
{24}

Figure 2 Forest Plot: Effectiveness - Primary Retinal Reattachment Rate -All Studies
# 3. Subgroup Analyses
Subgroup analyses were conducted on the most recent literature published from January 1, 2012 to May 16, 2022. All recent literature were included in the above results from the original search starting January 1, 1980. The focus of these subgroup analyses were to identify changes in outcomes or treatments in the last 10 years. The search resulted in an observed trend of more publications reported using $\mathrm{C_3F_8}$ during vitrectomy surgery rather than pneumatic retinopexy and fewer publications reported elevated IOP rates. Regarding effectiveness, the overall primary retinal reattachment rate in the analyzed vitrectomy surgery studies $(n = 8)$ was $86.5\%$ (95% CI: $81.1\%$ , $90.5\%$ ) and the overall primary retinal reattachment rate in the analyzed pneumatic retinopexy studies $(n = 8)$ was $67.6\%$ (95% CI: $62.5\%$ , $72.4\%$ ). The rate of elevated IOP in the analyzed vitrectomy surgery studies $(n = 3)$ was $35.1\%$ (95% CI: $10.6\%$ , $71.1\%$ ) and there were no reports of elevated IOP in the analyzed pneumatic retinopexy studies $(n = 3)$ .
No analyses were performed for sex, gender, age, race, ethnicity, or other subgroups.
# 4. Pediatric Extrapolation
PMA P220030: FDA Summary of Safety and Effectiveness Data
{25}
In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.
## XI. FINANCIAL DISCLOSURE
N/A
## XII. SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION
N/A
## XIII. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Ophthalmic Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XIV. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
### A. Effectiveness Conclusions
The overall primary retinal reattachment rate reported ≥3-months postoperatively in the quantitative effectiveness analysis was 81.9% (95% CI: 76.4%, 86.3%), which exceeds the prespecified threshold of 72%. Therefore, the success criteria for the primary effectiveness outcome is met and the objective clinical evidence supports the effectiveness of the subject device due to the comparable C₃F₈ gas chemical profiles to the marketed/studied C₃F₈ gas.
### B. Safety Conclusions
The risks of the device are based on nonclinical laboratory and animal studies (as summarized in Section IX above) as well as data collected in a clinical literature review conducted to support PMA approval as described above. Literature reported a tolerable side-effect profile, specifically in terms of adverse event rates, which were within acceptable postoperative safety limits. The overall rate of elevated IOP >25 mmHg reported 1 to 7 days postoperatively in the quantitative safety analysis was 13% (95% CI: 2.6%, 46.2%), which is less than the prespecified threshold of 34%. Therefore, the success criteria for the primary safety outcome is met and the objective clinical evidence supports the safety of the subject device.
### C. Benefit-Risk Determination
PMA P220030: FDA Summary of Safety and Effectiveness Data
25 of 27
{26}
The probable benefits of the device are also based on data reported in clinical literature published since 1980 to support PMA approval as described above. The literature shows that this device is one of several options for treating uncomplicated retinal detachment and is considered the standard of care. This device allows patients to be treated for uncomplicated retinal detachment with primary success rates of reattachment of over 80% and acceptable safety profiles of elevated IOP known to occur in ocular surgeries.
The probable risks of the device are also based on data reported in clinical literature published since 1980 to support PMA approval as described above. Risks reported were in the form of adverse events and complications reported in literature for the surgical procedure of uncomplicated retinal detachment repair. Literature reported a tolerable side-effect profile, specifically in terms of adverse event rates, which were within acceptable postoperative safety limits. Possible adverse events could include, but are not limited to, elevated intraocular pressure, retinal redettachment, subretinal migration of gas, escape of gas through surgical incisions, cataract, retinal artery occlusion, and macular hole.
1. Patient Perspective
This submission either did not include specific information on patient perspectives or the information did not serve as part of the basis of the decision to approve or deny the PMA for this device.
In conclusion, given the available information above, the data support that for the use of UNIPURE C₃F₈ Ophthalmic Gas in the UNIFEYE and UNIPEXY Gas Delivery Systems in the treatment of uncomplicated retinal detachments the probable benefits outweigh the probable risks.
D. Overall Conclusions
The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. The literature search reports did not introduce any new risks and conclude that the benefits of the device outweigh the risks when used as intended.
XV. CDRH DECISION
CDRH issued an approval order on 7/02/2024.
The applicant’s manufacturing facilities have been inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
XVI. APPROVAL SPECIFICATIONS
PMA P220030: FDA Summary of Safety and Effectiveness Data
26 of 27
{27}
Directions for use: See device labeling.
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.
Post-approval Requirements and Restrictions: See approval order.
## XVII. REFERENCES
1. J. P. Dieckert, P. S. O'Connor. D. E. Schacklett et al.,: Air travel and intraocular gas. Ophthalmology, vol. 93, no. 5, pp. 642-645, 1986.
2. Fuller D: Flying and intraocular gas bubble [letter]. Am J Ophthalmol 1981;91:276.
3. Houston S, Graf J, Sharkey J. Commercial air travel after intraocular gas injection. Aviat Space Environ Med. 2012 Aug;83(8):809-10.
4. Kokame GT, Ing MR: Intraocular gas and low altitude flight. Retina 1994, 14:356-358.
5. Mills MD, Devenyi RG, Lam WC, Berger AR, Beijer CD, Lam SR. An assessment of intraocular pressure rise in patients with gas-filled eyes during simulated air flight. Ophthalmology. 2001 Jan;1 08(1):40-4.
6. Muzychuk AK, Adatia FA, Ford BA, Kherani AM. Commercial air travel with a small intravitreal gas bubble. Arch Ophthalmol. 2011 Jun;1 29(6):811-3.
7. Noble J, Kanchanaranya N, Devenyi RG, Lam WC. Evaluating the safety of air travel for patients with scleral buckles and small volumes of intraocular gas. Br J Ophthalmol. 2014 Sep;98(9):1226-9.
8. Hanscom, TA, and Diddle, KR: Mountain travel and intraocular gas bubbles, AM J Ophthalmology 104:546, 1987.
PMA P220030: FDA Summary of Safety and Effectiveness Data
27 of 27
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.