DiamondTemp Ablation System
P200028 · Medtronic, Inc. · OAE · Jan 28, 2021 · Cardiovascular
Device Facts
| Record ID | P200028 |
| Device Name | DiamondTemp Ablation System |
| Applicant | Medtronic, Inc. |
| Product Code | OAE · Cardiovascular |
| Decision Date | Jan 28, 2021 |
| Decision | APPR |
| Device Class | Class 3 |
| Attributes | Therapeutic |
Indications for Use
The DiamondTemp Catheter is indicated for use in cardiac electrophysiological mapping (stimulation and recording) and for treatment of drug refractory, recurrent, symptomatic paroxysmal atrial fibrillation when used in conjunction with the DiamondTemp RF Generator and accessories (DiamondTemp Catheter-to-RF Generator Cable, DiamondTemp GenConnect Cable, DiamondTemp EGM Cable, DiamondTemp Irrigation Pump, DiamondTemp Irrigation Tubing Set) and compatible mapping system.
Device Story
DiamondTemp Ablation System delivers radiofrequency (RF) energy to cardiac tissue for treatment of paroxysmal atrial fibrillation. System includes 7.5Fr open-irrigated ablation catheter (unidirectional/bidirectional), RF generator, irrigation pump, and cables. Catheter tip features diamonds for rapid cooling and thermocouples for temperature sensing. Physician operates catheter under fluoroscopic/mapping guidance to create endocardial lesions. Generator operates in temperature-control mode, automatically adjusting power based on thermocouple feedback to maintain user-selected tip-to-tissue temperature. Irrigation pump delivers saline to cool tip-tissue interface. Output includes real-time ablation parameters (temperature, power, impedance, duration, flow rate) displayed on generator. System enables pulmonary vein isolation; clinical benefit includes reduction in AF recurrence and improved quality of life.
Clinical Evidence
DIAMOND-AF study: prospective, randomized, single-blind, multicenter trial (N=482). Primary safety: freedom from composite SAEs/symptomatic PV stenosis at 6 months (96.7% DiamondTemp vs 93.4% Control; p<0.0001). Primary effectiveness: freedom from AF/AFL/AT recurrence at 12 months (79.1% DiamondTemp vs 75.7% Control; p<0.0001).
Technological Characteristics
7.5Fr quadripolar open-irrigated catheter; platinum/iridium electrodes; CVD diamond tip; Pebax shaft. Energy: RF. Connectivity: proprietary cables to generator/mapping system. Sterilization: EO gas. Software: temperature-control algorithm.
Indications for Use
Indicated for patients >18 years with drug-refractory, recurrent, symptomatic paroxysmal atrial fibrillation. Contraindicated in patients with systemic infection, prosthetic valves, intracardiac thrombus/myxoma, inability to receive anticoagulation, pregnancy, or hemodynamic instability.
Regulatory Classification
Identification
For the treatment of atrial fibrillation.
Predicate Devices
- TactiCath Quartz Contact Force Catheter (P130026)
Reference Devices
- ThermoCool Catheter (PMA-approved)
- Stockert Generator (PMA-approved)
Submission Summary (Full Text)
{0}
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
## I. GENERAL INFORMATION
Device Generic Name: Catheter, percutaneous, cardiac ablation, intended for treatment of atrial fibrillation
Generator, Irrigation pump and accessories
Device Trade Name: DiamondTemp™ Ablation System
DiamondTemp™ Ablation Catheter
(Models CEDT100S, CEDT200L, CEDTB300S, CEDTB400L)
DiamondTemp™ RF Generator
(Model CEDTG200)
DiamondTemp™ Irrigation Pump
(Model CEDTP100)
DiamondTemp™ Irrigation Tubing Set
(Model CEDTTS100)
DiamondTemp™ Catheter-to-RF Generator Cable
(Model CEDTC100)
DiamondTemp™ GenConnect Cable
(Model CEDTGC100)
DiamondTemp™ EGM Cable
(Model CEDTEGM100)
Device Procode: OAE
Applicant's Name and Address: Medtronic Inc.
8200 Coral Sea Street NE
Mounds View, MN 55112 USA
Date(s) of Panel Recommendation: None
Premarket Approval Application
PMA 200028: FDA Summary of Safety and Effectiveness Data
{1}
(PMA) Number: P200028
Date of FDA Notice of Approval: January 28 2021
## II. INDICATIONS FOR USE
The DiamondTemp Catheter is indicated for use in cardiac electrophysiological mapping (stimulation and recording) and for treatment of drug refractory, recurrent, symptomatic paroxysmal atrial fibrillation when used in conjunction with the DiamondTemp RF Generator and accessories (DiamondTemp Catheter-to-RF Generator Cable, DiamondTemp GenConnect Cable, DiamondTemp EGM Cable, DiamondTemp Irrigation Pump, DiamondTemp Irrigation Tubing Set) and compatible mapping system.
## III. CONTRAINDICATIONS
The DiamondTemp Catheter is contraindicated for use in:
- Patients with active systemic infection;
- Patients with prosthetic valves;
- Patients with intracardiac thrombus or myxoma, or interatrial baffle or patch via transseptal approach;
- Patients unable to receive heparin or an acceptable alternative to achieve adequate anticoagulation;
- Pregnant women and children <18 years of age;
- Patients who are hemodynamically unstable
## IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the individual DiamondTemp Ablation Catheter (Unidirectional/Bidirectional), DiamondTemp RF Generator, DiamondTemp Irrigation Pump, DiamondTemp Catheter-to-RF Generator Cable, DiamondTemp GenConnect Cable, DiamondTemp EGM Cable, and DiamondTemp Irrigation Tubing Set labeling.
## V. DEVICE DESCRIPTION
The DiamondTemp Ablation System includes the DiamondTemp Ablation Catheter (Unidirectional/Bidirectional), DiamondTemp RF Generator, DiamondTemp Irrigation Pump, DiamondTemp Catheter-to-RF Generator Cable, DiamondTemp GenConnect Cable, DiamondTemp EGM Cable and DiamondTemp Irrigation Tubing Set. The DiamondTemp Ablation System is designed to deliver radiofrequency (RF) energy to the cardiac anatomy via
PMA 200028: FDA Summary of Safety and Effectiveness Data
{2}
the DiamondTemp Catheter. Figure 1 shows the DiamondTemp System when connected to a compatible mapping system.

Figure 1. DiamondTemp Ablation System
The DiamondTemp ablation catheter is a 7.5Fr ablation tip quadripolar open-irrigated ablation catheter designed to deliver radiofrequency (RF) energy for cardiac ablation. The catheter is available with unidirectional or bidirectional steering and either small curve or large curve reach (Refer to Table 1 for model information).
Table 1. Catheter models and specifications
| Catheter shaft size (outer diameter) | 2.83mm (8.5 Fr) |
| --- | --- |
| Catheter ablation tip size | 2.50 mm (7.5 Fr) |
| Length (nominal) | 110 cm (43.3 in) |
| Model | Description |
| CEDT100S | Unidirectional, small curve (45 mm) |
| CEDT200L | Unidirectional, large curve (63 mm) |
| CEDTB300S | Bidirectional, small curve (45 mm) |
| CEDTB400L | Bidirectional, large curve (63 mm) |
The catheter tip includes diamonds to enable rapid cooling and thermocouples for temperature sensing during RF ablation.
The catheter, when connected to the tubing set and irrigation pump, delivers normal
PMA 200028: FDA Summary of Safety and Effectiveness Data
{3}
saline via a lumen and ports in the catheter tip to provide cooling of the catheter tip and tip-tissue interface. One luer connection at the proximal end of the handle connects to the tubing set, allowing the irrigation pump to generate the flow of normal saline to the catheter.
The DiamondTemp RF generator provides RF energy and temperature monitoring functions, as well as control and communication to the DiamondTemp irrigation pump and commercially available external devices, such as cardiac stimulators, electrophysiology recording systems, and compatible EP navigational and mapping systems.
The generator operates in temperature control mode. The desired catheter tip-to-tissue temperature is selected by the user. Thermocouples in the catheter tip provide temperature feedback and the generator automatically adjusts the power output to maintain the desired tip-to-tissue temperature.
The generator (Figure 2) has a touch-screen display, control buttons, and a control knob for modifying and controlling ablation parameters during the procedure. Ablation parameters such as temperature, power, impedance, duration, and irrigation flow rate are displayed on the front panel and can be recorded and saved by the generator in a format that can be downloaded to a computer or a USB flash drive.
A footswitch is also included with the generator and may be used as an option to start or stop RF energy delivery.

Figure 2. DiamondTemp RF Generator
PMA 200028: FDA Summary of Safety and Effectiveness Data
{4}
The DiamondTemp irrigation pump (Figure 3) delivers saline to the catheter when used in conjunction with the DiamondTemp tubing set. The irrigation pump has a touch screen display and flow control button that controls a two-flow-rate feature for easy selection of the appropriate irrigation flow rate. The rate can be changed between a low flow rate (1-5 mL/min) and a high flow rate (6-30 mL/min). Large numbers on the touch screen display and an LED light on the flow control button indicate the flow rate selected. The irrigation pump communicates with the DiamondTemp generator and may be operated independently or under control of the generator

Figure 3. Epix Therapeutics Irrigation Pump
A transparent pump head door (4, Figure 3) protects the rotating pump head (3, Figure 3), while allowing visibility of the entire tubing set during pump operation.
The tubing set is placed in the path and around the pump head for operation. The irrigation pump uses twin ultrasonic air bubble detectors (5, Figure 3) for added safety in preventing air infusion.
Audible or visual indicators and informational messages displayed on the touch-screen panel (1, Figure 3) warn of air in the tubing, an open pump head door, or other operational conditions.
The DiamondTemp irrigation tubing set consists of the following components (Figure 4). The length of the tubing set assembly is $3.66\mathrm{m}$ (144 in).
PMA 200028: FDA Summary of Safety and Effectiveness Data
{5}
- A drip chamber with an intravenous (IV) spike for connection to an IV bag
- A pump head section with plastic retention clips that fit the slots for the air-bubble detectors (located inside the irrigation pump)
- A catheter end that terminates in a standard luer lock connector and connects to the DiamondTemp ablation catheter
- A 3-way stopcock (not shown)

Figure 4. Tubing Set Components
The DiamondTemp Catheter-to-RF Generator (RFG) cable is used to connect the DiamondTemp catheter to the RF generator. The distal end of the cable has a 19-pin connector that connects to the DiamondTemp catheter. The proximal end of the cable has a 26-pin connector that connects to the RF generator. The length of the cable is 2.5 m (8.2 ft).
The DiamondTemp GenConnect cable connects the DiamondTemp catheter to the DiamondTemp RF generator when a GenConnect device is used. The distal end of the GenConnect cable has a 26-pin female connector that connects to the catheter cable and the proximal end has a 26-pin male connector that connects to the generator. The length of the cable is 1.8 m (6.0 ft).
The DiamondTemp EGM Cable connects the RF generator to a hospital's compatible EP recording system. This Cable is used only with the DiamondTemp ablation system. The DiamondTemp EGM Cable is 3 meters (m) long and has 4 connectors. The one end of the Cable has a male, 9-pin connector that will connect with the DiamondTemp RF Generator and the other end of the Cable has male, 2.0 mm shrouded pin connectors (x4) that will connect with the hospital's compatible EP recording system. The DiamondTemp EGM Cable is provided non-sterile.
## VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several other alternatives for the treatment of drug refractory, recurrent, symptomatic paroxysmal atrial fibrillation, including:
- Commercially available PMA-approved ablation devices
PMA 200028: FDA Summary of Safety and Effectiveness Data
{6}
- Pharmacological therapy for rate and/or rhythm control
- Electrical or pharmacologic cardioversion
- Surgical intervention to create atrial lesions
- Implantable devices to control heart rate.
Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
## VII. MARKETING HISTORY
The DiamondTemp Ablation System has been marketed in select European countries (Belgium, Germany, Spain, Italy, France and United Kingdom) for ablation of cardiac arrhythmias. In Australia, the system received approval in 2019 but has not yet begun distribution.
There have been no product withdrawals from any country for any reason related to safety or effectiveness.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Potential adverse events (e.g., complications) associated with the use of the system during cardiac ablation therapy to treat arrhythmias include the following
- Abnormal vision
- Air embolism
- Anaphylaxis
- Anemia
- Aneurysm
- Angina
- Arrhythmia (including new or worsening of existing condition, or requiring cardioversion)
- Arterial or venous thrombus
- Atrial septal defect
- AV fistula
- Cardiac arrest
- Cardiac tamponade
- Catheter entrapment leading to valve or heart wall damage
- Catheter insertion site hematoma
- Chest pain (non-specific)
- Congestive heart failure exacerbation
PMA 200028: FDA Summary of Safety and Effectiveness Data
7 of 62
{7}
- Component damage to ICD or pacemaker
- Coronary artery dissection
- Death
- Dislodgement of implantable device or permanent pacing lead
- Dizziness
- Embolic events, including infarction of other tissues, coronary, pulmonary, and bowel structures
- Endocarditis
- Esophageal damage or necrosis
- Exacerbation of COPD
- Exacerbation of pre-existing atrial fibrillation
- Fluid overload
- Gastroparesis or GI event
- Hemorrhage
- Hemothorax
- Hypotension
- Hypoxia
- Inadvertent AV block
- Infection
- Myocardial infarction
- Neck, back, or groin pain
- Palpitations
- Perforation (cardiac)
- Pericardial effusion
- Pericarditis
- Peripheral venous thrombosis
- Phrenic nerve damage
- Pleural effusion
- Pneumonia
- Pneumothorax
- Pseudoaneurysm
- Pulmonary edema
- Pulmonary vein stenosis
- Radiation injury resulting in dermatitis, erythema, etc.
- Renal insufficiency or failure
- Respiratory failure
- Seizure
- Sepsis
- Skin burns
- Stroke or cerebrovascular incident
PMA 200028: FDA Summary of Safety and Effectiveness Data
8 of 62
{8}
- Syncope
- Thromboembolic event
- Transient ischemic attack
- Vasovagal reaction
- Ventricular arrhythmia
- Vessel wall or valvular damage or insufficiency
For the specific adverse events that occurred in the clinical study(ies), please see Section X below.
## IX. SUMMARY OF NONCLINICAL STUDIES
Nonclinical testing of DiamondTemp Ablation System included verification and validation (device, system, and software), biocompatibility of patient-contacting materials, sterilization, packaging, and shelf life testing, and animal studies. Performance testing was conducted to demonstrate design integrity. Tests that were identified in standards or guidance documents were performed based on product specification requirements. The following summarized testing was performed on devices representative of proposed commercial devices manufactured by trained operators. "Pass" denotes the devices and systems met established product specification and/or performance criteria, or were in conformance with the requirements of the standards tested to. Test results confirmed that the DiamondTemp Ablation System met product specifications.
## A. Laboratory Studies
The DiamondTemp Ablation Catheter passed design verification (functional) bench testing including dimensional, strength, reliability, mechanical, and electrical integrity. Testing including performance of the DiamondTemp RF Generator used in conjunction with the DiamondTemp Catheter and all other system components, which include the DiamondTemp Irrigation Pump and Tubing Set.
Table 2 below summarizes the bench testing for the Bidirectional and Unidirectional Catheters. This includes reliability, mechanical and electrical integrity and performance test results.
Table 2: Design Verification Testing of DiamondTemp Ablation Catheters (Unidirectional and Bidirectional)
| Test | Acceptance Criteria | Results |
| --- | --- | --- |
| Dimensional: Tip Size, Tip Length, Shaft Size, Ring Size, Effective Catheter Length, Distal Small/Large Curve | All physical dimensions identified in the product specifications must be met. | Pass |
PMA 200028: FDA Summary of Safety and Effectiveness Data
{9}
| Interface – connector cycling | After 10 engagement/disengagement cycles: - the catheter shall have no physical or mechanical failure from the handle - the engagement and disengagement forces of the catheter shall be ≤ 20.0lbf (89N) - when the catheter is straight, tip electrode (D1, D2) resistance shall be 4.4 Ω ± 15%. - when the catheter is straight, ring (R1,R2) electrode resistance shall be 4.4 Ω ± 20% - when the catheter is straight, thermocouple resistance shall be 144 Ω ±10% at room temperature - when the catheter is straight, there shall be no wire-to-wire shorting. The DC resistance > 5MΩ - when the catheter is straight, there shall be no electrical opens | Pass |
| --- | --- | --- |
| Introduction/Withdrawal – Mechanical Integrity | After 20 introduction/withdrawal cycles using a commercially available short sheath 8.5 F (2.67 mm), there shall be no mechanical failures such as bending, bond delamination, electrode movement, or adhesive dislodgment. | Pass |
| Introduction/Withdrawal – Thermal Shock Conditioning | At the end of 5 thermal/shock cycles and then 20 introduction/withdrawal cycles using a commercially available short sheath 8.5 F (2.67 mm), there shall be no mechanical failures such as bending, bond delamination, electrode movement, or adhesive dislodgment. | Pass |
| Introduction/Withdrawal – Electrical Integrity | After 10 introduction/withdrawal cycles using a commercially available short sheath 8.5 F (2.67 mm): -Tip, Ring, Thermocouples meet electrical requirements - no electrical opens | Pass |
| Steering Mechanism Acutuation – Unidirectional or Bidirectional | For unidirectional catheter - the catheter shall deflect in one direction. For bidirectional catheter - the catheter shall deflect in two directions. | Pass |
| Steering Life Cycle | After 100 steering cycles around a 4.0” diameter: - catheter curves shall meet its post-sterile curve template - catheter shall be removed from a commercially available steerable sheath | |
PMA 200028: FDA Summary of Safety and Effectiveness Data
10 of 62
{10}
| | - no bond failures, component malfunctions, or separations from catheter body
- tip electrode (D1, D2) resistance shall be 4.4 Ω ± 15%
- ring (R1,R2) electrode resistance shall be 4.4 Ω ± 20%
- thermocouple resistance shall be 144 Ω ±10% at room temperature
- DC resistance > 5MΩ | |
| --- | --- | --- |
| Twisting Reliability | After two 360° twist rotations:
- the catheter tip shall not separate from the catheter body; the butt bond shall not separate.
- tip electrode (D1, D2) resistance shall be 4.4 Ω ± 15%.
- ring (R1,R2) electrode resistance shall be 4.4 Ω ± 20%
- thermocouple resistance shall be 144 Ω ±10% at room temperature
- DC resistance > 5MΩ
- no electrical opens
After one 90° rotation, the cooling extension tubing shall have no physical or mechanical failure (external components only) | |
| Tensile strength | The Tensile strength of catheter joints shall be less than the defined maximum values per requirements documentation on each joint or juncture. | Pass |
| Curve degradation | The catheter shall meet its post-sterile curve template after simulated use. | Pass |
| Torque response | Torsional transmission along the catheter shaft shall be >1.22ozf-in | Pass |
| Curve range | Catheter shall fully deflect in two directions after use and knob shall operate appropriately. | Pass |
| Distal Bending Stiffness | Bending stiffness of the catheter shaft shall be < 50gf. | Pass |
| Distal Buckling | The catheter shaft shall buckle at < 272gf | Pass |
| Cooling Extension Tubing Integrity | After one 90° rotation the cooling extension tubing shall have no physical or mechanical failures. | Pass |
| Holding Pressure Integrity | The catheter shall hold 45psi for at least 30sec with ΔP ≤ 2 psi when the irrigation | Pass |
PMA 200028: FDA Summary of Safety and Effectiveness Data
11 of 62
{11}
Table 3 below summarizes the verification testing for the DiamondTemp Irrigation and Tubing Set.
Table 3: DiamondTemp Irrigation Pump and Tubing Set Design Verification Testing
| Test | Acceptance Criteria | Results |
| --- | --- | --- |
| Dimensions, connections and labeling | All physical dimensions, connections and labeling identified in the product specifications must be met. | Pass |
| Acoustic output limit | Acoustic level < 83 dBA | Pass |
| Mains power requirements including | Mains current < 5A | Pass |
| Software control of pump flow rates | Interface between microprocessor and motor controller processor | Pass |
| Air bubble detection | Bubble detection at various flow rates | Pass |
PMA 200028: FDA Summary of Safety and Effectiveness Data
{12}
| Flow Performance - Flow rate range, Back Pressure, Air purge flow rate, Flow rate operating duty life cycle and life time | Flow rate range, accuracy and tolerances at 1, 3, 5, 8, 15, 30, and 60mL/min; Specified flow rate verified into a back pressure up to 45psi | Pass |
| --- | --- | --- |
## Software Validation
The DiamondTemp RF Generator firmware (Version 1.30US) and DiamondTemp Irrigation Pump firmware (Version 1.20) have primary responsibility for the operator interface, procedure settings, control of RF power of the DiamondTemp RF Generator, and saline irrigation for the DiamondTemp Irrigation Pump. The DiamondTemp Irrigation Pump firmware also communicates with the DiamondTemp RF Generator to react to the ablation state of the generator. When used with the DiamondTemp Ablation Catheters, the generator operates in temperature-control mode in order to maintain the defined temperature input. The generator computes and displays data related to its operation and the procedure and interfaces with all other peripheral equipment.
Firmware testing included a full suite of safety and performance tests. The firmware was evaluated through unit, integration, verification and validation testing to demonstrate that the performance and safety of the DiamondTemp RF Generator and the DiamondTemp Irrigation Pump conform to specifications.
## Accessories
Additional design verification and validation (bench) testing was completed for the DiamondTemp Ablation System, which includes the DiamondTemp RF Generator, DiamondTemp Irrigation Pump, DiamondTemp Tubing Set, DiamondTemp Catheter-to-RFG Cable, EGM cable, and DiamondTemp GenConnect Cable. Testing includes electrical safety, electromagnetic compatibility (EMC), system design validation testing and system usability testing.
The DiamondTemp Catheter-to-RFG Cable and GenConnect Cable bench testing includes functionality after ten (10) cycles of cleaning for the Catheter-to-RFG Cable and (1) cycle of cleaning for the GenConnect Cable, functionality after ten (10) cycles of autoclave sterilization (Catheter-to-RFG Cable only), life cycle flexibility, connector tensile strength, mechanical connection force, connector reliability, and conformance to resistance, impedance, capacitance, and isolation resistance requirements.
## Biocompatibility
Biocompatibility testing of the DiamondTemp Ablation Catheter (Unidirectional/Bidirectional) and the DiamondTemp Irrigation Tubing Set was conducted in accordance with ISO 10993-1:2009 - Biological evaluation of medical devices – Part 1, and
PMA 200028: FDA Summary of Safety and Effectiveness Data
{13}
FDA/CDRH/ODE Blue Book Memorandum G95-1, Use of International Standard ISO-10993.
The DiamondTemp Ablation Catheter test samples are derived from the finished product. The catheter is classified according to ISO 10993-1 as follows:
- Category: Externally Communicating
- Contact Duration: < 24 hours (Limited)
- Device Body Contact: Circulating Blood Path
A summary of the results is provided in Table 4 below and demonstrates that the DiamondTemp Ablation Catheters are biocompatible per ISO 10993-1. These test results provide objective evidence that the catheter is biocompatible per its intended use.
Table 4. Biocompatibility Testing Summary – DiamondTemp Ablation Catheter (Unidirectional/Bidirectional)
| Test Performed / Applicable ISO 10993 Part No. | Test Performed | Results |
| --- | --- | --- |
| Cytotoxicity (10993-5) | MEM Elution, GLP | Pass |
| Sensitization (10993-10) | ISO Guinea Pig Maximization Sensitization Test, GLP | Pass |
| Irritation or Intracutaneous Reactivity (10993-10) | ISO Intracutaneous Irritation Test, GLP | Pass |
| Acute Systemic Toxicity (10993-11) | ISO Acute Systemic Injection Test, GLP | Pass |
| Material-Mediated Pyrogenicity (10993-11) | ISO Materials Mediated Rabbit Pyrogenicity, GLP | Pass |
| Hemocompatibility (10993-4) | ASTM Hemolysis, Direct Contact, GLP | Pass |
| Hemocompatibility (10993-4) | ASTM Hemolysis, Extract Method, GLP | Pass |
| Hemocompatibility (10993-4) | Complement Activation, GLP | Pass |
| Hemocompatibility (10993-4) | Thrombogenicity, GLP | Pass |
| Hemocompatibility (10993-4) | Partial Thromboplastin (PTT) Test, GLP | Pass |
| Latex | ASTM D6499 Inhibition ELISA | Pass |
PMA 200028: FDA Summary of Safety and Effectiveness Data
14 of 62
{14}
Patient contacting materials of the DiamondTemp Ablation Catheters are listed in Table 5 below.
Table 5: DiamondTemp Ablation Catheter (Unidirectional/Bidirectional) Blood/Fluid Contact Components and Materials
| Component | Material Name |
| --- | --- |
| Tip, Proximal Electrode and Ring Electrode | Platinum/Iridium |
| Tip, Distal/Proximal Diamond | Chemical Vapor Deposited (CVD) Diamond |
| Shaft | Pebax 7233, 5533, 4033, 3533 |
| Irrigation Bump Tube | Pebax 7233 |
| Luer | Polycarbonate |
| Irrigation Tube (Thin Tip) | 304 Stainless Steel |
| Adhesives | Cyanoacrylate, UV Cyanoacrylate, Epoxy, Accelerator |
| Solder/flux | Tin/Silver/Copper |
| Distal Thermocouple | Polyimide, Loctitie |
The DiamondTemp Irrigation Tubing Set test samples were derived from the finished product. The tubing set is classified according to ISO 10993-1 as follows:
- Category: Externally Communicating
- Contact Duration: < 24 hours (Limited)
- Device Body Contact: Blood Path Indirect
A summary of the results is provided in Table 6 below and demonstrates that the DiamondTemp Irrigation Tubing Set is biocompatible per ISO 10993-1. These test results provide objective evidence that the catheter is biocompatible per its intended use.
Table 6. Biocompatibility Testing Summary – DiamondTemp Tubing Set
| Test Performed / Applicable ISO 10993 Part No. | Test Performed | Results |
| --- | --- | --- |
| Cytotoxicity (10993-5) | MEM Elution, GLP | Pass |
| Sensitization (10993-10) | ISO Guinea Pig Maximization Sensitization Test, GLP | Pass |
PMA 200028: FDA Summary of Safety and Effectiveness Data
{15}
PMA 200028: FDA Summary of Safety and Effectiveness Data
16 of 62
| Irritation or Intracutaneous Reactivity (10993-10) | ISO Intracutaneous Irritation Test, GLP | Pass |
| --- | --- | --- |
| Acute Systemic Toxicity (10993-11) | ISO Acute Systemic Injection Test, GLP | Pass |
| Material-Mediated Pyrogenicity (10993-11) | ISO Materials Mediated Rabbit Pyrogenicity, GLP | Pass |
| Hemocompatibility (10993-4) | ASTM Hemolysis, Direct Contact, GLP | Pass |
| Hemocompatibility (10993-4) | ASTM Hemolysis, Extract Method, GLP | Pass |
Indirect patient contacting materials for the DiamondTemp Irrigation Tubing Set are listed in Table 7 below.
Table 7: DiamondTemp Irrigation Tubing Set Indirect Blood/Fluid Contact Components and Materials
| Component | Material Name |
| --- | --- |
| Spike, Vented | ABS |
| Spike Cap, Vented | Polyethylene |
| Tubing | Tygon, DEHP-free PVC |
| Drip Chamber | DEHP-free PVC |
| Adhesives | Cyanoacrylate, UV Loctite |
| StopCock | Polycarbonate |
| Luer | PVC |
## B. Animal Studies
A total of five (5) in vivo animal studies for the DiamondTemp Ablation System were conducted to support paroxysmal atrial fibrillation with the DiamondTemp Ablation Catheter. The purpose of Animal Studies is described as follows.
- Before undertaking a First-in-Man (FIM) clinical evaluation a confirmatory chronic animal study was conducted to characterize the safety profile and performance of the DiamondTemp Ablation System in an animal model and confirm readiness for human use.
- Two (2) confirmatory, comparative GLP animal evaluations were conducted to assess the safety and effectiveness of the DiamondTemp Ablation System prior to initiation of the DIAMOND-AF IDE clinical study. The first study compared the DiamondTemp Ablation System in sub-acute and chronic animal evaluations in a canine model to a commercial control device (ThermoCool Catheter) using two operators. The second study
{16}
characterized the DiamondTemp Ablation System performance as well as a commercial control device (ThermoCool Catheter/Stockert Generator) in a perfused swine tissue model ("thigh model") scientifically recognized for characterizing ablation systems.
- A confirmatory, comparative chronic GLP evaluation using three (3) clinical physician operators was undertaken to evaluate safety and effectiveness and to compare handling and performance of the DiamondTemp Ablation System to a commercial control device (TactiCath Quartz Contact Force Catheter).
- As part of design validation for the CEDTG200 Generator, a chronic study in the canine atria was conducted using 2 clinical physician operators was conducted to characterize the safety of the DiamondTemp Ablation System in the creation of atrial endocardial lesions.
A summary of the animal study is provided in Table 8.
Table 8: DiamondTemp Ablation System In Vivo Animal Summary
| Study Type | Number of Animals | Follow-up Duration | Relevant Findings |
| --- | --- | --- | --- |
| Non-GLP Chronic Canine Study | N=4 | 14 days | A Chronic Study to Assess the Safety and Performance of the DiamondTemp System in the Creation of Endocardial RF Ablation Lesions in the Canine Atria The confirmatory chronic study characterized the safety profile and device performance of the DiamondTemp Ablation System in the creation of endocardial radiofrequency (RF) ablation lesions in the canine atria. The study also demonstrated the compatibility of the DiamondTemp Ablation System with ancillary devices and equipment. No significant physical, neurological, or pathologic abnormalities were observed. The gross anatomical and histo-pathological findings were representative of RF ablation procedures and the analysis of the lesions showed that most lesions created were transmural. No significant safety concerns were raised in the histopathology assessment. |
| GLP Chronic Canine Study | N=25 | 7 days (N=12) 30 days (N=13) | A Sub-Acute and Chronic GLP Study to Evaluate the Safety of the DiamondTemp Ablation System in the Creation of Endocardial RF Ablation Lesions in the Canine Atria |
PMA 200028: FDA Summary of Safety and Effectiveness Data
17 of 62
{17}
| | | | The DiamondTemp Ablation System was compared to the commercially available the ThermoCool/Stockert System approved for the treatment of atrial fibrillation.
There were no significant differences regarding adverse events and safety outcomes when the DiamondTemp was compared to the control device. The findings of the health report and clinical pathology and histopathology reports were comparable for both study groups. No incidences of microemboli in up/downstream organs and draining myocardium were reported, no clinically significant collateral injury or lesions were reported, no pulmonary vein stenosis were reported in either study group at any of the termination time points and no cardiac tamponade or perforation were reported. None of the findings were unanticipated.
The effectiveness of the DiamondTemp Ablation System was also established with the demonstration of electrical conduction block in 2 pulmonary veins at least 20 minutes after final ablation treatment in the vessel as determined by EGM in all treated animals. |
| --- | --- | --- | --- |
| Acute Swine GLP Study | N=4 | Acute
(same day) | A GLP Study to Compare the Ablation Lesions Created using the DiamondTemp Catheter/Ablation System to the Biosense Webster ThermoCool Catheter/Stockert System in the Swine Thigh Prep Model
There were no significant differences in lesion diameter and volume between the DiamondTemp test article for any lesions created under WC operating conditions. The lesion diameter and volumes achieved under IFU operating conditions resulted in smaller lesions for the DiamondTemp Ablation system when compared to the control.
There was no significant difference in the incidence rate of steam pops between the DiamondTemp test article and the control test articles in any combination of orientation and ablation settings except the Parallel WC condition. In that condition, the DiamondTemp test article yielded significantly fewer steam pops than the control (4/15, 27% vs. 11/15, 73%; p=0.027).
The presence of char on the catheter was not observed after any ablation for either the DiamondTemp or control test articles. |
PMA 200028: FDA Summary of Safety and Effectiveness Data
18 of 62
{18}
| | | | The presence of thrombus on the catheter or tissue represented a critical endpoint relative to the ability to characterize the safety profile of the DiamondTemp due to the ability of thrombotic material to cause thromboembolic events in critical organs including the ventricular myocardium, the kidneys, the brain or the lungs. For thrombus on the device and tissue, the DiamondTemp was equivalent to the control device at every test condition. |
| --- | --- | --- | --- |
| Chronic Canine GLP Study | N=12 | 30 days | A Chronic GLP Study to Evaluate the Safety of the DiamondTemp System in the Creation of Endocardial RF Ablation Lesions in the Canine Atria.
Safety of the DiamondTemp Ablation System (Test Article) was compared to the Abbott/St. Jude TactiCath Quartz Contact Force Control Catheter (Control Article).
There was no significant difference in handling and usability scores of the DiamondTemp to the control. There was no occurrence of catheter entrapment, valvular injuries, phrenic nerve injury or pulmonary vein stenosis in the test or control animals. All animals in both the test and control groups showed presence of electrical conduction block in at least two pulmonary veins. DiamondTemp resulted in a significantly shorter total ablation time, due to shorter individual RF delivery times. DiamondTemp also resulted in less fluid delivered during the procedures.
The safety and effectiveness criteria of the study were met. The DiamondTemp Ablation System demonstrated safety and effectiveness when compared to a commercial control device. The performance and usability of the DiamondTemp Ablation System was comparable or better than the control device. |
| Non-GLP Chronic Canine Study | N=6 | 30 days | A chronic study to characterize the safety of the DiamondTemp Ablation System in the creation of atrial endocardial ablation lesions in the canine model. Multiple focal ablations or drag lesions were applied in three animals each.
The findings of the health report and clinical pathology and histopathology reports were acceptable. No incidences of microemboli in up/downstream organs |
PMA 200028: FDA Summary of Safety and Effectiveness Data
19 of 62
{19}
PMA 200028: FDA Summary of Safety and Effectiveness Data
20 of 62
| | | and draining myocardium were reported, no clinically significant collateral injury or lesions were reported, no pulmonary vein stenosis were reported in either study group at any of the termination time points and no cardiac tamponade or perforation were reported.
The effectiveness of the DiamondTemp Ablation System was established with the demonstration of electrical conduction block in 2 pulmonary veins at least 20 minutes after final ablation treatment in the vessel as determined by EGM in all treated animals. |
| --- | --- | --- |
## C. Additional Studies
Sterilization, Packaging, and Shelf Life
The DiamondTemp Ablation Catheters (Unidirectional/Bidirectional), DiamondTemp Irrigation Tubing Set, and DiamondTemp Catheter-to-RFG Cable are supplied sterile, single use and ready to use. The catheters are sterilized via 20% ethylene oxide (EO) and 80% carbon dioxide gas, and the tubing set and catheter-to-RFG cable are sterilized via 100% ethylene oxide (EO) at qualified sterilization facilities using a validated sterilization cycle. The sterilization process validation was conducted to provide a sterility assurance level (SAL) of at least 10⁻⁶ in accordance with ISO 11135-1:2007, a recognized standard for EO terminally sterilized medical devices.
The DiamondTemp Catheter-to-RFG Cable is provided sterile, not patient-contacting, and can be re-sterilized up to ten (10) times after initial use. The GenConnect cable is not patient-contacting, can be reused multiple times and is not sterilized.
The packaged DiamondTemp Unidirectional and Bidirectional Ablation Catheters, DiamondTemp Catheter-to-RF Generator Cable, DiamondTemp GenConnect Cable, EGM Cable, and DiamondTemp Irrigation Tubing Set were evaluated to demonstrate product and packaging system performance after exposure to applicable conditioning. Device functionality after the following conditioning was assessed:
- Storage / climatic conditioning per ASTM D 4332 for exposure to extreme cold, tropical, and desert temperature and humidity
- Distribution conditioning per ASTM D4169 cycle 13 (variations of vibration, shock, and compression).
- Challenge sterilization of 2x Ethylene Oxide
Packaging System performance testing included maintenance of sterile barrier integrity from gross leaks per ASTM F2096, pouch seal strength tested per ASTM F288, and legibility of labeling and Instructions for Use for the subject devices.
The DiamondTemp Ablation Catheters, DiamondTemp Catheter-to-RFG Cable, and DiamondTemp Irrigation Tubing Set are labeled for 1-year shelf life.
{20}
PMA 200028: FDA Summary of Safety and Effectiveness Data
21 of 62
# X. SUMMARY OF PRIMARY CLINICAL STUDY
The applicant performed the DIAMOND-AF Study to establish a reasonable assurance of safety and effectiveness of radiofrequency ablation with the DiamondTemp Ablation System for the treatment of drug refractory, recurrent, symptomatic paroxysmal atrial fibrillation in the US under IDE # G170227. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is presented below.
## A. Study Design
Patients were treated between November 6, 2017 and October 26, 2018. The database for this original PMA reflected data collected through December 3, 2019 and included 482 patients. There were 23 investigational sites (14 US sites, 1 Canadian site, and 8 France/Italy/Czech Republic sites).
The study was a prospective, single-blind, 1:1 randomized, controlled, multicenter pivotal clinical trial. The study enrolled subjects with symptomatic paroxysmal atrial fibrillation refractory to one or more antiarrhythmic drugs (Class I-IV). Enrolled subjects were randomized to catheter ablation using the investigational DiamondTemp Ablation System or the control TactiCath Contact Force Ablation Catheter. The control device received FDA approval for the treatment of paroxysmal atrial fibrillation (P130026). Subjects were followed up prior to hospital discharge and at 7 days, 1 month, 3 months, 6 months, and 12 months post-ablation. Subjects were provided a cardiac event monitor at the hospital pre-discharge visit to be used throughout the duration of the study. This data was transmitted to and read at an ECG core lab.
The study success was originally defined by freedom from documented symptomatic atrial fibrillation (AF), atrial flutter (AFL), and atrial tachycardia (AT) recurrence following the blanking period through the end of the effectiveness evaluation period (from 3-month to 12-month follow-up post-ablation procedure) and freedom from Primary Safety Device- or Procedure-related serious adverse events (SAE) composite occurring within 30 days of the AF ablation procedure (or clinically symptomatic pulmonary vein stenosis through 6 months post-index ablation procedure). Prior to data lock, the sponsor and FDA agreed to change the primary effectiveness endpoint to include all-cause atrial arrhythmia recurrence. The study would be considered successful if the investigational device (DiamondTemp Ablation System) is considered non-inferior to the control device for the primary safety and effectiveness endpoints.
An independent ECG Core Lab adjudicated rhythm data for the primary effectiveness endpoint, and an independent Clinical Events Committee (CEC) adjudicated primary
{21}
safety endpoint events. Additionally, an independent Data Safety Monitoring Board (DSMB) reviewed all safety data throughout the course of the study.
1. Clinical Inclusion and Exclusion Criteria
Enrollment in the DIAMOND-AF Study was limited to patients who met the following inclusion criteria:
1) Above eighteen (18) years of age or of legal age to give informed consent specific to state and national law.
2) Subjects with a history of symptomatic, PAF who have had ≥ 2 episodes of PAF reported within the 6 months prior to index ablation procedure with a physician note indicating recurrent, self-terminating AF.
3) At least one episode of PAF documented by electrocardiographic data within 12 months prior to index ablation procedure.
4) Refractory to at least one Class I-IV anti-arrhythmic AAD for treatment of PAF.
5) Suitable candidate for intra-cardiac mapping and ablation of arrhythmia.
6) Subject agrees to comply with study procedures and be available (geographically stable) for follow-up visits for at least 12 months after enrollment.
7) Subject is willing and able to provide written consent.
Patients were not permitted to enroll in the DIAMOND-AF Study if they met any of the following exclusion criteria:
At time of enrollment and/or prior to procedure:
1) AF secondary to electrolyte imbalance, thyroid disease or reversible or noncardiac cause.
2) LA diameter > 5.5 cm.
3) Left ventricular ejection fraction (LVEF) < 35%.
4) Currently New York Heart Association (NYHA) Class III or IV or exhibits uncontrolled heart failure.
5) Body Mass Index (BMI) > 40 kg/m².
6) LA ablation, septal closure device or mitral valve surgical procedure at any time prior to enrollment.
7) Presence of intramural thrombus, tumor or abnormality that precludes vascular access, catheter introduction or manipulation.
8) Coagulopathy, bleeding diathesis or suspected procoagulant state.
PMA 200028: FDA Summary of Safety and Effectiveness Data
22 of 62
{22}
9) Sepsis, active systemic infection or fever (>100.5°F / 38°C) within a week prior to the ablation procedure.
10) Significant restrictive or obstructive pulmonary disease or chronic respiratory condition.
11) Renal failure requiring dialysis or renal compromise that in the investigator’s judgement would increase risk to the subject or deem the subject inappropriate to participate in the study.
12) Known allergies or intolerance to anticoagulant and antiplatelet therapies to be used in conjunction with the study or contrast sensitivity that cannot be adequately pre-treated prior to the ablation procedure.
13) Positive pregnancy test results for female subjects of childbearing potential or breast feeding.
14) Enrollment in a concurrent clinical study that in the judgement of the investigator would impact study outcomes.
15) Acute or chronic medical condition that in the judgment of the investigator would increase risk to the subject or deem the subject inappropriate to participate in the study.
16) Life expectancy < 12 months based on medical history or the medical judgement of the investigator.
Within 1 month of enrollment or just prior to procedure:
17) Documented LA thrombus upon imaging.
18) Creatinine >2.5mg/dl or creatinine clearance <30mL/min.
Within 2 months of enrollment:
19) Regularly (uninterrupted) prescribed amiodarone.
Within 3 months of enrollment:
20) Significant gastrointestinal (GI) bleed.
21) Myocardial infarction (MI), unstable angina, cardiac surgery or coronary intervention.
PMA 200028: FDA Summary of Safety and Effectiveness Data
23 of 62
{23}
Within 6 months of enrollment:
22) Coronary artery bypass graft (CABG) procedure.
23) Implantable Cardioverter Defibrillation (ICD), Cardiac Resynchronization Therapy (CRT) leads or pacemaker implant procedure.
24) Documented stroke, Cerebrovascular Accident (CVA), Transient Ischemic Attack (TIA) or suspected neurological event.
Within 12 months of enrollment:
25) An episode of AF lasting >7 days in duration.
2. Follow-up Schedule
All patients were scheduled to return for follow-up examinations at baseline, operative/discharge, 7 days, 1, 3, 6, and 12 months following the index procedure. All patients were followed per protocol in relation to the date of the index ablation procedure. Clinical data were required to be collected at all subject visit intervals. Adverse events and complications were recorded at all visits. Post-ablation rhythm monitoring included symptomatic and twice monthly symptomatic/asymptomatic event monitor transmissions during the evaluation period, ECG at 3, 6, and 12 months, and 24-hour Holter monitor at 6 and 12 months.
The key timepoints are shown below in Table 9 summarizing schedule of treatments and evaluations.
PMA 200028: FDA Summary of Safety and Effectiveness Data
24 of 62
{24}
Table 9. Schedule of Treatments and Evaluations
| Assessments/Activities | Baseline Evaluation | Ablation Procedure | Blanking Period (Ablation Procedure – 3 Months Post Procedure) | | | | Effectiveness Evaluation Period (3 Months -12 Months Post Procedure) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Pre-Discharge | 7-Day (± 3 days) | 1-M (± 14 days) | Repeat Ablation | 3-M (± 21 days) | 6-M (± 28 days) | 12-M (± 45 days) |
| Eligibility Screening | X | | | | | | | | |
| Informed Consent | X | | | | | | | | |
| Subject Demographics | X | | | | | | | | |
| Medical History | X | | | | | | | | |
| Physical Exam | X | | X | | X | X^{A} | X | X | X |
| 12-lead ECG | | X | X | | X | X^{A} | X | X | X |
| TEE or TTE | X^{B} | | | | | | | | |
| TEE or ICE^{G} | | X^{C} | | | | X^{C} | | | |
| Chest CT / MRI | | | | | | | X^{F} | X^{F} | X^{F} |
| NIH Stroke Scale | X | | X | | | X^{A} | | | X |
| Procedural Data | | X | | | | X^{A} | | | |
| Event Monitor Recording | | | X^{D} | X^{D} | X^{D} | | X^{E} | X^{E} | X^{E} |
| 24 hr. Holter Monitor | | | | | | | | X | X |
| Cardiac Medication Changes | X | X | X | X | X | X^{A} | X | X | X |
| Protocol Deviations | X | X | X | X | X | X^{A} | X | X | X |
| Adverse Events | X | X | X | X | X | X^{A} | X | X | X |
| AF Quality of Life Survey (AFEQT) | X | | | | | X^{A} | | X | X |
A Only required if a repeat ablation procedure performed. 1 repeat ablation procedure was allowed during Blanking Period.
B TEE or TTE only required if subject does not have imaging data to determine LA diameter, LVEF for eligibility within 180d of ablation procedure
C TEE or ICE required pre-procedure to rule out LA thrombus if any of the following were met: CHA2DS2-VASc score was $\geq 2$, if LA diameter $\geq 4.6$ or if pre-procedure anticoagulation requirements were not met
D Event monitor recording only required if subject was experiencing symptoms during blanking period
E Following the blanking period (3-12 month following ablation), subjects were required to take two 1-minute event monitor recordings per month regardless of whether they were experiencing symptoms.
F Only required if subject presents with symptoms associated with PV stenosis
## 3. Clinical Endpoints
### Primary Endpoints
The primary safety endpoint was defined as freedom from composite of serious adverse events (SAE) occurring within 30-days and clinically symptomatic pulmonary vein stenosis through 6-months post-index ablation procedure, as adjudicated by an independent Clinical Events Committee (CEC) for relatedness to the procedure or device.
The primary safety device- or procedure-related SAE composite was a combined rate of the following events:
PMA 200028: FDA Summary of Safety and Effectiveness Data
{25}
- Atrioesophageal fistula
- Bleeding complication
- Cardiac tamponade / perforation
- Death
- Extended hospitalization*
- Myocardial infarction
- Pericarditis
- Phrenic nerve paralysis
- Pulmonary edema
- Pulmonary vein stenosis
- Stroke post-ablation
- Thromboembolism
- Transient ischemic attack (TIA) post-ablation
- Vagal nerve injury
- Vascular access complications
* Extended hospitalization is defined as extended hospital stay or re-hospitalization that is related to the procedure or device.
The primary effectiveness endpoint was defined as freedom from documented AF, AFL* and AT episodes following the blanking period (3-month follow-up post-ablation procedure) through the end of the effectiveness evaluation period (12-month follow-up post-ablation procedure).
An effectiveness failure was defined by any of the following events:
- Inability to electrically isolate all accessible targeted pulmonary veins during the ablation procedure\*\*
- Documented episodes of AF, AFL or AT lasting ≥30 seconds in duration as evidenced by electrocardiographic data during the effectiveness evaluation period
- DC cardioversion for AF, AFL or AT during the effectiveness evaluation period
- A repeat ablation procedure to treat AF, AFL or AT during the effectiveness evaluation period
- Use of a new or modification to existing Class I-IV AAD regimen to treat AF, AFL or AT recurrence during the effectiveness evaluation period
- Use of a non-study device for ablation of any AF targets during the index or repeat ablation procedure during the blanking period
- More than one (1) repeat ablation procedure during the blanking period
PMA 200028: FDA Summary of Safety and Effectiveness Data
26 of 62
{26}
* Occurrence and/or ablation of cavotricuspid isthmus (CTI)-dependent AFL, as confirmed by entrainment maneuvers during EP testing at any time during this study was not a primary effectiveness failure because it was not considered an iatrogenic arrhythmia following a left atrial ablation procedure for AF.
**Electrical isolation as confirmed by demonstration of exit and/or entrance conduction block.
The Primary Effectiveness and Primary Safety Endpoint analyses were performed on the Intention-to-Treat (ITT) Cohort to demonstrate non-inferiority of the investigational group (INV) to the control group (CTRL). The ITT cohort was comprised of all randomized subjects regardless of whether they received study treatment, with the following analyses conducted according to the randomized treatment assignment.
## Primary Effectiveness Analysis:
The following primary effectiveness hypothesis was evaluated using the exact test for a binomial proportion at a one-sided significance level of 2.5%.
- H₀: πINV ≤ πCTRL - δ
- H₁: πINV > πCTRL - δ
where π was the population proportion for the corresponding treatment group and δ was the non-inferiority margin of 12.5%.
## Primary Safety Analysis
The following primary safety hypothesis was evaluated using the exact test for a binomial proportion at a one-sided significance level of 2.5%.
- H₀: πINV ≤ πCTRL - δ
- H₁: πINV > πCTRL - δ
where π was the population proportion for the corresponding treatment group and δ was the non-inferiority margin of 6.5%.
## Secondary Endpoints
Secondary endpoints to characterize the performance of the DiamondTemp Ablation System, relative to the control device, include:
1. Mean duration of individual RF ablations (seconds).
2. Mean cumulative RF time per procedure (minutes).
3. Freedom from a composite of SAE occurring within 7-days post-index ablation procedure as adjudicated by an independent CEC for relatedness to the procedure or device.
PMA 200028: FDA Summary of Safety and Effectiveness Data
27 of 62
{27}
4. Freedom from documented AF, AT and AFL episodes following the blanking period through 12-month follow-up post-ablation procedure in the absence of class I and III anti-arrhythmic drug therapy.
5. Rate of acute procedural success, defined as confirmation of electrical isolation of PVs via assessment of entrance block at least 20 minutes following the last ablation around the respective PV.
6. Rate of single procedure success defined as the rate of subjects treated with one single ablation procedure during study participation and with freedom from documented AF, AT and AFL at 12 months.
7. Rate of single procedure success defined as the rate of subjects treated with one single ablation procedure during study participation and with freedom from ALL primary effectiveness endpoint failure criteria.
8. Rate of occurrence of electrically reconnected PVs following a 20-minute waiting period assessed by entrance block at index procedure.
9. Accumulated changes in Quality of Life (QOL) using the AF QOL Survey (AFEQT Questionnaire) from baseline through 6 and 12 months following ablation procedure.
10. Neurological changes measured using the NIH stroke scale between baseline and post-ablation (pre-discharge visit) and at 12 months post-ablation procedure.
11. Total procedure time (minutes), defined as time of first assigned ablation catheter insertion into the vasculature to time of last procedural ablation catheter removed.
12. Time to achieve initial PVI at index procedure (minutes), defined as time of delivery of first RF ablation with the assigned ablation catheter until confirmation of PVI.
13. Total treatment device time (minutes), defined as time of delivery of first RF ablation with the assigned ablation treatment catheter to removal of the treatment catheter.
14. Total number of RF ablations per procedure.
15. Total fluid infused through the assigned ablation catheter (mL).
16. Total fluoroscopy time (minutes).
17. Number of re-hospitalizations due to atrial fibrillation recurrence after blanking period.
Of the seventeen (17) pre-defined secondary endpoints above, four (4) specific secondary endpoints were evaluated for superiority over Control. The analytical
PMA 200028: FDA Summary of Safety and Effectiveness Data
28 of 62
{28}
approach was one of a priori hierarchical hypotheses, with the pre-specified order of the endpoints as follows:
1. Mean duration of individual RF ablations (seconds).
2. Mean cumulative RF time per procedure (minutes).
3. Total fluoroscopy time (minutes).
4. Total procedure time (minutes), defined as time of first assigned ablation catheter insertion into the vasculature to time of last procedural ablation catheter removed.
The first secondary endpoint (i.e., RF ablation time) needs to be significant at the two-tailed $\alpha$ -level of 0.05, before the next one can be considered at the same threshold. Once an endpoint was determined to be non-significant, testing was to stop and any remaining hypotheses were not tested.
# B. Accountability of PMA Cohort
At the time of database lock, of 485 patients consented in the PMA study, 455 (93.8%) patients were available for analysis at the completion of the study, the 12-month postoperative visit. A total of 482 subjects were enrolled and randomized (DiamondTemp: 239, Control: 243) in the DIAMOND-AF Study at 23 US and OUS centers. Of these, 14 DiamondTemp and 13 Control subjects exited the study before completing the study. The last subject exited the study on 12/3/2019. Table 10 shows subject disposition. Table 11 shows an accounting of follow-up visit attendance during the study. Subject accountability is further summarized in Figure 5 below. The database was locked on 3/27/2020.
Table 10. Subject Disposition
| Subject Disposition | Number (N) |
| --- | --- |
| Number of Subjects with Signed Consent | 485 |
| Number of Subjects Not Randomized | 3 |
| Documented Stroke, CVA, TIA or Suspected Neurological Event [1] | 1 |
| Enrollment Cap Met [2] | 1 |
| Regularly Prescribed Amiodarone [3] | 1 |
| Number of Subjects Enrolled/Randomized (Intention-to-Treat Analysis Cohort) | 482 |
[1] Subject 15-021 met exclusion criterion #24
[2] Subject 15-010 met all criteria but was never randomized prior to study exit (30-Oct-2018), with a reason of "Subject's ablation was not scheduled and patient was not randomized prior to enrollment number being met"
[3] Subject 17-001 met exclusion criterion #19
PMA 200028: FDA Summary of Safety and Effectiveness Data
{29}
Table 11. Scheduled Visit Compliance
| Visit | Control (N=243) | DiamondTemp (N=239) | All Subjects (N=482) |
| --- | --- | --- | --- |
| Enrolled/Randomized | 243 (100%) | 239 (100%) | 482 (100%) |
| Ablation Procedure | 241 (99.2%) | 235 (98.3%) | 476 (98.8%) |
| Pre-Discharge Visit | 241 (99.2%) | 235 (98.3%) | 476 (98.8%) |
| 7 Day Visit | 238 (97.9%) | 234 (97.9%) | 472 (97.9%) |
| 1 Month Visit | 236 (97.1%) | 227 (95.0%) | 463 (96.1%) |
| 3 Month Visit | 230 (94.7%) | 226 (94.6%) | 456 (94.6%) |
| 6 Month Visit | 223 (91.8%) | 222 (92.9%) | 445 (92.3%) |
| 12 Month Visit | 230 (94.7%) | 225 (94.1%) | 455 (94.4%) |
| Study Completion | | | |
| Completion of Study as Planned | 230 (94.7%) | 225 (94.1%) | 455 (94.4%) |
| Discontinued Prematurely | 13 (5.3%) | 14 (5.9%) | 27 (5.6%) |
PMA 200028: FDA Summary of Safety and Effectiveness Data
30 of 62
{30}
Figure 5. Subject Accountability FlowChart

[a] Subject withdrew consent (06-013); Subject progressed to persistent AF (11-001); Enrollment closure (09-003, 13-015).
[b] Enrollment closure (10-004); Physician no longer believed the subject was a good candidate for the study (11-004).
[c] Treatment attempted but not delivered with TactiCath due to technical difficulties (09-001); Treatment attempted but not delivered with TactiCath due to procedure failure (22-005).
Follow-up visit compliance was $91.8\%$ or higher for all follow-up visits, with $94.7\%$ of control subjects and $94.1\%$ of DiamondTemp subjects completing the study as planned through the 12 month follow-up visit.
PMA 200028: FDA Summary of Safety and Effectiveness Data
{31}
The protocol specified analysis populations include:
- Intention-to-Treat (ITT): The 482 randomized subjects comprise the ITTpopulation.
- Safety Analysis: The 476 subjects who had a study ablation catheter inserted comprise the Safety Analysis cohort.
- Per Protocol (PP): The PP population consisted of 451 subjects who met all eligibility criteria, had no major protocol deviations, and were treated in accordance with the randomization treatment assignment.
# C. Study Population Demographics and Baseline Parameters
The demographics of the study population are typical for a paroxysmal atrial fibrillation catheter ablation study performed in the US. Tables 12, 13 and 15 summarize the demographics, baseline characteristics, and medical history for subjects by treatment group in the ITT cohort. Demographic data and baseline characteristics were balanced with no significant differences between the treatment groups except for history of non-AF/AFL arrhythmias or conduction disturbance.
Table 12 Demographic Characteristics, Intention-to-Treat Cohort, Control vs DiamondTemp
| Demographics | Control (N=243) | Diamond Temp (N=239) |
| --- | --- | --- |
| Age, years | | |
| Mean (SEM / SD) | 63.0 (0.67 / 10.42) | 62.3 (0.72 / 11.13) |
| Median | 64.0 | 65.0 |
| Min, Max | 27.0, 84.0 | 22.0, 82.0 |
| N (N Missing) | 243 (0) | 239 (0) |
| Sex, n (%) | | |
| Male | 143 (58.8%) | 136 (56.9%) |
| Female | 100 (41.2%) | 103 (43.1%) |
| Race, n (%) | | |
| American Indian or Alaska Native | 2 (0.8%) | 3 (1.3%) |
| Asian | 2 (0.8%) | 0 (0%) |
| Black or African American | 4 (1.6%) | 4 (1.7%) |
| Other, specify: Caribbean | 0 (0%) | 1 (0.4%) |
| Other, specify: Ecuadorian | 0 (0%) | 1 (0.4%) |
| Prefer Not to Say | 68 (28.0%) | 66 (27.6%) |
| Unknown | 6 (2.5%) | 5 (2.1%) |
PMA 200028: FDA Summary of Safety and Effectiveness Data
{32}
Table 13. Baseline Characteristics, Intention-to-Treat Cohort, Control vs DiamondTemp
| Baseline Characteristics | Control (N=243) | DiamondTemp (N=239) | p-value |
| --- | --- | --- | --- |
| Height, cm | | | 0.7480 |
| Mean (SEM / SD) | 172.7 (0.63 / 9.89) | 172.9 (0.63 / 9.69) | |
| Median | 172.7 | 172.7 | |
| Min, Max | 152.0, 196.0 | 147.3, 205.7 | |
| N (N Missing) | 243 (0) | 239 (0) | |
| Weight, kg | | | 0.2821 |
| Mean (SEM / SD) | 85.4 (1.03 / 16.01) | 84.1 (1.17 / 18.06) | |
| Median | 85.0 | 83.0 | |
| Min, Max | 51.0, 146.0 | 45.8, 131.7 | |
| N (N Missing) | 243 (0) | 239 (0) | |
| BMI, kg/m2 | | | 0.2422 |
| Mean (SEM / SD) | 28.6 (0.29 / 4.48) | 28.0 (0.32 / 5.00) | |
| Median | 28.1 | 27.5 | |
| Min, Max | 19.8, 42.9 | 14.2, 44.1 | |
| N (N Missing) | 243 (0) | 239 (0) | |
| Serum Creatinine, mg/dL | | | 0.9254 |
| Mean (SEM / SD) | 0.9 (0.01 / 0.22) | 0.9 (0.02 / 0.23) | |
| Median | 0.9 | 0.9 | |
| Min, Max | 0.1, 1.6 | 0.5, 2.2 | |
| N (N Missing) | 227 (16) | 223 (16) | |
| LVEF, % | | | 0.4905 |
| Mean (SEM / SD) | 60.1 (0.45 / 7.08) | 59.8 (0.47 / 7.19) | |
| Median | 60.0 | 60.0 | |
| Min, Max | 38.0, 80.0 | 44.0, 82.0 | |
| N (N Missing) | 243 (0) | 235 (4) | |
| LA Diameter, cm | | | 0.5014 |
| Mean (SEM / SD) | 1.0 (0.01 / 0.22) | 1.0 (0.02 / 0.23) | 0.9000 |
| Median | 1.0 | 1.0 | |
| Min, Max | 0.1, 1.0 | 0.5, 1.0 | |
| N (N Missing) | 227 (16) | 223 (16) | |
| LVA, mm | | | |
| Mean (SEM / SD) | 0.0 (0.01 / 0.01) | 0.0 (0.01 / 0.01) | 0.0000 |
| Median | 0.0 | 0.0 | |
| Min, Max | 0.0 | 0.0 | |
| N (N Missing) | 227 (16) | 223 (16) | |
| LVA, mm | | | |
PMA 200028: FDA Summary of Safety and Effectiveness Data
{33}
| Mean (SEM / SD) | 4.1 (0.04 / 0.67) | 4.0 (0.04 / 0.59) |
| --- | --- | --- |
| Median | 4.0 | 4.0 |
| Min, Max | 2.2, 5.5 | 2.5, 5.5 |
| N (N Missing) | 243 (0) | 233 (6) |
| NYHA Functional Class, n (%) | | |
| Class I | 36 (14.8%) | 32 (13.4%) |
| Class II | 20 (8.2%) | 22 (9.2%) |
| Class III | 0 (0%) | 0 (0%) |
| Class IV | 0 (0%) | 0 (0%) |
| NA (1) | 117 (48.1%) | 116 (48.5%) |
| Unknown (2) | 70 (28.8%) | 69 (28.9%) |
| Heart Rate, bpm | | |
| Mean (SEM / SD) | 69.3
(1.23 / 19.23) | 69.0
(1.12 / 17.37) |
| Median | 64.0 | 65.0 |
| Min, Max | 36.0, 169.0 | 35.0, 140.0 |
| N (N Missing) | 243 (0) | 239 (0) |
| Systolic BP, mmHg | | |
| Mean (SEM / SD) | 136.3
(1.25 / 19.41) | 138.1
(1.28 / 19.82) |
| Median | 133.0 | 135.0 |
| Min, Max | 92.0, 206.0 | 77.0, 199.0 |
| N (N Missing) | 243 (0) | 239 (0) |
| Diastolic BP, mmHg | | |
| Mean (SEM / SD) | 77.4
(0.70 / 10.94) | 77.5
(0.69 / 10.73) |
| Median | 78.0 | 78.0 |
| Min, Max | 46.0, 110.0 | 43.0, 104.0 |
| N (N Missing) | 243 (0) | 239 (0) |
| CHA2DS2-VASc Score | | |
| Mean (SEM / SD) | 2.11 (0.10 / 1.50) | 1.92 (0.09 / 1.38) |
| Median | 2.0 | 2.0 |
| Min, Max | 0.0, 7.0 | 0.0, 6.0 |
| N (N Missing) | 243 (0) | 239 (0) |
BMI=Body Mass Index; BP=Blood Pressure; LA=Left Atrium; LVEF=Left Ventricular Ejection Fraction; NYHA=New York Heart Association.
Min = Minimum, Max = Maximum; SD = Standard Deviation; SEM=Standard Error of the Mean.
(1) Subjects without heart failure, will have an NYHA result that is not applicable (NA).
PMA 200028: FDA Summary of Safety and Effectiveness Data
{34}
(2) NYHA score is missing/not available in source documents.
Table 14 shows years since first diagnosis of atrial fibrillation and history of AAD therapy for subjects by treatment group in the Intention-to-Treat cohort.
Table 14. History of Atrial Fibrillation and ADD Therapy, Intent-to-Treat Cohort, Control vs DiamondTemp
| Medical History | Control (N=243) | DiamondTemp (N=239) |
| --- | --- | --- |
| Years Since First Diagnosis (years) | | |
| Mean (SEM / SD) | 4.0 (0.34 / 4.85) | 3.5 (0.33 / 4.68) |
| Median | 2 | 2 |
| Min, Max | 0, 26 | 0, 28 |
| N (N Missing) | 205 (38) | 207 (32) |
| AAD Use History | | |
| Subjects with History of AAD Use and Failed/Not Tolerate, n(%) | 243 (100.0%) | 239 (100.0%) |
| Subjects with History of Class I/III AAD Use and Failed/Not Tolerate, (*), n(%) | 191 (78.6%) | 187 (78.2%) |
| Subjects with History of Class II/IV AAD Use and Failed/Not Tolerate, (*), n(%) | 121 (49.8%) | 117 (49.0%) |
AAD=Anti-arrhythmic drugs; AF=Atrial Fibrillation; PAF=Paroxysmal Atrial Fibrillation
Min = Minimum, Max = Maximum; SD=Standard Deviation; SEM=Standard Error of the Mean.
Notes: N = Number of subjects in the population. n = Number of subjects in the specific category. Percentages are calculated as 100 x (n/N).
All other percentages are calculated as $100 \times (\mathrm{n} / \mathrm{N}1)$ . N1 = Number of subjects in category.
(*) Categories are not mutually exclusive and subjects may count in more than one category.
Table 15. Medical History, Intention-to-Treat Cohort, Control vs DiamondTemp
| Medical History | Control (N=243) | DiamondTemp (N=239) |
| --- | --- | --- |
| Atrial Flutter | 51 (21.0%) | 46 (19.2%) |
| Hypertension Requiring Medication | 137 (56.4%) | 124 (51.9%) |
| Hypertension Regardless of Medications Required | 138 (56.8%) | 125 (52.3%) |
| Diabetes | 28 (11.5%) | 20 (8.4%) |
| Structural Heart Disease | 7 (2.9%) | 7 (2.9%) |
| Cerebrovascular Accident/Transient Ischemic Attack | 20 (8.2%) | 12 (5.0%) |
| Thromboembolic Events | 6 (2.5%) | 4 (1.7%) |
| Coronary Artery Disease | 29 (11.9%) | 27 (11.3%) |
| Myocardial Infarction | 2 (0.8%) | 4 (1.7%) |
| Non-PAF/AFL Arrhythmias or conduction disturbance | 42 (17.3%) | 22 (9.2%) |
| Vascular Disease | 20 (8.2%) | 13 (5.4%) |
| Congestive Heart Failure | 3 (1.2%) | 6 (2.5%) |
PMA 200028: FDA Summary of Safety and Effectiveness Data
{35}
| Previous CABG Procedure | 1 (0.4%) | 0 (0%) |
| --- | --- | --- |
| Previous ICD/CRT/Pacemaker Implant | 0 (0%) | 1 (0.4%) |
| Gastrointestinal (GI) Disease | 25 (10.3%) | 28 (11.7%) |
| Pulmonary Disease with Different Etiologies | 4 (1.6%) | 6 (2.5%) |
| Sleep Apnea | 45 (18.5%) | 24 (10%) |
| Other | 159 (65.4%) | 149 (62.3%) |
| Smoking History - Yes | 75 (30.9%) | 71 (29.7%) |
| Smoking History – Current Smoker | 16 (6.6%) | 17 (7.1%) |
| Smoking History – Previously Smoked | 59 (24.3%) | 54 (22.6%) |
| Class I or III AAD at Baseline | 141 (58.0%) | 125 (52.3%) |
CABG=Coronary Artery Bypass Graft; CRT=Cardiac resynchronization therapy; ICD=implantable cardioverter defibrillator; PAF=Paroxysmal Atrial Fibrillation
Notes: N = Number of subjects in the population. n = Number of subjects in the specific category. Percentages are calculated as 100 x (n/N). Categories are not mutually exclusive and subjects may count in more than one category.
## D. Procedure and Follow-up Data
### 1. Procedure Data
Table 16 presents the index ablation procedure characteristics by treatment group. Ablation time, fluoroscopy time and procedural duration were formally tested for superiority in the secondary analysis.
Table 16. Index Ablation Procedure Characteristics, Intention-to-Treat Control vs DiamondTemp
| Procedural Characteristics | Control (N=243) | DiamondTemp (N=239) |
| --- | --- | --- |
| Total Number of AF Index Ablation ProceduresN1 | 241 | 235 |
| TEE for LA Thrombus Screening Performed | 236 (97.9%) | 234 (99.6%) |
| Esophageal Monitoring/Protection: | | |
| Esophageal Deviation, n (%) | 22 (9.1%) | 19 (8.1%) |
| Use of Esophageal Temperature Probe, n (%) | 191 (79.3%) | 186 (79.1%) |
| Ablation Parameter Settings – Max Power Set Point | | |
| Mean (SEM/SD) | 33.3 (0.39/6.01) | 49.7 (0.24/3.74) |
| Median | 30.0 | 50.0 |
| Min, Max | 5.0, 70.0 | 0.0, 56.0 |
| N (N Missing) | 239 (2) | 233 (2) |
| Ablation Parameter Settings – Max Temperature Set Point | | |
| Mean (SEM/SD) | 44.1 (0.32/4.94) | 59.9 (0.10/1.60) |
| Median | 43.0 | 60.0 |
PMA 200028: FDA Summary of Safety and Effectiveness Data
{36}
| Min, Max | 30.0, 70.0 | 43.0, 65.0 |
| --- | --- | --- |
| N (N Missing) | 233 (8) | 234 (1) |
| Non-PVI Ablation Targets(*): | | |
| Type I CTI Flutter | 69 (28.6%) | 74 (31.5%) |
| Posterior Wall | 3 (1.2%) | 4 (1.7%) |
| Focal Triggers | 2 (0.8%) | 4 (1.7%) |
| Roof | 2 (0.8%) | 2 (0.9%) |
| SVC | 3 (1.2%) | 1 (0.4%) |
| CFAE | 2 (0.8%) | 1 (0.4%) |
| Flutter Line | 2 (0.8%) | 1 (0.4%) |
| Mitral Isthmus Line | 2 (0.8%) | 1 (0.4%) |
| AVRT/AVNRT | 1 (0.4%) | 1 (0.4%) |
| Incidence of Steam Pops, n (%) | 5 (2.1%) | 7 (3.0%) |
| Incidence of Char or Coagulum, n(%) | 0 (0%) | 0 (0%) |
AF=Atrial Fibrillation; CTI=Cavotricuspid Isthmus; LA=Left Atrium; PV=Pulmonary Vein; TEE=Transesophageal echocardiography.Min = Minimum, Max = Maximum; SD=Standard Deviation; SEM=Standard Error of the Mean.Notes: N = Number of subjects in the Intention-to-Treat Population.n = Number of subjects in the specific category. Percentages are calculated as 100 x (n/N1). N1 = Number of subjects in category. (*) Categories are not mutually exclusive and subjects may count in more than one category.
2. Treatment During Blanking period
Table 17 presents the additional intervention performed to maintain sinus rhythm during the blanking period. Comparable number of subjects between the two arms received repeat ablation or cardioversion before the evaluation period.
Table 17. Additional Treatment During the Blanking Period
| Subjects with Additional Treatment during Blanking Period | Control (N=243) | DiamondTem p (N=239) | P-value* |
| --- | --- | --- | --- |
| Repeat ablation | 11 (4.53%) | 10 (4.18%) | 1.00 |
| Cardioversion | 11 (4.53%) | 7 (2.93%) | 0.47 |
| *P-value is calculated using two-sided Fisher’s exact test | | | |
3. Rhythm Monitoring Compliance
Post-ablation rhythm monitoring included symptomatic and twice monthly symptomatic/asymptomatic event monitor transmissions during the evaluation period, ECG at 3, 6, and 12 months, and 24-hour Holter monitor at 6 and 12 months. Table 18 presents the rhythm monitoring compliance for each group. The two groups had similar adherence across different monitoring methods.
PMA 200028: FDA Summary of Safety and Effectiveness Data
{37}
Table 18. Rhythm Monitoring Compliance
| Rhythm Monitoring Method | Control (N=243) | DiamondTemp (N=239) | All Subjects (N=482) |
| --- | --- | --- | --- |
| 12-Lead ECGs | | | |
| 3 Month Visit 12-Lead ECG | 219/237 (92.4%) | 209/235 (88.9%) | 428/472 (90.7%) |
| 6 Month Visit 12-Lead ECG | 206/234 (88.0%) | 207/228 (90.8%) | 413/462 (89.4%) |
| 12 Month Visit 12-Lead ECG | 224/231 (97.0%) | 223/227 (98.2%) | 447/458 (97.6%) |
| 24-hour Holter Monitor | | | |
| 6 Month Visit 24-hour Holter Monitor | 202/234 (86.3%) | 199/228 (87.3%) | 401/462 (86.8%) |
| 12 Month Visit 24-hour Holter Monitor | 213/231 (92.2%) | 204/227 (89.9%) | 417/458 (91.0%) |
| TTMs | | | |
| (Total) Transmitted TTMs | 5419 | 4557 | 9976 |
| Expected TTMs | 4373 | 4288 | 8661 |
| Overall TTM Compliance (Subject) (%) | | | |
| Mean (SEM/SD) | 60.5 (2.02/31.01) | 61.3 (2.09/32.03) | 60.9 (1.45/31.49) |
| Median | 70.0 | 72.2 | 72.2 |
| Min, Max | 0.0, 100.0 | 0.0, 100.0 | 0.0, 100.0 |
| N (N Missing) | 236 (7) | 235 (4) | 471 (11) |
ECG= Electrocardiogram;
Notes: N = Number of subjects in the Intention-to-Treat Population. n = Number of subjects in the specific category. For ECG and Holter, percentages for populations are calculated as 100 x (n/expected number of measurements at that visit).
Six subjects were randomized/enrolled but did not undergo index ablation procedure (06-013, 09-003, 10-004, 11-001, 11-004, 13-015).
TTM= Trans-telephonic Monitor.
Six subjects were randomized/enrolled but did not undergo index ablation procedure (06-013, 09-003, 10-004, 11-001, 11-004, 13-015).
[1] Subject Expected TTM is 2 if the subject's study participation in a given month is longer than 15 days, otherwise it is 1 for that month.
[2] Overall compliance is defined on a per subject basis and is based on a subject average monthly compliance rates over months 1 through 10 after the blanking period. A subject's monthly compliance rate is defined as minimum (TTM transmitted for that month, TTM expected for that month)/TTM expected for that month, over months 1 through 10 after the blanking period
PMA 200028: FDA Summary of Safety and Effectiveness Data
38 of 62
{38}
PMA 200028: FDA Summary of Safety and Effectiveness Data
39 of 62
# E. Safety and Effectiveness Results
## 1. Safety Results
### i. Primary Safety Endpoint
The primary safety analysis includes all randomized ITT subjects (243 Control and 239 DiamondTemp). There were 16 (6.6%) Control subjects and 8 (3.3%) DiamondTemp subjects that experienced at least one CEC-adjudicated primary safety endpoint event that contributed to the primary safety endpoint. The primary safety event freedom rate was 96.7% for the DiamondTemp group and 93.4% for the Control group. The difference (DiamondTemp – Control) in the primary safety endpoint freedom was 3.24% (95% CI: -1.32%, 7.79%), and the lower 97.5% confidence bound of -1.32% exceeded the pre-specified non-inferiority margin (NIM) of -6.5%. The primary safety endpoint was met (p < 0.0001, Table 19).
Table 19. Primary Safety Result, Intention-to-Treat Cohort
| Primary Safety Endpoint: Freedom from Primary Safety Event as Adjudicated by the CEC | Control (N=243) Number (%) (95% CI) | DiamondTemp (N=239) Number (%) (95% CI) | Difference (95% CI) | Farrington-Manning p-value (Non-inferiority Test) |
| --- | --- | --- | --- | --- |
| Total, By Subject | 227 (93.4%) (89.5%, 96.2%) | 231 (96.7%) (93.5%, 98.5%) | 3.24% (-1.32%, 7.79%) | <0.0001 |
{39}

Figure 6. Primary Safety Freedom Rate; Between Treatment Difference
There were 24 Primary Safety Endpoint events reported in 24 subjects. Table 20 summarizes the events that met the primary safety endpoint. Specifically, there were 16 subjects that met the primary safety endpoint in the control group and 8 subjects that met the primary safety endpoint in the DiamondTemp group. The most common primary safety endpoint met in the control group was extended hospitalization (6 subjects) followed by vascular access site complication (4 subjects). No more than 2 subjects met any of the individual primary safety endpoint events in the DiamondTemp group.
Table 20 CEC Adjudicated Primary Safety Endpoint Events (Intention-to-Treat Cohort)
| CEC Adjudicated Adverse Events Contributing to the Primary Safety Endpoint | Control By Subject (Ns=243) n (%) | DiamondTemp By Subject (Ns=239) n (%) |
| --- | --- | --- |
| Atrioesophageal Fistula | 0 (0%) | 0 (0%) |
| Bleeding Complication | 0 (0%) | 0 (0%) |
| Cardiac Tamponade/Perforation | 2 (0.8%) | 2 (0.8%) |
| Cardiovascular-Related Death Post-Ablation | 0 (0%) | 0 (0%) |
| Clinically Symptomatic Pulmonary Vein Stenosis at 6 Months Post-Index Ablation Procedure | 0 (0%) | 0 (0%) |
| Extended Hospitalization | 6 (2.5%)* | 0 (0%) |
| Myocardial Infarction | 0 (0%) | 0 (0%) |
| Pericarditis | 1 (0.4%) | 0 (0%) |
| Phrenic Nerve Paralysis | 0 (0%) | 1 (0.4%) |
| Pulmonary Edema | 1 (0.4%) | 0 (0%) |
PMA 200028: FDA Summary of Safety and Effectiveness Data
{40}
| Stroke Post Ablation | 1 (0.4%) | 0 (0%) |
| --- | --- | --- |
| Thromboembolism | 0 (0%) | 0 (0%) |
| Transient Ischemic Attack | 1 (0.4%) | 2 (0.8%) |
| Vagal Nerve Injury | 0 (0%) | 1 (0.4%) |
| Vascular Access Complication | 4 (1.6%) | 2 (0.8%) |
| Total | 16 (6.6%) | 8 (3.3%) |
*Reasons for extended hospitalization include hematoma, pericardial effusion (< 1 cm), fever and chill, bladder outlet obstruction with UTI, hypotension, chest pain.
Notes: Ns = Number of subjects in the population. n = Number of subjects in the specific category.
Subject based percentages are calculated as 100 x (n/Ns).
Ne = Number of events in the Population. n2 = Number of events in the specific category.
Event based percentages are calculated as 100 x (n2/Ne).
For the 'by Subject' columns, subjects reporting a particular adverse event more than once are only counted once by the event category.
For the 'by Event' columns, all events are counted.
Table 20 and Figure 6 display the primary safety objective results for the ITT cohort. The primary safety event freedom rate was 96.7% for the DiamondTemp group and 93.4% for the control group. The DiamondTemp minus control group primary safety endpoint freedom rate was 3.24% with a two-sided 95% confidence interval of -1.32% to 7.79%. Since the lower two-sided 95% confidence limit of -1.32% exceeded the non-inferiority margin of -6.5%, the primary safety objective was met (p < 0.0001).
A Kaplan-Meier analysis was also performed to evaluate the primary safety endpoint as a sensitivity analysis. The Kaplan-Meier method allows subjects to be included in the analysis up until the time they fail the primary safety endpoint or are censored due to premature study exit.
Figure 7 displays the Kaplan-Meier estimates for the freedom from primary safety event through 6 months (180 days) post-index ablation procedure; the entire time period for which subjects were at risk for a primary safety event. Based on the Kaplan-Meier methodology the freedom from primary safety event at 6 months was 97% for the DiamondTemp group and 93% for the control group. The log-rank test indicated that there was no difference in the freedom from primary safety event between groups (p=0.11).
PMA 200028: FDA Summary of Safety and Effectiveness Data
{41}

Figure 7. Kaplan- Meier Survival: Time to Failure of the Primary Safety Endpoint, Intention-to-Treat Cohort
| Kaplan-Meier Survival: Time to Failure of the Primary Safety Endpoint, Intention-to-Treat Cohort | | | | |
| --- | --- | --- | --- | --- |
| | | Month 1 | Month 3 | Month 6 |
| Control | Number at Risk | 225 | 222 | 219 |
| | Kaplan-Meier | 0.9 | 0.9 | 0.9 |
| | Estimate | 4 | 3 | 3 |
| | Standard Error | 0.0155 | 0.0160 | 0.0160 |
| DiamondTemp | Number at Risk | 227 | 226 | 218 |
| | Kaplan-Meier | 0.9 | 0.9 | 0.9 |
| | Estimate | 7 | 7 | 7 |
| | Standard Error | 0.0117 | 0.0117 | 0.0117 |
Table 21 displays the primary safety endpoint status by treatment group and geography and indicates that the primary safety endpoint results were consistent by geography (Breslow-Day p-value = 0.54).
Table 21. Primary Safety Event Outcome: Relative Risk of Success; Overall and Stratified by Geographic Region and Treatment, Intention-to-Treat Cohort
| Geographic Region | Treatment | PSE Success | PSE Failure | Total | Relative Risk of Success | Breslow Day Test p-value |
| --- | --- | --- | --- | --- | --- | --- |
| Europe | DiamondTemp | 124 | 5 | 129 | 1.02 | 0.5425 |
| | Control | 124 | 8 | 132 | | |
| North America | DiamondTemp | 107 | 3 | 110 | 1.05 | |
| | Control | 103 | 8 | 111 | | |
| Overall | DiamondTemp | 231 | 8 | 239 | 1.03 | |
| | Control | 227 | 16 | 243 | | |
| Pooled / Adjusted (CMH)a | | | | | 1.03 | |
a CMH= Cochran-Mantel-Haenszel
PMA 200028: FDA Summary of Safety and Effectiveness Data
{42}
The DIAMOND-AF Clinical Study met its primary safety objective (Intention-to-Treat Cohort). Primary safety endpoint success was observed in 227 (93.4%) control (TactiCath) subjects and 231 (96.7%) DiamondTemp subjects (95% CI for difference: -1.3% to 7.8%; p<0.0001 for non-inferiority). There was no evidence of heterogeneity in primary effectiveness outcome between treatment groups by geography (p=0.54). The DiamondTemp Ablation System demonstrated a reasonable assurance of safety for the treatment of drug refractory, recurrent, symptomatic PAF.
## Summary of Adverse Events
Adverse events (AE) occurring during the study were continuously monitored and collected. There were no Unanticipated Adverse Device Effects or deaths reported in the DIAMOND-AF Clinical Study.
Table 22 summarizes all adverse events by seriousness and relatedness. In the ITT cohort, there were 171 adverse events reported in 98 (41.0%) of the 239 subjects randomized to the DiamondTemp group. Of these events, 21 events in 18 (7.5%) subjects were considered device or procedure related regardless of severity. There were 199 total adverse events reported in 103 (42.4%) of the 243 subjects randomized to the control group. Of these events, 35 events in 31 (12.8%) subjects were considered device or procedure related regardless of severity.
Table 22. AE Overall Summary Table, Intention-to-Treat Cohort, Control vs DiamondTemp
| | Number of Events (Number of subjects, % of Subjects) | |
| --- | --- | --- |
| Adverse Event Classification | Control (N=243) | DiamondTemp (N=239) |
| Total Adverse Events | 199 (103, 42.4%) | 171 (98, 41.0%) |
| Primary Safety Events[1] | 16 (16, 6.6%) | 8 (8, 3.3%) |
| Serious[2] | | |
| Yes | 61 (43, 17.7%) | 53 (34, 14.2%) |
| No | 138 (80, 32.9%) | 118 (79, 33.1%) |
| Relatedness[2], [4] | | |
| Device and/or Procedure Related[3] | | |
| Related | 35 (31, 12.8%) | 21 (18, 7.5%) |
| Possibly Related | 23 (18, 7.4%) | 16 (15, 6.3%) |
| Unknown | 2 (2, 0.8%) | 0 (0, 0%) |
| Not Related | 139 (79, 32.5%) | 134 (81, 33.9%) |
| Device Relatedness | | |
PMA 200028: FDA Summary of Safety and Effectiveness Data
43 of 62
{43}
All serious adverse events reported during this study are listed in Table 23 by each treatment group.
Table 23 Serious Adverse Events by System Organ Cla…