TACTICATH QUARTZ SET
P130026 · St Jude Medical · OAE · Oct 24, 2014 · Cardiovascular
Device Facts
| Record ID | P130026 |
| Device Name | TACTICATH QUARTZ SET |
| Applicant | St Jude Medical |
| Product Code | OAE · Cardiovascular |
| Decision Date | Oct 24, 2014 |
| Decision | APPR |
| Device Class | Class 3 |
| Attributes | Therapeutic |
Indications for Use
The TactiCath™ Quartz Contact Force Ablation Catheter is indicated for use in cardiac electrophysiological mapping and for the treatment of drug refractory recurrent symptomatic paroxysmal atrial fibrillation, when used in conjunction with a compatible radiofrequency generator and three-dimensional mapping system. TactiSys™ Quartz Equipment and accessories are indicated for use in conjunction with a TactiCath™ Quartz Contact Force Ablation Catheter. TactiSys™ Quartz Equipment allows the visualization of the force information coming from the catheter tip.
Device Story
TactiCath™ Quartz system comprises a 7F steerable, open-irrigated RF ablation catheter with an integrated tri-axial optical force sensor and thermocouple; TactiSys™ Quartz signal processing unit; and TactiSoft™ software. Used in cardiac electrophysiology labs by physicians for mapping and ablation of paroxysmal atrial fibrillation. Catheter tip transmits contact force data via optical sensors to the TactiSys™ unit, which processes signals and displays real-time force information on a dedicated PC via TactiSoft™. System integrates with third-party 3D mapping systems and RF generators. Real-time force visualization allows physicians to optimize contact force during lesion delivery, potentially improving lesion durability and reducing reablation rates compared to non-force-sensing catheters.
Clinical Evidence
Pivotal TOCCASTAR study: prospective, randomized, multicenter trial (n=317 enrolled, 295 SAF population) comparing TactiCath™ Set to a control device. Primary effectiveness: noninferiority of treatment success (freedom from recurrence) at 9 months post-blanking. Primary safety: noninferiority of device-related serious adverse events (SAEs) within 7 days. Results: TactiCath met noninferiority for both primary endpoints. Secondary analysis showed significant reduction in total RF application time (51.7 vs 58.3 min, p=0.002) and improved success rates (75.9% vs 58.1%, p=0.018) when using optimal contact force (>10g in ≥90% of lesions).
Technological Characteristics
7F steerable, open-irrigated RF ablation catheter; platinum-iridium electrodes; tri-axial optical force sensor; thermocouple temperature sensor. Materials: Polyether block amide, polyurethane, silicone, cyanoacrylate, epoxy, Inox 316L. Connectivity: Ethernet connection to dedicated PC. Sterilization: Ethylene oxide (EO). Software: TactiSoft™ for force visualization. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, and IEC 62366.
Indications for Use
Indicated for patients 18+ years with drug refractory recurrent symptomatic paroxysmal atrial fibrillation. Contraindicated in patients with recent ventriculotomy/atriotomy (within 4 weeks), prosthetic valves, active systemic infection, myxoma, intracardiac thrombus, interatrial baffle/patch, or prior aortic valve replacement (if using retrograde trans-aortic approach). Not for use in coronary vasculature.
Regulatory Classification
Identification
For the treatment of atrial fibrillation.
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
Device Generic Names:
- Contact Force Irrigated RF Ablation Catheter
- Contact Force Visualization Equipment and Accessories
Device Trade Names:
- TactiCath™ Quartz Contact Force Ablation Catheter
- TactiSys™ Quartz Equipment Composed of:
- TactiSys™ Quartz Unit
- TactiSoft™ Software
- TactiSys™ Quartz RF Cable
- TactiSys™ Quartz Ethernet Cable
- Equipotential Cable
- Mains Cord and Adapter
Optional Accessories:
- TactiSys™ Quartz Mount
- TactiSys™ Analog Output Cable
- TactiSys™ Quartz Integration Module
Device Product Code:
OAE
Applicant's Name and Address:
St. Jude Medical
One St. Jude Medical Drive
St. Paul, MN 55117
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P130026
Date of FDA Notice of Approval: 10/24/2014
Priority Review: Not Applicable
II. INDICATIONS FOR USE
The TactiCath™ Quartz Contact Force Ablation Catheter is indicated for use in cardiac electrophysiological mapping and for the treatment of drug refractory recurrent
Page 1 of 29
{1}
symptomatic paroxysmal atrial fibrillation, when used in conjunction with a compatible radiofrequency generator and three-dimensional mapping system.
TactiSys™ Quartz Equipment and accessories are indicated for use in conjunction with a TactiCath™ Quartz Contact Force Ablation Catheter.
TactiSys™ Quartz Equipment allows the visualization of the force information coming from the catheter tip.
## III. CONTRAINDICATIONS
The TactiCath™ Quartz Contact Force Ablation Catheter is contraindicated for:
- Patients who have had a ventriculotomy or atriotomy within the preceding four weeks as the recent surgery may increase the risk of perforation
- Patients with prosthetic valves as the catheter may damage the prosthesis.
- Patients with an active systemic infection as this may increase the risk for cardiac infection.
- The use in coronary vasculature due to risk of damage to the coronary arteries.
- Patients with a myxoma or an intracardiac thrombus as the catheter could precipitate an embolus.
- The transseptal approach in a patient with an interatrial baffle or patch because the opening could persist and produce an iatrogenic atrial shunt.
- The retrograde trans-aortic approach in patients who have had aortic valve replacement.
## IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the TactiCath™ Quartz Contact Force Ablation Catheter Instructions for Use (IFU), and the TactiSys™ Quartz Equipment User Manual.
## V. DEVICE DESCRIPTION
The TactiCath™ Quartz Set is comprised of the following components:
- TactiCath™ Quartz Contact Force Ablation Catheter: A sterile, 7F steerable, open irrigated radiofrequency (RF) ablation multi-electrode catheter with a deflectable tip incorporating a force sensor and a thermocouple temperature sensor. The catheter is designed to facilitate electrophysiology mapping of the heart chambers and to transmit RF current to the catheter tip electrode for intracardial ablation. The deflectable distal section of the catheter shaft includes four platinum-iridium electrodes (three ring electrodes and one tip electrode) and is available in 65 mm or 75 mm length, with a total useable length of 115 cm. At the proximal end of the catheter, a saline port is used to deliver isotonic saline solution to irrigate and cool the tip electrode and ablation site,
Page 2 of 29
{2}
through the 0.7 mm diameter catheter lumen and the six small holes on the distal electrode tip. The catheter features a tri-axial optical force sensor embedded in the tip section which transmits contact force information to the TactiSys™ Quartz Equipment.
- TactiSys™ Quartz Equipment: The contact force visualization equipment is composed of the TactiSys™ Quartz unit (hardware and associated cables) and TactiSoft™ (stand-alone software). The TactiSys™ Quartz unit is a non-sterile active signal and data processing unit that interconnects the TactiCath™ Quartz catheter to an external RF generator, and collects data from the catheter to compute contact force and related information transmitted from the catheter. The optical measurement system, which is compatible with the force sensing technology incorporated in the TactiCath™ Quartz catheter, is composed of three optical sensor analyzer modules. The TactiSoft™ software displays the contact force information transmitted from the catheter through an Ethernet connection on a dedicated, properly configured computer.
The following devices are required in addition to the TactiCath™ Quartz Set to perform catheter ablation procedures:
- RF generator and related disposable grounding pad (dispersive pad)
- Irrigation pump
- Electrophysiology recording system
- Dedicated Personal Computer with display screen provided by Endosense to run TactiSoft™ software.
Optional accessories include: choice of RF cable, Ethernet cables available in different lengths, an analog output cable, a mount and an integration module (Ethernet - switch) that will allow the integration of contact force information with third party equipment (e.g. integration in the screen of 3D mapping system for enhanced electrophysiology laboratory set-up).
VI. ALTERNATIVE PRACTICES AND PROCEDURES
Alternative therapy for symptomatic paroxysmal atrial fibrillation includes the following:
- Commercially available PMA-approved devices
- Pharmacological therapy for rate and/or rhythm control
- Electrical or pharmacologic cardioversion
- Surgical intervention to create atrial lesions
- Implantable devices to control heart rate
Page 3 of 29
{3}
Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his/her physician to select the method that best meets expectations and lifestyle.
## VII. MARKETING HISTORY
The TactiCath™ Quartz Set is currently marketed outside the United States in Europe, Australia, and New Zealand.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
The following adverse events have been documented for catheter ablation procedures:
| ·Air embolism
·Anesthesia reaction
·Aorto-right atrial fistula
·Arrhythmias; bradycardia and tachycardia
·Atrio-Ventricular fistula
·Cardiac perforation/tamponade
·Cardiac thromboembolism
·Cerebrovascular incident or Attack / Stroke
·Chest pain/discomfort
·Coronary artery dissection
·Coronary artery spasm
·Coronary artery thrombosis / occlusion
·Death
·Diaphragmatic paralysis
·Dislodgement of implantable cardioverter defibrillator or permanent pacing leads
·Endocarditis
·Gastroparesis
·Heart failure / pump failure
·Hemothorax
·Hospitalization (initial and prolonged)
·Increased creatinine phosphokinase (CPK) level
·Infections
·Laceration
·Leakage of air or blood into the lungs or other organs due to perforation
·Left atrial esophageal fistula
·Major bleeding, requiring surgery or transfusion
·Myocardial infarction
·Obstruction or perforation or damage to the vascular system | ·Pericarditis
·Pericardial effusion
·Phrenic nerve damage including diaphragmatic paralysis
·Pleural effusion
·Pneumonia
·Pneumothorax
·Pseudoaneurysm
·Pulmonary edema
·Pulmonary embolism
·Pulmonary vein dissection
·Pulmonary vein thrombus
·Pulmonary hypertension
·Respiratory depression
·Skin burns
·Severe pulmonary vein (PV) stenosis (>70%), or complete occlusion of a PV, even in the absence of symptoms
·Tamponade, potentially requiring surgery
·Temperature elevation or fever
·Transient Ischemic Attack (TIA)
·Thromboembolism
·Thrombosis
·Unintended complete or incomplete Atrioventricular (AV), Sinus node, or other heart block or damage
·Valvular damage
·Vascular bleeding / local hematomas / ecchymosis
·Vasovagal reactions
·Ventricular tachyarrhythmia
·Volume overload |
| --- | --- |
For the specific adverse events that occurred in the clinical studies, please see Section X below.
Page 4 of 29
{4}
Page 5 of 29
# IX. SUMMARY OF PRECLINICAL STUDIES
Pre-clinical testing of the TactiCath™ Quartz Contact Force Ablation Catheter and TactiSys™ Quartz Equipment included verification and validation testing (catheter testing and hardware/software testing), biocompatibility, shelf life testing, and animal studies. Performance testing was conducted to demonstrate design integrity. All tests performed which were identified in standards or guidance documents were based on the product specification requirements. In the tests described below, the catheters were manufactured by trained manufacturing operators. "Pass" as used below denotes that the devices and system met established product specifications and/or performance criteria, or were in conformance with the requirements of the standards identified. Testing results confirmed that the catheters meet product specifications.
## A. Laboratory Studies
In Vitro Bench Studies – TactiCath™ Quartz Catheter
In-vitro bench testing was conducted to ensure the TactiCath™ Quartz catheter met product requirements and applicable standards. Table 1 summarizes the bench testing for the catheters including mechanical and electrical integrity, and performance test results.
Table 1: TactiCath™ Quartz Catheter Bench Testing Results Summary
| | Test Description | Acceptance Criteria | Result |
| --- | --- | --- | --- |
| Mechanical Performance | Steering, planarity, handle, pullwire force | Length of bend within specification
Maximal / minimal angles of bend within specifications
Radius of curvature at maximum bending within specification
Steering section maintaining its position when the pull force is released
Steered tip planarity when the catheter is fully steered
Steering force over 150 cycles within specification | Pass |
| | Kink, stiffness, tip side force, hook force | No kink, no fracture, no visible peeling of catheter shaft
Distal tip side force within specification
Hook force within specification
Stiffness on non-steering section of catheter within specification | Pass |
| | External cable torsion resistance | No kinks or breaks of external cables. Catheter fully functional after several full rotations of cables | Pass |
| | External cable tension resistance | Catheter fully functional under tensile force | Pass |
| | Shaft torsion and buckling | No kinks or breaks after 360° rotation of distal part of catheter
Buckling force within specification | Pass |
{5}
| | Test Description | Acceptance Criteria | Result |
| --- | --- | --- | --- |
| | Force sensor mechanical integrity after high forces | No catheter part separation when extreme axial or lateral force applied | Pass |
| | Introducer extraction tests and force at break | Pullwire force resistance within specification
No loose part when extracting steered or unsteered catheter through introducer
Forces at break ≥ 15N for each catheter section (per ISO 10555-1 standard) | Pass |
| | Corrosion resistance | No corrosion after 48h of soaking, per ISO-10555-1 standard, Annex A | Pass |
| | Catheter removal from blister | No loose parts or detectable damage to catheter or force sensor when catheter removed from blister | Pass |
| Force Sensor Performance | Humidity drift | Force sensing drift due to humidity within specification | Pass |
| | Force sensor repeatability | Standard deviations of Delta force within specifications | Pass |
| | Force sensor accuracy | Delta force between measured Total force and reference force within specifications
Delta angle α between measured and reference angle within specifications
Delta angle θ between measured and reference angle within specifications | Pass |
| | Scale measurement of force | Return to zero of Total force after application of high forces within specifications | Pass |
| | Force reading sensitivity | Force reading standard deviation within specification | Pass |
| | Sensitivity to steering mechanism | Average value of delta Total force within specification | Pass |
| | Sensitivity to cable manipulation | Total force variations during cable manipulation within specification | Pass |
| | Distal tip thermocouple accuracy | Difference between measurement and reference within specification | Pass |
| | Force sensor temperature reading | Difference between reference temperature and force sensor thermocouple within specification | Pass |
| | Plug cold junction temperature reading | Difference between reference temperature and thermistor temperature within specification | Pass |
| | Drift during ablation | Means of the absolute value of the “Force total” drift during ablation within specifications | Pass |
| | Fatigue and high forces test | Force sensor fully functional after 200 cycles of axial and lateral force applied
Return to zero of Total force within specifications | Pass |
{6}
Page 7 of 29
| | Test Description | Acceptance Criteria | Result |
| --- | --- | --- | --- |
| | Trackability - Tortuous path | Catheter fully functional, including force sensor, after navigation in tortuous path
Electrical insulation/resistance measurements within specifications | Pass |
| | Operating temperature range and temperature compensated range | Force total drift of the force sensor in the limit conditions | Pass |
| | Transversal force effect | Accuracy of the force sensor and plot of the transversal force effect | Pass |
| | Heart beating simulation | Test the force sensor with a variable force applied on the tip in the axial direction and lateral direction | Pass |
| | Repeatability and reproducibility | Test the repeatability and the reproducibility of the calibration method and the inspection method with 3 catheters and 3 operators | Pass |
| Functional Performance | Sheath introduction test | Catheter fully functional, including force sensor, after passing 10 times back and forth through St Jude SL1 8.5F introducer | Pass |
| | Irrigation test | Catheter and Luer free from leakage
Tight and firm Luer connection
Pressure drop for different flow rates within specifications | Pass |
| | Kink Simulation | No kink of irrigation tube in hard conditions
Pressure drop for different flow rates within specifications | Pass |
| | Joint Seal Leakage | Catheter shall be free from leakage per ISO 10555-1 standard | Pass |
| Electrical Performance | Electrical test | Measurements of electrical insulation between all electrodes, thermocouples and memory in wet condition within specifications
Measurements of electrical resistance of all electrodes within specifications | Pass |
| Compatibility with external devices | Electrical Connector Separation Force | Axial force necessary to disconnect the electrical connector from the TactiSys Quartz within specifications | Pass |
| | RF Generator Compatibility | Power deviation for different ablation powers and load resistances within specifications | Pass |
In Vitro Bench Testing – TactiSys™ Quartz Equipment (Hardware & Software)
The TactiCath™ Quartz Set was evaluated in accordance with internationally recognized standards for electrical safety and electromagnetic compatibility. External third party agency testing was completed to ensure all applicable IEC 60601-1 and applicable collateral and particular standards requirements were
{7}
met. Passing results were concluded for testing conducted to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, and IEC 62366.
# Biocompatibility Testing
The TactiCath™ Quartz catheter biocompatibility testing was conducted in accordance with ISO 10993-1 for external communicating devices in contact with circulating bloods for less than 24 hours. Biocompatibility testing was performed in accordance with applicable parts of ISO 10993-1 and establishes biocompatibility and material safety of the proposed device. The biocompatibility results are summarized in Table 2.
Table 2: Biocompatibility Testing Summary
| Biological Test/Method | Result |
| --- | --- |
| Cytotoxicity – MEM Elution Assay (ISO 10993-5) | Pass |
| Sensitization – Guinea Pig Maximization (ISO 10993-10) | Pass |
| Intracutaneous Reactivity – Rabbit (ISO 10993-10) | Pass |
| Acute Systemic Toxicity – Mouse Systemic Injection (ISO 10993-11) | Pass |
| Pyrogenicity – Material Mediated Rabbit Pyrogen (ISO 10993-11) | Pass |
| Haemocompatibility – Blood Compatibility Test for Hemolysis (ISO 10993-4) | Pass |
| Haemocompatibility – Complement Activation (ISO 10993-4) | Pass |
| Thrombogenicity – Porcine (ISO 10993-4) | Pass |
Patient contacting materials of the TactiCath™ Quartz catheter are provided in Table 3.
Table 3. List of Patient Contact Materials and Components
| Description | Material | Patient Contact |
| --- | --- | --- |
| Tip and ring Electrodes | Platinum-iridium | Direct contact |
| Outer shaft | Polyether block amide
Polyurethane elastomer | Direct contact |
| Strain relief | Silicone elastomer | Direct contact |
| Adhesives | Cyanoacrylate ester | Direct contact |
| | Acrylate urethane | Direct and indirect contact |
| | Loctite epoxy | Indirect contact |
| Irrigation tubing | Inox 316L
Polyimide | Indirect contact |
| Female luer lock | Methyl Methacrylate Acrylonitrile
Butadiene Styrene (MABS) | Indirect contact |
| Single lumen | Polyether block amide | Indirect contact |
| Tip o-ring | Clear silicone | Indirect contact |
{8}
Page 9 of 29
## B. Animal Studies
Acute *in vivo* animal testing was conducted using the TactiCath™ Set to demonstrate safety and acute effectiveness in a setting that simulated a clinical ablation procedure. The animal study was conducted on a total of 8 animals (4 Yorkshire pigs and 4 Mongrel dogs). Applied contact force and tip temperatures at pre- and post-ablations, RF power, time and heart sites were recorded. Testing demonstrated that the catheters successfully delivered RF energy to target endocardial locations. The overall device performance from a gross pathological perspective was characteristic of cardiac ablation. No safety issues related to the device were recorded. All animals survived the procedures without any significant adverse events such as cardiac tamponade, respiratory distress, or uncontrollable bleeding. One animal experienced a pericardial effusion, which was interpreted as related to the procedure and not related to the investigational device.
Out of 183 ablations, 9 (4.9%) steam pops (4 audible pops and 3 impedance pops during procedure, and 2 through histology) were observed during and after ablation procedure. Eight (8) pops occurred with energy set at 50W while and one (1) pop occurred with energy set at 30W. During ablation, the median contact force was 20 grams, ranging between 4 and 74 grams.
## C. Other Studies
### Sterilization
The TactiCath™ Quartz catheter is supplied sterile. The catheters are sterilized using ethylene oxide (EO) gas to a sterility assurance level (SAL) of 10⁻⁶. The sterilization process was validated according to ISO 11135-1. Catheters meet the ISO allowable limits for EO/ECH gas residuals as set forth in ISO 10993-7. Catheters are routinely tested for pyrogens of non-material mediated origin and meet the USP criteria for devices in contact with blood. The catheters are single use only.
### Packaging, and Shelf Life
The TactiCath™ Quartz catheter is packaged in a coiled configuration, within a blister tray sealed with a Tyvek Lid. The sealed tray is packaged in a heat sealed pouch, which is then packaged in a rectangular cardboard box. The catheter has been validated for a two year shelf life. The catheters are labeled with the appropriate shelf life. Table 4 provides a summary of testing conducted to support the shelf life and packaging.
{9}
Table 4. TactiCath™ Quartz Catheter Shelf Life and Packaging Testing
| Test | Method | Acceptance Criteria | Result |
| --- | --- | --- | --- |
| Packaging Stability Testing | ISO 11607-1: 2006
ISO 11607-2: 2006 | • Packaging integrity after:
o Dye penetration test
o Burst test on blister
o Seal strength test on blister
o Seal strength test on pouch
• Sterility: The sterile barrier system shall maintain sterile
• Labels: remains legible, no smearing ink, adherence to the support, no deterioration
• Manual peel-off force (lid and pouch):
o Packaging easy to open
o Peel-off – Adhesive
o No material torn | Pass |
| Device Stability Testing | ASTM F1980-07 | • Functionality and sterility of packaged catheter after sterilization and aging:
o No catheter degradation
o No interaction with packaging which could adversely affect the device
o No degradation of catheter integrity: assembly tip and deformable body | Pass |
| Pre-shipment transport testing | ISTA 2A | Functionality and sterility of packaged catheter after simulated shipping testing | Pass |
Page 10 of 29
{10}
Page 11 of 29
# X. SUMMARY OF PRIMARY CLINICAL STUDY
A clinical study named TOCCASTAR was performed to establish a reasonable assurance of safety and effectiveness of the TactiCath™ Set to treat drug refractory recurrent symptomatic paroxysmal atrial fibrillation when compared with an approved control device in the US and outside of the US under IDE G100230. The clinical study was conducted on the TactiCath™ Set. The TactiCath Quartz Set is a product evolution relative to the original TactiCath Set, with minor changes to the irrigated RF ablation catheter and enhancements to the graphical user interface designed to enhance reliability and manufacturability. The indications for use and functionalities remain unchanged, as does the catheter basic design. The TactiCath™ Quartz Set, was evaluated through bench testing and a supplemental clinical study. Data from this clinical study were the basis for the PMA approval decision. A summary of the clinical study is provided below.
## A. Study Design
The TOCCASTAR clinical study was a prospective, randomized, multicenter, interventional clinical investigation conducted at 17 investigational sites (10 in the US and 7 outside of the US). The study was conducted on patients 18 or older.
Subjects were enrolled between January 11, 2011 and June 4, 2012. A total of 317 patients were enrolled in the study. Twelve month follow-up was completed on the last randomized patient June 25, 2013.
## B. Clinical Endpoints
### Primary Effectiveness
The primary effectiveness endpoint was a noninferiority comparison of treatment success between the TactiCath™ Set and the control device.
Acute success was defined as electrical isolation of all 4 PVs, or in the event of a common PV, the clinical equivalent of all PVs by the end of the index procedure.
Chronic success was defined as acute procedural success and freedom from recurrence of symptomatic paroxysmal atrial fibrillation (PAF), atrial flutter (AFL), and atrial tachycardia (AT) lasting longer than 30 seconds through 9 months of follow-up after a 3 month blanking period. Reablation for AF occurring 80 or more days after the index procedure or the use of class I or class III AADs after the 3-month follow-up constitutes treatment failure.
### Secondary Effectiveness
The secondary effectiveness endpoints were related to the use of the contact force sensor and assess the procedural effectiveness superiority of the
{11}
TactiCath™ Set over the control device by use of a hierarchical closed test procedure.
## Primary Safety
The primary safety endpoint was a noninferiority comparison of device-related early-onset primary serious adverse events (SAEs) between the TactiCath™ Set and the control device occurring within 7 days of the index procedure or hospital discharge, whichever was later, and diagnosed at any time during the follow-up period.
Primary SAEs that met the severity criteria described in Table 5 contributed to the primary safety endpoint only in subjects in whom a study device (TactiCath or control) was introduced. Hospitalizations solely for arrhythmia recurrence (without coexisting conditions such as thromboembolism, worsening heart failure, etc.) were not considered primary SAEs.
Table 5. Primary Serious Adverse Events and Severity Criteria
| Primary SAE | Severity criteria |
| --- | --- |
| Atrial perforation | Visible (either with radiographic and/or ultrasonographic imaging and/or direct visualization) movement of ablation catheter, needle or sheath through the atrial wall as evidenced by bleeding and the need for pericardial drainage or surgical intervention |
| AV block | New, persistent 2^{nd}or 3^{rd} degree atrioventricular (AV) block not attributable to a vasovagal reaction or medication effect and requiring permanent pacing |
| Death | Adverse event resulting in subject death |
| Diaphragmatic paralysis | Change in baseline diaphragmatic function as evidenced by elevation of a hemidiaphragm above its normal position or loss of normal respiratory excursion but not due to a pulmonary process such as atelectasis and persisting longer than the end of the procedure |
| Gastroparesis | Gastroparesis as a result of ablation requiring intervention or hospitalization |
| Hospitalization (initial or prolonged) | Adverse event leading to new hospital admission or extension of initial hospital stay beyond expected timeframe due to ablation procedure-related cause. Excludes hospitalization solely for arrhythmia recurrence. |
| Left atrial esophageal fistula | Creation of a direct communication (fistula) between the left atrium and esophagus necessitating surgical intervention or resulting in permanent impairment (eg, due to hemorrhage or septic emboli) |
| Myocardial infarction | Requires 2 of the following 3 criteria:
• Elevation of biochemical markers of myocardial necrosis (preferably troponin levels)
• Ischemic symptoms
• Development of pathologic Q waves on the electrocardiography (ECG) or persistent ECG changes indicative of ischemia (ST segment elevation or depression) |
| Pericarditis | Pleuritic chest discomfort associated with either pericardial rub and/or ECG changes that requires or prolongs hospitalization |
Page 12 of 29
{12}
| Primary SAE | Severity criteria |
| --- | --- |
| Pneumothorax | Identification of air in the pleural space which either prolongs hospital stay (for observation) or requires surgical intervention or chest tube placement |
| Pulmonary edema | Pulmonary alveolar fluid accumulation accompanied by typical symptoms (dyspnea), physical findings (rales, hypoxemia), radiologic findings, and response to diuretic therapy and requiring hospitalization |
| PV stenosis | Severe (≥70%), or complete occlusion of a PV, even in the absence of symptoms |
| Stroke | Brain disorder involving loss of brain functions (that persists for >24 hours) that occurs when the blood supply to any part of the brain is interrupted as determined by the consulting neurologist |
| Tamponade | Pericardial effusion of sufficient size to cause hemodynamic compromise and requiring drainage based on hypotension, echocardiographic findings, or other clinical factors |
| Thromboembolism | Deep vein thrombosis or pulmonary embolism |
| Transient ischemic attack | Acute episode of temporary (<24 hrs.) and focal loss of cerebral function of vascular (occlusive) origin as determined by the consulting neurologist |
| Vascular access complications | Vascular access complication requiring surgical repair, blood transfusion (eg, groin hematoma, AV fistula) or significant intervention such as thrombin injection (eg, pseudoaneurysm) |
## Secondary Safety
The secondary safety endpoint was to evaluate the incidence of all SAEs during the 12-month follow-up period.
## C. Statistical Analysis
### Analysis Populations
The following populations were analyzed for this study:
- Full analysis (FA): The FA population includes all randomized subjects.
- Safety (SAF): The SAF population includes all randomized subjects in whom a study device was introduced.
- Modified intent-to-treat (mITT): The mITT population includes all randomized subjects in whom a study device was introduced and PVI was attempted.
- Per-protocol (PP): The PP population includes all subjects in the mITT population who did not have major protocol deviations that would have compromised the effectiveness or safety evaluation of the study device.
- Roll-in population: Subjects recruited into the roll-in subject population – these subjects were not randomized but were followed for safety. Their data were summarized separately and were also included in the listings.
- Supplemental clinical study: Subjects treated with the TactiCath Quartz Set were not randomized and are reported separately.
- Extended follow-up data: Subject data provided after the end of the required 12-month follow-up period.
{13}
Page 14 of 29
# Sample Size Calculation
The treatment success rate and the incidence of primary SAEs within 7 days of the index procedure for the TOCCASTAR study was based on data from previous pivotal and feasibility studies in similar populations to estimate event rates.
Sample size was calculated based on the estimated treatment success rate and the incidence of primary SAEs. Let Pt denote the percent of treatment success or the incidence of primary SAEs in the TactiCath arm and Pc denote the corresponding rate in the control arm. Under these assumptions, a total of 274 subjects (137 subjects per treatment arm) would provide 80% power to reject the null hypothesis that (Pt - Pc < -0.15) with a one-sided 0.05 significance level for the primary effectiveness endpoint. A total of 274 subjects (137 subjects per treatment arm) would provide 83% power to reject the null hypothesis that (Pt - Pc > 0.09) with a one-sided 0.05 significance level for the primary safety endpoint. We further assumed that approximately 10% of the randomized subjects would be excluded from the denominator due to early terminations unrelated to a study device or excluded from the PP population due to major protocol violations. Therefore, the study was designed to include approximately 300 randomized subjects.
In order to prevent bias in the results due to different performance of a single site, no more than 15% of subjects were enrolled at any individual site.
# Endpoint Performance Goals
Study success was achieved when the protocol-defined primary effectiveness endpoint and primary safety endpoint were met, demonstrating noninferiority of the TactiCath Set to the control.
Based on these estimates, a noninferiority margin of 15% for effectiveness (treatment success rate at the end of the 9-month effectiveness assessment follow-up period post blanking) and 9% for safety (incidence of primary SAEs with onset within 7 days of the index procedure or prior to hospital discharge and diagnosed throughout the study) were utilized in the TOCCASTAR study analyses.
# Effectiveness Analyses
## Primary effectiveness endpoint analysis
The one-sided 95% lower confidence limit (LCL) was calculated for the difference between the 2 treatment groups (Pt - Pc) based on the PP population. If the 95% LCL was ≥ -15%, noninferiority was demonstrated in terms of effectiveness.
{14}
Page 15 of 29
# Secondary effectiveness endpoint analysis
For the 3 secondary effectiveness endpoints related to the use of the contact force sensor a superiority test was performed comparing the TactiCath Set to the control device using a hierarchical closed test procedure. It was hypothesized that the use of the contact force sensor information would result in added procedural effectiveness through a reduction in (1) the number of electrically reconnected PVs, (2) the time to achieve total pulmonary vein isolation (PVI), and (3) the total time of RF application. These endpoints were tested sequentially, in the order they appear above, using a one sided, $\alpha = 0.025$ test based on the PP population. Secondary endpoint (1) was tested using the Wilcoxon rank sum test; secondary endpoints (2) and (3) were tested using a log-rank test, stratified by site.
# Safety Analyses
## Primary safety endpoint analysis
For the primary safety endpoint analysis, the one-sided 95% upper confidence limit (UCL) was calculated using the normal approximation method for the difference between the 2 treatment arms (Pt - Pc) based on the SAF population. If the 95% UCL was ≤9%, noninferiority was demonstrated in terms of safety.
## Secondary safety endpoint analysis
The secondary safety endpoint was the incidence of all SAEs during the 12-month follow-up period. All non-primary SAEs were assessed by the CEC for relationship to study device and relationship to study procedure. SAEs related to arrhythmia recurrence and those adjudicated as effectiveness endpoints were described separately.
## D. External Evaluation Groups
Two independent governance committees were utilized for the study to ensure overall integrity of the study conduct and evaluation of study results.
### Clinical Events Committee (CEC)
An independent, standing CEC consisting of members not otherwise associated with the study, reviewed all serious adverse events (SAEs) and the effectiveness endpoints and adjudicated causality relative to the primary safety endpoints. The CEC was blinded to subject treatment assignments. The CEC is comprised of physicians with experience in clinical study event adjudication, including at least 1 physician with expertise in electrophysiology.
### Data Safety Monitoring Board (DSMB)
An independent DSMB comprised of members not otherwise associated with the study was convened to review the proposed research protocol and plans for data and safety monitoring and oversight prior to the start of the study. Thereafter, the DSMB met biannually to provide assessment of the overall study safety and
{15}
advise the sponsor on the continuation of the study. The DSMB members include 3 experts in clinical electrophysiology and ablation as voting members as well as a qualified statistician and medical monitor as nonvoting members.
## Core Laboratories
Three core labs were utilized during the study to evaluate and assess the transtelophonic event monitor (TTM) tracings, 24-hour Holter, and selected computed tomography (CT)/magnetic resonance imaging (MRI) scans.
## E. Study Design Discussion
Subjects undergoing elective catheter radio frequency (RF) ablation for symptomatic PAF who were refractory or intolerant to at least 1 anti-arrhythmic drug (AAD; class I-IV) were screened for enrollment. Subjects who met the study entry criteria and signed the informed consent form (ICF) were enrolled and treated in accordance with the 2007 HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation for Atrial Fibrillation.
Eligible subjects were randomly assigned 1:1 to receive ablation treatment with either the TactiCath Set or the control device. Use of 3-dimensional (3D) mapping equipment was required for both treatment arms. Subjects were blinded to treatment assignment.
After the index procedure, subjects were followed for a total of 12 months for chronic effectiveness assessment, beginning with a 3-month blanking period and ending with a 9-month effectiveness assessment period. During the blanking period, subjects were permitted to be prescribed an AAD and undergo up to 2 reablation procedures (no later than 10 days before the end of the blanking period) using the same device specified by the initial randomization. Subjects were evaluated before discharge; at 7 days; and at 3, 6, and 12 months after the index procedure for the primary study endpoint. After completion of the 12-month visit, subjects continue to be evaluated every 6 months under the current investigational device exemption (IDE) protocol until FDA approval of the Premarket Approval (PMA) application.
## Clinical Inclusion and Exclusion Criteria
Enrollment in the TOCCASTAR study was limited to patients who met the following inclusion criteria:
- A catheter ablation procedure was planned due to symptomatic paroxysmal atrial fibrillation (PAF) that was refractory or intolerant to at least 1 class I-IV AAD
- Minimum of 1 episode of PAF greater than 30 seconds in duration within 12 months prior to enrollment documented by 12-lead ECG, Holter monitor, TTM, telemetry strip, or implanted device
- Minimum of 3 episodes of PAF within the preceding 12 months documented by subject history
- Age 18 years or older
Page 16 of 29
{16}
- Willing and capable of complying unassisted with the study protocol requirements, including all specified follow-up visits
- Provided written informed consent prior to enrollment in the study
Subjects were not permitted enrollment in the TOCCASTAR study if they met any of the following exclusion criteria:
- Persistent or long-standing persistent atrial fibrillation (AF)
- Four or more cardioversions in the last 12 months
- Active systemic infection
- Presence of implantable cardiac defibrillator (ICD)
- Arrhythmia due to reversible causes including thyroid disorders, acute alcohol intoxication, and other major surgical procedures in the preceding 3 months
- Myocardial infarction (MI), acute coronary syndrome, percutaneous coronary intervention (PCI), or valve or coronary bypass grafting surgery within the preceding 3 months
- Left atrial diameter >5.0 cm
- Left ventricular ejection fraction <35%
- New York Heart Association (NYHA) class III or IV (Appendix G of the protocol)
- Previous left atrial ablation procedure, either surgical or catheter ablation
- Previous left atrial surgical procedure or incision with resulting scar
- Previous tricuspid or mitral valve replacement or repair
- Heart disease in which corrective surgery was anticipated within 12 months
- Bleeding diathesis or suspected procoagulant state
- Contraindication to long-term antithromboembolic therapy
- Presence of condition that precluded appropriate vascular access
- Renal failure requiring dialysis
- Known sensitivity to contrast media (if needed during the procedure) not controlled with premedication
- Contraindication to CT and MRI procedures
- Severe pulmonary disease (eg, restrictive pulmonary disease, constrictive or chronic obstructive pulmonary disease) or any other disease or malfunction of the lungs or respiratory system that produces severe chronic symptoms
- Positive pregnancy test results for female subject of childbearing potential
- Other anatomic or comorbid conditions that, in the investigator's opinion, limited the subject's ability to participate in the study or to comply with follow-up requirements, or impacted the scientific soundness of the study results
- Currently participating in another clinical study or had participated in a clinical study within 30 days prior to screening that could have interfered with this study
- Unlikely to survive the protocol follow-up period of 12 months
Page 17 of 29
{17}
# Treatment Procedure and Follow-up Schedule
Table 6 lists the protocol-required baseline, procedural, and follow-up assessments for all subjects.
Table 6. Schedule of Required Assessments
| Assessment | Baseline | Procedure | Predischarge | 7 days ±2 days Telephone | 3 months ±2 weeks | 6 months ±3 weeks | 12 months ±3 weeks | Extended every 6 mos ±4 weeks |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Informed consent | X | | | | | | | |
| Inclusion/exclusion criteria review | X | | | | | | | |
| Documentation of PAF episodes | X | | | | | | | |
| Medical history | X | | | | | | | |
| Thyroid and renal function (standard practice) | X | | | | | | | |
| PT/PTT or INR | X | | X | | | | | |
| Physical examination, vital signs, height, weight^{1} | X | | X | | X | X | X | |
| Pregnancy test (women of childbearing potential) | X | | | | | | | |
| 12-lead ECG | X | | X | | X | X | X | |
| Cardiac medications | X | | X | X | X | X | X | |
| NYHA functional class | X | | | | X | X | X | |
| Quality of life survey | X | | | | | | X | X |
| Randomization | X | | | | | | | |
| Activated clotting time | | X | | | | | | |
| Ablation procedure data | | X | | | | | | |
| Force sensing data (TactiCath arm only) | | X | | | | | | |
| Arrhythmia events | X | | X | X | X | X | X | |
| Adverse events | | X | X | X | X | X | X | X^{6} |
| Protocol deviation | X | X | X | X | X | X | X | |
| TEE (ICE)^{2} | X | (X) | | | | | | |
| TTE (recommended)^{3} | X | | (X) | | | | | |
| CT or MRI scan^{4} | X | | | | X | | | |
| Transtelephonic monitoring^{5} | | | | | X | X | X | |
| Holter monitoring | | | | | | | X | |
| Arrhythmia history | | | | | | | | X |
Abbreviations: CT, computed tomography; ECG, electrocardiogram; ICE, intracardiac echocardiography; INR, international normalized ratio; MRI, magnetic resonance imaging; NYHA, New York Heart Association; PAF, paroxysmal atrial fibrillation; PT, prothrombin time; PTT, partial thromboplastin time; TEE, transesophageal echocardiography; TTE, transthoracic echocardiography; TTM, transtelephonic monitoring.
1 Weight and height were assessed at baseline only.
2 TEE within 24 hours or ICE at time of procedure to exclude atrial thrombus.
3 TTE within 6 months to measure left atrial dimension. Recommended within 24 hours post procedure to exclude pericardial effusion.
4 Not required for supplemental study subjects. If evidence of stenosis was seen at 3 months, the investigator may have prescribed additional serial scans.
5 TTMs were required weekly for months 4 and 5, monthly for months 6-12, and as needed to document symptomatic episodes.
6 Survival and hospitalizations only.
Page 18 of 29
{18}
# F. Accountability of PMA Cohort
The study was initiated on 11 January 2011 at 10 United States (US) and 7 international sites with a total of 317 subjects enrolled, of which 300 were randomized. Subjects were followed for 12 months after the index ablation procedure. The last subject's 12-month visit was completed on 25 June 2013. The database lock date for this report was 02 August 2013. Table 7 provides a summary of the subject disposition for the study.
Table 7. Subject Accountability and Disposition
| Subject Accountability | TactiCath | Control | Total |
| --- | --- | --- | --- |
| Subjects enrolled in study | 172 | 145 | 317 |
| Subjects in Roll-in population | 17 | NA | 17 |
| Full Analysis Population-all subjects randomized (FA) | 155 | 145 | 300 |
| Subjects who had no study device introduced | 3 | 2 | 5 |
| Safety Analysis Population (SAF) | 152 | 143 | 295 |
| Subjects who did not have PVI attempted | 3 | 2 | 5 |
| Modified Intent-to-treat Population (mITT) | 149 | 141 | 290 |
| Subjects excluded from analysis due to major protocol deviation | 3 | 7 | 10 |
| Per Protocol Population (PP) | 146 | 134 | 280 |
# G. Study Population Demographics and Baseline Parameters
Tables 8 and 9 below summarize the demographic information. Subjects were randomized 1:1 upon signing the informed consent.
Table 8. Subject Demographics (SAF Population)
| Demographic | | TactiCath N=152 | Control N=143 | Total N=295 |
| --- | --- | --- | --- | --- |
| Age, years | 59.6 (9.32) | 61.0 (10.84) | 60.3 (10.09) | |
| | | 61.0 [53.0, 66.0] | 62.0 [53.0, 68.0] | 61.0 [53.0, 67.0] |
| | | 31, 78 | 28, 82 | 28, 82 |
| Sex, male | | 100 (65.8%) | 91 (63.6%) | 191 (64.7%) |
| Race | American Indian or Alaska Native | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| | Asian | 1 (0.7%) | 0 (0.0%) | 1 (0.3%) |
| | Black | 1 (0.7%) | 0 (0.0%) | 1 (0.3%) |
| | Native Hawaiian or Pacific Islander | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| | | | | |
{19}
Table 9. Medical History (SAF Population)
| Medical history | TactiCath N=152 | | Control N=143 | | Total N=295 | |
| --- | --- | --- | --- | --- | --- | --- |
| | All history | Currently active | All history | Currently active | All history | Currently active |
| Cardiovascular system | 43 (28.3%) | 31 (20.4%) | 57 (39.9%) | 44 (30.8%) | 100 (33.9%) | 75 (25.4%) |
| Head, eyes, ears, nose, throat | 8 (5.3%) | 4 (2.6%) | 10 (7.0%) | 5 (3.5%) | 18 (6.1%) | 9 (3.1%) |
| Respiratory system | 31 (20.4%) | 26 (17.1%) | 16 (11.2%) | 15 (10.5%) | 47 (15.9%) | 41 (13.9%) |
| Musculoskeletal system | 29 (19.1%) | 10 (6.6%) | 27 (18.9%) | 14 (9.8%) | 56 (19.0%) | 24 (8.1%) |
| Integumentary system | 4 (2.6%) | 2 (1.3%) | 3 (2.1%) | 1 (0.7%) | 7 (2.4%) | 3 (1.0%) |
| Gastrointestinal system | 35 (23.0%) | 16 (10.5%) | 29 (20.3%) | 17 (11.9%) | 64 (21.7%) | 33 (11.2%) |
| Genitourinary system | 23 (15.1%) | 7 (4.6%) | 16 (11.2%) | 4 (2.8%) | 39 (13.2%) | 11 (3.7%) |
| Nervous/neurological system | 7 (4.6%) | 5 (3.3%) | 3 (2.1%) | 2 (1.4%) | 10 (3.4%) | 7 (2.4%) |
| Endocrine system | 13 (8.6%) | 9 (5.9%) | 21 (14.7%) | 13 (9.1%) | 34 (11.5%) | 22 (7.5%) |
| Lymphatic system | 1 (0.7%) | 0 (0.0%) | 3 (2.1%) | 1 (0.7%) | 4 (1.4%) | 1 (0.3%) |
| Immunological system | 3 (2.0%) | 2 (1.3%) | 3 (2.1%) | 2 (1.4%) | 6 (2.0%) | 4 (1.4%) |
| Psychiatric | 6 (3.9%) | 6 (3.9%) | 3 (2.1%) | 3 (2.1%) | 9 (3.1%) | 9 (3.1%) |
| Allergic conditions | 2 (1.3%) | 2 (1.3%) | 2 (1.4%) | 2 (1.4%) | 4 (1.4%) | 4 (1.4%) |
| Other | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| Other | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
Continuous variables are presented as mean (SD), median [IQR], min, max; categorical variables are presented as n (%).
${}^{a}$ Height and weight were not reported for 1 TactiCath subject.
{20}
Page 21 of 29
# Pediatric Patient Population
Atrial fibrillation (AF) is rare in pediatric populations. When observed, the pediatric patient usually has a genetic abnormality and/or disease state that causes enlargement of the heart. Due to the infrequency of AF in pediatric populations, targeting directly to AF is rarely indicated or necessary. Therefore, the TactiCath Quartz Contact Force Ablation Catheter was not studied in patients less than 18 years of age.
Lone AF is defined as occurring in patients with AF who are less than 60 years of age and do not suffer from other forms of cardiovascular disease or hypertension. Lone AF in younger patients is rare and constitutes less than 5% of all types of AF [1].
## H. Procedural Data
Tables 10 and 11 below summarize the index procedural data and the TactiCath contact force results:
Table 10. Index Procedural Data (PP Population)
| Index Procedure data | TactiCath
N=146 | Control
N=134 |
| --- | --- | --- |
| Total fluoroscopy time (minutes) | 30.5 (17.6)
4.0, 131.8 | 27.3 (17.4)
3.8, 105.0 |
| Number of RF applications | 59.7 (32.6)
13.0, 204.0 | 64.7 (42.9)
10.0, 241.0 |
| Total RF time (seconds) | 3103 (1699)
709, 8728 | 3435 (1493)
1036, 10135 |
| Duration of ablation (seconds) | 56 (29)
24, 190 | 74 (62)
14, 464 |
| Power (Watts)ᵃ | 28.8 (4.4)
19.3, 41.1 | 28.5 (4.6)
19.0, 41.0 |
Results are presented as mean (SD), min, max. Includes data from all lesions delivered.
ᵃ sample size for Power, : TactiCath (n=148), Control (n=133).
## I. Safety and Effectiveness Results
### Primary Effectiveness Results
The primary effectiveness endpoint success criterion was met, to demonstrate noninferiority of the TactiCath™ device compared to the control.
Table 11. Summary of Treatment Success: Primary Effectiveness Analysis (PP Population)
| Endpoint | TactiCath
N=146 | Control
N=134 |
| --- | --- | --- |
| Acute procedural success | 146 (100%) | 134 (100%) |
| Acute procedural failure | 0 (0.0%) | 0 (0.0%) |
| Chronic success | 99 (67.8%) | 93 (69.4%) |
| Chronic failure | 47 (32.2%) | 41 (30.6%) |
{21}

Figure 1 below presents the plot for time-to-treatment failure for protocol-defined effectiveness for the PP population.
Figure 1. Kaplan-Meier Plot for Time-to-Treatment Failure (PP Population)

Figure 2 below presents the Kaplan-Meier plot for time-to-treatment failure due to the reablation procedure in the PP population. As shown, over time, more subjects in the control arm experienced reablation than in the TactiCath™ arm.
Figure 2. Kaplan-Meier Plot - Time-to-Treatment Failure Due to Re-Ablation Procedures (PP Population)
{22}
# Secondary Effectiveness Results
Table 12 provides a summary of the procedural efficiency. The descriptive effectiveness endpoints in the study were related to the use of the contact force sensor and demonstrated a significant reduction in mean total RF application time with the TactiCath Set versus the control (51.7 vs 58.3 min, $P = 0.002$ ).
Table 12. Summary of Procedural Efficiency (PP Population)
| Secondary effectiveness endpoint | TactiCath N=146 | Control N=134 | P value (1-sided) |
| --- | --- | --- | --- |
| Percentage of lesion sets with electrically reconnected PVs following a 30-minute waiting period assessed by entrance block (%) at index procedure | n=145 11.09 (22.738) 0.00 [0.00, 0.00] 0.0, 100.0 | n=134 14.61 (27.751) 0.00 [0.00, 25.00] 0.0, 100.0 | 0.209 |
| Time to achieve initial PV isolation at index procedure (minutes) | n=146 90.4 (43.71) 87.0 [55.0, 117.0] 16, 261 | n=134 86.4 (38.67) 85.5 [58.0, 108.0] 18, 224 | 0.941 |
| Total time of RF application at index procedure (minutes) | n=146 51.72 (27.671) 46.40 [28.30, 65.30] 13.0, 139.3 | n=133 58.31 (27.333) 53.50 [40.80, 72.80] 17.1, 196.0 | 0.002 |
Results are presented as mean (SD), median [IQR], min, max.
# Primary Safety Results
The primary safety endpoint of noninferiority when compared to the control device was met. The predefined analysis for the primary safety endpoint is based on the SAF cohort of 295 subjects (Tables 13 and 14).
Table 13. Primary Safety Endpoint (SAF Population)
| Primary safety endpoint | TactiCath N=152 | Control N=143 | Difference in primary SAE rates (TactiCath minus control) | |
| --- | --- | --- | --- | --- |
| | | | Point estimate | One-sided 95% UCLa |
| Subjects experienced primary SAEs | 3 (2.0%) | 2 (1.4%) | 0.6% | 3.0% |
a Noninferiority of TactiCath to the control device is demonstrated if the $95\%$ UCL for the difference in rates (TactiCath minus control) is less than 9
Table 14. Summary of Primary SAEs by Term (SAF Population)
| Event | TactiCath N=152 | | Control N=143 | | Total N=295 | |
| --- | --- | --- | --- | --- | --- | --- |
| | Subjects (%) | Events | Subjects (%) | Events | Subjects (%) | Events |
| Atrial perforation | 0 (0.0%) | 0 | 1 (0.7%) | 1 | 1 (0.3%) | 1 |
| Cardiac perforation/cardiac tamponade | 1 (0.7%) | 1 | 0 (0.0%) | 0 | 1 (0.3%) | 1 |
| Pericarditis | 2 (1.3%) | 2 | 0 (0.0%) | 0 | 2 (0.7%) | 2 |
| PV stenosis | 0 (0.0%) | 0 | 1 (0.7%) | 2 | 1 (0.3%) | 2 |
| Total subjects with at least 1 primary SAE | 3 (2.0%) | 3 | 2 (1.4%) | 3 | 5 (1.7%) | 6 |
{23}
# Secondary Safety Results
The secondary safety endpoint consists of an evaluation of all SAEs during the 12-month follow-up period. A total of 32 (10.8%) subjects experienced 39 safety-related SAEs in the study. Six (6) of these SAEs were primary safety events, the other 33 SAEs were non-primary events and are summarized in the Tables 15 and 16 below.
Table 15. Procedure-Related, Non-Primary SAEs (SAF Population)
| | TactiCath
N=152 | | Control
N=143 | | Total
N=295 | |
| --- | --- | --- | --- | --- | --- | --- |
| Serious adverse event | Subjects
n (%) | Events | Subjects
n (%) | Events | Subjects
n (%) | Events |
| Total number (%) subjects with at least 1 non-primary SAE^{a} | 10 (6.6) | 11 | 11 (7.7) | 11 | 21 (7.1) | 22 |
| Atrial perforation^{b} | 0 | 0 | 1 (0.7) | 1 | 1 (0.3) | 1 |
| Cardiac perforation/cardiac tamponade^{b} | 1 (0.7) | 1 | 0 | 0 | 1 (0.3) | 1 |
| Pulmonary edema | 2 (1.3%) | 2 | 2 (1.4) | 2 | 4 (1.4) | 4 |
| Vascular access complications | 3 (2.0) | 3 | 3 (2.1) | 3 | 6 (2.0) | 6 |
| Hospitalizations (initial or prolonged) | 4 (2.6) | 4 | 5 (3.5) | 5 | 9 (3.1) | 9 |
| GI bleeding | 1 (0.7) | 1 | 0 | 0 | 1 (0.3) | 1 |
| Anemia | 0 | 0 | 1 (0.7) | 1 | 1 (0.3) | 1 |
| Hematuria | 0 | 0 | 1 (0.7) | 1 | 1 (0.3) | 1 |
| Rule out esophageal injury | 0 | 0 | 1 (0.7) | 1 | 1 (0.3) | 1 |
| Secondary to groin hematoma | 1 (0.7) | 1 | 0 | 0 | 1 (0.3) | 1 |
| Secondary to pneumonia | 0 | 0 | 1 (0.7) | 1 | 1 (0.3) | 1 |
| Secondary to small bowel obstruction | 1 (0.7) | 1 | 0 | 0 | 1 (0.3) | 1 |
| Secondary to urinary retention | 0 | 0 | 1 (0.7) | 1 | 1 (0.3) | 1 |
| Vascular access complication | 1 (0.7) | 1 | 0 | 0 | 1 (0.3) | 1 |
| Non-primary SAE (other) | 1 (0.7) | 1 | 0 | 0 | 1 (0.3) | 1 |
| Secondary to cardiac arrest | 1 (0.7) | 1 | 0 | 0 | 1 (0.3) | 1 |
a Subjects reporting more than one SAE in each level are only counted once. Only SAEs adjudicated as non-primary SAE are included
b Was not related to the use of a study device
Page 24 of 29
{24}
Table 16. Not Procedure Related, Non-Primary SAEs (SAF Population)
| | TactiCath
N=152 | | Control
N=143 | | Total
N=295 | |
| --- | --- | --- | --- | --- | --- | --- |
| Serious adverse event | Subjects
n (%) | Events | Subjects
n (%) | Events | Subjects
n (%) | Events |
| Total number (%) subjects with at least 1 non-primary SAE^{a} | 6 (3.9) | 7 | 5 (3.5) | 5 | 11 (3.7) | 12 |
| Hospitalizations(initial or prolonged) | 3 (2.0) | 4 | 2 (1.4) | 2 | 5 (1.7) | 6 |
| Pericardial effusion at redo (not with study device) | 1 (0.7) | 1 | 0 | 0 | 1 (0.3) | 1 |
| Coronary artery disease | 1(0.7) | 1 | 1 (0.7) | 1 | 2 (0.7) | 2 |
| Mitral valve incompetence | 1 (0.7) | 1 | 0 | 0 | 1 (0.3) | 1 |
| Pneumonia | 0 | 0 | 1 (0.7) | 1 | 1 (0.3) | 1 |
| Pulmonary hypertension | 1 (0.7) | 1 | 0 | 0 | 1 (0.3) | 1 |
| Non-primary SAE (other) | 3 (2.0) | 3 | 3 (2.1) | 3 | 6 (2.0) | 6 |
| Pacemaker implant | 0 | 0 | 1 (0.7) | 1 | 1 (0.3) | 1 |
| Redo during blanking | 3 (2.0) | 3 | 2 (0.7) | 2 | 5 (1.7) | 5 |
aSubjects reporting more than one SAE in each level are only counted once. Only SAEs adjudicated as non-primary SAE are included
## TactiCath™ Set Contact Force Results
The TactiCath™ Set generates a summary of contact force parameters applied during the ablation procedure. A statistical summary of the distribution of contact force averaged over all subjects in the TactiCath™ arm is presented in Table 17.
Table 17. Distribution of Contact Force in the TactiCath Arm (PP Population)
| Contact Force Distribution | TactiCath
N=146^{a} |
| --- | --- |
| Number of ablations | n=145
61.1 (32.97)
55.0 [37.0, 76.0]
17, 204 |
| Average contact force (g) | n=144
22.2 (7.47)
22.0 [17.0, 27.0]
7, 44 |
| Standard deviation of contact force (per procedure) (g) | n=144
12.3 (9.88)
11.0 [8.0, 14.5]
3, 113 |
| Minimum contact force (g) | n=144
4.2 (3.46)
4.0 [2.0, 6.0]
0, 16 |
| 5^{th} Percentile average contact force (g) | n=144
7.8 (4.38)
7.0 [5.0, 10.0]
0, 30 |
Page 25 of 29
{25}
| Contact Force Distribution | TactiCath N=146a |
| --- | --- |
| 95thPercentile average contact force (g) | n=144 42.6 (14.99) 42.0 [33.0, 53.0] 2, 80 |
| Maximum average contact force (g) | n=144 66.5 (95.35) 53.0 [39.5, 69.5] 17, 1111 |
Results are presented as mean (SD), median [IQR], min, max.
aFor those subjects with evaluable data
Previous studies have suggested that optimizing contact force during lesion delivery yields better acute and chronic outcomes. Recent analysis of study subjects experiencing reablation procedures demonstrated a strong correlation between contact force parameters and site of reconnection. A significant determinant of durable PV isolation was the number of low contact force lesions.
A post hoc analysis of treatment success for all TactiCath™ subjects in the PP population who were treated with greater than 10 g of contact force (CF) in at least 90% of all lesions (n=83) is presented in Table 18.
Table 18. Success Rates for TactiCath™ Subjects with a High and Low Percentage of Lesions Above 10g (PP Population)
| Contact force | Subjects (n) | Success (%) | P value |
| --- | --- | --- | --- |
| Optimal (≥90% of lesions with CF >10 g) | 83 | 75.9% | 0.018 |
| Suboptimal (<90% of lesions with CF >10 g) | 62 | 58.1% | |
Using protocol-defined criteria for success, TactiCath subjects who were treated with "optimal" contact force ( $\geq 90\%$ of lesions with CF $>10\mathrm{g}$ ) were significantly more successful (75.9% vs 58.1%, P=0.018) than those treated with suboptimal contact force.
Subjects treated with optimal contact force who failed after 12 months $(n = 20)$ tended to fail for TTM and/or AAD use, and less for reablation procedures as shown in Table 19.
Table 19. Failure mode for TactiCath subjects treated with optimal contact force
| Reason for failure | Subjects with failure mode (n) |
| --- | --- |
| Acute failure | 0 |
| Reablation only | 1 |
| AAD only | 4 |
| TTM/ECG only | 8 |
| Reablation + AAD | 0 |
| Reablation + TTM/ECG | 1 |
| Reablation + AAD + TTM/ECG | 2 |
| AAD + TTM/ECG | 4 |
| Total | 20 |
{26}
Control subjects had a higher failure rate due to reablation procedures after the blanking period compared to TactiCath subjects (12.7% vs 7.5%, respectively). Only 4 of the 83 (4.8%) subjects treated with optimal contact force required treatment with repeat ablation during the effectiveness assessment period compared to 17 of 134 (12.7%) subjects in the control arm ( $P = 0.044$ ).
## Quality of Life Results
Quality of Life was measured at baseline and at 12 month follow-up, using the 'Atrial Fibrillation Effect on QualiTy-of-life' (AFEQT) questionnaire. From the 280 patients in the Per-Protocol (PP) Population, 254 patients answered to both questionnaires at baseline and 12 months. The results are represented Figure 3 below.

AFEQT score increase from Baseline to 12 Months - SAF
Figure 3. Atrial Fibrillation Effect of QualiTy-of-life Score from Baseline to 12 months (SAF Population)
## Gender Subgroup Analysis
A summary of treatment success by gender is provided in Table 20 below.
Table 20. Treatment Success (Protocol-Defined Effectiveness) by Gender (PP Population)
| Endpoint | Male | | Female | | Difference in success rates (TactiCath minus control) | |
| --- | --- | --- | --- | --- | --- | --- |
| | TactiCath N=97 | Control N=86 | TactiCath N=49 | Control N=48 | Male point estimate % [95% LCL] | Female point estimate % [95% LCL] |
| Acute procedural success | 97 (100%) | 86 (100%) | 49 (100%) | 48 (100%) | 0 [NA] | 0 [NA] |
| Acute procedural failure | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) | | |
{27}
| Chronic success | 66 (68.0%) | 56 (65.1%) | 33 (67.3%) | 37 (77.1%) | 2.9 [-8.6] | -9.7 [-24.6] |
| --- | --- | --- | --- | --- | --- | --- |
| Chronic failure | 31 (32.0%) | 30 (34.9%) | 16 (32.7%) | 11 (22.9%) | | |
Endpoint success/failures are reported as n (%), point estimate is reported as % [95% LCL].
## J. Financial Disclosure
The Financial Disclosure by Clinical Investigators regulation (21 CFR Part 54), requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation.
The clinical investigation included 104 investigators, of which none were full-time or part-time employees of the sponsor. Four investigators had disclosable financial interests/arrangements as defined in 21 CFR 54.2(a), (b), (c) and (f) and described below:
- Compensation to the investigator for conducting the study where the value could be influenced by the outcome of the study: no investigators.
- Significant equity interest held by investigator in sponsor of covered study: two investigators.
- Proprietary interest in the product tested held by the investigator: one investigator.
- Significant payment of other sorts: two investigators.
The applicant has adequately disclosed the financial interest/arrangements with clinical investigators. Statistical analyses were conducted by FDA to determine whether the financial interests/arrangements had any impact on the clinical study outcome. The information provided does not raise any questions about the reliability of the data.
## XI. PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Circulatory System Devices Advisory Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XII. CONCLUSIONS DRAWN FROM PRECLINICAL AND CLINICAL STUDIES
### A. Safety Conclusions
The adverse events of the device are based on data collected in the clinical study conducted to support PMA approval of the TactiCath™ Quartz Set as described
Page 28 of 29
{28}
above. The TOCCASTAR pivotal study demonstrated that the TactiCath™ Set had met the safety goal for its intended use as defined in the protocol.
## B. Effectiveness Conclusions
The TOCCASTAR pivotal study demonstrated that the TactiCath™ Set had met the efficacy performance goal for its intended use as defined in the protocol.
## C. Benefit-Risk Conclusion
The pre-clinical and clinical information presented support that the probable benefits outweigh the probable risks when the device is used for the treatment of drug refractory recurrent symptomatic paroxysmal atrial fibrillation when used in accordance with the product labeling.
## D. Overall Conclusions
The data in this application supports the reasonable assurance of safety and effectiveness of the TactiCath™ Quartz Set when used according to the product labeling.
## XIII. CDRH DECISION
CDRH issued an approval order on 10/24/2014. The final conditions of approval cited in the approval order are described below.
The applicant's manufacturing facilities were inspected and found to be in compliance with the device Quality System (QS) regulation (21 CFR 820).
## XIV. APPROVAL SPECIFICATIONS
Directions for use: See device labeling
Hazards to Health from Use of the Device: See Indications, Contraindications, Warnings, Precautions, and Adverse Events in the device labeling.
## XV. REFERENCES
1. "Lone Atrial Fibrillation in the Pediatric Population," by Lindsay C. Mills, MD, Robert M. Gow, MBBS, Kim Myers, MD, Michal J. Kantoch, MD, Gil J. Gross, MD, Anne Fournier, MD, and Shubhayan Sanatani, MD, Canadian Journal of Cardiology, Volume 29/Issue 10 (October 2013), DOI: http://dx.doi.org/10.1016/j.cjca.2013.06.014, published by Elsevier
Page 29 of 29