SHANGHAI GLOVMAX ENTERPRISE CO., LTD. PRE-POWDERED VINYL PATIENT EXAMINATION GLOVES

K993859 · Shanghai Glovmax Enterprise Co.,Ltd · LYZ · Jan 12, 2000 · General Hospital

Device Facts

Record IDK993859
Device NameSHANGHAI GLOVMAX ENTERPRISE CO., LTD. PRE-POWDERED VINYL PATIENT EXAMINATION GLOVES
ApplicantShanghai Glovmax Enterprise Co.,Ltd
Product CodeLYZ · General Hospital
Decision DateJan 12, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Pre-powdered vinyl patient examination glove; disposable barrier device. Worn on examiner's hand or finger; prevents contamination between healthcare personnel and patient body, fluids, waste, or environment. Used in clinical settings by healthcare personnel. Provides physical barrier protection; reduces risk of cross-contamination. Manufactured to ASTM D5250-92 standards.

Clinical Evidence

No clinical data. Bench testing only: physical and dimensions testing (ASTM D5250-92, AQL 4.0), 1000 ml water fill test (AQL 2.5), and biocompatibility testing (primary skin irritation and sensitization) showing no adverse reactions.

Technological Characteristics

Vinyl material; pre-powdered with USP absorbable dusting powder. Dimensions and physical properties meet ASTM D5250-92. Biocompatible. Non-sterile. Class I device.

Indications for Use

Indicated for use by healthcare personnel as a disposable barrier worn on the hand or finger to prevent cross-contamination between the examiner and the patient.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SHANGHAI GLOVMAX ENTERPRISE CO., LTD. Song Jiang Industrial Zone, Che Dun Branch Zone, Rong Chang Rd., #2, Shanghai, China C/O: Room B, 11F, No. 201, Nanking E. Rd., Sec.3, Taipei, Taiwan, R.O.C. JAN 1 2 2000 TEL: 886-2-25462479 FAX: 886-2-25147612 K 993859 ### 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. The assigned 510(k) number is: ### 1.Submitter's Identification: Mr. Mao-Sheng Lee SHANGHAI GLOVMAX ENTERPRISE CO., LTD. Song Jiang Industrial Zone, Che Dun Branch Zone, Rong Chang Rd., #2, Shanghai, China Date Summary Prepared: Nov. 5, 1999 ### 2. Name of the Device: SHANGHAI GLOVMAX ENTERPRISE CO., LTD. Pre-powdered Vinyl Examination Gloves ### 3. Predicate Device Information: Shanghai Huamao Gloves Co., Ltd. #K984461 ### 4. Device Description: Classified by FDA's General and Plastic Surgery Device Panel as Class I, 21 CFR 880.6250, Pre-powder Vinyl Patient Examination Glove, 80LYZ , powdered with Absorbable Dusting Powder, USP, Class III and meets all requirements of ASTM Standard D5250-92. ### 5.Intended Use: A glove is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment. {1}------------------------------------------------ # SHANGHAI GLOVMAX ENTERPRISE CO., LTD. Song Jiang Industrial Zone, Che Dun Branch Zone, Rong Chang Rd., #2, Shanghai, China C/O: Room B, 11F, No. 201, Nanking E. Rd., Sec.3, Taipei, Taiwan, R.O.C. > TEL: 886-2-25462479 FAX: 886-2-25147612 #### 6. Comparison to Predicate devices: SHANGHAI GLOVMAX ENTERPRISE CO., LTD. Pre-powdered Vinyl Patient Examination Gloves, is substantially equivalent in safety and effectiveness to the Shanghai Huamao Gloves Co., Ltd. Pre-powdered vinyl patient examination Gloves. # 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: The standards used for SHANGHAI GLOVMAX ENTERPRISE CO., LTD. Production are based on ASTM D-5250-92. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0. The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, Inspection Level S-4, meeting these requirements. Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions. There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels. SHANGHAI GLOVMAX ENTERPRISE CO., LTD. Operates in compliance with FDA's GMPs. ### 8. Discussion of Clinical Tests Performed: Not Applicable - There is not hypoallergenic claim. ### 9.Conclusions: SHANGHAI GLOVMAX ENTERPRISE CO., LTD. Vinyl Patient Examination Gloves conform fully to ASTM D-5250-92 standards as well as applicable 21 CFR references, and meets pinhole FDA requirements, bio-compatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the " substantial equivalence" products cited. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 2 2000 Mr. Mao-Sheng Lee, President Shanghai Glovmax Enterprise Co., Ltd. Room B, 11th Floor, No. 201 Nanking E. Road, Sec. 3 Taipei, Taiwan, R.O.C. Re: K993859 Shanghai Glovmax Enterprise Co., Ltd. Pre-Trade Name: Powdered Vinyl Patient Examination Gloves Regulatory Class: I Product Code: LYZ November 5, 1999 Dated: November 15, 1999 Received: Dear Mr. Lee: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any {3}------------------------------------------------ Page 2 - Mr. Lee obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Timothy A. Ulatowski Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # ATTACHMENT A Page _ 1__ of _ 1 _ **510(k) number (if known):** K993859 Shanghai Glovmax Enterprise Co., Ltd. Device name: Pre-powdered Vinyl Examination Gloves ## Indications For Use: A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | | |----------------------|--| | (Per 21 CFR 801.109) | | OR Over-The-Counter Use (Optional Format 1-2-96) Clive S. Lin (Division Sign-Off) Division of Dental, Infection Control, and General Hospit 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...