TERUMO U-100 INSULIN SYRINGE

K992802 · Terumo Medical Corp. · FMF · Sep 9, 1999 · General Hospital

Device Facts

Record IDK992802
Device NameTERUMO U-100 INSULIN SYRINGE
ApplicantTerumo Medical Corp.
Product CodeFMF · General Hospital
Decision DateSep 9, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TERUMO® U-100 Insulin Syringe, with fixed hypodermic single lumen needle, is a device intended for medical purposes for the manual aspiration of fluids, and for the injection of fluids into parts of the body below the surface of the skin. This device is intended particularly for the aspiration and injection of insulin.

Device Story

Terumo U-100 Insulin Syringe is a sterile, single-use, manual piston syringe with a fixed 30-gauge by 3/8-inch hypodermic single lumen needle. Available in 1/2cc and 3/10cc volumes. Operated manually by clinicians or patients for subcutaneous insulin delivery. Device functions via manual aspiration and injection. Biocompatible materials; validated for sterility per EN 556. Intended for insulin administration.

Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO-10993. Sterilization validated per EN 556 to SAL 10-6. Functional and sterility testing performed on production lots.

Technological Characteristics

Manual piston syringe; 30G x 3/8" fixed hypodermic single lumen needle; 1/2cc and 3/10cc volumes. Biocompatible materials per ISO-10993. Sterilization via validated process (EN 556).

Indications for Use

Indicated for manual aspiration and injection of fluids, specifically insulin, into subcutaneous tissue.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP g 1999 # SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION PERTAINING TO SUBSTANTIAL EQUIVALENCE TERUMO® U-100 Insulin Syringe, or similar proprietary name Device Name: Piston syringe with fixed hypodermic single lumen needle Classification Name: ## INTENDED USE The TERUMO® U-100 Insulin Syringe, with fixed hypodermic single lumen needle, is a device intended for medical purposes for the manual aspiration of fluids, and for the injection of fluids into parts of the body below the surface of the skin. This device is intended particularly for the aspiration and injection of insulin. ## DESCRIPTION The Terumo U-100 Insulin Syringe is a sterile, single use piston syringe with a fixed hypodermic single lumen needle, designed for manual use. The syringe is available in 1/2cc and 3/10cc volumes with a 30 gauge by 3/8 inch fixed hypodermic single lumen needle. ## SUBSTANTIAL EQUIVALENCE The Terumo U-100 Insulin Syringe submitted in this 510k, is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared Terumo Medical Corporation U-100 Insulin syringe (K822083). There is no intended change in how this device is used. With respect to the 30G x 3/8" cannula, the Terumo U-100 Insulin Syringe submitted in this 510k is substantially equivalent to the cleared Becton Dickinson® Ultra Fine II Insulin Syringe with 30G x 5/16" cannula (K955235). ## PRINCIPLE OF OPERATION/TECHNOLOGY The Terumo U-100 Insulin Syringe is operated manually. ## MATERIALS The materials used in the manufacture of the Terumo U-100 Insulin Syringe are the same materials used in the manufacture of the predicate device. ## PERFORMANCE The performance of the TERUMO® U-100 Insulin Syringe submitted in this 510k is equivalent to the performance of the cleared TERUMO U-100 Insulin Syringe (K822083) and to the Becton-Dickinson® Ultra Fine II Insulin Syringe (K955235). {1}------------------------------------------------ Testing was performed to demonstrate the substantial equivalence of the TERUMO U-100 Insulin Syringe with 30G x 3/8" cannula to the cleared Terumo U-100 Insulin Syringe and the B-D Ultra Fine II Insulin Syringe with 30G x 5/16" cannula. #### ADDITIONAL SAFETY INFORMATION Sterilization conditions have been validated according to the European Standard, EN 556: Sterilization of Medical Devices - Requirements for Medical Devices to Be Labeled Sterile, to provide a Sterility Assurance Level (SAL) of 106. Manufacturing control test methods include functional and sterility tests. LAL testing is performed on production samples from of every lot number. Blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." [External Communicating Devices, Blood Path Indirect, Limited Duration of Contact ( < 24 hours).] The blood contacting materials were found to be biocompatible. The expiration dating for the Terumo U-100 Insulin Syringe has been established to be 5 years, or 60 months. ## CONCLUSION The Terumo U-100 Insulin Syringe with 30 gauge x 3/8 inch cannula submitted in this 510k is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared Terumo U-100 Insulin Syringe (K822083), and the B-D Ultra Fine II Insulin Syringe with 30 gauge x 5/16 inch cannula. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness. Terumo's statement that these devices are substantially equivalent to any other device is done solely to comply with the requirements of the Federal Food, Drug and Cosmetic Act and is not intended whatsoever to be the basis for a patent infringement action. | Date Prepared: | July 30, 1999 | |----------------|----------------------------------------| | Prepared by: | Kristine Wagner | | | Regulatory Affairs Specialist | | Prepared for: | Terumo Medical Corporation | | | 125 Blue Ball Road | | | Elkton, MD 21921 | | | Phone (410) 392-7241 or (410) 392-7231 | | | Fax (410) 398-6079 | {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP g 1999 Ms. Kristine Wagner Requlatory Affairs Specialist Terumo Medical Corporation Regulatory Affairs Department 125 Blue Ball Road Elkton, Maryland 21921 Re : K992802 Terumo U-100 Insulin Syringe Trade Name: Requlatory Class: I Product Code: FMF Dated: August 18, 1999 Received: August 19, 1999 Dear Ms. Wagner: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of {3}------------------------------------------------ Page 2 -Ms. Wagner the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4690. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely Yours Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): K992802 #### TERUMO® U-100 INSULIN SYRINGE Device Name: #### Indications For Use: The TERUMO® U-100 Insulin Syringe, with fixed hypodermic single lumen needle, is a device intended for medical purposes for the manual aspiration of fluids, and for the injection of fluids into parts of the body below the surface of the skin. This device is intended particularly for the aspiration and injection of insulin. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of/CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|---------| | (Division Sign-Off) | | | Division of Dental, Infection Control, | | | and General Hospital Devices | | | 510(k) Number | K992802 | Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use V
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...