IMMULITE METHAMPHETAMINE, MODEL LKMA1, LKMA5

K992634 · Diagnostic Products Corp. · LAF · Jan 20, 2000 · Clinical Toxicology

Device Facts

Record IDK992634
Device NameIMMULITE METHAMPHETAMINE, MODEL LKMA1, LKMA5
ApplicantDiagnostic Products Corp.
Product CodeLAF · Clinical Toxicology
Decision DateJan 20, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3610
Device ClassClass 2

Intended Use

IMMULITE® Methamphetamine is a solid-phase, chemiluminescent enzyme immunoassay for use with the IMMULITE Automated Analyzer and designed for the qualitative measurement of methamphetamine in urine. It is intended strictly for in vitro diagnostic use in clinical laboratories, in the context of a program involving an established confirmatory test for methamphetamine. A cutoff of 500 ng/mL is used by the IMMULITE Methamphetamine assay to identify positive and negative results.

Device Story

IMMULITE® Methamphetamine is a solid-phase, chemiluminescent enzyme immunoassay used with the IMMULITE Automated Analyzer. It detects methamphetamine in urine samples. The device uses a polystyrene bead coated with monoclonal antibodies specific to methamphetamine. Patient samples and alkaline phosphatase-conjugated methamphetamine are incubated with the bead; methamphetamine in the sample competes for antibody binding sites. After a centrifugal wash, a phosphate ester of adamantyl dioxetane substrate is added. Alkaline phosphatase hydrolyzes the substrate, producing light. The luminometer measures photon output, which is inversely proportional to the methamphetamine concentration in the sample. Results are determined by comparing the sample's light output to an 'Adjustor' sample representing the 500 ng/mL cutoff. The device is operated by laboratory personnel in a clinical setting. Results assist clinicians in diagnosing and treating methamphetamine use or overdose and monitoring therapy. It provides a qualitative positive/negative result relative to the established cutoff.

Clinical Evidence

Method comparison study performed on 214 urine samples comparing IMMULITE Methamphetamine to the Coat-A-Count (CAC) predicate, showing 97% agreement (92% relative sensitivity, 100% relative specificity). A second comparison against GC/MS on 37 samples showed 87% agreement (88% sensitivity, 86% specificity).

Technological Characteristics

Solid-phase chemiluminescent enzyme immunoassay. Components: polystyrene bead coated with monoclonal anti-methamphetamine antibody, alkaline phosphatase-conjugated methamphetamine, and phosphate ester of adamantyl dioxetane substrate. Operates on the IMMULITE Automated Analyzer. Detection principle: competitive binding followed by chemiluminescent signal measurement via luminometer. Qualitative output based on 500 ng/mL cutoff.

Indications for Use

Indicated for the qualitative measurement of methamphetamine in urine for clinical laboratory use. Used in the context of a program involving an established confirmatory test for methamphetamine. Patient population includes individuals being screened for methamphetamine use or overdose.

Regulatory Classification

Identification

A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.

Special Controls

*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K992634 JAN 2 0 2000 ## 510 (k) Summary Safety and Effectiveness This summary of safety and effectiveness information has been prepared in accordance This sammary of SMDA 1990 and 21 CFR Part 807.92. | Name: | Diagnostic Products Corporation | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | 5700 West 96th Street<br>Los Angeles, CA 90045 | | Telephone Number: | (310) 645-8200 | | Facsimile Number: | (310) 645-9999 | | Contact Person: | Edward M. Levine, Ph.D.<br>Director of Clinical Affairs | | Date of Preparation: | November 29, 1999 | | Device Name: | | | Trade: | IMMULITE® Methamphetamine | | Catalog Number: | LKMA1 (100 tests), LKMA5 (500 tests) | | CFR: | An amphetamine test system is a device intended to<br>measure amphetamine, a central nervous system<br>stimulating drug, in plasma and urine. Measurements<br>obtained by this device are used in the diagnosis and<br>treatment of amphetamine use or overdose and in<br>monitoring levels of amphetamine to ensure appropriate<br>therapy. | | Common: | Reagent system for the determination of methamphetamine<br>in urine. | | Classification: | Class II device, 91-DKZ (21 CFR 862.3100) | | Panel: | Toxicology | | CLIA Complexity Category: | We believe the category to be moderate, based on previous<br>classification of analogous tests. | | Manufacturer: | Diagnostic Products Corporation (DPC)<br>5700 West 96th Street<br>Los Angeles, CA 90045-5597 | {1}------------------------------------------------ Establishment Registration #: DPC's establishment Registration No. is 2017183 | Substantially Equivalent<br>Predicate Device: | DPC's Coat-A-Count Methamphetamine (K885071) | |-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Description of Device: | IMMULITE® Methamphetamine is a solid-phase,<br>chemiluminescent enzyme immunoassay for use with the<br>IMMULITE® Automated Analyzer. | | Intended Use of the<br>Device: | IMMULITE® Methamphetamine is a solid-phase,<br>chemiluminescent enzyme immunoassay for use with the<br>IMMULITE Automated Analyzer and designed for the<br>qualitative measurement of methamphetamine in urine. It<br>is intended strictly for <i>in vitro</i> diagnostic use in clinical<br>laboratories, in the context of a program involving an<br>established confirmatory test for methamphetamine. A<br>cutoff of 500 ng/mL is used by the IMMULITE<br>Methamphetamine assay to identify positive and negative<br>results. | #### Summary and Explanation of the test: Methamphetamine is a potent sympathomimetic amine capable of stimulating the central nervous system with less peripheral actions than amphetamine. It is structurally similar to amphetamine and methylphenidate and, like them, acts through increasing the release of norepinephrine as well as through direct stimulation of postsynaptic norepinephrine receptors. There are two stereoisomers: d-methamphetamine is 10 times more potent as a CNS stimulant than I-methamphetamine, while I-methamphetamine (I-desoxyephedrine) is used in nasal inhalers for peripheral vasoconstriction Amphetamines induce euphoria, irritability and anxiety. Tolerance and dependence have been observed. Methamphetamine is metabolized to the active metabolite, amphetamine, hydroxylated metabolites and deaminated to hippuric and benzoic acid. It is a metabolite of benzphetamine and selegiline. About 70% of a dose is excreted in the urine in 24 hours. Methamphetamine is excreted primarily unchanged (44% to 76%) with a small fraction as 4-hydroxymethamphetamine (15%) or amphetamine (6%). The half-life of methamphetamine varies from 12 to 34 hours and is a function of urinary pH. The excretion of unchanged methamphetamine is increased in acidic urine, while in alkaline urine it is reduced to less than 5% of the dose. After a single 10 mg dose, urine concentrations ranged from 0.5 to 4.0 µg/mL in the first 24 hours. After chronic administration, methamphetamine concentrations of 25 to 300 µg/mL and amphetamine concentrations of 1 to 90 ug/mL are found in urine. However, the length of time {2}------------------------------------------------ following drug use for which a positive result may occur is dependent upon several factors including the frequency and amount of drug, metabolic rate, excretion rate, drug half-life, and the drug user's age, weight, activity, and diet. ### Technological Comparison to Predicate: Methamphetamine is a solid-phase, chemiluminescent enzyme IMMULITE immunoassay. The solid-phase, a polystyrene bead enclosed within an IMMULITE Test Unit, is coated with a monoclonal antibody specific for methamphetamine. The patient sample and alkaline phosphatase-conjugated methamphetamine are simultaneously introduced into the Test Unit and incubated for approximately 30 minutes at 37 ℃ with intermittent agitation. During this time, methamphetamine in the sample competes with enzyme-labeled methamphetamine for a limited number of antibody binding sites on the bead. Unbound enzyme conjugate is then removed by a centrifugal wash, after which substrate is added and the Test Unit is incubated for a further 10 minutes. The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to vield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving precision by providing a window for multiple readings. The bound complex and thus also the photon output, as measured by the luminometer - is inversely proportional to the concentration of methamphetamine in the sample. A qualitative result is then obtained by comparing the counts per second (cps) of the patient sample to those of a sample - the Adjustor supplied with the kit - representing the assay's 500 ng/mL cutoff. The Coat-A-Count Methamphetamine procedure is a solid-phase radioimmunoassay, wherein 1331-labeled methamphetamine competes for a fixed time with methamphetamine in the patient sample for methamphetamine-specific antibody sites. Because the antibody is immobilized to the wall of a polypropylene tube, simply decanting the supernatant suffices to terminate the competition and to isolate the antibody-bound fraction of the radiolabeled methamphetamine, which is then counted in a gamma counter. The Qualitative Procedure vields positive and negative results relative to a reference chosen as a cutoff. {3}------------------------------------------------ #### Performance Equivalence: Diagnostic Products Corporation asserts that the IMMULITE® Methamphetamine substantially equivalent results to other commercially marketed produces methamphetamine assays, such as the Coat-A-Count Methamphetamine assay. Each product is designed for the qualitative measurement of methamphetamine in urine. Each product is intended strictly for in vitro diagnostic use in the context of a program involving an established confirmatory test for methamphetamine. #### Method Comparison: IMMULITE Methamphetamine was compared to DPC's Coat-A-Count (CAC) Methamphetamine on 214 urine samples from volunteer donors, presumed not to be drug abusers, and a reference lab. The samples ranged from undetectable to approximately 1500 ng/mL (3 times the cutoff value). A cutoff of 500 ng/mL was used for both procedures. #### IMMULITE Methamphetamine | CAC Methamphetamine | Positive | Negative | Relative sensitivity: | 92% | |---------------------|----------|----------|-----------------------|------| | Positive | | | Relative Specificity: | 100% | | Negative | | ાર્ ર | Agreement: | 97% | IMMULITE Methamphetamine was also compared to GC/MS on 37 urine samples obtained similarly as above. The samples ranged from 212 ng/mL to 972 ng/mL on GC/MS. A cutoff of 500 ng/mL was used for both procedures. #### IMMULITE Methamphetamine | GC/MS | Positive | Negative | Sensitivity: | 88% | |----------|----------|----------|--------------|-----| | Positive | 14 | 2 | Specificity: | 86% | | Negative | 3 | 18 | Agreement: | 87% | #### Conclusion: The data presented in this summary of safety and effectiveness is the data that the Food and Drug Administration used in granting DPC substantial equivalence for IMMULITE® Methamphetamine. Edward Levin Edward M. Levine, Ph.D. Director of Clinical Affairs 1/29/99 Date {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing right, with flowing lines beneath them, possibly representing water or movement. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 # JAN 2 0 2000 Edward M. Levine, Ph.D. Director of Clinical Affairs Diagnostic Products Corporation 5700 West 96th Street Los Angeles, California 90045-5597 Re: K992634 Trade Name: IMMULITE® Methamphetamine Regulatory Class: II Product Code: LAF Dated: November 29, 1999 Received: November 30, 1999 Dear Dr. Levine: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {5}------------------------------------------------ #### Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure .............................................................................................................................................................................. {6}------------------------------------------------ # 16992634 510(k) Number (if known): Device Name: IMMULITE® Methamphetamine Indications For Use: IMMULITE® Methamphetamine is a solid-phase, chemiluminescent enzyme immunoassay for use with the IMMULITE Automated Analyzer and designed for the qualitative measurement of methamphetamine in urine. It is intended strictly for in vitro diagnostic use in clinical laboratories, in the context of a program involving an established confirmatory test for methamphetamine. A cutoff of 500 ng/mL is used by the IMMULITE Methamphetamine assay to identify positive and negative results ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) # Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...