ACON MAMP ONE STEP METHAMPHETAMINE TEST STRIP AND TEST DEVICE
K033299 · ACON Laboratories, Inc. · LAF · Dec 12, 2003 · Clinical Toxicology
Device Facts
Record ID
K033299
Device Name
ACON MAMP ONE STEP METHAMPHETAMINE TEST STRIP AND TEST DEVICE
Applicant
ACON Laboratories, Inc.
Product Code
LAF · Clinical Toxicology
Decision Date
Dec 12, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3610
Device Class
Class 2
Intended Use
The ACON® mAMP 500 One Step Methamphetamine Test Strip and Test Device are rapid immunochromatographic assays for the qualitative detection of methamphetamine, a central nervous stimulating drug, in urine. Measurements obtained by these devices are used in the diagnosis and treatment of methamphetamine use or overdose. This assay provides only a preliminary result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmation method.
Device Story
Rapid immunochromatographic assay; detects methamphetamine in human urine. Principle: competitive binding; drug in urine competes with drug conjugate for antibody binding sites. Input: urine specimen applied via dipping (strip) or dropper (device). Output: visual colored lines in control and test regions. Negative result: colored line in test region; positive result: no line in test region. Used in physician offices/clinical settings; operated by healthcare personnel. Provides preliminary analytical result; requires confirmation via GC/MS or alternative chemical method. Assists clinicians in diagnosing/treating methamphetamine use or overdose.
Clinical Evidence
Clinical evaluation of 300 urine specimens compared subject devices against a predicate device and GC/MS. Total agreement with GC/MS was 98% for both the test strip and test device. Positive agreement with GC/MS was 100% (strip) and 99% (device); negative agreement was 97% (strip) and 98% (device).
Indicated for qualitative detection of methamphetamine in urine for diagnosis and treatment of methamphetamine use or overdose in patients. Intended for use by healthcare professionals, including at point-of-care sites.
Regulatory Classification
Identification
A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.
Special Controls
*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
K011672 — ACON MAMP ONE STEP METHAMPHETAMINE TEST STRIP, ACON MAMP ONE STEP METHAMPHETAMINE TEST DEVICE · ACON Laboratories, Inc. · Jul 31, 2001
K994406 — INSTANT-VIEW METHAMPHETAMINE URINE DIP STRIP TEST · Alfa Scientific Designs, Inc. · Aug 14, 2000
Submission Summary (Full Text)
{0}------------------------------------------------
# DEC 1. 2 2003
#### 8. 510(k) SUMMARY
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The Assigned 510(k) number is _ K 033299
## Submitter:
ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121
Tel.: 858-535-2030 Fax: 858-535-2038
#### Date:
October 7, 2003
## Contact Person:
Edward Tung, Ph.D.
## Product Names:
ACON® mAMP-500 One Step Methamphetamine Test Strip ACON® mAMP-500 One Step Methamphetamine Test Device
#### Common Name:
Immunochromatographic test for the qualitative detection of Methamphetamine in urine.
## Regulation Name:
Methamphetamine test system.
## Product Code:
LAF
{1}------------------------------------------------
## Classification Number:
21 CFR, 862.3610
#### Device Classification:
The Methamphetamine test systems have been classified as Class II devices with moderate complexity. The ACON mAMP-500 One Step Methamphetamine Test Strip and ACON mAMP-500 One Step Methamphetamine Test Device are similar to another FDA-cleared device for the qualitative detection of Methamphetamine in urine specimens. These tests are used to provide only a preliminary analytical result. All positive test results obtained with these devices must be confirmed by another test method, preferably GC/MS analysis.
#### Intended Use:
The ACON mAMP-500 One Step Methamphetamine Test Strip and ACON mAMP-500 One Step Methamphetamine Test Device are rapid chromatographic immunoassays for the qualitative detection of Methamphetamine in urine at a cutoff concentration of 500 ng/mL. These tests are used to provide only a preliminary analytical result. All positive test results obtained with these devices must be confirmed by another test method, preferably GC/MS analysis. They are intended for healthcare professionals including professionals at point-of-care sites.
#### Description:
The ACON mAMP-500 One Step Methamphetamine Test Strip and ACON mAMP-500 One Step Methamphetamine Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative screening of Methamphetamine in a urine sample. The test is based on the principle of antibody immunochemistry. It utilizes the mouse monoclonal antibody to selectively detect elevated levels of Methamphetamine and its metabolite in urine at a cutoff concentration of 500 ng/mL. These tests can be performed without the use of an instrument.
A drug-positive urine specimen will not generate a colored-line in the designated test region, while a negative urine specimen or a urine specimen containing Methamphetamine at the concentration i below the cutoff level will generate a colored-line in the test region. To serve as a procedural control, a colored-line should always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.
## Comparison to a Predicate Device:
A comparison of the features of the ACON mAMP-500 One Step Methamphetamine Test Strip and ACON mAMP-500 One Step Methamphetamine Test Device versus a FDA-cleared Methamphetamine test with 500 ng/mL Methamphetamine cutoff is shown below:
{2}------------------------------------------------
- Both tests are assays intended for the qualitative detection of Methamphetamine in urine . samples.
- Both tests are intended as a screening method that provides a preliminary analytical test ● result.
- Both tests are immunochromatographic, lateral flow assays for the rapid detection of . Methamphetamine with a visual, qualitative end result.
- Both tests utilize the same basic immunoassay principles that rely on antigen/ antibody . interactions to indicate a positive or negative result.
- Both tests have a cutoff Methamphetamine concentration of 500 ng/mL. .
#### Safety and Effectiveness Data:
#### Accuracy
A clinical evaluation was conducted using 300 clinical urine specimens including approximately 10% of the specimens containing Methamphetamine concentration fell between -25% cutoff to +25% cutoff range. This evaluation compared the test results between ACON mAMP-500 One Step Methamphetamine Test Strip and ACON mAMP-500 One Step Methamphetamine Test Device with a FDA-cleared Methamphetamine test; as well as compared against data obtained from the customary Gas Chromatography/Mass Spectrometry analysis. These comparisons of data yielded the following results:
#### ACON mAMP-500 One Step Methamphetamine Test Strip versus a FDA-cleared mAMP 500 Test:
Positive Agreement: 108 / 108 =>99% (97 % - 100 %*) Negative Agreement: 153 / 192 = 80% (73 % = 85%*) Overall Agreement: 261 / 300 = 87% (83 % - 91 %*) * 95% confidence intervals
#### ACON mAMP-500 One Step Methamphetamine Test Device versus a FDA-cleared mAMP 500 Test:
Positive Agreement: 108 / 108= 100% (97 % - 100 %*) Negative Agreement: 158 / 192= 82% (76 % - 87 %*) Overall Agreement: 266 / 300 = 89% (84 % - 92 %*) * 95% confidence intervals
{3}------------------------------------------------
ACON mAMP-500 One Step Methamphetamine Test Strip versus data obtained with GC/MS at the cutoff concentration of 500 ng/mL:
| | | Specimen Cutoff Range by GC/MS Data | | | | % Agreement with<br>GC/MS Data | |
|-----------------------------------|----------|-------------------------------------|------------------|-------------------|-------------------|--------------------------------|---------------------------------|
| | | Negative | < -25%<br>Cutoff | -25% to<br>Cutoff | Cutoff to<br>+25% | > +25%<br>Cutoff | |
| ACON<br>mAMP<br>500 Test<br>Strip | Positive | 0 | 0 | 7 | 8 | 132 | 100% (140/140)<br>(97% - 100%)* |
| | Negative | 150 | 0 | 3 | 0 | 0 | 97% (153/160)<br>(91% - 98%)* |
# ACON mAMP-500 One Step Methamphetamine Test Strip versus GC/MS
Total agreement with GC/MS : 293/300 = 98% (95%- 99%)* * Denotes 95% confidence intervals.
#### ACON mAMP-500 One Step Methamphetamine Test Device versus GC/MS
| | | Specimen Cut off Range by GC/MS Data | | | | % Agreement with GC/MS Data | |
|---------------------------------|----------|--------------------------------------|--------------|----------------|----------------|-----------------------------|-------------------------------|
| | | Negative | <-25% Cutoff | -25% to Cutoff | Cutoff to +25% | >+25% Cutoff | |
| ACON<br>mAMP-500<br>Test Device | Positive | 0 | 0 | 4 | 8 | 131 | 99% (139/140)<br>(96% - 99%)* |
| ACON<br>mAMP-500<br>Test Device | Negative | 150 | 0 | 6 | 0 | 1 | 98% (156/160)<br>(94% - 99%)* |
Total agreement with GC/MS : 295/300 = 98% (96% - 99%)* * Denotes 95% confidence interval.
#### Performance Characteristics and Other information:
The performance characteristics of ACON mAMP-500 One Step Methamphetamine Test Strip, ACON mAMP-500 One Step Methamphetamine Test Device were verified by analytical sensitivity study, specificity and cross reactivity study, interference studies, precision study, read time flex study, temperature flex study, specimen storage and stability study. Study results indicate that these test devices are robust and can perform satisfactorily when used according to the "Indication for Use" statement specified in their package inserts.
#### Conclusion:
These clinical studies demonstrated substantial equivalency on performance between the ACON mAMP-500 One Step Methamphetamine Test Strip, ACON mAMP-500 One Step Methamphetamine Test Device and a FDA-cleared Methamphetamine test with the same Methamphetamine cutoff concentration. It is also demonstrated that these tests are safe and effective in qualitatively detecting Methamphetamine at a concentration of 500 ng/mL. The POL study demonstrated that these tests are suitable for healthcare professionals including professionals at point-of-care sites.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or other bird in flight, represented by three curved lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 1 2 2003
Edward Tung, Ph.D. Regulatory Affairs ACON Laboratories, Inc. 4108 Sorrento Valley BLvd. San Diego, CA 92121
Re: k033299
> Trade/Device Name: ACON mAMP-500 One Step Methamphetamine Test Strip ACON mAMP-500 One Step Methamphetamine Test Device Regulation Number: 21 CFR 862.3610 Regulation Name: Methamphetamine test system Regulatory Class: Class II Product Code: LAF Dated: October 7, 2003 Received: October 14, 2003
Dear Dr. Tung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
# Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device; or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### 11. INDICATIONS FOR USE
510(k) Number: K033299
Device Name: ACON mAMP-500 One Step Methamphetamine Test Strip
ACON mAMP-500 One Step Methamphetamine Test Device
Indications for Use:
The ACON® mAMP 500 One Step Methamphetamine Test Strip and Test Device are rapid immunochromatographic assays for the qualitative detection of methamphetamine, a central nervous stimulating drug, in urine. Measurements obtained by these devices are used in the diagnosis and treatment of methamphetamine use or overdose.
This assay provides only a preliminary result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmation method.
Albert Sarr
Joan
(opper
Division Sign-Off for
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K033299
Prescription USE ✓
Over the Counter
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.