ENSITE 3000 SYSTEM

K992479 · Endocardial Solutions, Inc. · DQK · Dec 2, 1999 · Cardiovascular

Device Facts

Record IDK992479
Device NameENSITE 3000 SYSTEM
ApplicantEndocardial Solutions, Inc.
Product CodeDQK · Cardiovascular
Decision DateDec 2, 1999
DecisionSESU
Submission TypeTraditional
Regulation21 CFR 870.1425
Device ClassClass 2

Intended Use

The EnSite Multi-electrode Diagnostic Catheter used with the EnSite 3000 Electrophysiology Workstation is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters).

Device Story

System consists of EnSite Multi-electrode Diagnostic Catheter (MEA) and EnSite 3000 Electrophysiology Workstation. Catheter is 9 French, single-use, percutaneous, deployed in right atrium; features distal balloon/braid MEA and ring electrodes. Workstation includes Silicon Graphics Octane processor and Patient Interface Unit. System collects, stores, and displays intracardiac electrograms. Unlike conventional catheters, MEA does not require direct endocardial contact. Proprietary software algorithms reconstruct and display right atrial endocardiograms. Used in EP laboratory by physicians. Output aids in identification of complex arrhythmias. Warning: use with radiofrequency ablation may increase risk of cardiac perforation, myocardial infarction, air embolism, and hematoma.

Clinical Evidence

Bench testing only. Device validation testing conducted in accordance with in-house procedures.

Technological Characteristics

9 French percutaneous catheter with polyurethane sheath and balloon/braid MEA. Workstation uses Silicon Graphics Octane processor. Non-contact sensing principle; does not require direct endocardial contact. Proprietary software algorithms for signal reconstruction. Biocompatible materials for blood-contacting surfaces.

Indications for Use

Indicated for patients with complex arrhythmias in the right atrium that are difficult to identify using conventional linear mapping catheters.

Regulatory Classification

Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Endocardial Solutions, Inc. | DEC - 2 1999 | 510(k) Summary | |-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | James W. Bullock<br>President and CEO<br>Endocardial Solutions<br>1350 Energy Lane, Suite 110<br>St. Paul, MN 55108 | | Contact: | James W. Bullock<br>President and CEO | | Date Prepared: | July 22, 1999 | | Trade Name: | EnSite 3000® System | | | a) Model EC 1000 EnSite® Multi-electrode Diagnostic Catheter<br>b) EnSite 3000® Electrophysiology Workstation | | Common name: | Electrophysiology cardiac mapping system | | | a) Electrode recording catheter or electrode recording probe<br>(21CFR 870.1220) | | | b) Programmable diagnostic computer (21 CFR 870.1425) | | Equivalence to: | a) Endocardial Solutions EnSite 3000® System | | Description: | EnSite 3000® System Components<br><br>EnSite® Multi-electrode Diagnostic Catheter<br><br>The EnSite® Multi-electrode Diagnostic Catheter (EnSite® catheter) is a single use, 9 French, percutaneous catheter. The EnSite® catheter is designed for use only with the EnSite 3000® System in an electrophysiology laboratory and for deployment in the right atrium. The proximal end contains the patient cable electrical connector, an inflation port for the distal balloon/ braid multi-electrode array (MEA), luer port compatible with a 0.035" guidewire, and a push shaft to facilitate expansion and deployment of the MEA. The shaft is a coaxial design with a polyurethane outer sheath. At the distal end in addition to the MEA, there are three ring electrodes, one distal and two proximal mounted at | {1}------------------------------------------------ specific locations to the MEA. The tip of the catheter is a pigtail shape to minimize trauma to the endocardium. Biocompatible materials are used for all blood contacting surfaces. EnSite 3000® Electrophysiology Workstation The EnSite 3000® Electrophysiology Workstation is a computerized storage and display system for use in electrophysiology studies of the human heart. It is designed for use in the EP laboratory with the EnSite Multi-electrode Diagnostic Catheter and in conjunction with other standard equipment found in the EP laboratory. This allows for the collection , storage, and display of intracardiac clectrograms. The EnSite 3000® may be used in conjunction with standard electrode mapping catheters, programmable cardiac stimulators, ECG leads and other analog inputs. The EnSite 3000® System is comprised of a display workstation consisting of a Silicon Graphics Octane processor containing proprietary software and a 21" monitor and a Patient Interface Unit which accepts the signals from the patient and from other accessories and converts them to digital signal and sends them to the display workstation for processing. ## Intended use: The EnSite® Multi-electrode Diagnostic Catheter used with the EnSite 3000® Electrophysiology Workstation is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters). # Technological Characteristics: Unlike currently available electrode recording catheters, the EnSite® catheter does not require direct contact with the endocardium for the detection of intracardiac electrograms. The EnSite 3000 Flectrophysiology Workstation connected to the EnSite® catheter utilizes proprietary software algorithms to reconstruct and display right atrial endocardiograms detected by the EnSite® catheter's MFA. ### Non-clinical performance data: July 22, 1999 Endocardial Solutions. Inc. EnSite 30000 System {2}------------------------------------------------ The changes made to the 3000° System underwent a battery of bench and user tests. Device validation testing was conducted in accordance with in-house procedures. # Conclusion: An evaluation of the device changes indicates that this EnSite 3000® System is substantially equivalent to the currently marketed system. July 22, 1999 Endocardial Solutions, Inc. EnSite 3000® System {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The eagle is positioned to the right of the department's name, which is arranged in a circular fashion around the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Public Health Service 2 1688 DEC Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. James W. Bullock President and Chief Executive Officer Endocardial Solutions, Inc. 1350 Energy Lane, Suite 110 Saint Paul, MN 55108-5254 K992479 Re: R92479 Trade Name: EnSite ™ Multi-Electrode Diagnostic Catheter and EnSite 3000 ™ Electrophysiology Workstation Regulatory Class: II (two) Product Code: MTD, DQK Dated: October 15, 1999 Received: October 15, 1999 Dear Mr. Bullock: We have reviewed your Section 510(k) notification of intent to market nt have reviews for above and we have determined the device is che device is equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate encrosure, eo rogarij mainent date of the enactment date of the Medical Device Commerce prior co hay boy bean reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls may sions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual eenstration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause Therefore, in accordance with Section 513(i)(1)(E) of the Act, harm. the following limitation must appear in the Warnings section of the device's labeling: The use of this device in conjunction with radiofrequency WARNING: ablation, as a part of the diagnosis and treatment of cardiac arrhythmias, may pose an increased risk of adverse events such as cardiac perforation, myocardial infarction, air embolism, and hematoma requiring surgical repair and/or blood transfusion. The warning should appear in a black box, and the font size of the text should be at least 2 points larger than any surrounding text. The Warning must be present on the first page of your Catheter Instructions for Use and on the packaging for each individual catheter. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device {4}------------------------------------------------ #### Page 2 - Mr. James W. Bullock can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification if the limitation statement above is added to your labeling, as described. Please note that the above labeling limitations are required by Section 513(i)(l)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, David W. Feigal, Jr., M.D., M.P.H. fice of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 3 - Mr. James W. Bullock Page of 1 . T 510(k) Number (if known) K992479. Device Name: EnSite ™ Multi-Electrode Diagnostic Catheter and EnSite 3000 ™ Electrophysiology Workstation FDA's Statement of the Indications For Use for device: The EnSite Multi-electrode Diagnostic Catheter used with the EnSite 3000 Electrophysiology Workstation is intended to be used in the right atrium of patients with complex arrhythmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters) . Nancy C brogdon rdiovascular, Respiratory, and Neurological I OR 510(k) Number N77277 Prescription Use xx (Per 21 CFR 801.109) Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...