ADDITIONAL IMS ASSAYS FOR THE BAYER ADVIA IMS SYSTEM
Applicant
Bayer Corp.
Product Code
CIX · Clinical Chemistry
Decision Date
Sep 15, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1035
Device Class
Class 2
Indications for Use
The Baver ADVIA IMS Albumin assay is an in vitro diagnostic device intended to measure Albumin in human serum or plasma. Measurements of albumin are used in the diagnosis, monitoring and treatment of a variety of diseases involving the liver and kidney. The Bayer ADVIA IMS Total Bilirubin assay is an in vitro diagnostic device intended to measure bilirubin in human serum or plasma. Measurements of direct or total bilirubin, organic compounds formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis, monitoring and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder disorders. The Bayer ADVIA IMS Carbon Dioxide (CO2) assay is an in vitro diagnostic method intended to measure CO2 in human serum. Measurements of CO2 are used as an aid in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. The Bayer ADVIA IMS Digoxin assay is an in vitro diagnostic device intended to measure Digoxin, a cardioactive drug, in human serum. Measurements of Digoxin are used as an aid in the diagnosis of Digoxin overdose and in the monitoring therapeutic levels of Digoxin to ensure appropriate therapy. The Baver ADVIA IMS HDL Cholesterol assay is an in vitro diagnostic device intended to measure HDL Cholesterol (a lipoprotein) in human serum and plasma. Such measurements are used as an aid in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. The Bayer ADVIA IMS Inorganic Phosphorus assay is an in vitro diagnostic device intended to measure phosphorus in human serum, plasma or urine. Measurements of inorganic phosphorus are used in the diagnosis, monitoring and treatment of a variety of diseases involving the parathyroid gland and kidney, and vitamin D imbalance. The Bayer ADVIA IMS Ion Selective Electrode assays for Sodium,Potassium,and Chloride are in vitro diagnostic devices intended to measure these analytes in human serum or plasma. Measurements of sodium are used as an aid in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension. Addison's disease, dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Measurements of potassium are used as an aid in the diagnosis and treatment of renal tubular disease, hyperaldosteronism, metabolic alkolosis, adrenocortical disease and diabetic ketoacidosis. Measurements of chloride are used as an aid in the diagnosis and treatment of acid-base and water imbalance. It is especially important to measure chloride during the correction of hypokalemic alkalosis and also during severe, prolonged vomiting which lowers serum chloride levels.
Device Story
ADVIA IMS is an automated in vitro diagnostic system for clinical chemistry analysis; processes human serum, plasma, or urine samples; utilizes reagent-based assays and Ion Selective Electrode (ISE) technology to measure specific analytes; provides quantitative results to clinicians; used in clinical laboratory settings; results assist in diagnosis, monitoring, and treatment of various metabolic, renal, hepatic, and electrolyte disorders; system automates sample handling and measurement to improve diagnostic efficiency and patient management.
Clinical Evidence
Bench testing only. Performance validated via precision studies (Total CVs typically 1-7%) and method correlation studies against predicate systems (N=48-107 per analyte, r=0.957-0.999). Interference testing conducted for common substances (bilirubin, hemoglobin, lipids).
Technological Characteristics
Automated clinical chemistry analyzer; utilizes colorimetric/enzymatic reagent assays and Ion Selective Electrode (ISE) potentiometric sensing; serum/plasma/urine matrices; standalone system; standard laboratory reagents and calibrators.
Indications for Use
Indicated for patients requiring diagnostic testing of serum, plasma, or urine for albumin, total bilirubin, CO2, digoxin, HDL cholesterol, inorganic phosphorus, sodium, potassium, and chloride to manage liver, kidney, metabolic, hematological, lipid, and electrolyte disorders.
Regulatory Classification
Identification
An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
{1}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Albumin method for ADVIA® IMS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K992396
# 1. Intended Use
This in vitro diagnostic method is intended to measure albumin in human serum or plasma on the Baver ADVIA IMS system. Measurements of albumin are used in the diagnosis, monitoring and treatment of a variety of diseases involving the liver and kidneys.
# 2. Predicate Device
| Product Name | Reagent Part # | Calibrator Part # |
|---------------------------|----------------|-------------------|
| Technicon CHEM 1 Albumin | T01-1458-53 | T03-1291-62 |
# 3. Device / Method
| Product Name | Reagent Part # | Calibrator Part # |
|------------------|----------------|-------------------|
| ADVIA IM Albumin | B41-3717-26 | T03-1291-62 |
# A. Imprecision(SERUM)
| ADVIA IMS | |
|--------------|-------------|
| Level (g/dL) | Total CV(%) |
| 2.1 | 1.9 |
| 3.4 | 1.6 |
| 4.9 | 1.5 |
#### Correlation (Y=ADVLA IMS. X=comparison system) 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
| Specimen type | Comparison System (X) | N | Regression Equation | Syx (g/dL) | r | Sample Range (g/dL) |
|---------------------|-----------------------|----|---------------------|------------|-------|---------------------|
| Serum | CHEM 1+ | 53 | Y=1.01X+0.55 | 0.13 | 0.990 | 1.4 - 5.3 |
| Plasma(y), Serum(x) | ADVIA IMS | 59 | Y=0.96X+0.12 | 0.06 | 0.978 | 3.9 - 5.2 |
# Interfering Substances
| Interfering<br>Substance | Interfering Sub.<br>Conc. (mg/dL) | Albumin Conc.<br>(g/dL) | Effect<br>(% change) |
|--------------------------|-----------------------------------|-------------------------|----------------------|
| U. Bilirubin | 25 | 4.2 | -4.8 |
| Hemoglobin | 1000 | 3.5 | +5.7 |
| Lipids (Triglycerides) | 1000 | 3.5 | -2.9 |
**Analytical Range**
Scrum/Plasma: 1 10 6 g/dL
7/16/99
Gabriel J. Murray, Jr.
Manager Reg Affairs.
{2}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS T.Bilirubin method for ADVIA® IMS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
# 1. Intended Use
This in vitro method is intended to quantitatively measure total bilirubin in human serum and plasma on the Bayer ADVIA IMS systems. Measurements of total bilirubin are used in the diagnosis and treatment of liver, bemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder disorders.
#### 2. Predicate Device
| Product Name | would be and the country of the country of the country of the country of the country of the country of the country of the country of the country of the country of the country<br> Reagent Part # Calibrator Part # <br>for and and the program and and the many of the many of the many of the many of the many of the may be the may be the may be the may be the may be the may be the may be the b |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technicon CHEM 1 T.Bilirubin T01-1522-53 T03-1291-62 | |
# 3. Device / Method
| Product Name | | Reagent Part # Calibrator Part # | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| www.rangestion.com/production/compress/from the managements of the comments of the comments of the comments of the comments of the comments of the comments of the comments of<br>ADVIA IMS T.Bilirubin | B41-3723-46 T03-1291-62 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
#### Imprecision(SERUM)
| ADVIA IMS | CHEM 1 | | |
|------------------|----------------|------------------|----------------|
| Level<br>(mg/dL) | Total<br>CV(%) | Level<br>(mg/dL) | Total<br>CV(%) |
| 1.1 | 7.0 | 1.1 | 9.1 |
| 4.5 | 5.1 | 5.0 | 4.8 |
| 17.0 | 2.4 | 20.0 | 2.6 |
#### Correlation (Y=ADVIA IMS, X=comparison system)
| Specimen type | Comparison<br>System (X) | N | Regression Equation | Syx<br>(mg/dL) | r | Sample Range<br>(mg/dL) |
|---------------------|--------------------------|----|---------------------|----------------|-------|-------------------------|
| Serum | CHEM 1+ | 58 | Y=1.08X-0.11 | 0.26 | 0.998 | 0.2 - 26.1 |
| Plasma(y), Serum(x) | ADVIA IMS | 57 | Y=1.01X-0.01 | 0.02 | 0.994 | 0.16 - 0.92 |
#### Interfering Substances
| Interfering<br>Substance | Interfering Sub.<br>Conc. (mg/dL) | T.Bilirubin<br>Conc. (mg/dL) | Effect<br>(% change) |
|--------------------------|-----------------------------------|------------------------------|----------------------|
| Hemoglobin | 500 | 6.9 | +1 |
| Lipids (Triglycerides) | 500 | 6.7 | -34 |
Gabriel J. Murray Jr.
Manager Reg. Affairs
#### Analytical Range
Serum/Plasma: 0 to 45 mg/dL
{3}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
# CO2 Method for the Bayer ADVIA Integrated Modular System (IMS)
Listed below is a comparison of the performance between the Bayer ADVIA IMS CO2 method and a similar device that was granted clearance of substantial equivalence (Technicon CHEM 1 Carbon Dioxide, Enzymatic). The information used in the Summary of Safety and Effectiveness was extracted from the Bayer ADVIA IMS CO2 method sheet and the CHEM 1 CO2 (Enzymztic) method sheet.
## INTENDED USE
This in-vitro diagnostic method is intended to measure carbon dioxide (CO2) in human serum. Measurement of CO, is used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.
| METHOD | | ADVIA IMS | CHEM 1 |
|-------------------|-----------------------------|----------------------------------------------------------------------------|-------------------------------------------------------|
| Part No. | Reagents | B41-3726-43 | T01-3253-53 |
| | Calibrators/<br>CO2 diluent | T03-1291-62/<br>T23-1291-52 | T01-1291-62/<br>T23-1291-52 |
| Analytical Range | | 10 to 40 mEq/L | 10 to 40 mEq/L |
| Precision (Total) | | 5.3% @ 11.98 mEq/L<br>4.3% @ 19.10 mEq/L<br>3.0% @ 28.39 mEq/L | 4.5% @ 14 mEq/L<br>3.8% @ 22 mEq/L<br>3.3% @ 28 mEq/L |
| Correlation | | Y=0.99X+0.41 mEq/L<br>Where<br>Y=ADVIA IMS<br>X=CHEM 1<br>N=106 (53 pairs) | 7/16/99<br>Gabriel J. Mno |
r=0.993
Sy.x=0.999 mEq/L
# Interfering Substances
Bilirubin (unconjugated) 25 mg/dL Bilirubin (conjugated) 25 mg/dL Hemoglobin (hemolysate) 500 mg/dL Lipemia (Triglycerides) 500 mg/dL
-2.0% effect change @ 27.84 mEq/L CO2 -37% effect change @ 28.83 mEq/L CO2 -10.0% effect change @ 28.67 mEq/L CO2
{4}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Digoxin Method for ADVIA IMS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.
The assigned 510(k) number is: (leave blank)
#### 1. Intended Use
This in vitro method is intended to quantitatively measure digoxin in human serum on the Bayer ADVIA IMS systems. Measurements of digoxin are used to aid in attaining optimum therapy in patients treated with the drug.
#### Predicate Device 2.
| Product Name | Reagent Part # | Calibrator Part # |
|------------------------|----------------|-------------------|
| Bayer Immuno 1 Digoxin | T01-2875-51 | T01-2864-01 |
#### Device / Method 3.
| Name | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| ----------------------------------------------------------------------<br>AMC T<br>unxoxobili<br>the call a the collection of the consideration of the are of | A 40 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 1 | 11<br> |
# A. Imprecision
| ADVIA IMS | | Immuno 1 | |
|------------------|-----------------|------------------|-----------------|
| Level<br>(ng/mL) | Total<br>CV (%) | Level<br>(ng/mL) | Total<br>CV (%) |
| 0.82 | 6.8 | 0.7 | 8.2 |
| 1.80 | -5.1 | 2.2 | 4.2 |
| 3.81 | 4.8 | 3.4 | 3.6 |
# B. Correlation (Y=ADVIA IMS, X=Comparison system)
| Specimen type | Comparison System (X) | N | Regression Equation | Syx (ng/mL) | R | Sample Range (ng/mL) |
|---------------|-----------------------|----|---------------------|-------------|------|----------------------|
| Serum | Immuno 1 | 63 | $Y=0.93X - 0.11$ | 0.17 | 0.98 | 0.15-3.90 |
7/14/99 Gabriel J. Murray Jr. - Manager Reg. Affairs
{5}------------------------------------------------
| C. Interfering Substances | | | |
|-----------------------------|-----------------------------------|--------------------------|----------------------|
| Interfering<br>Substance | Interfering Sub.<br>Conc. (mg/dL) | Digoxin Conc.<br>(ng/mL) | Effect<br>(% change) |
| Bilirubin | 25 | 2 | 3.9 |
| Bilirubin<br>(unconjugated) | 25 | 2 | -3.9 |
| Hemoglobin | 500 | 2 | -0.3 |
| Lipids<br>(Triglycerides) | 500 | 2 | -11.7 |
1. 2011
. ..........
. . . . . . . . . . . . . . . .
. . . . .
: · ·
Analytical Range: 0.1-5.0 ng/mL Minimum Detectable Concentration: 0.1 ng/mL
Date 7/16/99
Bayer Corporation, Business Group Diagnostics Tarrytown, NY. Gabriel J. M. Gabriel J. Muraca, Jr. Manager Regulatory Affairs
914-524-3494 (fax 914-524-2500)
{6}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
# HDL cholesterol Method for the ADVIA IMS Systems
Listed below is a comparison of the performance of the Bayer ADVIA HDL cholesterol method and a similar device that was granted clearance of substantial equivalence (Bayer Technicon Chem 1 HDL cholesterol method). The information was extracted from the Baver ADVIA IMS HDL cholesterol method sheet.
# INTENDED USE
The Bayer ADVIA IMS HDL cholesterol assay is an in-vitro diagnostic device intended to measure HDL cholesterol in human serum and plasma. Such measurements are used in the diagnosis, monitoring and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases. This diagnostic method is not intended for use on any other diagnostic system.
# SERUM
| HDL CHOLESTEROL<br>METHOD: | ADVIA IMS | CHEM 1 |
|----------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Part Number: | Cholesterol Reagent<br>B41-3727-46<br>HDL precipitating reagent<br>B21-3733-01<br>Calibrators T03-1291-62 | cholesterol reagent<br>T01-2869-01<br>HDL Precipitating reagent<br>T01-1897-56<br>Calibrators T03-1291-62 |
| Analytical Range: | 0 to 250 mg/dL | 10 to 100 mg/dL |
| Precision (Total): | Mean (mg/dL) | % CV |
| Level 1 | 18 | 3.2 |
| Level 2 | 42 | 2.6 |
| Level 3 | 121 | 2.8 |
| | mean (mg/dL) | % CV |
| | 20 | 8.5 |
| | 46 | 6.7 |
| | 58 | 4.8 |
| Correlation to existing system | |
|----------------------------------------|-----------------------------------|
| Regression Equation: y = 0.997x + 1.95 | |
| where: | y = ADVIA IMS |
| | x = Chem 1 |
| | n = 107 (54 samples in duplicate) |
| | r = 0.987 |
| | Sy.x = 3.58 |
| | range = 19 to 123 mg/dL |
7/16/94 Gabriel J. Muraca Jr. - Muraca By. Affairs.
{7}------------------------------------------------
| Plasma Qualification | | |
|-------------------------------------|-------|--------------------------------|
| Regression Equation: y=0.995X +0.12 | | |
| where: | y | = plasma |
| | x | = serum |
| | n | = 90 (47 samples in duplicate) |
| | r | = 0.996 |
| | Sy.x | = 1.3 |
| | range | = 18 to 71 mg/dL |
Interference
| | Interfering Substance<br>Concentration | HDL<br>cholesterol<br>Concentration | Effect<br>% Change |
|--------------------------|----------------------------------------|-------------------------------------|--------------------|
| Hemoglobin | 250 mg/dL | 47.2 mg/dL | -19 |
| Bilirubin (conjugated) | 6.25 mg/dL | 45.2 mg/dL | -27 |
| Bilirubin (unconjugated) | 25 mg/dL | 43.1 mg/dL | -2 |
| Lipemia (Triglycerides) | 1000 mg/dL | 44.3 mg/dL | +1 |
Gabriel J. Munaco Jr.
Andrews and
.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs. .................................................................................................................................................. Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
7/6/99
Date
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
: 上一篇: 上一篇:
{8}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Inorganic Phosphorus method for ADVIA® IMS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _
#### 1. Intended Use
This in vitro diagnostic method is intended to measure Inorganic Phosphorus in luman serum, plasma or urine on the Bayer ADVIA IMS system. . Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases, and vitamin D imbalance.
# 2. Predicate Device
| Product Name | Reagent Part # | Calibrator Part # |
|-------------------------------------------|----------------|-------------------|
| Technicon CHEM 1® Inorganic<br>Phosphorus | T01-1609-53 | T03-1291-62 |
# 3. Device / Method
| Product Name | Reagent Part # | Calibrator Part # |
|-----------------------------------|----------------|-------------------|
| ADVIA IMS Inorganic<br>Phosphorus | B41-3734-46 | T03-1291-62 |
## A. Imprecision(SERUM)
| ADVIA IMS | | |
|------------------|-------------|---------|
| Level<br>(mg/dL) | Total CV(%) | L<br>(m |
| 2.5 | 1.7 | |
| 5.0 | 2.1 | |
| 8.8 | 2.7 | |
| CHEM 1 | |
|------------------|----------------|
| Level<br>(mg/dL) | Total<br>CV(%) |
| 3.1 | 3.4 |
| 6.6 | 2.7 |
| 7.7 | 2.8 |
#### B. Imprecision(URINE)
| ADVIA IMS | |
|------------------|----------------|
| Level<br>(mg/dL) | Total<br>CV(%) |
| 24.8 | 1.9 |
| 69.8 | 2.2 |
| 96.3 | 1.9 |
| CHEM 1 | |
|------------------|----------------|
| Level<br>(mg/dL) | Total<br>CV(%) |
| 37 | 2.6 |
| 55 | 2.7 |
#### Correlation (Y=ADVIA IMS, X=comparison system)
| Specimen type | Comparison System (X) | N | Regression Equation | Syx (mg/dL) | r | Sample Range (mg/dL) |
|---------------------|-----------------------|----|---------------------|-------------|-------|----------------------|
| Serum | CHEM 1+ | 50 | Y = 0.94X + 0.15 | 0.15 | 0.997 | 1.3 - 10.0 mg/dL |
| Plasma(y), Serum(x) | ADVIA IMS | 58 | Y=0.96X-0.17 | 0.12 | 0.977 | 1.9-4.4 mg/dL |
| Urine | CHEM 1 | 48 | Y = 1.01X + 0.15 | 1.17 | 0.999 | 4.5 - 99.9 mg/dL |
Gabriel J. Menaca Jr. - Manager Reg. Affairs.
7/16/99
{9}------------------------------------------------
# Interfering Substances
| Interfering<br>Substance | Interfering Sub.<br>Conc. (mg/dL) | Inorganic<br>Phosphorus<br>Conc. (mg/dL) | Effect<br>(% change) |
|--------------------------|-----------------------------------|------------------------------------------|----------------------|
| Bilirubin | 20 | 3.7 | +2.2 |
| Hemoglobin | 500 | 3.6 | +19.4 |
| Lipids (Triglycerides) | 500 | 3.8 | +21.6 |
| Acetaminophen | 50 | 8.0 | -3.8 |
| Ascorbic Acid | 200 | 18.4 | +1.3 |
| Salicylate | 500 | 18.4 | +5.0 |
:
.
:
Analytical Range Serum/Plasma: Urine:
0 to 15 mg/dL 0 to 100 ing/dL
: : :
.
: 上一篇:
Gabriel J. Munoz Jr.
7/16/99
.
1
:
:
:
{10}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
#### Sodium,Potassium,and Chloride ISE methods for ADVIA® IMS I.
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
# 1. Intended Use
This in vitro diagnostic method to measure Sodium (Na), Potassium (K) and Chloride (CI) in human serum, plasma on the Bayer ADVIA IMS system
## 2. Device / Method
| Product Name | Reagent Part # | Calibrator Part # |
|------------------------------------------|----------------|--------------------------------------|
| ADVIA IMS Sodium, Potassium,<br>Chloride | B48-4054-54 | T21-4055-01 (Hi)<br>T21-4056-01 (Lo) |
# Imprecision
| ANALYTE | Level (mmol/L) | Total CV(%) |
|-----------|----------------|-------------|
| Sodium | 113 | 1.0 |
| | 139 | 1.0 |
| Potassium | 3.4 | 1.3 |
| | 5.9 | 1.2 |
| Chloride | 93 | 1.2 |
| | 113 | 1.2 |
# Correlation (Y=ADVIA IMS, X=comparison system)
| ANALYTE (Serum) | Comparison System (X) | N | Regression Equation | Syx (mmol/L) | R | Sample Range (mmol/L) |
|-----------------|-----------------------|----|--------------------------------------|--------------|-------|-----------------------|
| Sodium | IL Flame | 53 | $0.94 x Flame + 7.08 mmol/L$ | 1.68 | 0.957 | 120.0-147.1 |
| Potassium | IL Flame | 54 | $1.03 x Flame - 0.15 mmol/L$ | 0.17 | 0.990 | 2.55-8.72 |
| Chloride | Chloridometer | 51 | $0.84 x Chloridometer + 16.0 mmol/L$ | 1.1 | 0.986 | 83.6-115.5 |
#### Plasma/Serum Comparison
| | Sodium | Potassium | Chloride |
|-------------------|--------|-----------|----------|
| Mean Serum Level | 137.2 | 3.97 | 105.2 |
| Mean Plasma Level | 137.2 | 3.76 | 105.2 |
| Mean Difference | 0.0 | 0.21 | 0.0 |
| % Difference | 0.0 | 5.6 | 0.0 |
Gabriel J. Muraca Jr.
G. J. Muraca Jr.
Manager RA Bayer Corporation
7/6/99
---
Date
{11}------------------------------------------------
# ADVIA IMS ISE Na, K and Cl Report
#### Interfering Substances
Sodium: No significant effect on serum has been demonstrated from bromide, ammonium and iodide.
Potassium: Since the concentration of potassium inside erythrocytes is much greater than that in extracellular fluid, hemolysis should be avoided, and the serum/plasma should be separated from the cells as soon as possible after collection.
Chloride: Sodium Salicylate at a level of 2.2 mmol/L (30 mg/dL) will elevate the chloride result by less than 2.0 mmol/L. The presence of bromide or iodide salts will falsely elevate chloride results.
Analytical Range Serum or Plasma: Sodium 40 to 205 mmol/L Potassium 1.5 to 15 mmol/L 50 to 200 mmol/L Chloride
Gabriel J. Munoz, Jr.
7/16/99
{12}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
SEP 1 5 1999
Mr. Gabriel J. Muraca, Jr. Manager. Regulatory Affairs Baver Corporation Business Group Diagnostics 511 Benedict Avenue Tarrytown, New York 10591-5097
K992399 Re:
> Trade Name: 9 Additional IMS Assays for the Bayer ADVIA® IMS™ System Regulatory Class: II Product Code: CIX, CIG, CHS, KXT, CHH, CEO JGS, CEM, CGZ Dated: July 16, 1999 Received: July 19, 1999
Dear Mr. Muraca:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{13}------------------------------------------------
# Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{14}------------------------------------------------
# 510(k) Number (if known):
# Device Name: Bayer ADVIA® Integrated Modular System (IMS)
Indications For Use:
The Baver ADVIA IMS Albumin assay is an in vitro diagnostic device intended to measure Albumin in human serum or plasma. Measurements of albumin are used in the diagnosis, monitoring and treatment of a variety of diseases involving the liver and kidney.
The Bayer ADVIA IMS Total Bilirubin assay is an in vitro diagnostic device intended to measure bilirubin in human serum or plasma. Measurements of direct or total bilirubin, organic compounds formed during the normal and abnormal destruction of red blood cells, are used in the diagnosis, monitoring and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder disorders.
The Bayer ADVIA IMS Carbon Dioxide (CO2) assay is an in vitro diagnostic method intended to measure CO2 in human serum. Measurements of CO2 are used as an aid in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance.
The Bayer ADVIA IMS Digoxin assay is an in vitro diagnostic device intended to measure Digoxin, a cardioactive drug, in human serum. Measurements of Digoxin are used as an aid in the diagnosis of Digoxin overdose and in the monitoring therapeutic levels of Digoxin to ensure appropriate therapy.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|----|--------------------------------------------------|
|------------------------------------------|----|--------------------------------------------------|
| | <div> <img alt="signature" src="signature.png" width="200"/> <div>(Division Sign-Off)</div> <div>Division of Clinical L</div> <div>510(k) Number K992399</div> </div> |
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{15}------------------------------------------------
Page 2 of 2
# 510(k) Number (if known):
# Device Name: Bayer ADVIA® Integrated Modular System IMS
Indications For Use:
The Baver ADVIA IMS HDL Cholesterol assay is an in vitro diagnostic device intended to measure HDL Cholesterol (a lipoprotein) in human serum and plasma. Such measurements are used as an aid in the diagnosis and treatment of lipid disorders (such as diabetes mellitus), atherosclerosis, and various liver and renal diseases.
The Bayer ADVIA IMS Inorganic Phosphorus assay is an in vitro diagnostic device intended to measure phosphorus in human serum, plasma or urine. Measurements of inorganic phosphorus are used in the diagnosis, monitoring and treatment of a variety of diseases involving the parathyroid gland and kidney, and vitamin D imbalance.
The Bayer ADVIA IMS Ion Selective Electrode assays for Sodium,Potassium,and Chloride are in vitro diagnostic devices intended to measure these analytes in human serum or plasma. Measurements of sodium are used as an aid in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension. Addison's disease, dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Measurements of potassium are used as an aid in the diagnosis and treatment of renal tubular disease, hyperaldosteronism, metabolic alkolosis, adrenocortical disease and diabetic ketoacidosis. Measurements of chloride are used as an aid in the diagnosis and treatment of acid-base and water imbalance. It is especially important to measure chloride during the correction of hypokalemic alkalosis and also during severe, prolonged vomiting which lowers serum chloride levels.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) .
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.