The AutoSPECT Plus with InStill Motion Correction option to the ADAC Gamma Camera Systems produces images which depict anatomical density of the patient. This system is intended to provide enhancements of gamma camera emission images by automating previously manual image processing functions and providing manual and automated motion correction.
Device Story
AutoSPECT Plus with InStill Motion Correction is a software option for ADAC Gamma Camera Systems. It processes cardiac and non-cardiac SPECT, gated SPECT, Vantage SPECT, MCD, and MCD-AC data sets. The device automates image processing tasks—specifically reconstruction and reorientation—to enhance emission images. It offers three automated modes: 'Auto All' (reconstruction limits, reconstruction, and reorientation), 'Auto Recon' (reconstruction limits and reconstruction), and 'Auto Reorient' (reorientation). It also supports manual processing and batch mode for multiple data sets. Used in clinical settings by technicians or physicians, the software streamlines workflows by reducing manual interaction. The output consists of processed short axis, horizontal long axis, and vertical long axis data sets, which assist clinicians in visualizing anatomical density for diagnostic decision-making.
Clinical Evidence
Bench testing only. Images were processed using the AutoSPECT Plus with and without InStill Motion Correction to verify the functionality of the automated processing and motion correction features.
Technological Characteristics
Software-based image processing system for gamma camera emission data. Features automated and manual reconstruction and reorientation algorithms. Operates on ADAC Gamma Camera Systems. Connectivity includes batch processing of SPECT data sets. Software utilizes established algorithms from previously cleared ADAC products.
Indications for Use
Indicated for patients undergoing gamma camera emission imaging to produce images depicting anatomical density. Intended for use by clinicians to automate image processing and provide motion correction for cardiac and non-cardiac SPECT data sets.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
SMV STATISTM
SMV MyoSPECT with ReOrient Express
Submission Summary (Full Text)
{0}------------------------------------------------
K992317
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. General Information
| A. | Submitted By: | ADAC Laboratories<br>540 Alder Drive<br>Milpitas, CA 95035<br>Tel: (408) 468-3989<br>Fax: (408) 435-7427 |
|----|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Dennis W. Henkelman at address above |
| B. | Device Trade Name:<br>Common Name:<br>Classification Name: | AutoSPECT Plus with InStill Motion Correction<br>Gamma Camera Systems<br>System, Emission Computed Tomography |
| C. | Predicate Device: | SMV STATISTM<br>SMV MyoSPECT with ReOrient Express |
## D. Device Description:
The AutoSPECT Plus with InStill Motion Correction software option enables automatic processing of one or more cardiac SPECT, gated SPECT, Vantage SPECT, MCD, or MCD-AC cardiac data sets. In addition. AutoSPECT Plus with InStill Motion Correction enables the manual processing of one or more SPECT. gated SPECT, Vantage SPECT, MCD, or MCD-AC data sets. The basic data processing algorithms utilized by AutoSPECT Plus with InStill Motion Correction are ones which have been utilized in previously cleared ADAC products.
For automatically processing cardiac data sets, AutoSPECT Plus with InStill Motion Correction provides three options. The first of these options, "Auto All", automatically determines the reconstruction limits, reconstructs and reorients the cardiac data sets and creates the short axis, horizontal long axis and vertical long axis data sets. The second option, "Auto Recon", automatically determines the reconstruction limits and reconstructs the cardiac data sets. The third option. "Auto Reorient", automatically reorients the cardiac transverse data sets. If necessary, AutoSPECT Plus with InStill Motion Correction reconstructs the data sets prior to re-orientation.
{1}------------------------------------------------
AutoSPECT Plus with InStill Motion Correction software also allows the user to process noncardiac SPECT, Vantage, MCD, or MCD-AC data sets. In this case, the operator manually positions the reconstruction limit lines to reconstruct transverse data sets. If necessary, the data set can be reoriented by manually positioning the Azimuth and elevation lines to the desired location.
The AutoSPECT Plus with InStill Motion Correction software will also process groups of SPECT data sets in a batch mode fashion. Once the operator has selected the data sets and determined the processing method (i.e. Auto All, Auto Recon, Reorient, or Reconstruct), AutoSPECT Plus with InStill Motion Correction processes the first data set followed by all remaining data sets without further interaction from the operator.
- E. Indications for Use:
The AutoSPECT Plus with InStill Motion Correction option to the ADAC Gamma Camera Systems produces images which depict anatomical density of the patient. This system is intended to provide enhancements of gamma camera emission images by automating previously manual image processing functions and providing manual and automated motion correction.
- F. Technological Comparison:
The AutoSPECT Plus with InStill Motion Correction, SMV STASIS™ and SMV MyoSPECT with ReOrient Express devices have the same indications for use and operating principles.
- II. Testing
Images were processed using the AutoSPECT Plus with and without InStill Motion Correction.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is an abstract image of an eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
CHI 1 1999
Dennis W. Henkelman, R.A.C. Director, Regulatory Affairs & Quality Assurance ADAC Laboratories 540 Alder Drive Milipitas, California 95035
Re:
K992317 AutoSPECT Plus with InStill Motion Correction Dated: July 8, 1999 Received: July 9, 1999 Regulatory Class: II 21 CFR 892.1200/Procode: 90 KPS
Dear Mr. Henkelman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been redassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for vour device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
K992317
AutoSPECT Plus with InStill Motion Correction Device Name:
ADAC Laboratories Sponsor Name:
Indications For Use:
The AutoSPECT Plus with InStill Motion Correction option to the ADAC Gamma Camera Systems produces images which depict anatomical density of the patient. This system is intended to provide enhancements of gamma camera emission images by automating previously manual image processing functions and providing manual and automated motion correction.
Do Not Write Below This Line - Continue on Another Page if Needed
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
Over-The-Counter Use
Evinl A. Seaman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K992317
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.