ADDITIONAL ASSAYS FOR THE BAYER ADVIA 1650 CHEMISTRY SYSTEM

K991576 · Bayer Corp. · CIX · Jun 30, 1999 · Clinical Chemistry

Device Facts

Record IDK991576
Device NameADDITIONAL ASSAYS FOR THE BAYER ADVIA 1650 CHEMISTRY SYSTEM
ApplicantBayer Corp.
Product CodeCIX · Clinical Chemistry
Decision DateJun 30, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1035
Device ClassClass 2

Indications for Use

The ADVIA 1650 Chemistry System is an automated, clinical chemistry analyzer that can run tests on human serum, plasma, or urine in random access, batch, and STAT modes at a throughput rate of 1200 photometric tests per hour and 450 electrolyte tests per hour. The photometric analyzer performs clinical chemistry and immuno-turbidimetric methods. The electrylyte portion of the analyzer measures the sodium, and chloride concentrations in serum, plasma or urine samples based on a potentiometric procedure that uses ion-selective electrodes. The ADVIA 1650 is intended for use in conjunction with certain reagents to measure a variety of analytes contained in human fluids.

Device Story

The ADVIA 1650 is an automated clinical chemistry analyzer for use in clinical laboratories by trained personnel. It accepts human serum, plasma, or urine samples. The device utilizes photometric and immuno-turbidimetric methods for general chemistry analytes and ion-selective electrodes (potentiometric) for electrolytes (sodium, chloride, potassium). It operates in random access, batch, or STAT modes with a throughput of 1200 photometric and 450 electrolyte tests per hour. The system processes samples and reagents to produce quantitative analyte concentrations. Healthcare providers use these results to diagnose and monitor conditions such as liver disease, renal failure, diabetes, and electrolyte imbalances, facilitating clinical decision-making regarding patient treatment and management.

Clinical Evidence

Bench testing only. Performance was validated through comparative studies against predicate devices (e.g., Technicon DAX, Beckman Synchron CX series, Hitachi, Cobas Fara). Studies evaluated imprecision (Total CVs typically 1-5%) and correlation (N=30-193 per analyte; R values 0.89-1.00). Interfering substances (bilirubin, hemoglobin, lipids) were tested for each assay to determine concentration-dependent effects on analyte measurement.

Technological Characteristics

Automated clinical chemistry analyzer; photometric and immuno-turbidimetric detection; potentiometric sensing via ion-selective electrodes for electrolytes. Throughput: 1200 photometric tests/hr, 450 electrolyte tests/hr. Sample types: serum, plasma, urine. System is a floor-standing clinical analyzer. Connectivity and software details not specified beyond automated operation.

Indications for Use

Indicated for use in clinical laboratories to measure various analytes in human serum, plasma, or urine for the diagnosis and monitoring of diseases including liver, kidney, bone, and inflammatory disorders, diabetes, pancreatitis, and electrolyte imbalances.

Regulatory Classification

Identification

An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K991576 # SUMMARY OF SAFETY AND EFFECTIVENESS This section contains performance comparisons for 21 clinical methods. For each method, Advia 1650 was compared to a similar device (or devices) that was granted clearance of substantial equivalence. The table below lists the predicate devices and part numbers. | | ADVIA 1650 | Specimen | Predicate Device | Predicate Device | |----------------------------|----------------|-------------|------------------------------|------------------| | Method | Reagent Part # | Type | Name | Reagent Part # | | Albumin | B01-4121-01 | Serum | Technicon DAX® | T01-1337-B3 | | Alkaline Phosphatase - AMP | B01-4134-01 | Serum | Technicon DAX® | T01-1814 | | Amylase | B01-4136-01 | Serum | Beckman Synchron CX® | 442775 | | | | Urine | Beckman Synchron CX7® | 442775 | | Aspartate Aminotransferase | B01-4139-01 | Serum | Technicon DAX® | T01-1750 | | Calcium | B01-4145-01 | Serum | Technicon DAX® | T01-1476 | | | | Urine | DuPont Dimension | DF23A | | Chloride | B01-4171-01 | Urine | Beckman Synchron CX3® | 450214, 472095 | | | | Serum | Technicon DAX® | T152-17 | | | | Serum | Beckman Synchron CX7® | 450214, 472095 | | | | Urine | Beckman Synchron CX3/7® | 450214, 472095 | | Creatine Kinase | B01-4137-01 | Serum | Technicon DAX® | T01-1882 | | | | Serum | Beckman Synchron CX7® | 443794 | | Creatinine, Jaffe | B01-4126-01 | Serum | Technicon DAX® | T01-1927 | | | | Urine | Hitachi® | 1040847 | | Creatinine, Enzymatic | B01-4127-01 | Serum | Technicon DAX® | T01-1927 | | Bilirubin, Direct | B01-4123-01 | Serum | Technicon DAX® | T01-1565 | | | | Serum | Hitachi® | 1109774 | | Glucose - Hexokinase | B01-4129-01 | Serum | Technicon DAX® | T11-1832 | | | | Urine | Hitachi® | 1447521 | | Glucose - Oxidase | B01-4130-01 | Serum | Technicon DAX® | T01-1492-56 | | | | Urine | Beckman Synchron CX3/7® | 443355 | | HDL Cholesterol | B01-4125-01 | Serum | Boehringer Mannheim on RA-XT | 543004 | | Inorganic Phosphorus | B01-4144-01 | Serum | Technicon DAX® | T01-1303 | | | | Serum | Beckman Synchron CX7® | 465145 | | | | Urine | Cobas Fara® | 44031 | | Iron | B01-4147-01 | Serum | Technicon DAX® | 150-26E, 150-26F | | | | Serum | Sigma RA1000 | 565-1, 565-3 | | Magnesium | B01-4148-01 | Serum | Technicon DAX® | T01-2878 | | | | Serum/Urine | Hitachi® | 1273582 | | Potassium | B01-4171-01 | Serum | Technicon DAX® | T01-3161 | | | | Urine | Beckman Synchron CX3® | 443325, 443315 | | Sodium | B01-4171-01 | Serum | Technicon DAX® | T01-3161 | | | | Urine | Beckman Synchron CX3® | 450124, 472095 | | Bilirubin, Total | B01-4122-01 | Serum | Technicon DAX® | T01-1963 | | Uric Acid | B01-4131-01 | Serum | Technicon DAX® | T01-2577 | | | | Urine | Beckman Synchron CX7® | 442785 | | | | Urine | Cobas Fara® | 828475 | | Urea Nitrogen | B01-4132-01 | Serum | Technicon DAX® | T01-1823 | | | | Urine | Beckman Synchron CX3/7® | 443359 | Gabriel J. Muraca, Jr. Gabriel J. Muraca, Jr. Manager, Regulatory Affairs Bayer Corporation 511 Benedict Ave. Tarrytown, NY 10591-5097 4/29/99 --- Date {1}------------------------------------------------ ## 1. ALB ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Albumin Method for the Bayer ADVIA 1650 (g/dL) #### Intended Use This in vitro diagnostic assay is intended to measure albumin concentration in human serum and plasma on the ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of chronic inflammatory diseases, collagen diseases, and liver and kidney disorders. ## Imprecision | Advia 1650 | Technicon | | | |---------------|-----------|--------------|-------| | Specimen type | Level | Total CV (%) | Level | | Serum | 2.1 | 2.4 | 2.2 | | Serum | 3.4 | 1.8 | 3.4 | | | | | 4.8 | ## DAX Total CV (%) 3.3 2.5 2.0 ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type: Site | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|------|-------|--------------| | Serum: MSK | DAX | 156 | Y=0.86X+0.55 | 0.19 | 0.969 | 1.6-5.3 | | Plasma(y), Serum(x) | ADVIA 1650 | 58 | Y=0.96X+0.29 | 0.05 | 0.978 | 4.5-5.7 | ## Interfering Substances | Interfering | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |---------------|--------------------------------------|------------------|--------|-----| | Substance | | | conc. | % | | Bilirubin | 25 | 4.2 | 0.0 | 0.0 | | Hemoglobin | 500 | 3.7 | 0.3 | 8.1 | | Lipids (Trig) | 500 | 4.1 | 0.3 | 7.3 | ## Analytical Range 1 to 6 g/dL ## Expected Values 3.4 - 4.8 g/dL {2}------------------------------------------------ ## 2. ALP-AMP ## SUMMARY OF SAFETY AND EFFECTIVENESS # Alkaline Phosphatase AMP Method for the Bayer ADVIA 1650 (U/L) #### Intended Use This in vitro diagnostic assay is intended to measure alkaline phosphatase activity in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used mainly in the diagnosis and treatment of hepatobiliary and bone disease. #### Imprecision | Specimen type | Level | Total CV (%) | |---------------|-------|--------------| | Serum | 69 | 3.9 | | Serum | 130 | 3.1 | | Technicon DAX | | |---------------|--------------| | Level | Total CV (%) | | 79 | 2.5 | | 186 | 2.2 | | 571 | 1.9 | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type: Site | Comparison<br>System (X) | N | Regression<br>Equation | Syx | R | Sample<br>Range | |---------------------|--------------------------|----|------------------------|------|-------|-----------------| | Serum: TRYTN | DAX | 43 | $Y=1.03X-2.1$ | 3.12 | 0.999 | 22-346 | | Plasma(y), Serum(x) | ADVIA 1650 | 54 | $Y=1.06X+1.42$ | 5.37 | 0.972 | 33-149 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|------| | | | | conc. | % | | Bilirubin | 25 | 69.3 | 1.7 | 2.5 | | Hemoglobin | 500 | 100.3 | -5.6 | -5.6 | | Lipids (Trig) | 500 | 71.0 | 1.3 | 1.8 | #### Analytical Range 0-1100 U/L ## Expected Values 25 to 100 U/L {3}------------------------------------------------ ## 3. AMY ## SUMMARY OF SAFETY AND EFFECTIVENESS ### Amylase Method for the Bayer ADVIA 1650 (U/L) #### Intended Use This in vitro diagnostic assay is intended to measure amylase activity in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used primarily in the diagnosis and monitoring of acute pancreatitis (inflammation of the pancreas). ## Imprecision | Advia 1650 | | | Beckman CX7=Serum, Urine | | |------------------|-------|-----------------|--------------------------|--------------| | Specimen<br>type | Level | Total<br>CV (%) | Level | Total CV (%) | | Serum | 60 | 1.2 | 85.7 | 5.3 | | Serum | 242 | 1.3 | 240 | 5.3 | | Serum | 322 | 1.2 | 320 | 5.3 | | Urine | 792 | 7.2 | 800 | 5.3 | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type: Site | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|------|-------|--------------| | Serum: MSK | CX4 | 102 | Y=0.72X+1.91 | 3.5 | 0.998 | 6-549 | | Plasma(y), Serum(x) | ADVIA 1650 | 53 | Y=1.08X+1.46 | 3.9 | 0.985 | 18-110 | | Urine: MSK | CX7 | 82 | Y=0.75X+0.74 | 13.4 | 0.993 | 2-667 | Bias expected in serum and urine correlation when run against Beckman instrument due to the different substrates used in each instrument. #### Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|------| | Bilirubin | 25 | 68.3 | -1.6 | -2.3 | | Hemoglobin | 500 | 174.3 | -2.3 | -1.3 | | Lipids (Trig) | 500 | 66.3 | 1.4 | 2.1 | ## Analytical Range | Serum/Plasma: | 0 to 1500 U/L (from reagent stability study) | |---------------|----------------------------------------------| | Urine | 0 to 1500 U/L (from reagent stability study) | | Serum: | 20-104 U/L | |--------|------------------| | Urine: | 1-17 U/h (timed) | {4}------------------------------------------------ ## 4. AST ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Aspartate Aminotransferase Method for the Bayer ADVIA 1650 (U/L) #### Intended Use This in vitro diagnostic assay is intended to measure aspartate aminotransferase activity in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used mainly to determine the progress and prognosis of patients with myocardial infarction and the diagnosis and monitoring of liver disease. #### Imprecision | Advia 1650 | | | |---------------|-------|--------------| | Specimen type | Level | Total CV (%) | | Serum | 32 | 3.9 | | Serum | 146 | 1.7 | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type: Site | Comparison<br>System (X) | N | Regression<br>Equation | Syx | R | Sample<br>Range | |---------------------|--------------------------|-----|------------------------|------|-------|-----------------| | Serum: BERLIN | DAX | 111 | $Y=0.99X-6.3$ | 3.63 | 0.999 | 9.8-607.2 | | Plasma(y), Serum(x) | ADVIA 1650 | 54 | $Y=1.00X-2.9$ | 3.01 | 0.954 | 16-85 | #### Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|-------| | Bilirubin | 25 | 31.0 | -3.7 | -11.9 | | Hemoglobin | 500 | 74.7 | 18.6 | 24.9 | | Lipids (Trig) | 500 | 29.7 | -4.4 | -14.8 | Hemolyzed samples should not be used due to effect by RBC. ### Analytical Range 0-1000 U/L (from reagent stability study) ## Expected Values M: 15 - 40 U/L. F: 13 - 35 U/L. {5}------------------------------------------------ ## 5. Ca ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Calcium Method for the Bayer ADVIA 1650 (mg/dL) #### Intended Use This in vitro diagnostic assay is intended to measure calcium concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease, and tetany. ### Imprecision | Advia 1650 | | | Technicon DAX = Serum | | Dupont Dimension=Urine | | |------------------|-------|-----------------|-----------------------|--------------|------------------------|--------------------| | Specimen<br>type | Level | Total<br>CV (%) | Level | Total CV (%) | Level | Between Day CV (%) | | Serum | 5.9 | 2.7 | 7.7 | 2.3 | 8.8 | 0.6 | | Serum | 10.8 | 2.9 | 10.3 | 1.6 | 14.5 | 2.6 | | Serum | 12.0 | 3.5 | 12.5 | 1.5 | | | | Urine | 6.2 | 2.4 | | | | | | Urine | 21.2 | 2.5 | | | | | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|-------|-------|--------------| | Serum - ARI | DAX | 100 | $Y=0.99X + 0.13$ | 0.223 | 0.971 | 7.0 - 13.2 | | Plasma(y), Serum(x) | ADVIA 1650 | 59 | $Y=0.94X + 0.82$ | 0.094 | 0.963 | 9.0 - 10.8 | | Urine - ARI | Dimension | 32 | $Y=1.2X - 2.25$ | 0.165 | 0.999 | 3.7 - 14.5 | | Urine - MSK | CX3 | 63 | $Y=1.07X + 0.03$ | 0.558 | 0.988 | 2.1 - 14.7 | ### Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect<br>conc. | % | |--------------------------|--------------------------------------|------------------|-----------------|-----| | Bilirubin | 25 | 8.0 | 0.0 | 0.0 | | Hemoglobin | 500 | 8.8 | 0.2 | 2.3 | | Lipids (Trig) | 500 | 8.0 | 0.3 | 3.8 | ## Analytical Range | Serum/Plasma: | 1 to 15 mg/dL | |---------------|---------------| | Urine: | 1 to 15 mg/dL | | Serum/Plasma: | 8.6 to 10.0 mg/dL | |---------------|-------------------| | Urine: | 100 to 300 mg/d | {6}------------------------------------------------ ## SUMMARY OF SAFETY AND EFFECTIVENESS ### Chloride Method for the Bayer ADVIA 1650 (mmol/L) #### Intended Use This in vitro diagnostic assay is intended to measure chloride concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used for their inferential value and are helpful in diagnosing disorders of acid-base and water balance. It is especially important to measure chloride during the correction of hypokalemic alkalosis and also during severe, prolonged vomiting, which can lower the serum chloride level. ### Imprecision | Advia 1650 | | | |------------------|-------|-----------------| | Specimen<br>type | Level | Total<br>CV (%) | | Serum | 88.4 | 1.6 | | Serum | 111.8 | 2.0 | | Urine | 86.8 | 1.7 | | Urine | 290.4 | 3.8 | Beckman CX7=Serum,Urine Total CV (%) Level 80 1.3 118.8 1.1 89.4 3.4 250 ાં ર Technicon DAX=Serum | Level | Total CV (%) | |-------|--------------| | 97 | 1.5 | | 112 | 1.8 | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|------|-------|--------------| | Serum - ARI | CX7 | 115 | $Y=1.00X + 1.3$ | 1.6 | 0.985 | 76 - 128 | | Serum - MSK | DAX | 154 | $Y=0.98X + 3.3$ | 2.2 | 0.914 | 86 - 122 | | Plasma(y), Serum(x) | ADVIA 1650 | 69 | $Y=0.86X+15.9$ | 1.20 | 0.944 | 102-118 | | Urine - ARI | CX7 | 76 | $Y=1.05X - 0.9$ | 4.9 | 0.996 | 29 - 280 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|-----| | Bilirubin | 25 | 85.0 | 0.4 | 0.5 | | Hemoglobin | 500 | 82.7 | 0.0 | 0.0 | | Lipids (Trig) | 500 | 83.3 | 2.4 | 2.9 | ## Analytical Range | Serum/Plasma: | 15 to 200 mmol/L | |---------------|------------------| | Urine: | 15 to 400 mmol/L | | Serum: | 98 - 107 mEq/L | |--------|------------------| | Urine: | 110 - 250 mmol/d | {7}------------------------------------------------ # 7. СК ## SUMMARY OF SAFETY AND EFFECTIVENESS ### Creatine Phosphokinase Method for the Bayer ADVIA 1650 (U/L) #### Intended Use This in vitro diagnostic assay is intended to measure creatine kinase activity in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used mainly in the diagnosis and treatment of myocardial infarction and muscle diseases such as Duchenne progressive muscular dystrophy. ### Imprecision | Advia 1650 | Technicon DAX | | | Beckman CX7 | | | |------------------|---------------|-----------------|-------------|-----------------|-------|-----------------| | Specimen<br>type | Level | Total<br>CV (%) | Level (U/L) | Total<br>CV (%) | Level | Total<br>CV (%) | | Serum | 139 | 2.7 | 177 | 4.8 | 144 | 5.3 | | Serum | 449 | 3.0 | 590 | 2.5 | 460 | 5.3 | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type: Site | Comparison<br>System (X) | N | Regression<br>Equation | Syx | R | Sample<br>Range | |---------------------|--------------------------|-----|------------------------|------|-------|-----------------| | Serum: MSK | DAX | 151 | Y=0.97X+1.19 | 3.3 | 0.999 | 6-700 | | Serum: ARI | CX7 | 145 | Y=1.01X+0.8 | 6.50 | 1.000 | 13-1187 | | Plasma(y), Serum(x) | ADVIA 1650 | 56 | Y=1.05X-9.02 | 7.88 | 0.997 | 41-507 | #### Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|------| | Bilirubin | 25 | 70.0 | 2.7 | 3.9 | | Hemoglobin | 500 | 98.0 | 24.3 | 24.8 | | Lipids (Trig) | 500 | 65.0 | 6.0 | 9.2 | ## Analytical Range 0-1300 U/L ## Expected Values M: 38 - 174 U/L. F: 26 - 140 U/L. {8}------------------------------------------------ # 8. CREAT ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Creatinine Method for the Bayer ADVIA 1650 (mg/dL) #### Intended Use This in vitro diagnostic assay is intended to measure creatinine concentration in human serum, plasma and urine on the ADVIA® 1650 system. Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis. #### Imprecision | Advia 1650 | | | | Technicon DAX=Serum | | | |---------------|-------|------------|--|---------------------|--------------|--| | Specimen type | Level | Total CV % | | Level | Total CV (%) | | | Serum | 1.8 | 3.8 | | 1.8 | 4.3 | | | Serum | 8.4 | 3.7 | | 5.0 | 4.5 | | | Urine | 84.6 | 2.9 | | 10.5 | 3.7 | | | Urine | 209.3 | 4.0 | | | | | ## Correlation (Y=ADVIA 1650, X=comparison system) | | Comparison | Regression | | Sample | | | |---------------------|------------|------------|--------------|--------|-------|----------| | Specimen type: Site | System (X) | N | Equation | Syx | R | Range | | Serum: BERLIN | DAX | 112 | Y=0.92X+0.06 | 0.15 | 0.995 | 0.33-8.3 | | Plasma(y), Serum(x) | ADVIA 1650 | 58 | Y=1.02X+0.04 | 0.02 | 0.992 | 0.9-1.4 | | Urine: BERLIN | HITACHI | 86 | Y=1.13X+0.40 | 2.61 | 0.995 | 21-160 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect<br>conc. | Effect<br>% | |--------------------------|--------------------------------------|------------------|-----------------|-------------| | Bilirubin | 25 | 1.0 | -0.8 | -80.0 | | Hemoglobin | 500 | 7.3 | 0.1 | 1.4 | | Lipids (Trig) | 500 | 1.0 | 0.0 | 4.5 | ## Analytical Range | Serum/Plasma: | 0 - 25 mg/dL | |---------------|----------------| | Urine: | 17 - 160 mg/dL | | Serum: | M: 0.9 - 1.3 mg/dL | |--------|---------------------| | | F: 0.6 - 1.1 mg/dL. | | Urine: | M: 14 - 26 | | | F: 11 - 20 | {9}------------------------------------------------ ## 9. CREAT-E ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Creatinine-Enzymatic Method for the Bayer ADVIA 1650 (mg/dL) #### Intended Use This in vitro diagnostic assay is intended to measure creatinine concentration in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of renal diseases, and in monitoring renal dialysis. ## Imprecision | Advia 1650 | | | | Technicon DAX | | |------------------|-------|-----------------|-------|-----------------|--| | Specimen<br>type | Level | Total<br>CV (%) | Level | Total<br>CV (%) | | | Serum | 0.9 | 6.4 | 1.8 | 4.3 | | | Serum | 6.1 | 2.3 | 5.0 | 4.5 | | | Serum | 9.5 | 2.0 | 10.5 | 3.7 | | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type: Site | Comparison<br>System (X) | N | Regression<br>Equation | Syx | R | Sample<br>Range | |---------------------|--------------------------|-----|------------------------|------|-------|-----------------| | Serum: MSK | DAX | 155 | Y=0.99X-0.06 | 0.16 | 0.955 | 0.3-3.9 | | Plasma(y), Serum(x) | ADVIA 1650 | 57 | Y=0.96X+0.16 | 0.07 | 0.925 | 0.7-1.4 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|--------| | | | | conc. | % | | Bilirubin | 25 | 0.7 | -0.2 | -28.6 | | Hemoglobin | 500 | 7.8 | -0.1 | -1.3 | | Lipids (Trig) | 500 | 0.7 | -1.2 | -171.4 | ### Analytical Range 0-30 mg/dL ## Expected Values M: 0.9 - 1.3 mg/dL, F: 0.6 - 1.1 mg/dL. {10}------------------------------------------------ ## 10. DBILI ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Direct Bilirubin Method for the Bayer ADVIA 1650 (mg/dL) ### Intended Use This in vitro diagnostic assay is intended to measure direct bilirubin concentration in human serum and plasma on the ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis of common bile duct obstruction caused by a stone and of patients with Dubin-Johnson syndrome. ### Imprecision A do ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . | Advia 1650 | | Technicon DAX | | | |------------------|-------|-----------------|-------|-----------------| | Specimen<br>type | Level | Total<br>CV (%) | Level | Total<br>CV (%) | | Serum | 0.3 | 4.6 | 0.2 | ** | | Serum | 1.6 | 4.4 | 1.5 | 5.7 | | | | | 4.0 | 2.3 | ** Value not significant when level approaches zero. ### Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |-----------------------|-----------------------|-----|---------------------|------|-------|--------------| | Serum - TRYTN | DAX | 145 | Y=1.34X - 0.10 | 0.18 | 0.994 | 0 - 6.12 | | Serum - BERLIN | HITACHI | 92 | Y=1.07X + 0.06 | 0.14 | 0.997 | 0.02 - 9.26 | | Plasma (y), Serum (x) | ADVIA 1650 | 49 | Y=0.98X-0.01 | 0.02 | 0.988 | 0.1-0.6 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | % | |--------------------------|--------------------------------------|------------------|--------|---| | Hemoglobin | 200 | 0.3 | -0.6 | . | | Hemoglobin | 500 | 0.3 | -1.5 | . | | Lipids (Trig) | 500 | 0.2 | -0.5 | . | ## Analytical Range 0 - 10 mg/dL (from reagent stability study) ## Expected Values <0.2 mg/dL {11}------------------------------------------------ ## 11. GLU-HEX ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Glucose-Hexokinase Method for the Bayer ADVIA 1650 (mg/dL) #### Intended Use This in vitro diagnostic assay is intended to measure glucose in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and insulin overdose. ## Imprecision | Advia 1650 | | | | Technicon DAX=Serum | |------------|-------|-------|-------|---------------------| | Specimen | | Total | | Total | | type | Level | CV | Level | CV (%) | | Serum | 77 | 2.4 | 71 | 4.2 | | Serum | 279 | 3.3 | 262 | 2.1 | | Urine | 42 | 3.5 | 367 | 1.7 | | Urine | 285 | 3.6 | | | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison<br>System (X) | N | Regression Equation | Syx | R | Sample Range | |----------------------|--------------------------|-----|---------------------|------|-------|--------------| | Serum - BERLIN | DAX | 109 | Y=1.05X - 2.8 | 5.42 | 0.995 | 19-262 | | Urine - BERLIN | HITACHI | 81 | Y=0.99X - 11.2 | 9.88 | 0.997 | 4-476 | | Plasma (y) Serum (x) | ADVIA 1650 | 60 | Y=1.03X - 6.6 | 9.00 | 0.986 | 36 - 384 | #### Interfering Substances | Analyte | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |---------|--------------------------|--------------------------------------|------------------|--------|-------| | | | | | conc. | % | | GLU-HEX | Bilirubin | 10 | 95.5 | -8.2 | -8.6 | | GLU-HEX | Bilirubin | 25 | 95.5 | -24.4 | -25.5 | | GLU-HEX | Hemoglobin | 200 | 95.9 | -22.2 | -23.1 | | GLU-HEX | Hemoglobin | 500 | 95.9 | -59.9 | -62.5 | | GLU-HEX | Lipids (Trig) | 900 | 189.0 | 2.2 | 1.2 | ## Analytical Range | Serum/Plasma: | 0 - 700 mg/dL | |---------------|---------------| | Urine: | 0 - 700 mg/dL | | Serum/Plasma | 74 -- 106 mg/dL | |--------------|-----------------| | Urine: | <0.5 g/day | {12}------------------------------------------------ # 12. GLU-OX ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Glucose-Oxidase Method for the Bayer ADVIA 1650 (mg/dL) #### Intended Use This in vitro diagnostic assay is intended to measure glucose in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia and insulin overdose. #### Imprecision | Advia 1650 | | | Technicon DAX=Serum | | |------------------|-------|-----------------|---------------------|-----------------| | Specimen<br>type | Level | Total<br>CV (%) | Level | Total<br>CV (%) | | Serum | 77 | 1.5 | 74 | 2.1 | | Serum | 294 | 1.4 | 293 | 1.5 | | Urine | 48 | 1.5 | | | | Urine | 277 | 1.4 | | | #### Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type: Site | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|------|-------|--------------| | Serum: MSK | DAX | 155 | Y=0.98X-3.0 | 3.4 | 0.996 | 41-323 | | Plasma(y), Serum(x) | ADVIA 1650 | 48 | Y=0.99X+2.9 | 2.21 | 0.999 | 60-381 | | Urine: MSK | CX3 | 87 | Y=1.06X-1.2 | 4.8 | 0.989 | 0-249 | #### Interfering Substances | Analyte | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect<br>conc. | Effect<br>% | |---------|--------------------------|--------------------------------------|------------------|-----------------|-------------| | GLU-OX | Bilirubin | 10 | 91.9 | -3.5 | -3.8 | | GLU-OX | Bilirubin | 25 | 91.9 | -13.6 | -14.8 | | GLU-OX | Hemoglobin | 500 | 86.7 | 6.1 | 7.0 | | GLU-OX | Lipids (Trig) | 900 | 187.0 | -5.8 | -3.1 | ## Analytical Range | Serum/Plasma: | 0 - 750 mg/dL | |---------------|---------------| | Urine: | 0 - 750 mg/dL | | Serum: | 74 - 106 mg/dL | |--------|----------------| | Urine: | <0.5 g/day | {13}------------------------------------------------ ## 13. HDL ## SUMMARY OF SAFETY AND EFFECTIVENESS ## High Density Lipoprotein Method for the Bayer ADVIA 1650 (mg/dL) #### Intended Use This in vitro diagnostic assay is intended to measure HDL Cholesterol in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used in the risk assessment of coronary artery disease. #### Imprecision | Advia 1650 | | | RA-XT | | |---------------|-------|--------------|-------|--------------------| | Specimen type | Level | Total CV (%) | Level | Between Day CV (%) | | Serum | 28 | 3.8 | 15.5 | 5.1 | | Serum | 53 | 3.2 | 52.3 | 1.9 | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type: Site | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|------|-------|--------------| | Serum: MSK | RA-XT | 121 | $Y=1.03X+3.91$ | 2.5 | 0.987 | 12-94 | | Plasma(y), Serum(x) | ADVIA 1650 | 79 | $Y=1.09X-1.9$ | 4.50 | 0.960 | 26-98 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect<br>conc. | Effect<br>% | |--------------------------|--------------------------------------|------------------|-----------------|-------------| | Bilirubin | 25 | 29.7 | -0.4 | -1.3 | | Hemoglobin | 500 | 29.2 | -5.6 | -19.2 | ## Analytical Range 10-135 mg/dL ## Expected Values (years: mg/dL) Male, Ages 30 - 50: 27.8 to 52.89 mg/dL Female, Ages 30 - 50: 33.9 to 86.87 mg/dL {14}------------------------------------------------ ## 14. IPHOS ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Inorganic Phosphorus Method for the Bayer ADVIA 1650 (mg/dL) #### Intended Use This in vitro diagnostic assay is intended to measure inorganic phosphorus concentration in human serum, plasma and urine on the ADVIA® 1650 system. Such measurements are used in the diagnosis and treatment of kidney diseases, parathyroid gland disorders, and Vitamin D imbalance. ### Imprecision | Advia 1650 | | | Technicon DAX=Serum | | |---------------|-------|--------------|---------------------|--------------| | Specimen type | Level | Total CV (%) | Level | Total CV (%) | | Serum | 3.0 | 1.7 | 3.2 | 3.0 | | Serum | 5.7 | 2.6 | 6.7 | 2.3 | | Serum | 7.7 | 2.0 | 9.4 | 2.1 | | Urine | 32.0 | 1.6 | | | | Urine | 170.6 | 0.7 | | | ### Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type: Site | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|------|-------|--------------| | Serum: MSK | DAX | 154 | Y=0.93X+0.20 | 0.18 | 0.978 | 1.1-7.3 | | Serum: ARI | CX7 | 152 | Y=1.01X+0.30 | 0.25 | 0.983 | 1.4-10.0 | | Plasma(y), Serum(x) | ADVIA 1650 | 50 | Y=0.98X-0.23 | 0.11 | 0.975 | 2.0-4.1 | | Urine: MSK | Cobas Fara | 92 | Y=1.02X-0.2 | 1.3 | 0.998 | 5-87 | ## Interfering Substances | Analyte | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | conc. | Effect<br>% | |---------|--------------------------|--------------------------------------|------------------|-------|-------------| | IPHOS | Bilirubin | 25 | 3.4 | 0.0 | -0.6 | | IPHOS | Hemoglobin | 500 | 2.7 | 0.9 | 33.3 | | IPHOS | Lipids (Trig) | 500 | 3.4 | -0.2 | -5.9 | ## Analytical Range Serum/Plasma: 0 - 20 mg/dL Urine: 5 - 100 mg/dL | Serum: | 2.7 - 4.5 mg/dL | |--------|-----------------| | Urine: | 0.4 - 1.3 g/d | {15}------------------------------------------------ ## 15. IRON ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Iron Method for the Bayer ADVIA 1650 (µg/dL) #### Intended Use This in vitro diagnostic method is intended to measure iron concentration in human serum and plasma on an ADVIA® 1650 Chemistry System. Measurements are used in the diagnosis and treatment of iron deficiency anemias and hemochromatosis. ### Imprecision | Advia 1650 | | | Technicon DAX | | | |------------------|-------|-----------------|---------------|-----------------|--| | Specimen<br>type | Level | Total<br>CV (%) | Level | Total<br>CV (%) | | | Serum | 64 | 3.1 | 81 | 5.5 | | | Serum | 175 | 2.6 | 194 | 3.2 | | | Serum | 295 | 2.2 | 322 | 3.1 | | ### Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|----|---------------------|-----|-------|--------------| | Serum - MSK | DAX | 40 | $Y=0.88X + 4.5$ | 4.8 | 0.997 | 6 - 253 | | Plasma(y), Serum(x) | ADVIA 1650 | 57 | $Y=0.96X+15.5$ | 6.9 | 0.978 | 40 - 186 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|-------| | Bilirubin | 25 | 111.6 | 1.9. | 1.7 | | Hemoglobin | 500 | N/A | N/A | N/A | | Lipids (Trig) | 500 | 107.9 | -11.1 | -10.3 | ## Analytical Range 0 to 1000 µg/dL ## Expected Values Males: 65 to 175 µg/dL Females: 50 to 170 µg/dL {16}------------------------------------------------ ## 16. Mg ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Magnesium Method for the Bayer ADVIA 1650 (mg/dL) ### Intended Use This in vitro diagnostic assay is intended to measure magnesium concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of hypermagnesemia and monitoring of patients receiving prolonged magnesium-free intravenous therapy. ### Imprecision | Advia 1650 | | Technicon DAX= Serum | | | |------------------|-------|----------------------|-------|-----------------| | Specimen<br>type | Level | Total<br>CV (%) | Level | Total<br>CV (%) | | Serum | 1.9 | 2.3 | 2.4 | 5.6 | | Serum | 3.4 | 2.0 | 3.9 | 3.6 | | Urine | 8.7 | 11.9 | | | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|-------|-------|--------------| | Serum - TRYTN | DAX | 193 | Y=0.91X + 0.10 | 0.131 | 0.973 | 0.9 - 5.3 | | Serum - BERLIN | HITACHI | 112 | Y=0.95X + 0.25 | 0.124 | 0.916 | 1.5 - 3.1 | | Plasma(y), Serum(x) | ADVIA 1650 | 51 | Y=0.97X + 0.02 | 0.032 | 0.963 | 1.6 - 2.1 | | Urine - BERLIN | HITACHI | 97 | Y=0.99X + 0.31 | 0.253 | 0.997 | 1.4 - 15.0 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect<br>conc. | Effect<br>% | |--------------------------|--------------------------------------|------------------|-----------------|-------------| | Bilirubin | 25 | 2.6 | 0.4 | 14.0 | | Hemoglobin | 500 | 2.7 | 0.2 | 7.4 | | Lipids (Trig) | 500 | 1.8 | 1.0 | 55.6 | ## Analytical Range | Serum/Plasma: | 1.6 to 6 mg/dL | |---------------|-----------------| | Urine: | 1.6 to 25 mg/dL | ## Expected Values (values in Tietz converted to mg/dL) | Serum/Plasma: | 1.6 – 2.6 mg/dL | |---------------|-----------------| | Urine: | 7.3 – 12.2 mg/d | {17}------------------------------------------------ # 17. K ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Potassium Method for the Bayer ADVIA 1650 (mmol/L) #### Intended Use This in vitro diagnostic assay is intended to measure potassium concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used mainly to monitor electrolyte balance in the diagnosis and treatment of primary aldosteronism, metabolic alkalosis, diarrhea, severe vomiting, diuretic administration, diabetic ketoacidosis, and other diseases. #### Imprecision | Advia 1650 | | | Technicon DAX=Serum | | |---------------|-------|--------------|---------------------|--------------| | Specimen type | Level | Total CV (%) | Level | Total CV (%) | | Serum | 3.0 | 3.0 | 3.4 | 1.6 | | Serum | 6.4 | 2.7 | 7.5 | 1.7 | | Urine | 28.5 | 6.1 | | | | Urine | 102.9 | 4.8 | | | #### Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|------|-------|--------------| | Serum - MSK | DAX | 156 | Y=1.09X - 0.2 | 0.08 | 0.995 | 2.2 - 8.1 | | Plasma(y), Serum(x) | ADVIA 1650 | 49 | Y=0.92X-0.01 | 0.16 | 0.890 | 3.4-4.8 | | Urine - MSK | CX3 | 99 | Y=1.01X-0.2 | 1.2 | 0.999 | 7 - 116 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|-----| | Bilirubin | 25 | 3.3 | 0.0 | 0.3 | | Hemoglobin | 500 | N/A | N/A | N/A | | Lipids (Trig) | 500 | 3.3 | 0.2 | 6.1 | ## Analytical Range | Serum/Plasma: | 1 to 10 mmol/L | |---------------|-----------------| | Urine: | 3 to 300 mmol/L | | Serum: | 3.5 – 5.1 mmol/L | |-----------------|----------------------------------| | Plasma Males: | 3.5-4.5 mmol/L | | Plasma Females: | 3.4-4.4 mmol/L | | Urine: | 25-125 mmol/d (varies with diet) | {18}------------------------------------------------ ## 18. Na ## SUMMARY OF SAFETY AND EFFECTIVENESS ### Sodium Method for the Bayer ADVIA 1650 (mmol/L) ### Intended Use This in vitro diagnostic assay is intended to measure sodium concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used mainly in the diagnosis and treatment of gross changes in water and salt balance, aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic hormone secretion, diabetic acidosis, severe diarrhea, or other diseases involving electrolyte imbalance. ### Imprecision | Advia 1650 | | | | Technicon DAX=Serum | |------------------|-------|--------------|-------|---------------------| | Specimen<br>type | Level | Total CV (%) | Level | Total CV (%) | | Serum | 119 | 1.7 | 115 | 0.9 | | Serum | 143 | 1.8 | 141 | 0.9 | | Urine | 63 | 1.9 | 166 | 1.2 | | Urine | 208 | 2.0 | | | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|------|-------|--------------| | Serum - MSK | DAX | 156 | $Y=1.02X - 0.2$ | 1.08 | 0.970 | 121 - 154 | | Plasma(y), Serum(x) | ADVIA 1650 | 69 | $Y=0.92X+11.6$ | 1.45 | 0.964 | 139 - 156 | | Urine - MSK | CX3 | 97 | $Y=0.95X + 6.2$ | 3 | 0.997 | 20 - 202 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|-----| | Bilirubin | 25 | 122.8 | 0.0 | 0.0 | | Hemoglobin | 500 | 128.7 | 2.3 | 1.8 | | Lipids (Trig) | 500 | 117.3 | 3.1 | 2.6 | ## Analytical Range | Serum/Plasma: | 100 to 200 mmol/L | |---------------|-------------------| | Urine: | 10 to 400 mmol/L | | Serum/Plasma: | 136 - 145 mEq/L | |---------------|-----------------| | Urine: | 40 - 220 mmol/d | {19}------------------------------------------------ ## 19. TBILI ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Total Bilirubin Method for the Bayer ADVIA 1650 (mg/dL) #### Intended Use This in vitro diagnostic assay is intended to measure total bilirubin concentration in human serum and plasma on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of hemolytic, biliary, and liver disorders, including hepatitis and cirrhosis. ### Imprecision | Advia 1650 | | Technicon DAX | | | |------------------|-------|-----------------|-------|-----------------| | Specimen<br>type | Level | Total<br>CV (%) | Level | Total<br>CV (%) | | Serum | 1.0 | 7.6 | 0.7 | ** | | Serum | 5.0 | 4.1 | 5.1 | 2.3 | | Serum | 7.7 | 2.2 | 12.5 | 2.2 | ** Value not significant when level approaches zero. #### Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------|-----------------------|-----|---------------------|-------|-------|--------------| | Serum - MSK | DAX | 156 | $Y=1.10X + 0.08$ | 0.194 | 0.999 | 0.1 - 29.5 | | Serum/Plasma | ADVIA 1650 | 51 | $Y=0.94X + 0.16$ | 0.071 | 0.954 | 0.3 - 1.4 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|--------| | Hemoglobin | 200 | 0.9 | -0.5 | -61.6 | | Hemoglobin | 500 | 0.9 | -1.1 | -122.1 | | Lipids (Trig) | 500 | 0.9 | 0.5 | 55.6 | ## Analytical Range 0 to 30 mg/dL #### Expected Values 0.3 to 1.2 mg/dL {20}------------------------------------------------ ## SUMMARY OF SAFETY AND EFFECTIVENESS ## Uric Acid Method for the Bayer ADVIA 1650 (mg/dL) #### Intended Use This in vitro diagnostic assay is intended to measure uric acid concentration in human serum, plasma and urine on an ADVIA® 1650 Chemistry System. Such measurements are used in the diagnosis and treatment of renal failure, gout, and eclampsia. #### Imprecision | Advia 1650 | | | |------------------|-------|-----------------| | Specimen<br>type | Level | Total<br>CV (%) | | Serum | 3.9 | 1.9 | | Serum | 8.6 | 1.6 | | Serum | 10.0 | 2.3 | | Urine | 12.4 | 2.3 | | Urine | 23.9 | 5.2 | | Technicon DAX=Serum | | Beckman CX7=Urine | | |---------------------|--------------|-------------------|--------------| | Level | Total CV (%) | Level | Total CV (%) | | 5.2 | 3.0 | 12 | 3 | | 8.5 | 2.7 | 24 | 3 | | 10.0 | 2.6 | | | ## Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|-----|---------------------|-------|-------|--------------| | Serum – MSK | DAX | 154 | $Y=1.05X + 0.48$ | 0.27 | 0.994 | 0.2 – 18.0 | | Plasma(y), Serum(x) | ADVIA 1650 | 61 | $Y=1.03X + 0.02$ | 0.063 | 0.999 | 2.4 – 8.9 | | Urine – ARI | CX7 | 30 | $Y=1.03X - 0.5$ | 3.06 | 0.989 | 8 – 91 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|------| | Bilirubin | 25 | 4.5 | -0.2 | -4.2 | | Hemoglobin | 500 | 6.2 | 0.2 | 3.2 | | Lipids (Trig) | 500 | 4.3 | -0.4 | -9.3 | ## Analytical Range | Serum/Plasma: | 0 to 20 mg/dL | |---------------|----------------| | Urine: | 0 to 180 mg/dL | | Males/Serum: | 3.5 to 7.2 mg/dL | |----------------|------------------| | Females/Serum: | 2.6 to 6.0 mg/dL | | Urine: | 250 to 750 mg/dL | {21}------------------------------------------------ ## 21. UN # SUMMARY OF SAFETY AND EFFECTIVENESS ## Urea Nitrogen Method for the Bayer ADVIA 1650 (mg/dL) #### Intended Use This in vitro diagnostic assay is intended to measure urea nitrogen (an end product of nitrogen metabolism) concentration in human serum, plasma and urine on a ADVIA® 1650 Chemistry system. Such measurements are used in the diagnosis and treatment of kidney disease, urinary tract obstruction, and acute or chronic renal failure. #### Imprecision | Specimen<br>type | Level | Total<br>CV | |------------------|-------|-------------| | Serum | 18 | 2.4 | | Serum | 50 | 2.2 | | Serum | 86 | 1.7 | | Urine | 461 | 4.6 | | Urine | 618 | 2.3 | | Technicon DAX=Serum | | Beckman CX3=Urine | | |---------------------|--------------|-------------------|--------------| | Level | Total CV (%) | Level | Total CV (%) | | 21 | 3.5 | 460 | 7.5 | | 59 | 1.8 | 620 | 7.5 | | 102 | 1.7 | | | ### Correlation (Y=ADVIA 1650, X=comparison system) | Specimen type | Comparison System (X) | N | Regression Equation | Syx | R | Sample Range | |---------------------|-----------------------|----|---------------------|------|-------|--------------| | Serum - MSK | DAX | 34 | Y=1.12X - 0.8 | 1.01 | 0.997 | 8 - 64 | | Plasma(y), Serum(x) | ADVIA 1650 | 54 | Y=1.0X - 0.2 | 0.22 | 0.997 | 7 - 22 | | Urine - MSK | CX3 | 79 | Y=1.03X + 24.8 | 34.8 | 0.990 | 67 - 988 | ## Interfering Substances | Interfering<br>Substance | Interfering<br>Sub. Conc.<br>(mg/dL) | Analyte<br>conc. | Effect | | |--------------------------|--------------------------------------|------------------|--------|------| | | | | conc. | % | | Bilirubin | 25 | 16.3 | -0.2 | -1.2 | | Hemoglobin | 500 | 20.2 | -0.6 | -3.0 | | Lipids (Trig) | 500 | 15.8 | 0.7 | 4.4 | #### Analytical Range | Serum/Plasma: | 5 to 150 mg/dL | |---------------|------------------| | Urine: | 35 to 1000 mg/dL | ## Expected Values 6 to 20 mg/dL {22}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/22/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them. JUN 30 1999 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mr. Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation Business Diagnostic Division 511 Benedict Avenue Tarrytown, New York 10591-5097 Re: K991576 Trade Name: Bayer ADVIA® 1650 Chemistry System Regulatory Class: II Product Code: CIX, CJK, KHM, CIT, CIC, CGX, JFY, CGS, CIG, CFR, CGA, LBS, CDQ Regulatory Class: I Product Code: CEO, JIY, CFO, KNK Dated: April 29, 1999 Received: May 6, 1999 Dear Mr. Muraca: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {23}------------------------------------------------ #### Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the GMice of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {24}------------------------------------------------ Attachment 8 Page 1 of 1 510(k) Number (if known): K991576 Device Name: Bayer ADVIA® 1650 Chemistry System Indications For Use: The ADVIA 1650 Chemistry System is an automated, clinical chemistry analyzer that can run tests on human serum, plasma, or urine in random access, batch, and STAT modes at a throughput rate of 1200 photometric tests per hour and 450 electrolyte tests per hour. The photometric analyzer performs clinical chemistry and immuno-turbidimetric methods. The electrylyte portion of the analyzer measures the sodium, and chloride concentrations in serum, plasma or urine samples based on a potentiometric procedure that uses ion-selective electrodes. The ADVIA 1650 is intended for use in conjunction with certain reagents to measure a variety of analytes contained in human fluids. Sean Cooper (Division Sign-Off) Division of Clinical Laboratory Devices 991576 310(k) Number (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...