TERUMO SURFLO FLEX I.V. CATHETER AND TERUMO SURFLASH I.V. CATHETER
K991406 · Terumo Medical Corp. · FOZ · Jun 16, 1999 · General Hospital
Device Facts
Record ID
K991406
Device Name
TERUMO SURFLO FLEX I.V. CATHETER AND TERUMO SURFLASH I.V. CATHETER
Applicant
Terumo Medical Corp.
Product Code
FOZ · General Hospital
Decision Date
Jun 16, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TERUMO® SURFLO® FLEX I.V. Catheters and the TERUMO® SURFLASH™ I.V. Catheters are devices consisting of a slender, flexible, radio-opaque, plastic catheter with a hub that is inserted into the patient's vascular system for short term use to withdraw samples, administer fluid intravenously or through which to place monitoring equipment such as blood pressure monitors. The stainless steel cannula, placed in the catheter to maintain rigidity, is withdrawn after the catheter is placed in the vascular system.
Device Story
Device is a sterile, single-use, over-the-needle peripheral intravenous catheter. Components include a polyurethane catheter, stainless steel cannula, catheter hub, and filter cap with air filter. SURFLASH variant features a grooved cannula to allow visual flashback detection at the catheter tip. Device is inserted manually by a clinician into the patient's vascular system; cannula is withdrawn after placement. Catheter remains in the vein for short-term use (24 hours to 30 days). Output is vascular access for fluid delivery, blood withdrawal, or monitoring. Benefits include ease of insertion, flexibility during retention, and radio-opacity for detection. Used in clinical settings.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included catheter body tensile strength, stiffness, elongation, burst pressure, collapse, flexural fatigue, radio-detectability, flow rate, corrosion resistance, fitting strength, and flashback detection. Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
Materials: Polyurethane catheter, stainless steel cannula/caulking pin, polypropylene hub, polystyrene filter cap, polyester air filter. Radio-opaque barium sulfate stripes. Dimensions per ISO 10555-5. Sterilization: Ethylene Oxide (EtO) per EN 550. Manual operation; no software or electronic components.
Indications for Use
Indicated for short-term vascular access in patients requiring intravenous fluid administration, blood sampling, or placement of monitoring equipment (e.g., blood pressure monitors).
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
JUN 16 1999
K99/406
## SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION PERTAINING TO SUBSTANTIAL EQUIVALENCE
Proprietary Device Name: TERUMO® SURFLO® FLEX I.V. Catheter TERUMO® SURFLASH™ I.V. Catheter, or similar proprietary name
Classification Name: Intravascular Catheter
### INTENDED USE
The TERUMO® SURFLO® FLEX I.V. Catheters and the TERUMO® SURFLASH™ I.V. Catheters are devices consisting of a slender, flexible, radio-opaque, plastic catheter with a hub that is inserted into the patient's vascular system for short term use to withdraw samples, administer fluid intravenously or through which to place monitoring equipment such as blood pressure monitors. The stainless steel cannula, placed in the catheter to maintain rigidity, is withdrawn after the catheter is placed in the vascular system.
#### DESCRIPTION
,
The TERUMO SURFLO FLEX I.V. CATHETER AND THE TERUMO SURFLASH I.V. CATHETER are sterile, single use devices consisting of a slender tube (catheter) made of polyurethane. The devices are inserted into the patient's vascular system for short-term use to withdraw samples, administer fluids intravenously, or through which to place monitoring equipment such as blood pressure monitors.
Each device consists of a catheter assembly-catheter, caulking pin, catheter hub; a needle assembly-cannula, needle hub; and a filter cap with an air filter
The catheters are made of polyurethane, which allows appropriate stiffness for proper insertions into the vein, and flexibility during retention in the vein. Each catheter contains barium sulfate radio-opaque stripes along the length of the catheter to allow radio detectability. A stainless steel caulking pin fastens the catheter to the catheter hub.
Each catheter hub is color coded according to the outside diameter as specified in ISO 10555-5, "Sterile, single-use intravascular catheters - Part 5: Over-needle peripheral catheters".
{1}------------------------------------------------
The needle assembly has the same specifications as the Terumo® Surflo® I.V. Catheter, which is the subject of K891087. The transparent needle hubs permit flashback detection when the cannula enters the vessel.
The filter cap with an air filter is provided as a vent fitting.
The Terumo Surflo Flex I.V. Catheter is available in 14 sizes with different combinations of 6 catheter diameters and 5 catheter lengths (see following specifications table). The Terumo Surflash I.V. Catheter is available in 24G by 19mm currently. Additional sizes will be implemented at a later date.
The cannula of the Surflash I.V. Catheter are grooved from the heel of the cannula tip bevel to the end of the catheter tip, which allows flashback detection at the catheter tip after the catheter tip enters the vessel.
The only difference between the TERUMO SURFLO FLEX and the TERUMO SURFLASH I.V. catheters is the grooved cannula feature of the SURFLASH. All other specifications are the same.
## SPECIFICATIONS/DIMENSIONS
| Catheter gauge | Color code | Product code | Catheter length | Catheter O.D. | Catheter I.D. | Cannula gauge | Flow rate |
|----------------|-------------|--------------|-----------------|---------------|---------------|---------------|-----------|
| 14G | Orange | SR*OF1451 | 2" (51mm) | 2.17mm | 1.73mm | 16G | 316ml/min |
| | | SR*OF1464 | 2-1/2" (64mm) | | | | 282ml/min |
| 16G | Medium Grey | SR*OF1632 | 1-1/4"(32mm) | 1.70mm | 1.30mm | 18G | 202ml/min |
| | | SR*OF1651 | 2"(51mm) | | | | 184ml/min |
| | | SR*OF1664 | 2-1/2" (64mm) | | | | 177ml/min |
| 18G | Deep Green | SR*OF1832 | 1-1/4"(32mm) | 1.30mm | 0.95mm | 20G | 99ml/min |
| | | SR*OF1851 | 2" (51mm) | | | | 88ml/min |
| | | SR*OF1864 | 2-1/2" (64mm) | | | | 80ml/min |
| 20G | Pink | SR*OF2025 | 1"(25mm) | 1.10mm | 0.80mm | 22G | 73ml/min |
| | | SR*OF2032 | 1-1/4"(32mm) | | | | 67ml/min |
| | | SR*OF2051 | 2"(51mm) | | | | 59ml/min |
| 22G | Deep Blue | SR*OF2225 | 1"(25mm) | 0.85mm | 0.60mm | 24G | 38ml/min |
| | | SR*OF2232 | 1-1/4"(32mm) | | | | 30ml/min |
| 24G | Yellow | SR*OF2419 | 3/4" (19mm) | 0.67mm | 0.47mm | 27G | 15ml/min |
SURFLO FLEX Specification
SURFLASH Specification
| Catheter gauge | Color code | Product code | Catheter length | Catheter O.D. | Catheter I.D. | Cannula gauge | Flow rate |
|----------------|------------|--------------|-----------------|---------------|---------------|---------------|-----------|
| 24G | Yellow | SR*FF2419 | 3/4" (19mm) | 0.67mm | 0.47mm | 27G | 15ml/min |
{2}------------------------------------------------
### SUBSTANTIAL EQUIVALENCE
The TERUMO® SURFLO® FLEX I.V. Catheters and the TERUMO® SURFLASH™ I.V. Catheters, submitted in this 510k, are substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared Becton Dickinson (B-D®) Insyte® I.V. Catheter and the B-D Insyte (N) I.V. Catheter, which are the subject of K971339 (unable to confirm); and to the Terumo® Surflo® I.V. Catheter which is the subject of K891087.
## PRINCIPLE OF OPERATION/TECHNOLOGY
The Terumo SURFLO FLEX and SURFLASH I.V. catheters, the B-D® Insyte and Insyte (N) I.V. catheters, and the Terumo Surflo I.V. catheters are all operated manually .
| COMPONENT | PROPOSED<br>(SURFLO FLEX &<br>SURFLASH) | PREDICATE<br>TERUMO SURFLO | PREDICATE INSYTE<br>& INSYTE (ND) |
|------------------------|-----------------------------------------|----------------------------|-----------------------------------|
| Cannula | Stainless steel | Stainless steel | Stainless steel |
| Catheter | Polyurethane | Teflon® | Polyurethane |
| Radio opaque<br>medium | Barium sulfate | Barium sulfate | Barium sulfate |
| Catheter hub | Polypropylene | Polypropylene | Unknown |
| Caulking pin | Stainless steel | Stainless steel | Unknown |
| Filter cap | Polystyrene | Polystyrene | Unknown |
| Air filter | Polyester | Polyester | Unknown |
#### MATERIALS
#### PERFORMANCE
The performance of the TERUMO® SURFLO® FLEX and the SURFLASH I.V. Catheters is equivalent to the performance of the cleared Becton-Dickinson® (B-D) Insyte® I.V. Catheter, the cleared Becton-Dickinson® (B-D) Insyte® (N) I.V. Catheter (K971339-unable to confirm), and the cleared Terumo Medical Corporation Surflo I.V. Catheter (K891087).
2007 - 100 - 100 - 100
{3}------------------------------------------------
The following tests were performed demonstrating the substantial equivalence of the TERUMO SURFLO FLEX and the SURFLASH I.V. Catheters submitted in this 510k to the B-D Insyte I.V. Catheter, the B-D Insyte (N) I.V. Catheter and the Terumo Surflo I.V. Catheter.
- Catheter body tensile strength (force at break of catheter) 1.
- Stiffness 2.
- Elongation 3.
- Burst pressure 4.
- Collapse ડ.
- Flexural fatigue tolerance Q.
- Radio detectability 7.
- 8. Flow rate
- Corrosion resistance of caulking pin 9.
- 10. Catheter body to catheter hub tensile strength (force at break of catheter/hub)
- 11. Catheter/needle attachment (fitting strength)
- 12. Vent fitting
- 13. Detection of flashback at catheter tip and needle hub
- 14. Detection of flashback at needle hub
- 15. Strength of union between needle hub and needle
#### ADDITIONAL SAFETY INFORMATION
The sterility of the product is assured using a sterilization method validated in accordance with the European Standard, EN 550 (1994): "Sterilization of Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization", to provide a sterility assurance level (SAL) of 10°.
Ethylene oxide residual levels resulting from EtO sterilization will not exceed the maximum residue levels proposed in the Federal Register Notice issued June 23, 1978, and indicated as follows:
| Ethylene oxide | 25 ppm |
|----------------------|---------|
| Ethylene chlorhydrin | 25 ppm |
| Ethylene glycol | 250 ppm |
Manufacturing test methods include functional, chemical stability and sterility tests.
Blood contacting materials were tested in accordance with the tests recommended in FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard AAMI/ANSI/ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: Guidance on Selection of Tests".
{4}------------------------------------------------
The SURFLO FLEX and SURFLASH I.V. Catheters are classified as external communicating device, circulating blood, prolonged exposure (24 hrs to 30 days).
The expiration dating for the TERUMO® SURFLO® FLEX and the TERUMO® SURFLASH™ will be 60 months.
#### CONCLUSION
The TERUMO® SURFLO® FLEX I.V. Catheters and the TERUMO® SURFLASH™ I.V. Catheters, submitted in this 510k, are substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared Becton Dickinson (B-D®) Insyte® I.V. Catheter and the B-D Insyte (N) I.V. Catheter, which are the subject of K971339 (unable to confirm); and to the Terumo® Surflo® I.V. Catheter which is the subject of K891087. Differences between the devices cited in this section are not significant and do not raise any new issues of safety or effectiveness.
Terumo's statement that these devices are substantially equivalent to any other device is done solely to comply with the requirements of the Federal Food, Drug and Cosmetic Act and is not intended whatsoever to be the basis for a patent infringement action.
| Date Prepared | April 19, 1999 |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Prepared by | Kristine Wagner<br>Regulatory Affairs Specialist |
| Prepared for | Terumo Medical Corporation<br>125 Blue Ball Road<br>Elkton, MD 21921<br>Phone (410) 392-7241 or (410) 392-7231<br>Fax (410) 398-6079 |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the department's name encircling an abstract symbol. The symbol resembles a stylized caduceus, featuring a staff with three intertwined snakes. The logo is rendered in black and white.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 6 1999
Ms. Kristine Wagner Regulatory Affairs Specialist Terumo Medical Corporation Requlatory Affairs Department 125 Blue Ball Road Elkton, Maryland 21921
Re : K991406 TERUMO® SURFLO® FLEX I.V. CATHETER and Trade Name: TERUMO® SURFLASH™ I.V. CATHETER Requlatory Class: II FOZ Product Code: Dated: April 19, 1999 Received: April 22, 1999
Dear Ms. Wagner:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister.
{6}------------------------------------------------
Page 2 - Ms. Wagner
this response to your premarket notification Please note: submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Directbr Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# 510(k) Number (if known): K991406
#### TERUMO® SURFLO® FLEX I.V. CATHETER and TERUMO® Device Name: SURFLASH I.V. CATHETER
## Indications For Use:
The TERUMO® SURFLO® FLEX I.V. Catheters and the TERUMO® SURFLASH™ I.V. Catheters are devices consisting of a slender, flexible, radioopaque, plastic catheter with a hub. The plastic catheter is inserted into the patient's vascular system for short term use to withdraw samples, administer fluid intravenously or through which to place monitoring equipment such as blood pressure monitors. The stainless steel cannula, placed in the catheter to maintain rigidity, is withdrawn after the catheter is placed in the vascular system.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patrice Cucenti
**Prescription Use** X
(Per 21 CFR 801.109)
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.