K991133 · Impac Medical Systems, Inc. · IYE · Mar 27, 2000 · Radiology
Device Facts
Record ID
K991133
Device Name
MLC FIT
Applicant
Impac Medical Systems, Inc.
Product Code
IYE · Radiology
Decision Date
Mar 27, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
MLC Fit is to be used to define leaf plans for use with radiation treatment machines equipped with multileaf collimators manufactured by Siemens Medical Systems, Varian Associates, and With manned conniness Inc. This method of defining the geometric parameters associated with treatment fields can be used whenever a conformal treatment field is desired.
Device Story
MLC Fit is software for cancer radiotherapy departments; used by clinicians to define multileaf collimator (MLC) leaf positions and geometric parameters for conformal treatment fields. Input: user-defined treatment area shape and collimator characteristics. Processing: software calculates leaf positions based on input shape, replacing manual hand-calculation methods. Output: hardcopy reports and database files containing treatment field definitions. The device does not control the radiotherapy machine or move leaves directly; it serves as a planning aid. Benefits: eliminates slow, error-prone manual calculations; improves efficiency in treatment field setup. Used in clinical settings to assist in radiotherapy planning workflows.
Clinical Evidence
No clinical data. Bench testing only, including software requirements verification, design description documentation, and system hazard analysis performed under the IMPAC Quality System.
Technological Characteristics
Software-based planning tool for radiotherapy. Operates by processing user-defined geometric shapes against collimator-specific characteristics. Supports Siemens, Varian, and Elekta multileaf collimators. Designed under 21 CFR 820 quality system regulations.
Indications for Use
Indicated for defining multileaf collimator (MLC) leaf plans for conformal radiation therapy treatment fields in patients requiring radiotherapy, using equipment from Siemens, Varian, or Elekta.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
K962335 — MLC FIT · Impac Medical Systems, Inc. · Feb 6, 1997
K961902 — SIEMENS BEAM SHAPER · Siemens Medical Solutions USA, Inc. · Aug 8, 1996
K964206 — NUCLETRON PLATO EXTERNAL BEAM PLANNING RTS V2 AND RTS 3D V2 · Nucletron Corp. · Jun 20, 1997
K984532 — CADPLAN HELIOS OPTION 6.0 · Varian Medical Systems, Inc. · Aug 9, 1999
K982549 — RADIONICS MINI MULTI-LEAF COLLIMATOR, MODEL RADIONICS MMLC · Radionics, Inc. · Dec 30, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
MLC Fit - 510(k) - Response to Request for Additional Information
8/25/99
1991133
MAR 2 7 2000
## MLC Fit Non-Confidential Summary of Safety and Effectiveness
The primary function of MLC Fit is to provide a means to define multileaf collimator leaf plans based on a user defined shape of a desired treatment area for use with a cancer radiotherapy treatment machines equipped with multileaf collimators manufactured by Siemens Medical Systems, Varian Associates, or Elekta Oncology Systems, Inc. Users may create, view, and edit MLC leaf data as well as other geometric parameters associated with treatment field definitions. An intended use statement for MLC Fit is contained in Exhibit 2a of this submission (Exhibit 3 of the "Response").
The goal of MLC Fit is to provide the leaf positions for a given treatment shape in a manner that eliminates the slow and error prone method of hand calculating the position for each leaf and manually setting up the leaf plans. MLC Fit allows the users to define treatment field shapes and generate multileaf collimator leaf plans based on that input shape. Leaf positions and other geometric parameters can be adjusted to meet the user's clinical requirements. MLC Fit is not capable of moving the leaves or other parameters to the calculated positions. It only provides the definition of the treatment field as a hardcopy and to a database file.
The primary functions of MLC Fit are, in effect, the same as those of MLC Fit product (K962335) currently being marketed by IMPAC Medical Systems, Inc. (see Exhibit 3b). Both systems supplement the MLC treatment field definition process only, and do not attempt to control the movement of the defined parameters. The main difference is that the updated MLC Fit product has been expanded to support Elekta multileaf collimators. The user describes the characteristics of the collimator and MLC Fit uses this characterization to create the treatment definition.
IMPAC Medical Systems, Inc., has gained a reputation of providing high quality software products which serve the cancer therapy community. One reason we have earned this reputation is that we strive to provide end-user process oriented solutions with our products. Another reason is that software development for cancer therapy is our only business. This allows us to invoke a development system designed specifically for a cancer therapy software business. Strict adherence to these processes will ensure that MLC Fit is a safe and effective product suitable for use as intended in a cancer therapy department.
MLC Fit was designed and developed under the IMPAC Quality System. The IMPAC Quality System governs the processes by which system and software development are to be defined, implemented, tested, released, installed, and supported. In addition, the IMPAC Quality System demonstrates how IMPAC Medical Systems, Inc., conforms to the Quality System Regulations, 21 CFR 820, and other applicable regulations.
IMPAC Medical Systems, Inc. Copyright 1999
L
{1}------------------------------------------------
Under the processes set forth in the IMPAC Quality System, MLC Fit was developed per Software Requirements Specifications and documented by Software Design Descriptions. Samples of these are shown in Exhibits 4 and 5 for your review.
In addition, a Hazard Analysis was performed to determine and evaluate the areas which represent potentials hazards during MLC Fit operation. For hazards within the scope of the MLC Fit product, the hazard, effect, and protection implemented were documented and reviewed. The System Hazard Analysis is included in Exhibit 6.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus, which is a symbol often associated with medicine and healthcare.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 27 2000
Thomas Faris Regulatory Affairs Manager IMPAC Medical Systems, Inc. 215 Castro Street Mountain View, CA 94041
Re:
K991133 MLC Fit (Multi Leaf Collimator Planning Software) Dated: January 18, 2000 Received: January 19, 2000 Regulatory class: II 21 CFR 892.5050/Procode: 90 IYE
Dear Mr. Faris:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Aproval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Daniel C. Schultz, M.D.
Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{3}------------------------------------------------
## MLC Fit Indications for Use Statement
MLC Fit is to be used to define leaf plans for use with radiation treatment machines equipped with multileaf collimators manufactured by Siemens Medical Systems, Varian Associates, and With manned conniness Inc. This method of defining the geometric parameters associated with treatment fields can be used whenever a conformal treatment field is desired.
David A. Leymm
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number
**Prescription Use**
(Per 21 CFR 801.109)
IMPAC Medical Systems, Inc. Copyright 1999
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.