K984532 · Varian Medical Systems, Inc. · MUJ · Aug 9, 1999 · Radiology
Device Facts
Record ID
K984532
Device Name
CADPLAN HELIOS OPTION 6.0
Applicant
Varian Medical Systems, Inc.
Product Code
MUJ · Radiology
Decision Date
Aug 9, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
To automatically design the beam profiles that most closely generate the desired dose distribution.
Device Story
CadPlan Helios is an inverse planning software tool for Intensity Modulated Radiation Therapy (IMRT). It takes clinical constraints for target volumes and critical organs as input. The device uses an inverse planning algorithm to compute optimum field intensities for multiple beams to achieve a desired dose distribution. These intensity profiles are converted into dynamic Multi-Leaf Collimator (MLC) motion patterns. The system is used by clinicians to design conformal radiation plans. The output allows for modulated dose delivery, enabling precise dose conformation to anatomical structures. This assists healthcare providers in optimizing radiation therapy delivery, potentially improving treatment accuracy and sparing healthy tissue.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Inverse planning software for IMRT; supports dynamic Multi-Leaf Collimator (MLC) motion; computes intensity profiles based on clinical constraints; compatible with linear accelerators, teletherapy devices, and brachytherapy systems.
Indications for Use
Indicated for planning radiation therapy treatments using linear accelerators, teletherapy devices, and x-ray energies from 4 to 50 MV, Cobalt-60, electron energies from 4 to 50 MeV, and brachytherapy treatments.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
K021268 — ECLIPSE, HELIOS OPTION · Varian Medical Systems, Inc. · May 22, 2002
K071938 — MONACO RTP SYSTEM · Computerized Medical Systems, Inc. · Oct 1, 2007
K250963 — PlanOne 1 · Cosylab Jsc, Control System Laboratory · Nov 20, 2025
K964206 — NUCLETRON PLATO EXTERNAL BEAM PLANNING RTS V2 AND RTS 3D V2 · Nucletron Corp. · Jun 20, 1997
K090808 — PINNACLE3 SMARTARC · Philips Medical Systems, Inc. · Apr 3, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
AUG = 9 1999
Image /page/0/Picture/1 description: The image shows a handwritten string of characters. The string is "K984532". The characters are written in a cursive style. The image is in black and white.
# varian@ oncology systems
## Premarket Notification [510K] Summary as required by 21 CFR 807.92
Date Summary was prepared; December 18, 1998
#### Submitter's Name:
Varian Oncology Systems 3045 Hanover Street Palo Alto, CA 94304
#### Contact Person:
Linda S. Nash Regulatory Compliance & Radiation Safety Manager Phone (650) 424-6990 FAX (650) 424-4830 E-Mail linda.nash@os.varian.com
Device Name: CadPlan Helios Option 6.0
Classification Name: Accelerator, Linear, Medical
Predicate Device: Peacock Plan CORVUS, K940663
#### Product Description:
CadPlan Helios is an inverse planning tool for creating highly conformal plan using intensitymodulated fields. Inverse planning refers to the process of computing the optimum field intensities for the required numbers of beams to achieve the desired dose of distribution with specified complaints. It allows for creation of in-field intensity profiles that produce the desired dose distribution as precisely as possible.
Image /page/0/Picture/14 description: The image shows a logo with the words "PARTNERS FOR LIFE" stacked on top of each other. The word "PARTNERS" is in a bold, sans-serif font, while the word "FOR" is in a smaller, serif font. The word "LIFE" is written in a cursive script that extends to the right of the other words. The logo is black and white.
Page 1 of 2 Varian Associates, Inc. 3045 Hanover Street Palo Alto, California 94304-1129 U.S.A. 650.493.4000 FAX: 650.424.4830
{1}------------------------------------------------
Helios computes the in-field intensities on the basis of specified clinical constraints for the desired dose distribution inside target volumes and critical organs. The calculated intensity profiles are then converted to a pattern of dynamic Multi-Leaf Collimator (MLC) motions. The combination of modulated dose delivery and dynamic MLC motion allows dose to be conformed precisely to anatomical structures. Helios implements the inverse planning algorithm and support of dynamic MLC for Intensity Modulated Radiation Therapy.
#### Intended Use:
To automatically design the beam profiles that most closely generate the desired dose distribution.
### Technological Characteristics:
See the attached "Comparison of Characteristics and Specifications Table".
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of three overlapping shapes that resemble birds in flight or abstract human figures.
#### g lagg AUG
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Linda Nash Regulatory Affairs & Ouality Assurance Manager Varian Associates, Inc. 3045 Hanover Street Palo Alto, California 94304-1129
Re:
K984532 CadPlan Helios Option 6.0 Dated: May 7, 1999 Received: May 12, 1999 Regulatory Class: II 21 CFR 892.5050/Procode: 90 MUJ
Dear Ms. Nash:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been redassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive. Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Varian Oncology Systems. The word "varian" is in bold, and the words "oncology systems" are in a smaller font below it. There is a circled logo to the right of the word "varian".
### Indications For Use
510(k) Number (if known):
Device Name: CadPlan Helios Option 6.0
Indications for Use:
CadPlan is used to plan radiation therapy treatments employing linear accelerators and other similar teletherapy devices and x-ray energies from 4 to 50 MV, as well as Cobalt-60, and electron energies from 4 to 50 MeV, and to plan brachytherapy treatments.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(per 21 CFR 801.109)
OR
Over-The-Counter Use
(Division Sign-Off)
Division of Reproductive, Abdominal, EN and Radiological De 510(k) Number
Image /page/3/Picture/9 description: The image shows a logo for "Partners for Life." The word "Partners" is stacked on top of "For" in a bold, sans-serif font. The word "Life" is written in a cursive font and is placed to the right of "For." The logo is black and white.
Varian Associates, Inc. 3045 Hanover Street Palo Alto, California 94304-1129 U.S.A. 650.493.4000 FAX: 650.424.4830
printed on recycled pap
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.