K990899 · Osborn Group, Inc. · LCP · Dec 21, 1999 · Hematology
Device Facts
Record ID
K990899
Device Name
HEMOCHEK-A1C SAMPLE COLLECTION KIT
Applicant
Osborn Group, Inc.
Product Code
LCP · Hematology
Decision Date
Dec 21, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7470
Device Class
Class 2
Indications for Use
The HemoChek-A1c Sample Collection Kit is indicated for over-the-counter sale for use in the measurement of HbA1c on blood specimens which can be collected at the patient's home or at a physician's office on filter paper and delivered to the laboratory by mail. The HbA1c test is used in the assessment of the average blood glucose over a 10-12 week period. The results are to be evaluated by the patient and their physician. The product is not indicated for the diagnosis of diabetes mellitus.
Device Story
HemoChek-A1c Sample Collection Kit facilitates remote HbA1c testing. Kit includes sample card with proprietary filter paper, sterile lancet, alcohol swab, gauze, bandage, and mailing envelopes. Patient collects capillary blood sample via lancet, applies to filter paper, and mails to Osborn Group, Inc. laboratory. Laboratory performs HbA1c measurement using existing assay methods; results mailed to patient. Intended for patient/physician evaluation of long-term glucose control. Enables convenient home-based sample collection without immediate clinical supervision.
Clinical Evidence
No clinical data. Performance testing demonstrates the device performs in the same manner as the HemoChek predicate device.
Technological Characteristics
Kit includes proprietary filter paper sample card, sterile lancet, alcohol swab, gauze, and adhesive bandage. Analysis performed via existing laboratory assay methodology. No electronic components, software, or connectivity.
Indications for Use
Indicated for over-the-counter sale for patients at home or in physician offices requiring HbA1c measurement for assessment of average blood glucose over 10-12 weeks. Not indicated for diabetes mellitus diagnosis.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
{1}------------------------------------------------
1990899
DEC 2 1 1999
Image /page/0/Picture/2 description: The image shows the logo for Osborn Group, INC. The logo is in black and white and features a curved line above the word "OSBORN". Below the company name is the text "A ChoicePoint Company" in a smaller font. The logo is simple and professional.
### 510(k) SUMMARY
### Osborn Group, Inc.
## HemoChek-A1c Sample Collection Kit
March 17, 1999
### Submitter Information:
Osborn Group, Inc. 19401 West 117th Street Olathe, Kansas 66062
Submitter's Name: Phone:
Gilbert P. Bourk III (913) 390-7146
#### Device Name:
HemoChek-A1c Sample Collection Kit
Common Name: Hemoglobin A1c blood sample collection kit
Classification Name: Glycosvlated Hemoglobin Assay
### Predicate Device Equivalence:
Substantial equivalence is claimed to the EZCHEK™/HbA1c Sample Collection Kit, and to the HemoChek Sample Collection Kit, cleared for commercial distribution per K971919 and K984529, respectively.
### Device Description:
The HemoChek-A1c Sample Collection Kit is a kit which is purchased by a patient a pharmacy or other retail store. The kit consists of the following:
- A Sample Card containing the proprietary filter paper that the blood sample is placed on and a place for the patient to print his/her name, address, social security number and the sample was collected
- A pamphlet containing detailed instructions about how to obtain a blood sample and mail it to Osborn -Group, Inc.
- A specimen envelope with the words "After blood has dried, insert completed HemoChek test in this envelope. Seal flap and place in envelope" printed on it
- A self-adhesive envelope in which the card is inserted and then mailed to Osborn Group, Inc. -
Image /page/0/Picture/23 description: The image shows the logo for Osborn Laboratories, a ChoicePoint company. The logo features a stylized letter "L" formed by a grid pattern within a circle. To the right of the circular element, the words "OSBORN LABORATORIES" are printed in a simple, sans-serif font, with the tagline "A ChoicePoint Company" appearing in a smaller font size below.
Image /page/0/Picture/24 description: The image shows the logo for Intellisys, an Osborn Company. The word "INTELLISYS" is written in a bold, sans-serif font, with a curved line above it. Below the word "INTELLISYS" is the phrase "AN OSBORN COMPANY" in a smaller, sans-serif font. The logo is simple and modern, and it is likely used to represent the company's brand.
{2}------------------------------------------------
- A sterile lancet -
- An alcohol swab -
- A gauze pad -
- An adhesive bandage -
- -An outer package containing all of the items listed above
After purchasing the HemoChek-A1c Sample Collection Kit, the patient then prints the required information on the Sample Card and collects a blood sample, using a lancet (either the one provided in the kit, one provided by the patient's physician or one supplied by the patient). The blood sample is placed on both circles on the right hand side of the Sample Card, as described in the instructions. The Sample Card is then placed in the specimen envelope which in turn is placed in the mailing envelope provided in the kit and mailed to Osborn Group, Inc. When the blood sample is received by Osborn Group, Inc., the patient's HbA1c level is measured using existing assay methods. The results are then mailed to the patient.
### Intended Use:
The HemoChek-A1c Sample Collection Kit is indicated for over-the-counter sale for use in the measurement of HbA1c on blood specimens which can be collected at the patient's home or at a physician's office on filter paper and delivered to the laboratory by mail. The HbA1c test is used in the assessment of the average blood glucose over a 10-12 week period. The results are to be evaluated by the patient and their physician. The product is not indicated for the diagnosis of diabetes mellitus.
#### Comparison of Technological Characteristics:
Essentially, the devices use the same basic technology, i.e., collecting a blood sample and analyzing it using an existing assay methodology. However, the physical size of the device is different than the EZCHEK predicate device. Also, the existing assay methodologies used are different for the device and the EZCHEK predicate device, but are the same as for the HemoChek predicate device.
### Summary of Performance Testing:
Information contained in this submission demonstrates that the HemoChek-Atc performs in the same manner as the HemoChek predicate device.
#### Conclusions:
Based on the above, we concluded that the HemoChek-A1c Sample Collection Kit is substantially equivalent to legally marketed predicate devices and iffective for its intended use.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 21 1999
Mr. Gilbert P. Bourk III Vice President and General Counsel Osborn Group, Inc. 14901 West 117th Street Olathe, Kansas 66062
K990899 Re:
> Trade Name: HemoChek-A1c™ Sample Collection Kit Regulatory Class: II Product Code: LCP Dated: October 29, 1999 Received: November 1, 1999
Dear Mr. Bourk:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
## Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Osborn Group, Inc. The logo is in black and white and features the company name in a bold, sans-serif font. Below the company name is the text "A ChoicePoint Company" in a smaller font. A curved line extends from the top of the "O" in Osborn to the bottom of the "N".
## Device Name:
HemoChek-A1c™ Sample Collection Kit
# Indications for Use:
The HemoChek-A1c Sample Collection Kit is indicated for over-the-counter sale for use in the measurement of HbA1c on blood specimens which can be collected at the patient's home or at a physician's office on filter paper and delivered to the laboratory by mail. The HbA1c test is used in the assessment of the average blood glucose over a 10-12 week period. The results are to be evaluated by the patient and their physician. The product is not indicated for the diagnosis of diabetes mellitus.
Peter E. Makris
(Division Sign-Off)
Division of Clinical Laboratory Devices Ka90899
510(k) Number
ver the counter.
Image /page/4/Picture/11 description: The image shows the logo for Osborn Laboratories. The logo consists of a circular design on the left, made up of small squares, with a larger square in the center. To the right of the circular design, the text "OSBORN LABORATORIES" is displayed in a simple, sans-serif font. Below this, the text "A ChoicePoint Company" is written in a smaller font size.
Image /page/4/Picture/9 description: The image shows the logo for Intellisys, an Osborn Company. The word "INTELLISYS" is in large, bold, sans-serif font, with a line underneath. Below the line, in a smaller font, it says "AN OSBORN COMPANY". There is a curved line above and to the right of the word "INTELLISYS".
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.