K984529 · Osborn Laboratories, Inc. · LCP · Jan 28, 1999 · Hematology
Device Facts
Record ID
K984529
Device Name
HEMOCHEK SAMPLE COLLECTION KIT
Applicant
Osborn Laboratories, Inc.
Product Code
LCP · Hematology
Decision Date
Jan 28, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7470
Device Class
Class 2
Indications for Use
The HemoChek Sample Collection Kit is indicated for use in the measurement of HbA1c on blood specimens which can be collected at the patient's home or at a physician's office on a filter paper and delivered to the laboratory by mail. The results are to be evaluated by the patient's physician. The product is not indicated for the diagnosis of diabetes mellitus.
Device Story
HemoChek Sample Collection Kit facilitates remote HbA1c monitoring; patient collects capillary blood sample via lancet at home or physician's office; blood deposited onto proprietary filter paper card; card mailed to Osborn Laboratories; laboratory performs HbA1c measurement using existing assay methods; physician evaluates results to assess average blood glucose over 8-12 weeks; kit includes sample card, instructions, and mailing envelope; intended to support clinical management of patients; not for diabetes diagnosis.
Clinical Evidence
Bench testing only. Accuracy and repeatability validated by comparing HbA1c levels from whole blood specimens versus samples collected on HemoChek cards using identical assay methods. Testing included specimens with high HbA1c levels and environmental stress testing (temperature/humidity) to simulate mail transit. Results confirmed variation between whole blood and card-collected samples remained within acceptable accuracy limits.
Technological Characteristics
Kit consists of proprietary filter paper sample card, instructions, and mailing envelope. Utilizes capillary blood collection via lancet. Analysis performed via existing laboratory assay methodologies. No electronic components or software algorithms.
Indications for Use
Indicated for patients requiring HbA1c measurement for assessment of average blood glucose over 8-12 weeks. Specimens collected at home or physician's office via filter paper and mailed to laboratory. Not indicated for diabetes diagnosis.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Osborn Laboratories, which includes a stylized "L" formed by a grid pattern. The text "OSBORN LABORATORIES" is displayed next to the logo, followed by the text "A ChoicePoint Company". Above the logo, the date "JAN 28 1999" is printed.
JAN 28 1999
A ChoicePoint™ Company
# 510(k) SUMMARY
## Osborn Laboratories
## HemoChek Sample Collection Kit
# December 18, 1998
### Submitter Information:
Osborn Laboratories 19401 West 117th Street Olathe, Kansas 66062
Submitter's Name: Phone:
Gilbert P. Bourk III (913) 390-7146
### Device Name:
Osborn Laboratories HemoChek Sample Collection Kit
Hemoglobin A1c blood sample collection kit Common Name: Classification Name: Glycosylated Hemoglobin Assay
### Predicate Device Equivalence:
Substantial equivalence is claimed to the EZCHEK™/HbA1c Sample Collection Kit, cleared for commercial distribution per K971919.
### Device Description:
The HemoChek Sample Collection Kit is a kit which is sent to the patient's home after the patient has been enrolled in the HemoChek program by the patient's physician. The kit consists of the following:
- A Sample Card containing the proprietary filter paper that the blood sample is deposited on. the i patient's name and address, instructions on how to obtain the sample and send it to Osborn Laboratories and a place to make changes to any of the patient's information.
- A pamphlet containing a description of the HemoChek program and detailed instructions about how to l obtain a blood sample and mail it to Osborn Laboratories.
- A self-adhesive envelope in which the card is inserted and then mailed to the preprinted address on the envelope.
- The mailing envelope which is used to mail the above three items to the patient. -
{1}------------------------------------------------
After receiving the HemoChek Sample Collection Kit in the mail from Osborn Laboratories, the patient then collects a blood sample, using a lancet. The blood sample is placed on all three circles on the right hand side of the sample collection card, as described in the instructions. Then the card is placed in the mailing envelope provided in the kit and mailed to Osborn Laboratories. When the blood sample is received by Osborn Laboratories the patient's HbA1c level is measured using existing assay methods.
#### Intended Use:
The HemoChek Sample Collection Kit is indicated for use in the measurement of HbA1c on blood specimens which can be collected at the patient's home or at a physician's office on a filter paper and delivered to the laboratory by mail. The results are to be evaluated by the patient's physician. The product is not indicated for the diagnosis of diabetes mellitus.
### Comparison of Technological Characteristics:
Essentially, the two devices use the same basic technology, i.e., collecting a blood sample and analyzing it using an existing assay methodology. However, the physical size of the two devices is different. Also, the existing assay methodologies used are different for the two devices.
### Summary of Performance Testing:
To validate the accuracy and repeatability of the results obtained using the HemoChek Sample Cards. whole blood specimens were selected to obtain a range of HbA1c levels. Samples of whole blood from each specimen were analyzed using an existing assay method. Next, additional samples were obtained from each specimen using HemoChek Sample Cards and analyzed using the same assay method.
To further validate the HemoChek Sample Cards, blood specimens were taken at a local hospital from a number of persons with high levels of HoA1c. For each specimen, a sample of whole blood and a sample collected using a HemoChek Sample Card were analyzed using the same methodology used in the initial validation discussed in the previous paragraph. The variation between the whole blood value and the value obtained using a HemoChek Sample Card was within acceptable accuracy limits.
To obtain assurance that the blood samples would still provide an accurate assessment of the patients' hemoglobin A 1c levels even after undergoing the extreme temperatures and humidity conditions that can be experienced by any object sent by U.S. Mail, the HemoChek Sample Collection Kit was subiected to environmental testing. The test results were considered to be within acceptable accuracy limits.
To obtain patient comments about the HemoChek Sample Collection Kit. Osborn Laboratories conducted a survey. Overall, the users were very pleased with the product.
### Conclusions:
Based on the above, we concluded that the HemoChek Sample Collection Kit is substantially equivalent to a legally marketed predicate device and is safe and effective for its intended use.
{2}------------------------------------------------
JAN 28 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Gilbert P. Bourk, III Vice President and General Counsel Osborn Laboratory 14901 West 117th Street Olathe, Kansas 66062
Re: K984529
Trade Name: Osborn Laboratories HemoChek Sample Collection Kit Regulatory Class: II Product Code: LCP Dated: December 18, 1998 Received: December 21, 1998
Dear Mr. Bourk:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
# Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Steven Butman
Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page / of /
510(k) Number (if known): K984529
Device Name:
HemoChek Sample Collection Kit
Indications for Use:
The HernoChek Sample Collection Kit is indicated for use in the measurement of HbA1c on blood specimens which can be collected at the patient's home or at a physician's office on a filter paper and delivered to the laboratory by mail. The HbA1c test is used in the assessment of the average blood glucose over an 8-12 week period. The results are to be evaluated by the patient's physician. The product is not indicted for the diagnosis of diabetes mellitus.
Scan Cooger
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number 984529
(PLEASE DO NOT WRITE BELOW THIS LÌNE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The-Counter Use
by Prescription
(Optional Format 1-2-96)