LAMITRODE LEADS

K990469 · Advanced Neuromodulation Systems · GZB · Jun 3, 1999 · Neurology

Device Facts

Record IDK990469
Device NameLAMITRODE LEADS
ApplicantAdvanced Neuromodulation Systems
Product CodeGZB · Neurology
Decision DateJun 3, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5880
Device ClassClass 2
AttributesTherapeutic

Intended Use

Advanced Neuromodulation Systems, Inc.'s Lamitrode Leads are intended to be used with Advanced Neuromodulation Systems extensions and/or receivers, transmitters, and antennae to electrically stimulate spinal cord fiber tracts for treatment of chronic pain of the extremities and/or trunk. The proposed device modification does not affect the original intended use of the legally marketed device.

Device Story

Lamitrode leads are implantable spinal cord stimulator leads consisting of spaced electrodes connected by wires within a cover sheath. Leads are introduced into the epidural space via hemilaminectomy, positioned superior to the spinal segment responsible for pain impulse transmission. The device connects to an external or implanted radio-frequency (RF) receiver or pulse generator to deliver electrical stimulation to spinal cord fiber tracts. Used by physicians to treat chronic pain of the trunk and limbs. The device serves as a conduit for electrical signals; it does not process data or contain software. Benefits include pain management through electrical neuromodulation.

Clinical Evidence

Bench testing only. Biostability testing of polyurethane and silicone materials provided in manufacturer's Master File. Comparative bond strength testing demonstrated that the polyurethane lead insulator to platinum-cured paddle bond is superior to the current product.

Technological Characteristics

Implantable multi-electrode catheter; Platinum/Iridium electrodes; Stainless Steel 304 contact terminals; Polyurethane lead insulator; Braided wire cable; 4, 8, or 16 electrode configurations; 58-80 cm length; Sterile, prescription-only; No software or electronic processing components.

Indications for Use

Indicated for treatment of chronic pain of the trunk and limbs, either as sole mitigating agent or as adjunct to other therapies in multidisciplinary approach. Intended for use with ANS extensions, receivers, transmitters, and antennae.

Regulatory Classification

Identification

An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 3 JUN Image /page/0/Picture/1 description: The image shows the logo for ANS, which is a company or organization. The logo consists of a grid of small squares on the left, followed by the letters "ANS" in a bold, sans-serif font. The grid is made up of 7 rows and 7 columns of squares, and the letters are arranged horizontally to the right of the grid. 990469 Page 1 of 3 # 510(k) Summary of Safety and Effectiveness This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | Date Prepared: | February 1, 1999 | |----------------|----------------------------------------------------------------------------------------| | Company: | Advanced Neuromodulation Systems, Inc<br>One Allentown Parkway<br>Allen, TX 75002-4211 | | Contact: | Drew Johnson | | Phone Number: | 972-390-9800 Ext. 327 | | Fax Number: | 972-390-2881 | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for ANS, which is a company that specializes in technology. The logo consists of a square grid of smaller squares on the left, followed by the letters "ANS" in a bold, sans-serif font. The grid is likely meant to represent the interconnectedness of technology, while the letters "ANS" are the company's initials. # Lamitrode Spinal Cord Stimulator Lead 510(k) Summary of Safety and Effectiveness ## Device Information: | Trade Names: | Lamitrode® Model 1916L Lead | |----------------------|--------------------------------------------------| | | Lamitrode® Model 1944L Lead | | | Lamitrode® Model 1992LS Lead | | | Lamitrode® Model 1995L Lead | | | Lamitrode® Model 1998L Lead | | Common Name: | Spinal Cord Stimulator | | Classification Name: | Implanted Spinal Cord Stimulator for Pain Relief | ### Predicate Device: Advanced Neuromodulation Systems, Inc.,., currently markets spinal cord stimulator systems with Silicone lead insulators under 510(k) # K853643 and K905255 and K960728. ## Device Description: Advanced Neuromodulation Systems, Inc.'s Lamitrode Leads are implantable devices consisting of spaced electrodes connected by wires within a cover sheath. These lamitrode leads are introduced into the epidural space via a hemilaminectomy, superior to the spinal segment responsible for pain impulse transmission, and connected to a radio-frequency (RF) receiver or pulse generator. #### Intended Use: Advanced Neuromodulation Systems, Inc.'s Lamitrode Leads are intended to be used with Advanced Neuromodulation Systems extensions and/or receivers, transmitters, and antennae to electrically stimulate spinal cord fiber tracts for treatment of chronic pain of the extremities and/or trunk. The proposed device modification does not affect the original intended use of the legally marketed device. ## Comparison To Predicate Device: The following table illustrates the comparison between the modified device, the original legally marketed device and other legally marketed devices. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the letters 'ANS' in a bold, sans-serif font. To the left of the letters is a square grid pattern made up of smaller squares. The grid is also in black, matching the color of the letters. | | ANS Predicate Device<br>510(k) K853643,<br>K905255, K960728 | Medtronic Predicate<br>Device 510(k)<br>K884948, K811655,<br>and K905407 | ANS Modified Device<br>K# Under Review | |----------------------------|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------| | Intended Use: | Stimulation of spinal<br>cord for treatment of<br>chronic pain of the<br>trunk and limbs | Stimulation of spinal<br>cord for the treatment<br>of chronic pain of the<br>trunk and limbs | Stimulation of spinal<br>cord for treatment of<br>chronic pain of the<br>trunk and limbs | | Materials:<br>• Electrode: | Platinum/Iridium | Same | Platinum/Iridium | | • Contact Terminal: | Stainless Steel 304 | | Stainless Steel 304 | | • Insulator: | Silicone<br>(K960728)<br>Polyurethane | Silicone<br>Polyurethane | Polyurethane | | Design Features: | Multi-electrode<br>Catheter | | Multi-electrode<br>Catheter | | | Braided Wire Cable | | Braided Wire Cable | | | Platinum/Iridium<br>Electrode | Platinum/Iridium<br>Electrode | Platinum/Iridium<br>Electrode | | | 4,8,or 16 electrodes | 4 and 8 staggered<br>electrodes | 4,8, or 16 electrodes | | | Laminectomy<br>Introduction<br>Percutaneous<br>Introduction<br>(K960728) | Laminectomy<br>Introduction | Laminectomy<br>Introduction | | Dimensions:<br>• Length: | 58 - 80 cm | 10 - 100 CM | 58 - 80 cm | | Packaging: | Tray w/ Tyvek Lid | Tray w/ Tyvek Lid | Tray w/ Tyvek Lid | | Labeling: | Labeled as sterile, | Labeled as sterile, | Labeled as sterile,<br>prescription device | ## Non-clinical Testing: Biostability testing of the polyuethane and silicone material is provided in the manufacturer's Master File. Master File biocompatibility information demonstrates that the change to polyurethane and platinum cured silicone raises no significant safety or effectiveness questions relating to biocompatibility. Comparative testing of product bond strength indicates that the polyurethane lead insulator to the platinum cured paddle has a better bond strength than the current product. {3}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human figures connected at the shoulders, representing health and human services. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 3 1999 JUN Mr. Drew Johnson Director, Regulatory Affairs Advanced Neuromodulation Systems, Inc. One Allentown Parkway 75002-4211 Allen, Texas K990469 Re : Lamitrode Spinal Cord Stimulation Leads, Trade Name: Lamitrode® 1916L, 1944L, 1992LS, 1995L, and 1998L Requlatory Class: II Product Code: GZB May 17, 1999 Dated: May 18, 1999 Received: Dear Mr. Johnson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸੈ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ # Page 2 - Mr. Drew Johnson This letter will allow you to begin marketing your device as described in your 510 k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ~Celia M. Witten, JPh.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): K990469 Device Name: Lamitrode Spinal Cord Stimulation Lead Indications For Use: Advanced Neuromodulation Systems Lamitrode Spinal Cord Stimulation Leads are indicated for the treatment of chronic pain of the trunk and limbs, either as the sole mitigating agent, or as an adjunct to other modes of therapy used in a multidisciplinary approach and is intended to be used with Advanced Neuromodulation Systems extensions and/or receivers, transmitters, and antennae. The proposed device modification does not affect the original intended use of the legally marketed device. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) b coello Restorative Devices Prescription Use . (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...