POLYETHER URETHANE MATERIAL CHANGE IN NEUROSTIMULATION LEADS
Applicant
Medtronic Vascular
Product Code
GZB · Neurology
Decision Date
Mar 25, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Pisces Quad Model 3487A, Pisces Quad Compact Model 3887, Pisces Quad Plus Model 3888, Pisces Z Quad Model 3890, Pisces Z Quad Compact Model 3891, Pisces Z Quad Plus Model 3892, Specify Model 3998, Pisces Octad Model 3898, SymMix Model 3982A, Temporary Screening Lead Model 3861, and Verify Model 3862 are indicated for Spinal Cord Stimulation to aid in the management of chronic intractable pain of the trunk and/or limbs. Resume II Model 3587A and Resume TL Model 3986A are indicated for Spinal Cord Stimulation to aid in the management of chronic intractable pain of the trunk and/or limbs. They are also indicated for Peripheral Nerve Stimulation. The peripheral nerve stimulators are used to stimulate electrically a peripheral nerve in patients to relieve severe intractable pain. On-Point Model 3987A is indicated for Peripheral Nerve Stimulation. The peripheral nerve stimulators are used to stimulate electrically a peripheral nerve in patients to relieve severe intractable pain.
Device Story
Implantable neurostimulation leads; deliver electrical stimulation to spinal cord or peripheral nerves; used for management of chronic intractable pain. Device types include percutaneous (quadripolar, octapolar, bipolar) and surgical leads. Leads connect to neurostimulator; physician implants leads to target nerve tissue; electrical pulses modulate pain signals. Benefits include pain relief for patients unresponsive to conservative therapies. Used in clinical settings by physicians.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Implantable neurostimulation leads; configurations include quadripolar, octapolar, and bipolar; percutaneous and surgical lead types; in-line connectors; low impedance variants available.
Indications for Use
Indicated for patients with chronic intractable pain of the trunk and/or limbs requiring spinal cord stimulation, or patients with severe intractable pain requiring peripheral nerve stimulation.
Regulatory Classification
Identification
An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.
{1}------------------------------------------------
MAR 2 5 2004
K040568
1 of 3
## 510(k) Summary Medtronic Neurostimulation Leads (as required by 21 CFR 807.92)
Medtronic Neurological
710 Medtronic Parkway NE
Minneapolis, Minnesota 55432-5604 USA
## A. Submitter information
Submitter's Name:
Address:
Telephone Number
Contact Person:
Doug Atkins
763.505.0204
Date Submission Prepared:
February 27, 2004
### B. Device Information
| Device Trade Name: | Pisces Z Quad® lead kit for Spinal Cord Stimulation (SCS) |
|-----------------------|----------------------------------------------------------------------------------------------|
| | Pisces Z Quad Compact™ lead kit for Spinal Cord Stimulation (SCS) |
| | Pisces Z Quad Plus® lead kit for Spinal Cord Stimulation (SCS) |
| | Pisces Quad® lead kit for Spinal Cord Stimulation (SCS) |
| | Pisces Quad® Compact lead kit for Spinal Cord Stimulation (SCS) |
| | Pisces Quad® Plus lead kit for Spinal Cord Stimulation (SCS) |
| | Pisces Octad® lead kit for Spinal Cord Stimulation (SCS) |
| | Specify™ lead kit for Spinal Cord Stimulation (SCS) |
| | Resume II® lead kit for Spinal Cord Stimulation (SCS) and Peripheral Nerve Stimulation (PNS) |
| | Resume® TL lead kit for Spinal Cord Stimulation (SCS) and Peripheral Nerve Stimulation (PNS) |
| | SymMix® lead kit for Spinal Cord Stimulation (SCS) |
| | On-Point® lead kit for Peripheral Nerve Stimulation (PNS) |
| | Verify® lead kit for Spinal Cord Stimulation (SCS) |
| | Temporary Screening Lead for Spinal Cord Stimulation (SCS) |
| Common or Usual Name: | Spinal Cord Stimulation Lead |
| | Peripheral Nerve Stimulation Lead |
| Classification Name: | Implanted spinal cord stimulator for pain relief (21 CFR 882.5880) |
| | Implanted peripheral nerve stimulator for pain relief (21 CFR 882.5870) |
Image /page/0/Picture/14 description: The image shows the Medtronic logo with the word "Medtronic" in bold, black letters. To the left of the word is a circular logo with a stick figure inside. Below the word "Medtronic" is the word "Confidential" in a smaller font.
{2}------------------------------------------------
K040568
2023
Classification Code:
Predicate Device:
Device Description:
GZB-Spinal Cord Stimulators GZF—Peripheral Nerve Stimulators
Pisces Quad 3487A-K923931 Pisces Quad Compact 3887-K923931 Pisces Quad Plus 3888-K923567 Pisces Z Quad 3890-K033016 Pisces Z Quad Compact 3891-K033016 Pisces Z Quad Plus 3892—K033016 Specify 3998-K971756 Pisces Octad 3898-K934065 Resume II 3587A-K032561 SymMix 3982A-K032561 Resume TL 3986A-K032561 On-Point 3987A-K032561 Temporary Screening Lead 3861-K912764 Verify 3862--K932202
Resume II 3587A, Resume TL 3986A, On-Point 3987A, and SymMix 3982A are quadripolar implantable neurostimulation surgical leads with in-line connector.
Pisces Quad 3487A, Pisces Quad Compact 3887, and Pisces Quad Plus 3888 are percutaneous quadripolar implantable neurostimulation leads.
Pisces Z Quad 3890, Pisces Z Quad Compact 3891, and Pisces Z Quad Plus 3892 are low impedance percutaneous quadripolar implantable neurostimulation leads.
Pisces Octad 3898 is a percutaneous octapolar implantable neurostimulation lead.
Temporary screening lead 3861 is a percutaneous bipolar implantable neurostimulation lead to be used for no more than 10 days.
Verify 3862 is a percutaneous quadripolar implantable neurostimulation lead to be used for no more than 10 days.
Specify 3998 is an octapolar implantable ncurostimulation surgical lead.
Pisces Quad Model 3487A, Pisces Quad Compact Model 3887, Pisces Quad Plus Model 3888, Pisces Z Quad Model 3890, Pisces Z Quad Compact Model 3891, Pisces Z Quad Plus Model 3892, Specify Model 3998, Pisces Octad Model 3898, SymMix Model 3982A, Temporary Screening Lead Model 3861, and Verify Model 3862 are indicated for Spinal Cord Stimulation to aid in the management of chronic intractable pain of the trunk and/or limbs.
Image /page/1/Picture/13 description: The image shows the Medtronic logo. The logo consists of a symbol on the left and the word "Medtronic" on the right. Below the word "Medtronic" is the word "Confidential". The symbol on the left appears to be a person in motion inside of a circle.
Indications for Use:
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows handwritten text that appears to be a combination of letters and numbers. The text includes 'K040568' and '3 of 3'. The writing is in black ink and has a somewhat rough or潦草 appearance.
Resume II Model 3587A and Resume TL Model 3986A are indicated for Spinal Cord Stimulation to aid in the management of chronic intractable pain of the trunk and/or limbs. They are also indicated for Peripheral Nerve Stimulation. The peripheral nerve stimulators are used to stimulate electrically a peripheral nerve in patients to relieve severe intractable pain.
On-Point Model 3987A is indicated for Peripheral Nerve Stimulation. The peripheral nerve stimulators are used to stimulate electrically a peripheral nerve in patients to relieve severe intractable pain.
### C. Comparison of Required Technological Characteristics
The technological characteristics of the modified leads for neurostimulation are substantially equivalent to the noted predicate devices.
#### D. Performance Data
Performance data that supports the safety and effectiveness of the modified neurostimulation leads are included n this 510(k) premarket notification.
#### E. Conclusion
Medtronic neurostimulation leads (Pisces Quad 3487A, Pisces Quad Compact 3887, Pisces Quad Plus 3888, Pisces Z Quad 3890, Pisces Z Quad Compact 3891, Pisces Z Quad Plus 3892, Specify 3998, Pisces Octad 3898, SymMix 3982A, Temporary Screening Lead 3861, Verify 3862, Resume II 3587A, Resume TL 3986A, and On-Point 3987A) are substantially equivalent to the noted predicate devices based on the similarities of technological characteristics, the identical indications for use, and the results of the testing.
Image /page/2/Picture/9 description: The image shows the Medtronic logo with the word "Medtronic" in bold, black letters. To the left of the word is a circular symbol with abstract shapes inside. Below the word "Medtronic" is the word "Confidential" in a smaller, less bold font.
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 5 2004
Doug Atkins Regulatory Affairs Specialist 710 Medtronic Parkway NE Minneapolis, Minnesota 55432-5604
Re: K040568
Trade/Device Name: Pisces Z Quad®, Pisces Z Quad Compact™, Pisces Z Quad Plus®, Pisces Quad®, Pisces Quad® Compact, Pisces Quad® Plus, Pisces Octad®, Specify™, Resume® II, Resume® TL, SymMix®, Verify® and Temporary Screening Lead
Regulation Number: 21 CFR 882.5880 Regulation Name: Implanted Spinal Cord Stimulator for Pain Relief Regulatory Class: Class II Product Code: GZB
Trade/Devicc Name: Resume® II, Resume® TL and On-Point® Regulation Number: 21 CFR 882.5870 Regulation Name: Implanted Peripheral Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: GZF
Dated: March 3, 2004 Received: March 5, 2004
Dear Mr. Atkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{5}------------------------------------------------
#### Page 2 - Mr. Doug Atkins
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincercly vours.
Mark N. Milliman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Pisces 7 Quad® lead
Pisces Z Quad Compact™ lead Pisces Z Quad Plus® lead Pisces Quad® lead Pisces Quad® Compact Icad Pisces Quad® Plus lead Pisces Octad® lead Specify™ lead Resume II® lead Resume® TL lead SymMix® lead On-Point® lead
Indications for Use:
Pisces Quad Model 3487A, Pisces Quad Compact Model 3887, Pixees Quad Plus Model 3888, Pisces Z Quad Model 3890, Pisces Z Quad Compact Model 3891, Pisces Z Quad Plus Model 3892, Specify Model 3998, Pisces Octad Model 3898, SymMix Model 3982A, Temporary Screening Lead Model 3861, and Verify Model 3862 are indicated for Spinal Cord Stimulation to aid in the management of chronic intractable pain of the trunk and/or limbs.
Resume II Model 35874 and Resume TL Model 3986A are indicated for Spinal Cord Stimulation to aid m the management of chronic intractable pain of the trunk and/or limbs. They are also indicated for Peripheral Nerve Stimulation. The peripheral nerve stimulators are used to stimulate electrically a peripheral nerve in patients to relieve severe intractable pain.
On-Point Model 39874 is indicated for Peripheral Nerve Stimulation. The peripheral nerve stimulators are used to stimulate electrically a peripheral nerve in patients to relieve severe intractable pain.
Prescription Use × (Part 21 CFR 801 Subpart D)
Confidential
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER
PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for Mark N. Milkeran
(Division Sign-off)
Division of General, Restorative,
and Neurological Devices
510(k) Number. K040568
Page 1 of 1
Medtronic
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.