PISCES Z QUAD LEAD MODEL, 3890, PISCES Z QUAD COMPACT LEAD MODEL, 3891, PISCES Z QUAD PLUS LEAD MODEL, 3892

K033016 · Medtronic Neurological · GZB · Oct 30, 2003 · Neurology

Device Facts

Record IDK033016
Device NamePISCES Z QUAD LEAD MODEL, 3890, PISCES Z QUAD COMPACT LEAD MODEL, 3891, PISCES Z QUAD PLUS LEAD MODEL, 3892
ApplicantMedtronic Neurological
Product CodeGZB · Neurology
Decision DateOct 30, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5880
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Medtronic Pisces Z Quad® Model 3890, Pisces Z Quad Compact™ Model 3891, and Pisces Z Quad Plus® Model 3892 Leads for Spinal Cord Stimulation (SCS) are indicated as an aid in the management of chronic, intractable pain of the trunk and/or limbs.

Device Story

Percutaneous quadripolar implantable leads; deliver electrical stimulation to spinal cord; used with Medtronic Mattrix or X-trel neurostimulation systems. Leads feature four platinum iridium electrodes at distal end; proximal end provides in-line four-conductor contacts. Variable electrode length and spacing across models 3890, 3891, 3892. Model 3892 includes radio-opaque marker band for fluoroscopic identification. Leads available in 10-100 cm lengths. Single-use disposable devices. Physician-implanted to provide pain relief via neurostimulation.

Clinical Evidence

Bench testing only.

Technological Characteristics

Percutaneous quadripolar leads; platinum iridium electrodes; in-line four-conductor contacts; 10-100 cm length; radio-opaque marker (Model 3892); disposable; single-use; sterile.

Indications for Use

Indicated for management of chronic, intractable pain of trunk and/or limbs in patients requiring spinal cord stimulation.

Regulatory Classification

Identification

An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # OCT 3 0 2003 K033016 10F2 ## 510(k) Summary Medtronic Models 3890, 3891, and 3892 Leads ### A. General Provisions | Submitter's Name: | Medtronic, Inc. | |-----------------------|-------------------------------------------------------------------------------------------------------------------| | Submitter's Address: | Sullivan Lake Facility<br>800 53rd Avenue Northeast<br>Columbia Heights, MN 55421 | | Contact Person: | Pam Schaub<br>Principal Regulatory Affairs Specialist | | Classification Name: | Implanted Spinal Cord Stimulator for Pain<br>Relief<br>21 CFR Section 882.5880 | | Common or usual Name: | N/A | | Proprietary Names: | Pisces Z Quad® Model 3890 Lead,<br>Pisces Z Quad Compact™ Model 3891 Lead,<br>Pisces Z Quad Plus® Model 3892 Lead | ## B. Name of Predicate Devices | Medtronic Neurological | Models 3487A and 3887 Leads | K923931 | |------------------------|-----------------------------|---------| | Medtronic Neurological | Model 3888 Lead | K923567 | #### C. Device Description The function of the Medtronic Mattrix and X-trel® Neurostimulation Systems is accomplished with a power source, extension (for X-trel only) and lead (electrode). The power source generates and controls the stimulation, which is delivered to the spinal cord via electrodes at the end of the lead. The Models 3890, 3891 and 3892 Leads are lower impedance percutaneous quadripolar, implantable leads. The proximal end provides in-line fourconductor contacts that connect to the Mattrix receiver or Medtronic extension (which connects to the X-trel receiver). The proximal ends and lead bodies of all three leads are identical. Each lead model has four platinum iridium electrodes on the distal end but with variable electrode length and spacing. The electrode spacing and electrode lengths of the Models 3890, 3891, and 3892 Leads are identical to the current Models 3487A, 3887, 3888 Leads, respectively. A radio-opaque marker band at the tip of the Model 3892 Lead enhances identification when viewed in fluoroscopy. The Models 3890, 3891 and 3892 Leads are available in lengths from 10 cm to 100 cm. The Medtronic Models 3890, 3891, and 3892 Leads are packaged, sterilized, and labeled for single use only (disposable). {2}------------------------------------------------ Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is simple and recognizable, conveying the department's authority and commitment to public health. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 3 0 2003 Pam Schaub Principal Regulatory Affairs Specialist 710 Medtronic Parkway NE Minneapolis, Minnesota 55432-5604 Re: K033016 Trade/Device Name: Models 3890, 3891 and 3892 Leads Regulation Number: 21 CFR 882.5880 Regulation Name: Implanted Spinal Cord Stimulator for Pain Relief Regulatory Class: Class II Product Code: GZB Dated: September 25, 2003 Received: September 26, 2003 Dear Ms. Schaub: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. Pam Schaub forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html Sincerely yours, fo Mark N Wilkerson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use | 510(k) Number (if known): | K033016 | |---------------------------|---------| |---------------------------|---------| Medtronic Models 3890, 3891, and 3892 Leads Device Name: INDICATIONS FOR USE: The Medtronic Pisces Z Quad® Model 3890, Pisces Z Quad Compact™ Model 3891, and Pisces Z Quad Plus® Model 3892 Leads for Spinal Cord Stimulation (SCS) are indicated as an aid in the management of chronic, intractable pain of the trunk and/or limbs. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) #### Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use______________________________________________________________________________________________________________________________________________________________ OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109) Mark N. Millerson sion Sign-Off) Division of General, Restorative and Neurological Dev ್ 10(k) Number_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...