K984618 · Bci Intl., Inc. · DXN · Mar 25, 1999 · Cardiovascular
Device Facts
Record ID
K984618
Device Name
BCI MINI-TORR PLUS, MODEL 6004
Applicant
Bci Intl., Inc.
Product Code
DXN · Cardiovascular
Decision Date
Mar 25, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The BCI Mini-Torr Plus monitor is a portable noninvasive blood pressure (NIBP) monitor for spot checking or monitoring of a patient's systolic, diastolic, and mean arterial (MAP) blood pressures, and pulse rate. Optional oximeter, electronic thermometer and/or integral printer are available. The device will provide fast, reliable NIBP measurements on patients ranging form neonate to adults when using the appropriate BCI blood pressure cuff. The monitor has two modes for NIBP monitoring: neonate and adult. The oximetry option works with all BCI oximetry probes, providing SpO2 and pulse rate on all patients from neonate to adult. The electronic thermometry option requires Welch Allyn thermometry probes and probe covers, and provides oral or rectal temperature information for neonate through adult patients. The device is intended for use in both clinical and ground EMS environments by health care professionals. It is not intended for home use. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals for the NIBP and oximetry functions.
Device Story
Portable NIBP monitor; includes optional SpO2 sensor, electronic thermometer, and internal printer. Inputs: NIBP cuff pressure, SpO2 sensor signals, temperature probe data. Operation: NIBP algorithm calculates blood pressure; SpO2 and temperature modules process physiological signals. Output: LED display of systolic/diastolic/MAP, pulse rate, SpO2, temperature; audible/visible alarms. Used in clinical/EMS settings by healthcare professionals. Provides real-time patient monitoring; aids clinical decision-making via adjustable alarm limits and spot-check data. Benefits: portable, multi-parameter monitoring across patient populations from neonate to adult.
Clinical Evidence
Clinical study conducted at two hospitals on neonates (n=size range 0.52-8.60 kg). Compared NIBP monitor algorithm measurements to arterial line pressure transducer (gold standard). Results: Mean differences of -0.4 mmHg (systolic), -0.2 mmHg (diastolic), and -0.6 mmHg (MAP), meeting ANSI/AAMI SP10-1992 requirements (±5 mmHg limit). Standard deviations were 7.6 mmHg (systolic), 5.3 mmHg (diastolic), and 5.1 mmHg (MAP), within 8 mmHg limit.
Technological Characteristics
Portable tabletop monitor with LED display, serial port, and keypad. NIBP, SpO2, and thermometry sensing. Compatible with Welch Allyn temperature probes. Software-controlled NIBP algorithm with adult and neonatal modes. Powered by internal battery/charger. Designed for clinical/EMS use.
Indications for Use
Indicated for spot checking or monitoring systolic, diastolic, and mean arterial blood pressure, pulse rate, SpO2, and temperature in patients ranging from neonates to adults. Intended for use by healthcare professionals in clinical and ground EMS environments. Not for home use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
3/25/99
Image /page/0/Picture/1 description: The image shows a sequence of handwritten characters and numbers. The sequence starts with the letter 'K', followed by the numbers '9', '8', '4', '6', '1', and '8'. The handwriting style appears cursive, with some characters connected to each other.
## Summary of Safety and Effectiveness
| Submitter: | BCI International, Inc. |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | N7 W22025 Johnson Road<br>Waukesha, WI 53186 |
| Telephone: | (414) 542-3100 |
| Contact: | VP Regulatory Affairs |
| Prepared: | December 7, 1998 |
| Proprietary Name: | BCI Mini-Torr Plus (model 6004) with new<br>neonate NIBP mode |
| Common/Classification Name: | Noninvasive blood pressure measurement system |
| Predicate Devices: | BCI 6004 NIBP Monitor (K970801)<br>BCI 6004 Monitor with Temperature Option,<br>510(k) pending (K983796)<br>BCI 6200 Vital Signs Monitor (K953415) |
## New Device Description:
The BCI Mini-Torr Plus noninvasive blood pressure (NIBP) monitor with optional oximeter, electronic thermometer, and printer has been updated to include a new neonatal NIBP mode which uses the same technology as existing legally marketed devices. This device is designed to provide full featured monitoring capabilities in a light weight, transportable design. The system consists of a small table top NIBP monitor with a desk top charger. Features include an NIBP cuff hose connection, an optional SpO2 sensor interface, an optional temperature probe interface and holder, an optional internal printer, display of patient data via an LED display (systolic, diastolic, and mean arterial pressure, interval timer, SpO2, pulse rate, pulse strength, temperature), display of patient NIBP mode via an LED, system status LEDs (battery, sensor, alarm silence, alarm, and alert), and the function keypad area consisting of eleven keys (power, start, cancel, stat, up and down arrows, interval, recall, manual/auto, alarm set, and alarm silence). The monitor has a serial port that is used for data communication.
## Intended Use:
The BCI Mini-Torr Plus monitor is a portable noninvasive blood pressure (NIBP) monitor for spot checking or monitoring of a patient's systolic, and mean arterial (MAP) blood pressures, and pulse rate. Optional oximeter, electronic thermometer and/or integral printer
{1}------------------------------------------------
are available. The device will provide fast, reliable NIBP measurements on patients ranging from neonate to adults when using the appropriate BCI blood pressure cuff. The oximetry option works with all BCI oximetry probes, providing SpO2 and pulse rate on all patients from neonate to adult. The electronic thermometry option requires Welch Allyn thermometry probes and probe covers and provides oral or rectal temperature information for patients neonate through adult.
The device is intended for use in both clinical and ground EMS environments by health care professionals. It is not intended for home use. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals for the NIBP and oximetry functions.
## Performance Data:
The design of this device utilizes currently available technology found in many legally marketed devices. Testing was done previously to ensure that the BCI 6004 monitor would perform within the environment(s) for which it is to be marketed (510[k] pending, K983796). As only software has changed, further environmental testing was not necessary. The previous testing was performed in accordance with the guidelines and standards found in the reviewer's guides for respiratory devices and electronic thermometers. Electromagnetic compatability (EMC), electrical, mechanical durability, safety (operator and patient), and temperature/humidity testing has been completed. The results demonstrated that the BCI 6004 monitor was in compliance with the guidelines and standards referenced in the reviewer's guides and that it performed within its specifications and functional requirements.
Additional safety testing of the new neonatal NIBP mode was performed in accordance with IEC601-2-30: 1995. The monitor passes all of these tests and met all requirements of the standard.
Comparison testing of the 6004 with the new neonate NIBP mode and the predicate 6004 was done to show that the performance of the NIBP adult mode, oximetry, and electronic thermometry parameters of the two devices are the same (systolic, and mean arterial pressures, NIBP heart rate, SpO2, oximetry heart rate, and temperature). Using simulators, measurements were made by both devices. The tests were run with simulator settings spanning the 6004's entire specification range. The difference between measurements made by both devices (predicate - new) was calculated. The average difference between readings were 1.29 mmHg for adult mode systolic pressure, 0.71 mmHg for adult mode diastolic pressure, 0.93 mmHg for adult mode MAP, 0.43 bpm for heart rate measured with the adult mode NIBP parameter, 0.04% for SpO2, 0 bpm for heart rate measured with the oximetry parameter, and -0.05 °F for oral and rectal temperature. All measurements were within the specified tolerances of the monitors and simulators. These data support substantial equivalence of the adult mode NIBP, oximetry, and electronic thermometry parameters of the new 6004 monitor to the predicate 6004 monitor.
{2}------------------------------------------------
Finally, clinical testing of the BCI 6004 NIBP monitor with new neonatal NIBP mode demonstrated compliance to the ANSI/AAMI SP10-1992 standard (American National Standard for Electronic or automated sphygmomanometers ) and ANSI/AAMI/ISO SP10A-1996 Amendment to ANSI/AAMI SP10-1992. The SP-10 testing was conducted after receiving Institutional Review Board (IRB) approvals at two local hospitals on neonates ranging in size from 0.52 to 8.60 kilograms. Measurements of systolic, diastolic, and mean arterial blood pressures made by the 6004 monitor's algorithm were compared to measurements made from an arterial line pressure transducer. The means of these differences are: -0.4 mmHg for systolic pressures, -0.2 mmHg for diastolic pressures, and -0.6 mmHg for mean arterial pressures. These fall within the ±5 mmHg limit imposed by the SP-10 standard. The standard deviations of these differences are: 7.6 mmHg for systolic pressures, 5.3 mmHg for diastolic pressures, and 5.1 mmHg for mean arterial pressures. These fall within the 8 mmHg limit imposed by the SP-10 standard. Further testing demonstrated that the clinical software used for data acquisition and the final software for the monitor obtain the same measurements when tested using simulators.
The testing described above indicate that there is no functional difference between the operation of the 6004 NIBP monitor with new neonatal NIBP mode and the predicate 6004 NIBP monitor for adult mode NIBP, oximetry, and electronic thermometry measurements. Also, the clinical data provide information on the accuracy of the new neonatal NIBP function. On the basis of these results, it is our determination that the device is safe, effective, and performs as well as the legally marketed predicate device(s).
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Respectfully.
Donald Alexander
Donald Alexander VP Regulatory Affairs
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features the department's symbol, which consists of three stylized human profiles facing right, stacked on top of each other. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 5 1999
Mr. Donald J. Alexander BCI International N7 W22025 Johnson Road Waukesha, WI 53186
Re: K984618 BCI Mini-Torr Plus Monitor, Model 6004 with Neonatal NIBP Mode Regulatory Class: II (two) 74 DXN Product Code: December 22, 1998 Dated: December 29, 1998 Received:
Dear Mr. Alexander:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Donald J. Alexander
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number: K984618
Device Name: BCI Mini-Torr Plus Monitor, Model 6004, NIBP monitor with optional oximeter, electronic thermometer, and printer
Indications For Use:
The BCI Mini-Torr Plus monitor is a portable noninvasive blood pressure (NIBP) monitor for spot checking or monitoring of a patient's systolic, diastolic, and mean arterial (MAP) blood pressures, and pulse rate. Optional oximeter, electronic thermometer and/or integral printer are available. The device will provide fast, reliable NIBP measurements on patients ranging form neonate to adults when using the appropriate BCI blood pressure cuff. The monitor has two modes for NIBP monitoring: neonate and adult. The oximetry option works with all BCI oximetry probes, providing SpO2 and pulse rate on all patients from neonate to adult. The electronic thermometry option requires Welch Allyn thermometry probes and probe covers, and provides oral or rectal temperature information for neonate through adult patients.
The device is intended for use in both clinical and ground EMS environments by health care professionals. It is not intended for home use. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals for the NIBP and oximetry functions.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) of litted Shanker
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number_k984
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.