K970801 · Bci Intl., Inc. · DXN · Nov 12, 1997 · Cardiovascular
Device Facts
Record ID
K970801
Device Name
BCI MODEL 6004 NIBP MONITOR (6004)
Applicant
Bci Intl., Inc.
Product Code
DXN · Cardiovascular
Decision Date
Nov 12, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The 6004 NIBP monitor is a portable NIBP monitor for spot checking or monitoring of a patient’s systolic, diastolic and mean arterial (MAP) blood pressures, and pulse rate, with optional SpO₂ and/or integral printer. The device will provide fast, reliable measurements on patients ranging from children (pediatric) to adults when using the appropriate BCI blood pressure cuff. The oximetry option works with all BCI oximetry probes, providing SpO₂ and pulse rate on all patients from neonate to adult. The device is intended for use in both clinical and ground EMS environments by health care professionals. It is not intended for home use. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.
Device Story
BCI 6004 NIBP Monitor is a portable, tabletop device for non-invasive blood pressure (NIBP) and optional pulse oximetry (SpO2) monitoring. Inputs include NIBP cuff pressure signals and SpO2 probe optical signals. Device processes signals to calculate systolic, diastolic, mean arterial pressure, pulse rate, and SpO2 levels. Output is displayed on an integrated LED screen and can be printed via an optional internal printer. Operated by healthcare professionals in clinical and ground EMS settings. Device features adjustable alarm limits with audible/visible alerts. Benefits include rapid, reliable patient monitoring in transportable form factor. Data communication is supported via a serial port.
Clinical Evidence
Bench testing only. NIBP accuracy validated against ANSI/AAMI SP10-1992 standards; mean differences of -1.8 mmHg (systolic) and -2.6 mmHg (diastolic) observed. SpO2 performance validated using Biotek Index simulator and co-oximeter comparison; results within +/- 2% specification. Comparative testing against predicate BCI 6100 and BCI 9000 showed equivalent performance in SpO2 and heart rate measurements.
Technological Characteristics
Tabletop monitor with NIBP cuff interface and SpO2 probe interface. Utilizes MicroNIBP module and LOX board oximeter. Features LED display, serial port, and optional internal printer. Powered by desk top charger. Tested for EMC, electrical safety, and mechanical durability per respiratory device standards.
Indications for Use
Indicated for spot checking or monitoring systolic, diastolic, and mean arterial blood pressure and pulse rate in pediatric to adult patients. Optional SpO2 and pulse rate monitoring indicated for neonate to adult patients. Intended for use by healthcare professionals in clinical and ground EMS environments.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}
BCI INTERNATIONAL
KNIC801
# Summary of Safety and Effectiveness
NOV 12 1997
| Submitter: | BCI International, Inc. |
| --- | --- |
| Address: | W238 N1650 Rockwood Drive
Waukesha, WI 53188 |
| Telephone: | (414) 542-3100 |
| Contact: | VP Regulatory Affairs |
| Prepared: | March 3, 1997 |
| Proprietary Name: | BCI 6004 NIBP Monitor |
| Common/Classification Name: | Noninvasive blood pressure measurement system |
| Predicate Devices: | BCI 6100 Vital Signs Monitor |
## New Device Description:
The BCI 6004 NIBP monitor with optional pulse oximetry (SpO₂) and printer is a new monitor with the same parameters as existing devices legally marketed by BCI International. This device is designed to provide full featured monitoring capabilities in a light weight, transportable design. The system consists of a small table top non-invasive blood pressure monitor with a desk top charger. The system features an NIBP cuff hose connection, an SpO₂ probe interface, the optional internal printer, display of patient data via an LED display (Systolic, Diastolic, & Mean arterial pressure, Interval timer, SpO₂, Pulse Rate, Pulse Strength), system status LEDs (Battery, Probe, Alarm Silence, Alarm, & Alert), and the function keypad area consisting of eleven keys (O/I (off/on), START, CANCEL, STAT, Up and Down Arrows, INTRVL, RECALL, MANUAL/AUTO, ALARM SET, & Alarm Silence). The monitor has a serial port that is used for data communication. The model 6004 has two parameters, NIBP and SpO₂ plus the integrated printer.
## Intended Use:
The 6004 NIBP monitor is a portable NIBP monitor for spot checking or monitoring of a patient’s systolic, diastolic and mean arterial (MAP) blood pressures, and pulse rate, with optional SpO₂ and/or integral printer. The device will provide fast, reliable measurements on patients ranging from children (pediatric) to adults when using the appropriate BCI blood pressure cuff. The oximetry option works with all BCI oximetry probes, providing SpO₂ and pulse rate on all patients from neonate to adult. The device is intended for use in both clinical and ground EMS environments by health care professionals. It is not intended for home use. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.
BCI International
W238 N1650 Rockwood Drive
Waukesha, WI 53188-1199 USA
Phone: 414/542-3100 Fax: 414/542-0718
{1}
Performance Data:
The design of this device utilizes currently available technology found in many legally marketed devices. Testing was done to ensure that it would perform within the environment(s) for which it is to be marketed. The testing was performed in accordance with the guidelines and standards found in the reviewers guide for respiratory devices. This testing included EMC, electrical, mechanical durability, safety (operator and patient), and temperature/humidity. The results of the testing demonstrated that the device was in compliance with the guidelines and standards referenced in the reviewers guide and that it performed within its specifications and functional requirements.
SpO₂ & HR
Performance testing between the new 6004 with the LOX board oximeter option and the predicate 6100 was done to show that the performance of the two devices is the same (SpO₂ & HR).
The SpO₂ simulator was the Biotek Index (K933519). All the SpO₂ results were within one count (max) of the simulator and each other. The same was true of heart rate until 240 bpm was run. Both devices were within 0 to 3 counts of each other and the simulator. The result was still within the device specification of +/- 2% (@ 240 -> 4.8 counts).
A deep desaturation test was run on the LOX board oximeter using the Capnocheck Plus (9004) as the host at the VA Medical Center in Milwaukee under an approved IRB. The LOX values were compared to an OSM-3 co-oximeter. Over the SpO₂ range of 70% - 100% the standard deviation was 2.0 (spec = +/- 2%). Over the SpO₂ range of 50% - <70% the standard deviation was 2.7 (spec = +/- 3%). R squared = 0.97 (measure of how true the regression line is, one being perfect).
A test was run in-house comparing the 6004 with the LOX board to the 9004 with a LOX board to show that the device will give the same reading even if the host is different. Another monitor, the BCI 9000 Capnograph with Oximeter was also run at the same time. (BCI 9000 K873856). Again the SpO₂ simulator was the Biotek Index (K933519). All the SpO₂ results were within one count (max) of the simulator and each other. The same was true of heart rate until 240 bpm was run. All of the devices were within 0 to 3 counts of each other and the simulator. The result was still within the device specification of +/- 2% (@ 240 -> 4.8 counts).
NIBP
Performance testing for the NIBP function was composed of two sections. The first part was testing to the ANSI/AAMI SP10-1992 standard (American National Standard for Electronic or automated sphygmomanometers) & ANSI/AAMI/ISO SP10A-1996 Amendment to ANSI/AAMI SP10-1992. This test was to determine the accuracy of the NIBP technology using the MicroNIBP module. The next test showed that the 6004 monitor with the same hardware design, electronics and software as the MicroNIBP performed the same. The MicroNIBP design was moved into the 6004 design because of space and cost constraints.
The SP-10 testing was conducted under an approved IRB at the VA Medical Center in Milwaukee and at BCI International. When the MicroNIBP measurements are compared to the average of the manual readings the mean difference is -1.8 mmHg on systolic pressures and -2.6 mmHg on diastolic pressures. This meets the SP-10 requirement for a maximum mean difference of +/- 5 mmHg. The standard deviation of the difference between the MicroNIBP values and the average manual values was 7.1 mmHg for systolic pressures and 7.6 mmHg for diastolic
{2}
pressures. This falls within the 8 mmHg limit imposed by the SP-10 standard.
The next test showed that the MicroNIBP and the 6004 NIBP monitor will give the same NIBP readings. Using the Dynatech Nevada CuffLink NIBP Analyzer as the NIBP simulator a series of readings where taken over the specification range. The CuffLink repeatability specification is +/- 1% of the selected target value. The test was run for heart rates from 30 bpm to 200 bpm (over the spec) and a pressure range (SYS/DIA) of 60/30 to 250/200 mmHg. Three readings were taken at each setting for both devices. An average of the three readings was calculated for both devices. Then the difference between devices was calculated. The average difference of all the readings were 1.0 for SYS, 1.1 for DIA, 1.1 for MAP and 0 for HR. This shows that the MicroNIBP and the 6004 NIBP monitor operate the same.
The 6004 passed all the tests.
On the basis of these results and the above-referenced testing it is our determination that the device is safe, effective, and performs as well as or better than the legally marketed predicate device(s).
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Respectfully,

Donald Alexander
VP Regulatory Affairs
{3}
HUMAN SERVICES LTD.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
NOV 12 1997
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
NOV 12 1997
Mr. Donald J. Alexander
VP, Regulatory Affairs
BCI International, Inc.
W238 N1650 Rockwood Drive
Waukesha, Wisconsin 53188-1199
Re: K970801
BCI Model 6004 NIBP Monitor
Regulatory Class: II (Two)
Product Code: 74 DXN
Dated: September 22, 1997
Received: September 23, 1997
Dear Mr. Alexander:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2 - Mr. Donald J. Alexander
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
870.1130 - D
# Indications For Use
510(k) Number (if Known): K970801
Device Name: BCI 6004 NIBP Monitor
Indications For Use:
## Intended Use
The 6004 NIBP monitor is a portable NIBP monitor for spot checking or monitoring of a patient’s systolic, diastolic and mean arterial (MAP) blood pressures, and pulse rate, with optional SpO₂ and/or integral printer. The device will provide fast, reliable measurements on patients ranging from children (pediatric) to adults when using the appropriate BCI blood pressure cuff. The oximetry option works with all BCI oximetry probes, providing SpO₂ and pulse rate on all patients from neonate to adult. The device is intended for use in both clinical and ground EMS environments by health care professionals. It is not intended for home use. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K970801
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The_Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.