The Bayer Immuno 1™ System has been upgraded for Laboratory Automation capability. Software and Hardware will facilitate communication and robotic transfer of samples to and from a Laboratory Automation system. The Laboratory Automation System (LAS) will coordinate the scheduling, sample identification, and physical transfer of samples, as demonstrated for the following representative methods: CEA, FSH, Free T4, PSA, T3, T4, TSH, and T-Uptake.
Device Story
The Bayer Immuno 1 System Upgraded for Laboratory Automation (LAS) is an enhancement to the existing Bayer Immuno 1 diagnostic platform. It integrates hardware and software modifications to enable communication and physical interface with external laboratory automation systems. The system receives samples via robotic transfer from the LAS, which manages sample scheduling and identification. The device performs diagnostic assays on these samples and returns results to the laboratory information system. Used in clinical laboratory environments, it is operated by laboratory technicians. The upgrade ensures that the analytical performance of the Immuno 1 system remains consistent with the predicate device while allowing for high-throughput, automated workflow integration. This automation reduces manual handling, improves sample throughput, and maintains clinical decision-making accuracy for the specified analytes.
Clinical Evidence
Bench testing only. Performance was evaluated by comparing the upgraded system against the predicate Bayer Immuno 1 System using 360 samples for most assays (200 for PSA). Primary endpoints included precision (Total CV%) and method correlation (regression analysis). Correlation coefficients (r) ranged from 0.987 to 1.000 across all eight representative analytes. Precision data demonstrated consistent performance between the upgraded and predicate systems.
Technological Characteristics
The device is an automated in vitro diagnostic analyzer upgraded with hardware and software interfaces for robotic sample transfer. It utilizes the same analytical methods and reagents as the predicate Bayer Immuno 1 System. Connectivity is established via communication protocols for integration with Laboratory Automation Systems (LAS). The system is designed for clinical laboratory use, operating as a standalone analyzer integrated into a larger automated track system.
Indications for Use
Indicated for use in clinical laboratory settings to facilitate automated sample handling, scheduling, and identification for diagnostic assays including CEA, FSH, Free T4, PSA, T3, T4, TSH, and T-Uptake.
Regulatory Classification
Identification
A discrete photometric chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. Different models of the device incorporate various instrumentation such as micro analysis apparatus, double beam, single, or dual channel photometers, and bichromatic 2-wavelength photometers. Some models of the device may include reagent-containing components that may also serve as reaction units.
{0}------------------------------------------------
Attachment 3
#### 7 1998 DEC
## Summary of Safety and Effectiveness
# Bayer Immuno 1™ System Upgraded for Laboratory Automation Capability
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. Below is a comparison of performance, using eight representative assays, between the Bayer Immuno 1 System upgraded for Laboratory Automation capability and the Bayer Immuno 1™ System, already granted clearance of substantial equivalence for these representative and other assays.
#### Intended Use
The Bayer Immuno 1 System upgraded for laboratory automation capability will facilitate communication and robotic transfer of samples to and from a Laboratory Automation system. The Laboratory Automation system will coordinate the scheduling, sample identification, and physical transfer of samples.
#### Device
Bayer Immuno 1" System Upgraded for Laboratory Automation Capability
Gabriel J. Murace, Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
9/21/98
Date
{1}------------------------------------------------
| Part Numbers | | |
|--------------|-----------|---------------------------------------------------|
| | CEA: | Reagents: T01-3184-51<br>Calibrators: T03-3188-01 |
| | FSH: | Reagents: T01-3086-51<br>Calibrators: T03-3148-01 |
| | Free T4: | Reagents: T01-3360-51<br>Calibrators: T03-3401-01 |
| | PSA: | Reagents: T01-3450-51<br>Calibrators: T03-3541-01 |
| | T3: | Reagents: T01-2949-01<br>Calibrators: T03-2872-01 |
| | T4: | Reagents: T01-3260-51<br>Calibrators: T03-3174-01 |
| | TSH: | Reagents: T01-2942-51<br>Calibrators: T03-3568-01 |
| | T-Uptake: | Reagents: T01-3036-51<br>Calibrators: T03-3076-01 |
### Predicate Device
Bayer Immuno 1™ System
## Part Numbers
The part numbers are identical to those used for the Bayer Immuno 1 " System upgraded for laboratory automation capability.
.
{2}------------------------------------------------
| BAYER IMMUNO 1™ UPGRADED SYSTEM | | | BAYER IMMUNO 1™ METHOD SHEET | |
|---------------------------------|-----------------|-----------------|------------------------------|-----------------|
| METHOD | MEAN<br>(UNITS) | TOTAL<br>CV (%) | MEAN<br>(UNITS) | TOTAL<br>CV (%) |
| CEA | 1.54 ng/mL | 3.3 | 2.9 ng/mL | 2.3 |
| | 13.29 ng/mL | 1.2 | 8.9 ng/mL | 2.2 |
| | 63.43 ng/mL | 1.4 | 18.2 ng/mL | 1.8 |
| FSH | 9.4 mIU/mL | 1.6 | 5.5 mIU/mL | 3.2 |
| | 15.26 mIU/mL | 1 | 12.3 mIU/mL | 2.8 |
| | 47.51 mIU/mL | 0.7 | 30.3 mIU/mL | 2.8 |
| FREE T4 | 0.83 ng/dL | 4.1 | 0.36 ng/dL | 15.4 |
| | 1.43 ng/dL | 4.3 | 1.25 ng/dL | 6.0 |
| | 3.06 ng/dL | 4.7 | 4.35 ng/dL | 3.1 |
| PSA | 2.89 ng/mL | 3.1 | 0.05 ng/mL | 0.007 ng/mL |
| | 6.9 ng/mL | 2.4 | 0.92 ng/mL | 2.9 |
| | 26.41 ng/mL | 2.4 | 2.68 ng/mL | 3.1 |
| | | | 9.87 ng/mL | 2.2 |
| | | | 23.02 ng/mL | 3.4 |
| | | | 49.13 ng/mL | 2.0 |
| | | | 96.38 ng/mL | 2.5 |
| T3 | 0.66 ng/mL | 5.1 | 0.46 ng/mL | 13.3 |
| | 1.78 ng/mL | 2.4 | 1.34 ng/mL | 6.0 |
| | 2.9 ng/mL | 2 | 3.43 ng/mL | 3.9 |
| T4 | 3.31 µg/dL | 1.6 | 4.7 µg/dL | 3.6 |
| | 7.57 µg/dL | 2.2 | 8.2 µg/dL | 2.6 |
| | 14.12 µg/dL | 1.2 | 15.7 µg/dL | 2.5 |
| TSH | 0.55 µIU/mL | 6.4 | 0.1 µIU/mL | 13.0 |
| | 5.08 µIU/mL | 1 | 1.3 µIU/mL | 6.3 |
| | 32.73 µIU/mL | 1.2 | 9.0 µIU/mL | 2.0 |
| | | | 22.5 µIU/mL | 1.8 |
| T-UPTAKE | 0.97 | 3.3 | 0.71 | 2.8 |
| | 1.03 | 3.1 | 1.03 | 2.6 |
| | 1.24 | 2.2 | 1.41 | 2.4 |
.
- 1) Precision for Bayer Immuno 1™ System Upgraded for Laboratory Automation Capability:
* Standard deviation
:
.
:
ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﻋﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟ
{3}------------------------------------------------
## 2) Correlation:
| COMPARISON<br>METHOD | NUMBER<br>OF<br>SAMPLES | REGRESSION<br>EQUATION<br>y = | CORRELATION<br>COEFFICIENT<br>r | STANDARD<br>ERROR<br>Sy.x | RANGE OF<br>ANALYTE<br>CONCENTRATION |
|----------------------|-------------------------|-------------------------------|---------------------------------|---------------------------|--------------------------------------|
| CEA | 360 | 1.01x - 0.01 | 0.998 | 0.116 | 0.4 - 16.4 ng/mL |
| FSH | 360 | 0.99x + 0.02 | 0.999 | 0.841 | 0 - 145 mIU/mL |
| Free T4 | 355 | 0.99x + 0.04 | 0.988 | 0.042 | 0.26 - 1.98 ng/dL |
| PSA | 200 | 1.00x + 0.01 | 1.000 | 0.066 | 0 - 20.3 ng/mL |
| T3 | 360 | 0.99x + 0.02 | 0.987 | 0.046 | 0.48 - 2.16 ng/mL |
| T4 | 356 | 0.98x + 0.07 | 0.997 | 0.141 | 3.7 - 16.4 µg/dL |
| TSH | 360 | 0.99x + 0.02 | 1.000 | 0.108 | 0.23 - 77 µIU/mL |
| T-Uptake | 358 | 0.98x + 0.03 | 0.991 | 0.019 | 0.61 - 1.41 |
:
.
:
:
.
y = Bayer Immuno 1™ System upgraded for laboratory automation capability
x = Bayer Immuno 1™ System
: :
: .
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a single snake winding around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 7 1998
Gabriel J. Muraca, Jr. Manager Requlatory Affairs BAYER CORPORATION 511 Benedict Avenue Tarrytown, NY 10591-5097
Re: K983345 Trade Name: Bayer Immuno 1™ System Upgraded for Laboratory Automation Capability Requlatory Class: II Product Code: JJE September 22, 1998 Dated: Received: September 23, 1998
Dear Mr. Muraca:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls: Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1 1. 1. 1. 1.
### 510(k) Number (if known): K983345
# Device Name: Bayer Immuno 1™ System Upgrade for LAS
Indications For Use:
The Bayer Immuno 1™ System has been upgraded for Laboratory Automation capability. Software and Hardware will facilitate communication and robotic transfer of samples to and from a Laboratory Automation system. The Laboratory Automation System (LAS) will coordinate the scheduling, sample identification, and physical transfer of samples, as demonstrated for the following representative methods: CEA, FSH, Free T4, PSA, T3, T4, TSH, and T-Uptake.
_ .
. .. . .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--|--------------------------------------------------------|--|
|--|--------------------------------------------------------|--|
| Prescription Use | ✓ |
|----------------------|---|
| (Per 21 CFR 801.109) | |
OR
| Over-The-Counter Use | |
|----------------------|-------------------------|
| | Optional Format 1-2-96) |
(Division Sign-Off)
Division of Clinical Laboratos ! 510(k) Number ________________________________________________________________________________________________________________________________________________________________ 983345 1-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.