BIOSORBFX 2.0/2.4 ORTHOGNATHIC AND MANDIBULAR FIXATION SYSTEM
K982721 · Bionx Implants, Inc. · JEY · Feb 17, 1999 · Dental
Device Facts
Record ID
K982721
Device Name
BIOSORBFX 2.0/2.4 ORTHOGNATHIC AND MANDIBULAR FIXATION SYSTEM
Applicant
Bionx Implants, Inc.
Product Code
JEY · Dental
Decision Date
Feb 17, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BioSorbFX 2.0/2.4 Bioabsorbable Fixation System ("BioSorbFX 2.0/2.4 System") is intended for use in trauma and reconstructive procedures in the midface and maxilla. Specifically, the device is indicated for use in surgical repair procedures in the treatment of trauma to the midface and maxilla, and in orthognathic and reconstructive procedures of the midface or maxilla. An instrumentation set which is used to implant the fasteners to secure the plates to bone is included with the BioSorbFX 2.0/2.4 System. The BioSorbFX 2.0/2.4 System is not intended for use in mandibular applications.
Device Story
BioSorbFX 2.0/2.4 is a bioabsorbable bone fixation system comprising poly-L/DL-lactide (P(L/DL)LA) copolymer plates, mesh, and screws (2.0 mm, 2.4 mm, and 2.8 mm emergency). Used in midface and maxillary trauma/reconstructive surgery; implanted by surgeons. System stabilizes bone during healing; material degrades via hydrolysis over 2-4 years. Accessory instrumentation (drills, taps, screwdrivers, plate benders) made of 400-series stainless steel. Plates are cut/bent at room temperature. Provided sterile via gamma irradiation for single use. Clinical benefit includes bone stabilization without need for secondary removal surgery, as material resorbs in vivo.
Clinical Evidence
Clinical study of 10 patients (ages 17-33) undergoing bimaxillary surgery using 1.2 mm plates and 2.0 mm screws. Follow-up at 6 weeks and 3 months. Results: 96.1% of screws had good fit; no delayed bone union; jaws clinically stable at 6 weeks; no chronic inflammation or plate exposure. One infection occurred (resolved with antibiotics). Bench testing included mechanical strength (bending, shear, torsion, pull-out) and sterility validation. Animal studies in sheep mandibles also provided.
Technological Characteristics
Materials: Poly-L/DL-lactide (P(L/DL)LA) copolymer (70% L-lactide, 30% D,L-lactide); 400-series stainless steel instruments. Form factor: Plates (1.0-1.4 mm thick), mesh, and screws. Sterilization: Gamma irradiation (min 25 kGy). Degradation: Hydrolysis into lactate and glutate monomers; complete resorption in ~4 years. Mechanical: Self-reinforced copolymer matrix with axially oriented fibers.
Indications for Use
Indicated for trauma and reconstructive surgical repair procedures of the midface and maxilla. Contraindicated for mandibular applications.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
{0}------------------------------------------------
| | K982721 | | "SUBSTANTIAL EQUIVALENCE" (SE) DECISION MAKING DOCUMENTATION |
|--|---------|--|--------------------------------------------------------------|
|--|---------|--|--------------------------------------------------------------|
| REVIEWER: | Pamela D. Scott |
|------------------|-----------------|
| DIVISION/BRANCH: | DGRD/Dental |
| TRADE NAME: | BioSorbFX 2.0/2.4 Orthognathic Bioabsorbable Fixation System |
|--------------|--------------------------------------------------------------|
| COMMON NAME: | Bone Plates and Bone Screws |
| PRODUCT TO WHICH COMPARED: | LactoSorb® Trauma Plating System (K971870) |
|----------------------------|-----------------------------------------------------------|
| (510(k) NUMBER IF KNOWN) | BioSorbFX 1.5/2.0 Bioabsorbable Fixation System (K982139) |
| | YES | NO | |
|------------------------------------------------------------------------------------|-----|----|----------------------------------------|
| 1. IS PRODUCT A DEVICE? | X | | - IF NO STOP |
| 2. DEVICE SUBJECT TO 510(k)? | X | | - IF NO STOP |
| 3. SAME INDICATION STATEMENT? | X | | - IF YES GO TO 5 |
| 4. DO DIFFERENCES ALTER THE EFFECT OR RAISE NEW ISSUES OF SAFETY OR EFFECTIVENESS? | | | - IF YES STOP - NE |
| 5. SAME TECHNOLOGICAL CHARACTERISTICS? | X | | - IF YES GO TO 7 |
| 6. COULD THE NEW CHARACTERISTICS AFFECT SAFETY OR EFFECTIVENESS? | | | - IF YES GO TO 8 |
| 7. DESCRIPTIVE CHARACTERISTICS PRECISE ENOUGH? | X | | - IF NO GO TO 10<br>- IF YES STOP - SE |
| 8. NEW TYPES OF SAFETY OR EFFECTIVENESS QUESTIONS? | | | - IF YES STOP - NE |
| 9. ACCEPTED SCIENTIFIC METHODS EXIST? | | | - IF NO STOP - NE |
| 10. PERFORMANCE DATA AVAILABLE? | X | | - IF NO REQUEST DATA |
| 11. DATA DEMONSTRATE EQUIVALENCE? | X | | |
NOTE: IN ADDITION TO COMPLETING PAGE TWO, "YES" RESPONSES TO QUESTIONS
- NOTE: IN ADDITION TO COMPLETING PAGE TWO, "YES" RESPONSES TO QUESTIONS 4, 6, 8, AND 11, AND EVERY "NO" RESPONSE REQUIRES AN EXPLANATION ON PAGE THREE AND/OR FOUR.
{1}------------------------------------------------
## NARRATIVE DEVICE DESCRIPTION
- 1. INTENDED USE: The BioSorbFX Orthognathic and Mandibular (0/M) Bioabsorbable Fixation System is intended for use in trauma and reconstructive procedures in the midface, maxilla and mandible. Specifically, the device is indicated for use in surgical repair procedures in the treatment of trauma to the midface, maxilla and mandible, and in orthognathic and reconstructive procedures of the midface, maxilla or mandible. The BioSorbFX O/M System stabilizes bone during healing in conjunction with appropriate postoperative immobilization. The BioSorbFX O/M System is not intended for use in and is contraindicated for: 1) mandibular tumor resection; 2) significant comminuted fractures, including significant bone loss, of the mandible and 3) intermaxillary fixation.
- 2. DEVICE DESCRIPTION: Provide a statement of how the device is either similar to and/or different from other marketed devices, plus data (if necessary) to support the statement. The followinq should be considered when preparing the summary of the statement. Is the device life-supporting or life sustaining? Is the device implanted (short-term or long-term)? Does the device design use software? Is the device sterile? Is the device for single use? Is the device for home use or prescription use? Does the device contain drug or biological product as a component? Is this device a kit? Provide a summary about the devices design, materials, physical properties and toxicology profile if important.
#### SUMMARY :
The BioSorbFX O/M System is a system of bone plate and mesh configurations affixed to bone by 2.0 mm and 2.4 mm screws. The system also includes a 2.8 mm emergency screw and accessory instrumentation for implantation of the screws. The 2.0 mm screw has a minor (shaft) diameter of 1.45 mm and is available in lengths of 4, 6, 8, 10, 12, 14, 16, 18, 20 ,24, 28, 32, 36 and 40 mm. The 2.4 mm screw has a minor diameter of 1.55 mm and is provided in lengths in 6, 8, 10, 12, 14, 16, 18, 20 ,24, 28, 32, 36 and 40 mm. The 2.8 mm emergency screw has a minor diameter of 1.95mm and lengths of 6, 8, 10, 12, 14, 16, 18, 20 ,24, 28, 32, 36 and 40 mm. The 2.8 mm screw is intended to serve as an emergency screw for the 2.4 mm screw and the 2.4 mm screw can be used as an emergency screw for the 2.0 mm screw. The accessory instrumentation set contains bone drills, bone taps, a screwdriver and a plate bender. It is equivalent to the instrumentation set that is a part of the Bionx BioSorb Endobrow Screw. The only difference is that the components have been resized to correspond to the sizes of the screws and the plate bender has been added to the set. The accessory instruments are fabricated from stainless steel (400 series).
The bone plates will be available in 1.2 mm and 1.4 mm thick plates of various shapes. It should also be noted that the thickness of the mesh plates is 1 mm rather than 1.2 or 1.4 mm. The bone plates may be cut or bent at room temperature with being heated. The BioSorbFX O/M
{2}------------------------------------------------
System stabilizes bone during healing in conjunction with postoperative immobilization. The system also includes 5x5 hole and 5x10 hole rectangular meshes. The BioSorbFX O/M System will be provided sterile and pyrogen free. The components are to be sterilized via gamma radiation.
The components of the system are made from poly-L/DL-lactide {P(L/DL)LA} copolymer. This material has been used for other Bionx, Inc. predicate implant devices such as the Biodegradable Threaded Suture Anchor (K972783) and the BioSorb Endobrow Screw (K972919). The LactoSorb® Trauma Plating System, by Walter Lorenz Surgical, Inc., is also made from a bioabsorbable polylactide material. Bionx, Inc. has referenced the FDA Master File 357 for this material that was submitted by Boehringer Ingleheim; the Bionx product is the same as that described in MAF 357. The P(L/DL)LA copolymer is a bioabsorbable polylactide polymer that degrades within 2 to 3 years postoperatively; the material is generally completely resorbed in 4 years.
The P(L/DL)LA copolymer is a blend of 70 molar percent L-lactide and 30 molar percent D,L-lactide and maintains its strength for 12 to 22 weeks. This is longer than the healing periods generally required for the device's indications for use. The P(L/DL)LA copolymer is a selfreinforced copolymer, while the LactoSorb® is made of a synthetic polyester consisting of lactic and glycolic acids (PLA/PGA). The P(L/DL)LA copolymer degrades in vivo by hydrolysis into lactate and glutate monomers. L-lactate, the major conformation of lactic acid found in mammals, can be oxidized to form pyruvate, which is used in glucose synthesis or is metabolized into carbon dioxide and water.
Details regarding the chemical composition of the P(L/DL)LA copolymer were provided. The polymer contains less than 5 ppm of chromium, cobalt, manganese, molybedenum and tin and less than 20 ppm of aluminum, nickel, silicon and iron. Acetone is used as a solvent for processing the raw material. The final level of solvent contained in the product after processing the plates and fasteners is less than 0.1 wt% of the residue, toluene. The monomer content of P(L/DL)LA is less The average molecular weight of the raw material is than 0.5%. 260,000 - 430,000 Daltons (D) and the intrinsic viscosity of 5 -7dl/g. The molecular weight of the finished product ranges from 35,000 - 65,000 D with an intrinsic viscosity of 1.1 to 1.8 dl/q. The copolymer is amorphous. The polymer matrix is reinforced with P(L/DL)LA fibers that are oriented axially. The surface of the BioSorbFX O/M System plates is smooth, with an average roughness (Rg) of 2 - 8 pm.
The BioSorbFX O/M System is sold sterile and is for single use only. The devices are sterilized by gamma irradiation with a minimum dose of 25 kGy. The SAL is 10°. Each plate is packaged in an individual paper board holder. One to six screws are packaged in a paper board holder. Both the plates and screws are sealed in an aluminum foil pouch, which is then sealed in a Tyvek® pouch and placed in cardboard boxes. In case the device is accidentally contaminated by the user prior to insertion, the device is not to be resterilized by the user. After sterilization, samples from each lot are subjected to final
{3}------------------------------------------------
testing to validate conformance to specifications for sterility, mechanical strength and packaging integrity.
Animal and clinical studies have shown that the resorption rate is such that accumulation of the by-products of P(L/DL)LA hydrolysis is not sufficient to affect the blood pH levels.
## SUMMARY OF CLINICAL STUDIES
Clinical Study of Bionx BioSorbFX O/M P(L/DL)LA 2.0-2.4 mm Fixation System for Orthognathic Surgery:
- Treatment: bimaxillary surgery �
- Sample size: 10 patients (5 female, 5 male) �
- Age range: 17 33 (average age: 22) �
- Exclusions: patients with congenital deformities, local or systemic � diseases or a history of maxillofacial trauma
- � Surqical procedures: bimaxillary surgery and genioplasty
- Devices used: 1.2 mm plates and 2.0 mm screws �
- Total number of devices used: 62 plates, 40 in the maxilla, 20 in � the mandible and 2 in the chin; 305 screws
- Patients were followed for 6 weeks and 3 months postoperatively; � seven patients were evaluated at the 3 month postoperative time point
- � Results:
- None of plates broke during bending for surgical placement, A although internal delamination occurred as evidenced by opaque lines in the bent areas; the plates were able to function with sufficient strength and stability
- A 293 (96.1%) of 305 screws used had good fit; the slits in the screw head of 18 of these were damaged at the last twist, but this did not influence the fit in the bone
- No cases of delayed bone union A
- A At six weeks, all operate jaws clinically stable
- A At six weeks, no evidence of chronic inflammatory reaction, wound dehiscence or plate exposure
- A Six weeks postoperatively, occlusion was stable; there was no need to remove plates and screws because of malocclusion in any case
- A There were minimal changes in the occlusion between the six week and three month postoperative periods
- A The short term stability appeared to be better when compared to wire osteosynthesis and similar to the results of metal devices
- � Adverse events:
{4}------------------------------------------------
- A One patient developed an infection three weeks postoperatively, which resolved after intraoral incision and antibiotics
- A Twelve screws deficient in their fit: 6 screwheads broke off before full insertion; 6 screws had an insufficient fit; the screws were either replaced or holes drilled through the screw and a new screw inserted; there were no other adverse effects resulting from these screw failures
The remainder of the clinical information provided is based on data from clinical studies in which PLLA bone plates and/or screws were used for mandibular fixation. Animal studies were also submitted for both PLLA bone plates and screws. One animal study using the P(L/DL)LA bone plates and screws in the mandibles of sheep was submitted.
#### ANALYSIS:
The material used for this device appears to be well characterized chemically and in terms of toxicological biocompatibility. As stated by the company, the material has been used for several other bone fixation devices. In addition, the BioSorbFX 1.5/2.0 Bioabsorbable Fixation System was also recently cleared under K982139. This fixation system is similar to the present device except the screw hole size, the corresponding screw size and the indications for use of the device. The BioSorbFX 1.5/2.0 Bioabsorbable Fixation System is intended for midface and maxillary indications only. It is specifically contraindicated for mandibular use.
The BioSorbFX 2.0/2.4 O/M System is indicated for both midface/maxillary indications and mandibular indications. Sufficient biocompatibility information and mechanical strength retention information has been provided for the midface/maxillary indications. However, additional information was requested for the mandibular indication.
In response to the additional information requested regarding the mandibular indication, the sponsor has decided to remove this indication from the labeling. Additional information items 1, 2, and 3 were in reference to the mandibular indication; therefore, the information requested in these items is no longer necessary at this The applicant only seeks clearance for the midface and time. maxillary indication.
Based on the engineering drawings, it appeared as though the mesh plates receive a smaller screw. The applicant was asked to clarify the screw that is to be used with the mesh plates. It appears as though the 1.5 mm screw previously cleared in K982139 is to be used with the mesh plates. The applicant confirmed this by phone on February 16, 1999.
{5}------------------------------------------------
In a phone call to the company's consultant on December 24, 1998, the company was asked to explain the use and need for the longer screw lengths (i.e., those lengths above 28 mm) in midface and maxillary indications. They were also asked, in a subsequent telephone conversation, whether or not any testing had been performed on the longer screw lengths. The applicant responded that they would limit the screw lengths to the following.
| Diameter (mm) | Lengths (mm) |
|---------------|-------------------------------------|
| 2.0 mm | 4, 6, 8, 10, 12, 14, 16, 18 |
| 2.4 mm | 4, 6, 8, 10, 12, 14, 16, 18, 20 ,24 |
| 2.8 mm | 4, 6, 8, 10, 12, 14, 16, 18, 20 ,24 |
The 2.0 mm screw has the same length range as that of the previously cleared BioSorbFX 1.5/2.0 Bioabsorbable Fixation System. The applicant clarified that the bending and shear strength retention tests results previously submitted were performed on the 2.4 x 40 mm and 2.8 x 40 mm screws. In addition, torsional testing was performed on the 2.4 x 20 mm and 2.8 x 20 mm screws. The pull-out testing was performed on the 2.0 x 10 mm, 2.4 x 12 mm and the 2.8 x 12 mm screws. The results of this testing is acceptable.
## RECOMMENDATION
The BioSorbFX 2.0/2.4 Orthognathic Bioabsorbable Fixation System may be found substantially equivalent to other bioabsorbable fixation plating systems.
P.D. Scott 2/16/99
Sm
Shaka
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three abstract human figures connected to a single, flowing line.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 7 1999
Bionx Implants, Incorporated C/O Mr. Jonathan S. Kahan Hogan & Hartson Columbia Square 555 Thirteenth Street, NW Washington, D.C. 20004-1109
Re : K982721 BioSorbFX 2.0/2.4 Orthognathic and Trade Name: Mandibular Bioabsorbable Fixation System Regulatory Class: II Product Code: JEY Dated: November 19, 1998 November 19, 1998 Received:
Dear Mr. Kahan
We have reviewed your Section 510(k) notification of intent to we have reviewed four benced above and we have determined the marree is substantially equivalent (for the indications for device in babbeansearly oge) to devices marketed in interstate abe beaution to May 28, 1976, the enactment date of the Commondal Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of
{7}------------------------------------------------
Page 2 - Mr. Kahan
the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## INDICATIONS FOR USE
510(K) Number (if known): ___________________________________________________________________________________________________________________________________________________
Device Name:
BioSorbFX 2.0/2.4 Bioabsorbable Fixation System
The BioSorbFX 2.0/2.4 Bioabsorbable Fixation Indications for Use: System ("BioSorbFX 2.0/2.4 System") is intended for use in trauma and reconstructive procedures in the midface and maxilla. Specifically, the device is indicated for use in surgical repair procedures in the treatment of trauma to the midface and maxilla, and in orthognathic and reconstructive procedures of the midface or maxilla. An instrumentation set which is used to implant the fasteners to secure the plates to bone is included with the BioSorbFX 2.0/2.4 System. The BioSorbFX 2.0/2.4 System is not intended for use in mandibular applications.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-off) Division of Dental, Infection Control, and General Hospital Devices
| Prescription Use<br>(Per 21 CFR 801.109) | <div> <div style="display:inline-block;"> OR </div> <div style="display:inline-block;"> Over-The-Counter Use </div> </div> <div> (Division Sign-Off) </div> <div> Division of Dental, Infection Control, and General Hospital Devices </div> <div> 510(k) Number </div> <div> 10982721 </div> |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
\\\DC - 63547/14 - 0768567.01
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.