UniCAP Thyroglobulin ImmunoCAP™ is a device for the in vitro quantitative measurement of IgG antibodies specific for Thyroglobulin (TG) in human serum and plasma. UniCAP Thyroglobulin ImmunoCAP is intended to be used with the instrument UniCAP together with reagents as stated in the Directions For Use provided with UniCAP Specific IgG. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of certain thyroid diseases, such as autoimmune thyroiditis and Graves' Disease and is to be used in clinical laboratories, as well as, physician office laboratories. UniCAP Thyroid Peroxidase ImmunoCAP is a device for the in vitro quantitative measurement of IgG antibodies specific for Thvroid Peroxidase (TPO) in human serum and plasma. UniCAP Thyroid Peroxidase ImmunoCAP is intended to be used with the instrument UniCAP together with reagents as stated in the Directions For Use provided with UniCAP Specific IgG. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of certain thyroid diseases, such as autoimmune thyroiditis, and Graves' Disease and is to be used in clinical laboratories, as well as, physician office laboratories.
Device Story
UniCAP system is an integrated, automated immunodiagnostic platform for fluoroenzymeimmunoassays. Input: human serum or plasma samples. Process: patient IgG antibodies bind to specific antigens (Thyroglobulin or Thyroid Peroxidase) covalently coupled to a cellulose matrix (ImmunoCAP solid phase). The UniCAP 100 analyzer automates sample/reagent handling and processing over 2.5 hours. Output: quantitative measurement of antigen-specific IgG antibodies, provided via automatically printed laboratory reports. Used in clinical and physician office laboratories by laboratory personnel. Results are used in conjunction with other clinical findings to aid in the diagnosis of thyroid diseases like autoimmune thyroiditis and Graves' Disease, potentially facilitating timely clinical management.
Clinical Evidence
Correlation study performed using 100 serum samples for each assay (Thyroglobulin and Thyroid Peroxidase). Comparison against Varelisa predicate devices showed linear regression equations of Y = 1.04X + 11 (r=0.99) for Thyroglobulin and Y = 1.04X (r=0.99) for Thyroid Peroxidase. No prospective clinical diagnostic outcome data provided.
Technological Characteristics
Fully integrated, automated fluoroenzymeimmunoassay system. Solid phase: cellulose matrix with covalently coupled recombinant human Thyroid Peroxidase or purified human Thyroglobulin. Instrument: UniCAP 100 Analyzer. Connectivity: supports external PC software for data management and mainframe/network server integration. Calibration: stored curves valid for one month, validated by curve controls.
Indications for Use
Indicated for in vitro quantitative measurement of IgG antibodies specific for Thyroglobulin or Thyroid Peroxidase in human serum and plasma to aid in clinical diagnosis of thyroid diseases, including autoimmune thyroiditis and Graves' Disease. For use in clinical and physician office laboratories.
Regulatory Classification
Identification
A thyroid autoantibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the thyroid autoantibodies (antibodies produced against the body's own tissues). Measurement of thyroid autoantibodies may aid in the diagnosis of certain thyroid disorders, such as Hashimoto's disease (chronic lymphocytic thyroiditis), nontoxic goiter (enlargement of thyroid gland), Grave's disease (enlargement of the thyroid gland with protrusion of the eyeballs), and cancer of the thyroid.
Predicate Devices
Varelisa® Thyroglobulin Antibodies Assay
Varelisa Thyroid Peroxidase Antibodies Assay
Submission Summary (Full Text)
{0}------------------------------------------------
#### JUL 7 1998
UniCAP® Thyroglobulin IgG Assay / UniCAP® Thyroid Peroxidase IgG Assay 510(k) Submission
# Section 11. Summary of Safety and Effectiveness
K 98 1930\$\checkmark\$
ات
مست
### SUMMARY OF SAFETY AND EFFECTIVENESS 11.
This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safety Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92.
| Date of Summary Preparation: | June 23, 1998 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Distributor: | Pharmacia & Upjohn<br>Diagnostics Division, US Operation<br>7425-248-1<br>7000 Portage Road<br>Kalamazoo, MI 49001 |
| Manufacturer: | Pharmacia & Upjohn, Diagnostics AB<br>S-751 82 Uppsala, Sweden |
| Company Contact Person: | Karen E. Matis<br>Regulatory Affairs Manager<br>Diagnostics Division<br>US Operation<br>7000 Portage Road<br>7425-248-01<br>Kalamazoo, MI 49001<br>(614) 794-3324 (Phone)<br>(614) 794-0266 (Fax) |
| Device Names: | UniCAP® Thyroglobulin IgG Assay<br>UniCAP® Thyroid Peroxidase IgG Assay<br>UniCAP® Thyroglobulin IgG Antibodies Control NLH<br>UniCAP® Thyroid Peroxidase IgG Antibodies Control<br>NLH |
| Common Name: | Thyroid autoantibody immunological test system. |
{1}------------------------------------------------
### Classification:
| Product Name | Product Code | Class | CFR |
|---------------------------------------------------------|--------------|-------|----------|
| UniCAP® Thyroglobulin IgG Assay | 82JNL | II | 866.5870 |
| UniCAP Thyroid Peroxidase IgG<br>Assay | 82JZO | II | 866.5870 |
| UniCAP Thyroglobulin IgG<br>Antibodies Control NLH | 82JNL | II | 866.5870 |
| UniCAP Thyroid Peroxidase IgG<br>Antibodies Control NLH | 82JZO | II | 866.5870 |
### Substantial Equivalence to:
Varelisa® Thyroglobulin Antibodies Assay Varelisa Thyroid Peroxidase Antibodies Assay
## Intended Use Statement ;
UniCAP Thyroglobulin ImmunoCAP™ is a device for the in vitro quantitative measurement of IgG antibodies specific for Thyroglobulin (TG) in human serum and plasma. UniCAP Thyroglobulin ImmunoCAP is intended to be used with the instrument UniCAP together with reagents as stated in the Directions For Use provided with UniCAP Specific IgG. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of certain thyroid diseases, such as autoimmune thyroiditis and Graves' Disease and is to be used in clinical laboratories, as well as, physician office laboratories.
UniCAP Thyroid Peroxidase ImmunoCAP is a device for the in vitro quantitative measurement of IgG antibodies specific for Thvroid Peroxidase (TPO) in human serum and plasma. UniCAP Thyroid Peroxidase ImmunoCAP is intended to be used with the instrument UniCAP together with reagents as stated in the Directions For Use provided with UniCAP Specific IgG. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of certain thyroid diseases, such as autoimmune thyroiditis, and Graves' Disease and is to be used in clinical laboratories, as well as, physician office laboratories.
{2}------------------------------------------------
UniCAP® Specific IgG is an in vitro test system for the quantitative and qualitative measurement of antigen specific IgG antibodies. The corresponding antigen for the specific antibody to be measured by UniCAP Specific IgG is bound to the Antigen ImmunoCAP™ solid phase component of the UniCAP Specific IgG System. UniCAP Specific IgG assay is to be used with the instrument UniCAP. It is intended for in vitro diagnostic use in conjunction with other clinical findings, and is to be used in clinical laboratories, as well as physician office laboratories.
UniCAP Thyroglobulin IgG Antibodies Controls are intended for laboratory use in monitoring the performance of in vitro quantitative measurement of IgG antibodies specific for Thyroglobulin in human serum. UniCAP Thyroglobulin IgG Antibodies Controls are intended to be used with the instrument UniCAP.
UniCAP Thyroid Peroxidase IgG Antibodies Controls are intended for laboratory use in monitoring the performance of in vitro quantitative measurement of IgG antibodies specific for Thyroid Peroxicase in human serum. UniCAP Thyroid Peroxidase IgG Antibodies Controls are intended to be used with the instrument UniCAP.
## General Description
UniCAP Immunodiagnostic System is a fully integrated and automated system for immunodiagnostic testing. UniCAP System is comprised of Instruments (UniCAP 100 Analyzer, Test System Modules and Assay Products (Fluororenzymeimmunoassays for the measurement of IgE, IgG). ImmunoCAP™ Antigens (solid phase components which contain the specific antigens to be measured), and Software Accessories.
UniCAP 100 Analyzer is designed to handle all steps from sample and reagents handling to processing of results. Reagents, requests, samples and ImmunoCAP are loaded into the instrument and the process, which takes 2.5 hours is started. A laboratory report is automatically printed when the process is ended.
UniCAP 100 can store a calibration curve to be used for up to one month. After an initial calibration curve is accepted by the software, subsequent assay runs may use the stored calibration curve for calculation of results. In these runs, Curve Controls are included to validate that the run is on the same response level as the stored curve. Limits for the response of the Curve Controls are defined in the UniCAP 100 Operator and Panel Software.
{3}------------------------------------------------
UniCAP® 100 RM External Software is intended to be used with a Windows-based PC operating up to five UniCAP 100 instruments. The external software creates requests and assay runs, retrieves the test results from the instrument, and prints reports. It can also import requests from, and export requests to, a connected mainframe computer or network server.
UniCAP Specific IgG is a fluoroenzymeimmunoassay for the quantitative measurement of antigen specific IgG antibodies. The corresponding antigen for the specific antibody to be measured by UniCAP Specific IgG is bound to the Antigen ImmunoCAP™ solid phase component. Specific IgG antibodies in the patient serum or plasma specimen react with the antigens of interest, in this submission, Thyroglobulin and Thyroid Peroxidase, which are covalently coupled to ImmunoCAP.
The UniCAP Thyroid Peroxidase ImmunoCAP contains recombinant human Thyroid Peroxidase covalently coupled to a cellulose matrix. The UniCAP Thyroglobulin ImmunoCAP contains purified human Thyroglobulin covalently coupled to the cellulose matrix.
## Comparison Data:
## UniCAP Thyroglobulin IgG Assay
A comparison study was performend to generate correlation data between UniCAP Thyroglobulin IgG Assay and Varelisa® TG Antibody Assay. This study was performed to demonstrate that the performance of UniCAP Thyroglobulin IgG Assay is substantially equivalent to Varelisa TG Antibody Assay, which is the legally marketed predicate device in the United States. 100 serum samples were collected and were run on both assay systems with the following results.
Linear regression analysis gave the equation:
Y = 1.04 * X + 11 with correlation coefficient 0.99
This correlation study demonstrated that the new device, UniCAP Thyroglobulin IgG Assay is substantially equivalent to the legally marketed predicate device. Varelisa Thyroglobulin Antibodies Assay.
{4}------------------------------------------------
## UniCAP® Thyroid Peroxidase IgG Assay
A comparison study was performed to generate correlation data between UniCAP Thyroid Peroxidase IgG Assay and Varelisa® TPO Antibody Assay. This study was performed to demonstrate that the performance of UniCAP Thyroid Peroxidase IgG Assay is substantially equivalent to Varelisa TPO Antibody Assay, which is the legally marketed predicate device in the United States. 100 serum samples were collected and were run on both assay systems with the following results.
Linear regression analysis gave the following equation:
Y = 1.04 * X with correlation coefficient 0.99
This correlation study demonstrated that the new device. UniCAP Thyroid Peroxidase IgG Assay is substantially equivalent to the legally marketed predicate device, Varelisa Thyroid Peroxidase Antibodies Assay.
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three overlapping human profiles, resembling an abstract representation of people.
#### 7 1998 JUL
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Karen E. Matis Regulatory Affairs Manager Pharmacia & Upjohn Diagnostics Division, US Operations 5094 St. Andrews Drive Westerville, Ohio 43082
Re : K981930 UniCAP® Thyroid Peroxidase IgG Assay Trade Name: Requlatory Class: II Product Code: JZO Dated: April 30, 1998 Received: May 1, 1998
Dear Ms. Matis:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the You provisions of the Federal Food, Drug, and Cosmetic Act (Act). may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K981930
510(k) Number (if known):
Device Name: UniCAP® Thyroid Peroxidase IgG Assay
Indications For Use: UniCAP Thyroid Peroxidase ImmunoCAP™ is a device for the in vitro quantitative measurement of IgG antibodies specific for Thyroid Peroxidase (TPQ) in human serum and plasma. UniCAP Thyroid Peroxidase ImmunoCAP is intended to be used with the instrument UniCAP together with reagents as stated in the Directions For Use provided with UniCAP Specific IgG. It is intended for in vitro diagnostic use as an aïd in the clinical diagnosis of certain thyroid diseases, such as autoimmune thyroiditis and Graves' Disease and is to be used in clinical laboratories, as well as, physician office laboratories.
UniCAP Specific IgG is an in vitro test system for the quantitative measurement of antigen specific IgG antibodies. The corresponding antigen for the specific antibody to be measured by UniCAP Specific IgG is bound to the Antigen ImmunoCAP solid phase component of the UniCAP Specific IgG System. UniCAP Specific IgG assay is to be used with the instrument UniCAP. It is intended for in vitro diagnostic use in coniunction with other clinical findings, and is to be used in clinical laboratories, as well as physician office laboratories.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | | |
|--------------------------------------------------------|---------|--|
| <img alt="signature" src="signature.png"/> | | |
| (Division Sign-Off) | | |
| Division of Clinical Laboratory Devices | | |
| 510(k) Number | K981930 | |
| Prescription Use (Per 21 CFR 801.109) | <img alt="check" src="check.png"/> | OR | Over-The-Counter Use | |
|---------------------------------------|------------------------------------|----|----------------------|--|
|---------------------------------------|------------------------------------|----|----------------------|--|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.