UniCAP Thyroglobulin ImmunoCAP™ is a device for the in vitro quantitative measurement of IgG antibodies specific for Thyroglobulin (TG) in human serum and plasma. UniCAP Thyroglobulin ImmunoCAP is intended to be used with the instrument UniCAP together with reagents as stated in the Directions For Use provided with UniCAP Specific IgG. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of certain thyroid diseases, such as autoimmune thyroiditis and Graves' Disease and is to be used in clinical laboratories, as well as, physician office laboratories. UniCAP Thyroid Peroxidase ImmunoCAP is a device for the in vitro quantitative measurement of IgG antibodies specific for Thyroid Peroxidase (TPO) in human serum and plasma. UniCAP Thyroid Peroxidase ImmunoCAP is intended to be used with the instrument UniCAP together with reagents as stated in the Directions For Use provided with UniCAP Specific IgG. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of certain thyroid diseases, such as autoimmune thyroiditis, and Graves' Disease and is to be used in clinical laboratories, as well as, physician office laboratories.
Device Story
UniCAP System is a fully integrated, automated immunodiagnostic platform for fluoroenzymeimmunoassays. Input: human serum or plasma samples. Process: samples and reagents loaded into UniCAP 100 Analyzer; specific IgG antibodies bind to antigens (purified human Thyroglobulin or recombinant human Thyroid Peroxidase) covalently coupled to cellulose matrix (ImmunoCAP solid phase). System performs all handling and processing steps; 2.5-hour run time. Output: quantitative measurement of specific IgG antibodies; results printed automatically. Used in clinical and physician office laboratories by laboratory personnel. Calibration curves stored for up to one month; curve controls validate response levels. Results used by clinicians in conjunction with other clinical findings to aid in diagnosing thyroid conditions. Benefits: standardized, automated testing for thyroid autoantibodies.
Clinical Evidence
Bench testing only. Correlation studies performed on 100 serum samples comparing UniCAP Thyroglobulin IgG Assay and UniCAP Thyroid Peroxidase IgG Assay against respective Varelisa predicate assays. Linear regression analysis for TG assay: Y = 1.04 * X + 11 (r=0.99). Linear regression analysis for TPO assay: Y = 1.04 * X (r=0.99).
Technological Characteristics
Fluoroenzymeimmunoassay system. Solid phase: cellulose matrix with covalently coupled antigens (purified human Thyroglobulin or recombinant human Thyroid Peroxidase). Automated UniCAP 100 Analyzer handles sample/reagent processing. Connectivity: supports external PC software for request management and reporting; mainframe/network server integration. Calibration: stored curves validated by curve controls.
Indications for Use
Indicated for in vitro quantitative measurement of IgG antibodies specific for Thyroglobulin (TG) or Thyroid Peroxidase (TPO) in human serum and plasma to aid in the clinical diagnosis of thyroid diseases, including autoimmune thyroiditis and Graves' Disease. For use in clinical and physician office laboratories.
Regulatory Classification
Identification
A thyroid autoantibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the thyroid autoantibodies (antibodies produced against the body's own tissues). Measurement of thyroid autoantibodies may aid in the diagnosis of certain thyroid disorders, such as Hashimoto's disease (chronic lymphocytic thyroiditis), nontoxic goiter (enlargement of thyroid gland), Grave's disease (enlargement of the thyroid gland with protrusion of the eyeballs), and cancer of the thyroid.
K151799 — EliA anti-TG Immunoassay, EliA anti-TPO Immunoassay, EliA Thyroid Positive Control 250, EliA Thyroid Positive Control 2500/5000 · Phadia AB · Mar 25, 2016
K072661 — IMMUNOCAP THYRLOBULIN IMMUNOCAP · Phadia US, Inc. · Nov 20, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
# 7 1998 JULL
UniCAP® Thyroglobulin IgG Assay / UniCAP® Thyroid Peroxidase IgG Assay K981559 V 510(k) Submission Section 11. Summary of Safety and Effectiveness
#### SUMMARY OF SAFETY AND EFFECTIVENESS 11.
This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safety Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92.
| Date of Summary Preparation: | June 23, 1998 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Distributor: | Pharmacia & Upjohn<br>Diagnostics Division, US Operation<br>7425-248-1<br>7000 Portage Road<br>Kalamazoo, MI 49001 |
| Manufacturer: | Pharmacia & Upjohn, Diagnostics AB<br>S-751 82 Uppsala, Sweden |
| Company Contact Person: | Karen E. Matis<br>Regulatory Affairs Manager<br>Diagnostics Division<br>US Operation<br>7000 Portage Road<br>7425-248-01<br>Kalamazoo, MI 49001<br>(614) 794-3324 (Phone)<br>(614) 794-0266 (Fax) |
| Device Names: | UniCAP® Thyroglubulin IgG Assay<br>UniCAP® Thyroid Peroxidase IgG Assay<br>UniCAP® Thyroglobulin IgG Antibodies Control NLH<br>UniCAP® Thyroid Peroxidase IgG Antibodies Control<br>NLH |
| Common Name: | Thyroid autoantibody immunological test system. |
510k/utgtp11
{1}------------------------------------------------
# UniCAP® Thyroglobulin IgG Assay / UniCAP® Thyroid Peroxidase IgG Assay 510(k) Submission Section 11. Summary of Safety and Effectiveness
#### Classification:
| Product Name | Product Code | Class | CFR |
|---------------------------------------------------------|--------------|-------|----------|
| UniCAP® Thyroglobulin IgG Assay | 82JNL | II | 866.5870 |
| UniCAP Thyroid Peroxidase IgG<br>Assay | 82JZO | II | 866.5870 |
| UniCAP Thyroglobulin IgG<br>Antibodies Control NLH | 82JNL | II | 866.5870 |
| UniCAP Thyroid Peroxidase IgG<br>Antibodies Control NLH | 82JZO | II | 866.5870 |
#### Substantial Equivalence to:
Varelisa® Thyroglobulin Antibodies Assay Varelisa Thyroid Peroxidase Antibodies Assay
### Intended Use Statement :
UniCAP Thyroglobulin ImmunoCAP™ is a device for the in vitro quantitative measurement of IgG antibodies specific for Thyroglobulin (TG) in human serum and plasma. UniCAP Thyroglobulin ImmunoCAP is intended to be used with the instrument UniCAP together with reagents as stated in the Directions For Use provided with UniCAP Specific IgG. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of certain thyroid diseases, such as autoimmune thyroiditis and Graves' Disease and is to be used in clinical laboratories, as well as, physician office laboratories.
UniCAP Thyroid Peroxidase ImmunoCAP is a device for the in vitro quantitative measurement of IgG antibodies specific for Thyroid Peroxidase (TPO) in human serum and plasma. UniCAP Thyroid Peroxidase ImmunoCAP is intended to be used with the instrument UniCAP together with reagents as stated in the Directions For Use provided with UniCAP Specific IgG. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of certain thyroid diseases, such as autoimmune thyroiditis, and Graves' Disease and is to be used in clinical laboratories, as well as, physician office laboratories.
{2}------------------------------------------------
UniCAP® Specific IgG is an in vitro test system for the quantitative and qualitative measurement of antigen specific IgG antibodies. The corresponding antigen for the specific antibody to be measured by UniCAP Specific IgG is bound to the Antigen ImmunoCAP™ solid phase component of the UniCAP Specific IgG System. UniCAP Specific IgG assay is to be used with the instrument UniCAP. It is intended for in vitro diagnostic use in conjunction with other clinical findings, and is to be used in clinical laboratories, as well as physician office laboratories.
UniCAP Thyroglobulin IgG Antibodies Controls are intended for laboratory use in monitoring the performance of in vitro quantitative measurement of IgG antibodies specific for Thyroglobulin in human serum. UniCAP Thyroglobulin IgG Antibodies Controls are intended to be used with the instrument UniCAP.
UniCAP Thyroid Peroxidase IgG Antibodies Controls are intended for laboratory use in monitoring the performance of in vitro quantitative measurement of IgG antibodies specific for Thyroid Peroxidase in human serum. UniCAP Thyroid Peroxidase IgG Antibodies Controls are intended to be used with the instrument UniCAP.
### General Description
UniCAP Immunodiagnostic System is a fully integrated and automated system for immunodiagnostic testing. UniCAP System is comprised of Instruments (UniCAP 100 Analyzer. Test System Modules and Assay Products (Fluororenzymeimmunoassays for the measurement of IgE, IgG), ImmunoCAP™ Antigens (solid phase components which contain the specific antigens to be measured), and Software Accessories.
UniCAP 100 Analyzer is designed to handle all steps from sample and reagents handling to processing of results. Reagents, requests, samples and ImmunoCAP are loaded into the instrument and the process, which takes 2.5 hours is started. A laboratory report is automatically printed when the process is ended.
UniCAP 100 can store a calibration curve to be used for up to one month. After an initial calibration curve is accepted by the software, subsequent assay runs may use the stored calibration curve for calculation of results. In these runs. Curve Controls are included to validate that the run is on the same response level as the stored curve. Limits for the response of the Curve Controls are defined in the UniCAP 100 Operator and Panel Software.
{3}------------------------------------------------
UniCAP® 100 RM External Software is intended to be used with a Windows-based PC operating up to five UniCAP 100 instruments. The external software creates requests and assay runs, retrieves the test results from the instrument, and prints reports. It can also import requests from, and export requests to, a connected mainframe computer or network server.
UniCAP Specific IgG is a fluoroenzymeimmunoassay for the quantitative measurement of antigen specific IgG antibodies. The corresponding antigen for the specific antibody to be measured by UniCAP Specific IgG is bound to the Antigen ImmunoCAP™ solid phase component. Specific IgG antibodies in the patient serum or plasma specimen react with the antigens of interest, in this submission, Thyroglobulin and Thyroid Peroxidase, which are covalently coupled to ImmunoCAP.
The UniCAP Thyroid Peroxidase ImmunoCAP contains recombinant human Thyroid Peroxidase covalently coupled to a cellulose matrix. The UniCAP Thyroglobulin ImmunoCAP contains purified human Thyroglobulin covalently coupled to the cellulose matrix.
## Comparison Data:
### UniCAP Thyroglobulin IgG Assay
A comparison study was performend to generate correlation data between UniCAP Thyroglobulin IgG Assay and Varelisa® TG Antibody Assay. This study was performed to demonstrate that the performance of UniCAP Thyroglobulin IgG Assay is substantially equivalent to Varelisa TG Antibody Assay, which is the legally marketed predicate device in the United States. 100 serum samples were collected and were run on both assay systems with the following results.
Linear regression analysis gave the equation:
Y = 1.04 * X + 11 with correlation coefficient 0.99
This correlation study demonstrated that the new device. UniCAP Thyroglobulin IgG Assay is substantially equivalent to the legally marketed predicate device. Varelisa Thyroglobulin Antibodies Assay.
{4}------------------------------------------------
## UniCAP® Thyroid Peroxidase IgG Assay
A comparison study was performed to generate correlation data between UniCAP Thyroid Peroxidase IgG Assay and Varelisa® TPO Antibody Assay. This study was performed to demonstrate that the performance of UniCAP Thyroid Peroxidase IgG Assay is substantially equivalent to Varelisa TPO Antibody Assay, which is the legally marketed predicate device in the United States. 100 serum samples were collected and were run on both assay systems with the following results.
Linear regression analysis gave the following equation:
Y = 1.04 * X with correlation coefficient 0.99
This correlation study demonstrated that the new device, UniCAP Thyroid Peroxidase IgG Assay is substantially equivalent to the legally marketed predicate device, Varelisa Thyroid Peroxidase Antibodies Assay.
{5}------------------------------------------------
→
Image /page/5/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract design of an eagle with three heads, symbolizing health, hope, and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle emblem.
#### 7 1998 JUL
Ms. Karen E. Matis Regulatory Affairs Manager Pharmacia & Upjohn Diagnostics Division, US Operations 5094 St. Andrews Drive Westerville, Ohio 43082
Re: K981559 Trade Name: UniCAP® Thyroglobulin IgG Assay Regulatory Class: II Product Code: JZO Dated: April 30, 1998 May 1, 1998 Received:
Dear Ms. Matis:
We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labelinq, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in witro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
UniCAP® Thyroglobulin IgG Assay/ UniCAP® Thyroid Peroxidase IgG Assay 510(k) Submission Section 1. Intended Use Statements
K981559
510(k) Number (if known):
Device Name: UniCAP® Thyroglobulin IgG Assay
Indications For Use: UniCAP Thyrodlobulin ImmunoCAP™ is a device for the in vitro quantitative measurement of IgG antibodies specific for Thyroglobulin (TG) in human serum and plasma. UniCAP Thyroglobulin ImmunoCAP is intended to be used with the instrument UniCAP together with reagents as stated in the Directions For Use provided with UniCAP Specific IgG. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of certain thyroid diseases, such as autoimmune thyroiditis and Graves' Disease and is to be used in clinical laboratories, as well as, physician office laboratories.
UniCAP Specific IgG is an in vitro test system for the quantitative measurement of antigen specific IgG antibodies. The corresponding antigen for the specific antibody to be measured by UniCAP Specific IgG is bound to the Antigen ImmunoCAP solid phase component of the UniCAP Specific IgG System. UniCAP Specific IgG assay is to be used with the instrument UniCAP. It is intended for in vitro diagnostic use in conjunction with other clinical findings, and is to be used in clinical laboratories, as well/as physician office laboratories.
Titic. Maker
(Division Sign-Off) Division of Clinical Laboratory De 610(k) Number
(Please Do Not WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use
Over-The-Counter Use
(Per 21 CFR 801.109)
{8}------------------------------------------------
## UniCAP® Thyroglobulin IgG Assay/ UniCAP® Thyroid Peroxidase IgG Assay 510(k) Submission Section 1. Intended Use Statements
510(k) Number (if known):
Device Name: UniCAP® Thyroglobulin IgG Antibodies Control NLH
K981559
Indications For Use: UniCAP Thyroglobulin IgG Antibodies Controls are intended for laboratory use in monitoring the performance of in vitro quantitative measurement of IgG antibodies specific for Thyroglobulin in human serum. UniCAP Thyroglobulin IgG Antibodies Controls are intended to be used with the instrument UniCAP.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V
OR
Over-The-Counter Use
(Per 21 CFR 801.109)
00003 00
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.