FRESENIUS F SERIES HEMOCONCENTRATORS F400,F500,F700,F800,F400TS,F500TS,F700S,F800TS
K974584 · Fresenius USA, Inc. · KDI · May 14, 1998 · Gastroenterology, Urology
Device Facts
Record ID
K974584
Device Name
FRESENIUS F SERIES HEMOCONCENTRATORS F400,F500,F700,F800,F400TS,F500TS,F700S,F800TS
Applicant
Fresenius USA, Inc.
Product Code
KDI · Gastroenterology, Urology
Decision Date
May 14, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fresenius F Series Hemoconcentrators are indicated to relieve or mitigate overhydration in patients undergoing cardiopulmonary procedures and to increase the concentration of cells and proteins in the blood.
Device Story
Hollow fiber-type hemoconcentrator; polysulfone capillary fibers bundled/potted with polyurethane in polyurethane jacket. Input: diluted blood from cardiopulmonary bypass circuit. Operation: passive filtration of plasma water across semi-permeable polysulfone membrane; plasma water exits via filtrate port; concentrated blood exits via venous port. Used in operating room by perfusionists/clinicians. Output: concentrated blood returned to patient; excess fluid removed. Benefits: manages fluid balance; increases hematocrit/protein concentration. Tubing sets (PVC) optional. Four models (F400, F500, F700, F800) vary by fiber count/filtration capacity.
Clinical Evidence
Bench testing only. Comparative in vitro studies evaluated F400, F500, F700, and F800 against predicates (F40, F60). Endpoints: ultrafiltration rates, concentration rates, sieving coefficients (large/small proteins), hemolysis (plasma hemoglobin), and pressure drop. Results showed equivalent hemolysis and molecular weight cutoff (~65,000 Daltons). Performance characterized per EN 1283 standard.
Indicated for patients undergoing cardiopulmonary procedures requiring relief or mitigation of overhydration and concentration of blood cells and proteins.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
MAY 1 4 1998
Image /page/0/Picture/1 description: The image shows a combination of letters and numbers. The top row contains the characters "K974584", while the bottom row displays "P.1/3". The characters are handwritten and have a slightly rough, uneven appearance.
# 510(k) Summary for the Fresenius F Series Hemoconcentrators
Submitter's Name and Address:
Phone Number: Telefax Number: Contact Person:
Date Summary Prepared:
Device Trade Name:
Common name:
Classification Name:
Legally Marketed Device to which substantial equivalence is claimed: Fresenius USA, Critical Care Division 2637 Shadelands Drive Walnut Creek, CA 94598 (510) 295-0200 (510) 988-1932 Virginia Singer
December 5, 1997
Fresenius F Series Hemoconcentrators
High Flux Hemoconcentrator
High Permeability Hemodialysis System
Bard HC40 Hemoconcentrator (K971180) Amicon Diafilter 30 (K902837) Minntech Hemocor HPH 1000 (K923139) Baxter Bentley Quick Prime HQ 7000 (K903641) Research Medical Biofilter 140 (K951344 and K970739)
### Intended Use:
The Fresenius F Series Hemoconcentrators are indicated to relieve or mitigate overhydration in patients undergoing cardiopulmonary procedures and to increase the concentration of cells and proteins in the blood.
### Device Features:
A complete description of the Fresenius Hemoflow Series Dialyzer, the same product as the F Series Hemoconcentrators, has been submitted in previous premarket notifications. There have been no modifications to this device to adapt it for hemoconcentration. The Fresenius Hemoflow Series Dialyzer is a hollow fiber-type filter. Fresenius-produced polysulfone capillary fibers are bundled and potted with polyurethane into an artificial kidney jacket manufactured from polyurethane. Screw-type end caps, manufactured from polyurethane, have twist lock connectors for the connection of venous and arterial blood lines. Two filtrate ports are located on the filter adjacent to the filtrate chambers. The ports have Hansen-type fittings for connection of filtrate tubing. For hemoconcentration, only the filtrate port on the venous end of the filter is used; the other port on the arterial end is capped.
There are four (4) models within the F Series Hemoconcentrators family. The difference between the four models in the F Series Hemoconcentrators is the number of fibers contained within the artificial kidney jacket. As the number of fibers contained in the filter increases, the diameter of the filter and the filtration capacity increases proportionally. Each model will be manufactured with a tubing set (F400TS, F500TS, F700TS, F800TS); other models will be manufactured with tubing adapter, only (F400, F500, F700, F800).
{1}------------------------------------------------
8t
p. 2/3
# Technological Characteristics of the Subject Device Compared with Predicate Devices
The 510(k) "Substantial Equivalence Decision Making Process (Detailed)" decision tree (ODE Guidance Memo #K86-3) was used to make a determination of substantial equivalence (reference Exhibit IV-1 included in this section). The answers to questions identified on this decision tree lead to a determination of substantial equivalence.
#### Does the New Device Have the Same Indication Statements? 1.0
Yes. The F Series Hemoconcentrators (F400, F500, F700, and F800), Bard HC40 Hemoconcentrator, Amicon Diafilter 30, Minntech Hemocor HPH 1000, Baxter Bentley Quick Prime HQ 7000, and Research Medical Biofilter 140 are all indicated for the removal of excess fluid from blood in cardiopulmonary procedures.
#### Does New Device Have Same Technological Characteristics (e.g., design, materials etc.)? 2.0
Yes. There are virtually no differences in technology and operation between the F Series Hemoconcentrator devices and the predicate devices. The core of the technology is the filter material within the hemoconcentrators housing. With the exception of the filter manufactured for Research Medical, Inc. (Biofilter 140), the filter material is polysulfone. The diameter of the polysulfone fibers is 200 micrometers, with the exception of the Amicon Diafilter that is 250 micrometers. The filtration capacity of the hemoconcentrator is dependent upon the number of fibers contained in the filter; the more fibers contained, the faster plasma water can be removed. In all marketed filters, plasma water is filtered from the diluted blood and exits the filter from a filtration port located on the side of the filter. Diluted blood enters the filter via the arterial blood port and concentrated blood exits the filter at the venous blood port.
The tubing sets that accompany hemoconcentration filters are virtually identical. They consist of ¼" ID 3/8" OD PVC tubing with suitable proximal end connectors and line clamps. The tubing set may or may not include a filtrate line, clamps, and filtrate collection bag. These filtrate line items (tubing, adapters, clamps, collections containers) are typically available in an operating room and, therefore, are optional with respect to a tubing kit.
#### 3.0 Are the Descriptive Characteristics Precise Enough to Ensure Equivalence?
No. Although, the structure of the F Series Hemoconcentrator and tubing sets are very similar to the predicate devices manufactured by Fresenius (F40 and F60) and their respective tubing sets, there may be subtle differences in the materials used to manufacture assemblies of other predicate devices (Amicon Diafilter 30, Minntech Hemocor HPH 1000, and Research Medical Biofilter 140) and in the manufacturing methods employed to build these devices. These differences could impact the filtration characteristics and biocompatibility of the product.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "K974584 p. 3/3" in a handwritten style. The characters are bold and slightly uneven, suggesting they were written quickly or with a thick marker. The text is arranged in two lines, with the "K974584" on the top line and "p. 3/3" on the bottom line.
## Are Performance Data Available to Assess Equivalence?
Yes. All blood/fluid contacting materials of the Fresenius F Series Hemoconcentrators and tubing sets have been subjected to biocompatibility testing consistent with FDA's modified ISO standards for biological evaluation of medical devices.
Comparative in vitro testing of the Fresenius F40, F50, F70, and F80 has been performed to assess end-to-end pressure drop, ultrafiltration rates, concentration rates of cellular blood components, sieving coefficients of a large and a small molecular weight plasma protein, and hemolysis. This testing served to characterize and contrast the performance of the predicate products (F40 and F60) relative to those products for which clearance is being requested (F400, F500, F700 and F800). The testing also served to demonstrate that, at maximum recommended transmembrane pressure and flow rate, the production of plasma hemoglobin was statistically the same for all devices evaluated.
Performance specifications of other hemoconcentrators not manufactured by Fresenius (eg. Amicon Diafilter 30, Minntech Hemocor HP 1000, and Research Medical Biofilter 140) were also compared to Fresenius in vitro data to demonstrate that under similar operating conditions (TMP, blood flow rate, hematocrit, protein content, temperature) filtration performance was relatively the same.
#### Does Performance Data Demonstrate Equivalence? 5.0
Based on the results of the testing cited above, Fresenius has demonstrated that: Yes.
- The performance of the F Series Hemoconcentrators is consistent with the draft specifications . for these products. Performance has been characterized according to a recognized international standard, EN 1283, Haemodialyzers, Haemodiafilters, Haemoconcentrators and Their Extracorporeal Circuits.
- . The ultrafiltration and concentration rates of the F Series Hemoconcentrators (F400, F500, F700, and F800) are relative consistent with the rates of predicate devices (F40 and F60) as demonstrated in in vitro studies. Differences in rates may be attributed to differences in Active Surface Area of the filter itself.
- As with the predicate devices (F40 and F60), no cellular blood components were lost in the . ultrafiltrate in the F Series Hemoconcentrators.
- The production of plasma hemoglobin (hemolysis) was equivalent for all devices studied. .
- Performance data also demonstrates that the molecular weight cutoff for all devices studied is . approximately 65,000 Daltons. This fact is supported by data that shows that minimal albumin was lost from the blood product and that a relatively small protein, beta-2microglobulin, was filtered with the plasma water.
CONCLUSION: Based on the information and test results provided in this premarket notification, the Fresenius F Series Hemoconcentrators and tubing sets are substantially equivalent to currently marketed hemoconcentrators.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's three main goals: protecting the health of all Americans, providing essential human services, and fostering advances in medicine, public health, and social services. The text "DEPARTMENT OF HEALTH & HUMAN" is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 4 1998
Ms. Virginia Singer Regulatory Affairs Manager Fresenius USA, Inc. Critical Care Division 2637 Shadelands Drive Walnut Creek, CA 94598
Re: K974584
Fresenius F-series Hemoconcentrator Models F400, F400TS, F500, F500TS, F700 F700TS, F800, and F800TS Dated: April 3, 1998 Received: April 7, 1998 Regulatory Class: III 21 CFR 876.5860/Procode: 78 KDI
Dear Ms. Singer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely vours
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known);
Device Name:
Indications for Use:
K974584
Fresenius F Series Hemoconcentrators
The Fresenius F Series Hemoconcentrators are indicated to relieve or mitigate overhydration in patients undergoing cardiopulmonary procedures and to increase the concentration of cells and proteins in the blood.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert P. Sattinger
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 14584 510(k) Number .
**Prescription Use**
(Per 21 CFR 801.109)
OR
Over The Counter Use _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.