SYNTHES CANNULATED ANGLE BLADE PLATE (CABP) SYSTEM
Applicant
Synthes (Usa)
Product Code
KTT · Orthopedic
Decision Date
Feb 18, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Synthes CABP System is a plate and screw system intended to treat fractures of the proximal humerus and distal tibia. Proximal humerus fractures include two-part greater tubercle fractures and fracture dislocations, two-part surgical neck fractures and fracture dislocations, three-part fractures or fracture dislocations, fractures in osteopenic bone, and nonunions and malunions. Distal tibia fractures include acute fractures, fractures in osteopenic bone, and nonunions and malunions.
Device Story
Cannulated Angle Blade Plate (CABP) system for orthopedic fracture fixation; used in proximal humerus and distal tibia. Device consists of straight plate with integrated blade; cannulated blade allows guide wire placement and adjustment. Undersurface cuts reduce bone contact area to preserve cortical blood supply and minimize damage-induced porosis; undercuts also improve bending uniformity for contouring. System includes round and dynamic compression screw holes; accepts 4.5 mm and 6.5 mm screws. Used by surgeons in clinical/OR settings to stabilize bone fragments; facilitates healing in acute fractures, nonunions, and malunions.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, material, and intended use comparison to predicate device.
Technological Characteristics
Materials: stainless steel and commercially pure titanium. Design: straight plate with cannulated blade; undersurface undercuts for bone contact reduction and uniform bending. Fixation: accepts 4.5 mm and 6.5 mm screws; round and dynamic compression holes. Form factor: multiple blade lengths.
Indications for Use
Indicated for patients with proximal humerus fractures (two-part greater tubercle, two-part surgical neck, three-part fractures/dislocations, osteopenic bone, nonunions, malunions) and distal tibia fractures (acute, osteopenic bone, nonunions, malunions).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Synthes Cannulated Angle Blade Plate System (K954289)
Submission Summary (Full Text)
{0}
K974537
Feb. 18, 1998
# SYNTHES
## Attachment VIII:
Summary of Safety and Effectiveness Info. [510(k) Summary]
### SUBMITTER
Synthes (USA)
1690 Russell Road
Paoli, PA 19301
(610) 647-9700
Contact: Sheri L. Musgnung
### COMMON OR USUAL NAME:
Appliance, Fixation, Nail/Blade/Plate Combination, Single Component
### DEVICE CLASSIFICATION:
Class II, 21 CFR 888.3030; 888.3040
### PREDICATE DEVICE:
Synthes Cannulated Angle Blade Plate System (K954289)
### DESCRIPTION:
The CABP is a straight plate with a blade at the head to allow for better fixation in the head of the humerus or in the distal tibia. The blade of the plate is cannulated to fit over a guide wire, allowing for the adjustment of the wire placement several times without adversely affecting the final result. There are cuts in the undersurface of the plate to reduce the surface area of the plate in contact with bone. The area of contact between the plate and the bone is decreased in an effort to reduce damage to the cortical blood supply under the plate, and resultant reduction in damage-induced porosis and remodeling of the bone near the plate undersurface. Plate undercuts also make the bending properties of the plate more uniform, which facilitate contouring. The plate has round and dynamic compression screw holes, accepts 4.5 mm and 6.5 mm screws, is available in multiple blade lengths, and is manufactured from stainless steel and commercially pure titanium.
### INTENDED USE:
Synthes CABP System is a plate and screw system intended to treat fractures of the proximal humerus and distal tibia. Proximal humerus fractures include two-part greater tubercle fractures and fracture dislocations, two-part surgical neck fractures and fracture dislocations, three-part fractures or fracture dislocations, fractures in osteopenic bone, and nonunions and malunions. Distal tibia fractures include acute fractures, fractures in osteopenic bone, and nonunions and malunions.
1C
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
FEB 18 1998
Ms. Sheri L. Musgnung
Regulatory Affairs Associate
Synthes (USA)
1690 Russell Road
P.O. Box 1766
Paoli, Pennsylvania 19301
Re: K974537
Synthes (USA) Cannulated Angle Blade Plate (CABP) System
Regulatory Class: II
Product Code: KTT
Dated: December 2, 1997
Received: December 3, 1997
Dear Ms. Musgnung:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800-to-895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}
Page 2 - Ms. Sheri L. Musgnung
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and Restorative Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}
SYNTHES
SYNTHES (USA)
1690 Russell Road
Post Office Box 1766
Paoli, Pennsylvania 19301
Telephone 610-647-9700
Page 1 of 1
510(k) Number (if known): k974537
Device Name: Synthes (USA) Cannulated Angle Blade Plate System
Indications For Use:
The Synthes CABP System is a plate and screw system intended to treat fractures of the proximal humerus and distal tibia. Proximal humerus fractures include two-part greater tubercle fractures and fracture dislocations, two-part surgical neck fractures and fracture dislocations, three-part fractures or fracture dislocations, fractures in osteopenic bone, and nonunions and malunions. Distal tibia fractures include acute fractures, fractures in osteopenic bone, and nonunions and malunions.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐

Original Instruments and Implants of the Association for the Study of Internal Fixation—AO/ASIF
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.