GLUMA ONE BOND

K974390 · Heraeus Kulzer, Inc. · KLE · Feb 4, 1998 · Dental

Device Facts

Record IDK974390
Device NameGLUMA ONE BOND
ApplicantHeraeus Kulzer, Inc.
Product CodeKLE · Dental
Decision DateFeb 4, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3200
Device ClassClass 2

Indications for Use

GLUMA ONE BOND is a light curing single component enamel/dentin adhesive for bonding of all current brands of resin composite restorative materials and compomers. It also enhances adhesion to metals such as amalgam. GLUMA ONE BOND combines primer and adhesive in single bottle.

Device Story

GLUMA ONE BOND is a single-bottle, light-curing dental adhesive; used by dentists in clinical settings to bond resinous restorative materials to enamel, dentin, and dental alloys (e.g., CoCr, amalgam). Composition includes HEMA for wetting/penetration, 4-MET(A) for polymerization, urethane di-methacrylate for cross-linking, and acetone as a carrier/water chaser. Procedure involves applying to conditioned, moist tooth surface, air-drying, and light-curing for 20 seconds. Simplifies bonding process compared to conventional multi-step agents; improves reliability and ease of use for practitioners; facilitates secure restoration placement.

Clinical Evidence

Short-term clinical observation conducted by eight dentists under regular office conditions; evaluated over 300 restorations. Results confirmed ease of use, safety, and effectiveness. Supported by in-vitro investigations comparing performance against leading resinous bonding systems.

Technological Characteristics

Single-bottle light-curing monomer mixture. Composition: HEMA (hydrophilic monomer), 4-MET(A) (wetting/polymerization monomer), urethane di-methacrylate (cross-linking monomer), and acetone (solvent/carrier). Application requires phosphoric acid conditioning of tooth structure followed by light curing (20 seconds).

Indications for Use

Indicated for use as an enamel/dentin adhesive for bonding resin composite restorative materials and compomers to tooth structure, and for enhancing adhesion to metals such as amalgam.

Regulatory Classification

Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

Submission Summary (Full Text)

{0} Heraeus Kulzer, Inc. South Bend, IN. 46614 GLUMA ONE BOND 510(K) K974390 FEB 4 1998 # Statement of Safety and Effectiveness GLUMA ONE BOND is a one bottle adhesive, designed to bond resinous restorative materials to dental hard tissue. The product is a light curing monomer mixture dissolved in acetone which, following conditioning with phosphoric acid, is applied to enamel and dentin prior to co-polymerization with the restorative. When compared with conventional bonding agents the main advantage of such simplified bond mediators is their reliability and ease of use in the dental office. The composition of GLUMA ONE BOND is based on three monomers which individually have been used in marketed dental products for decades. The hydrophilic hydroxyethylmethacrylate (HEMA) is essential for perfect wetting of and penetration into the conditioned tooth structure. Similarly, 4-MET(A) has hydrophilic and moieties for wetting and polymerization, whereas the urethane di-methacrylate monomer is responsible for formation of a cross-linked polymer network. The organic solvent acetone has the function of a carrier for the monomers while being an effective water chaser at the same time. The application procedure is very easy. The solution is applied in small amounts to the conditioned moist tooth surface with two to three consecutive strokes, the water acetone is eliminated by a gentle air blast and finally, the resin is light cured for 20 seconds. Thorough in-vitro investigation has proven that the GLUMA ONE BOND performs better than or at least as good as the leading resinous bonding systems available in the market. GLUMA ONE BOND is also proven to bond to non-precious dental alloys, such as CoCr-based casting alloy and amalgam. The results of a short-term clinical observation done by eight dentists under regular office conditions, on some more than 300 restorations bonded with GLUMA ONE BOND confirms that the product is easy to use, safe, and effective. Based on this evidence and the very promising in vitro data, GLUMA ONE BOND is recommended as an effective enamel/dentin adhesive when used in accordance with the instructions for use. {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 4 1998 Ms. Cheryl V. Zimmerman Manager, Quality Operations & Regulatory Affairs Heraeus Kulzer, Incorporated Dental Products Division 4315 South Lafayette Boulevard South Bend, Indiana 46614-2517 Re: K974390 Trade Name: Gluma One Bond Regulatory Class: II Product Code: KLE Dated: November 17, 1997 Received: November 21, 1997 Dear Ms. Zimmerman: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 {2} Page 2 - Ms. Zimmerman through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} Page ___ of ___ 510(k) Number (if known): ___________________________ Device Name: GLUMA ONE BOND Indications For Use: GLUMA ONE BOND is a light curing single component enamel/dentin adhesive for bonding of all current brands of resin composite restorative materials and compomers. It also enhances adhesion to metals such as amalgam. GLUMA ONE BOND combines primer and adhesive in single bottle. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Runner (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number 8074390 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...