KERR ADVANCED ADHESIVE
Device Facts
| Record ID | K962515 |
|---|---|
| Device Name | KERR ADVANCED ADHESIVE |
| Applicant | Kerr Dental Materials Center |
| Product Code | KLE · Dental |
| Decision Date | Aug 1, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3200 |
| Device Class | Class 2 |
Intended Use
Kerr Advanced Adhesive is an all purpose dental adhesive consisting of a visible light curable methacrylate resin based mixture of monomers capable of forming both chemical and mechanical adhesive bonds to both natural tooth structures ( enamel and dentin ) and to commonly used restorative materials ( composite resins, porcelain and metals ). It is supplied as a one bottle, single step dental adhesive that is simple to use and produces fast, strong adhesion suitable for all commonly used restorative techniques.
Device Story
One-bottle, single-step dental adhesive; visible light-curable methacrylate resin monomer mixture. Applied by clinician to tooth structure (enamel/dentin) and restorative materials (composite, porcelain, metal). Organic solvent facilitates flow into micro-fissures; solvent evaporation leaves thin resin layer for chemical/mechanical bonding interface. Used in dental clinics for restorative procedures. Benefits include simplified application and strong adhesion.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993: Ames Mutagenicity Assay, Kligman Maximization Study, Cytotoxicity Study (MEM Elution Method USP 23), and Implantation Test. Effectiveness demonstrated via comparative bench testing against predicate devices.
Technological Characteristics
Visible light-curable methacrylate resin monomer mixture; organic solvent-based. One-bottle delivery system. Biocompatibility testing per ISO 10993.
Indications for Use
Indicated for use as a dental adhesive to bond restorative materials, including composite resins, porcelain, and metals, to natural tooth structures (enamel and dentin).
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
Predicate Devices
- PRIME & BOND
- One-Step
- Tenure QUIK
Related Devices
- K960823 — PRIME & BOND 2.0 MULTIPURPOSE DENTIN/ENAMEL BONDING AGENT · Dentsply Intl. · Apr 26, 1996
- K972858 — KERR NEXUS ONE BOTTLE ADHESIVE · Sybron Dental Specialties, Inc. · Sep 26, 1997
- K260682 — Bond-PR Universal Adhesive · Premier Dental Products Company · Mar 3, 2026
- K111980 — CLEARFIL TRI-S BOND PLUS · Kuraray America, Inc. · Oct 14, 2011
- K011908 — OPTIBOND SOLO PLUS 3 · Kerr Dental Materials Center · Sep 6, 2001