The indication for use of the air conductin hearing aids in this submissin is to amplify sound for individuals with imparied hearing.
Device Story
Finetone ITE Faceplate Assembly models (M-LIN, M-H LIN, M-D, M-PP, M-D/ASP, M-D/AGCI, M-K-6AMP, M-EQII) are in-the-ear (ITE) hearing aids. Devices capture ambient sound via microphone; process signals using various circuit technologies (Class A linear, Class B, Class D, hybrid amplifiers, automatic signal processing, input compression, or wide dynamic range dual-channel compression); and deliver amplified sound to the ear canal. Operated by the patient to assist with hearing impairment. Healthcare providers adjust settings (e.g., low/high cut, output, threshold kneepoint, gain) to match patient audiometric needs. Output aids speech intelligibility and sound perception. Potential for electromagnetic interference from digital cellular phones noted.
Clinical Evidence
No clinical data provided. Bench testing only.
Technological Characteristics
ITE faceplate assembly hearing aids. Circuit types: Class A linear, Class B, Class D, hybrid amplifier, automatic signal processing (ASP), input compression, and wide dynamic range dual-channel compression. User-adjustable controls: low cut, high cut, output, threshold kneepoint, low/high frequency gain, and crossover. Powered by standard hearing aid batteries.
Indications for Use
Indicated for individuals with impaired hearing, ranging from slight to severe hearing loss, including high frequency precipitously sloping, gradually sloping, and flat configurations, with varying tolerances to loudness.
Regulatory Classification
Identification
An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.
Submission Summary (Full Text)
{0}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure. The symbol is composed of three curved lines that converge at the bottom.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Paul Menezes Quality Engineer Finetone Hearing instruments, Inc 301 US Route 1 Scarborough, Maine 04074-9701
K974219 Re:
ITE Faceplate Assembly Aid Models: M-LIN, M-H LIN, M-D, M-PP, M-D/ASP, M-D/AGCI, M-K-6AMP and M-EQII Dated: February 9, 1998 Received: February 17, 1998 Regulatory Class: I 21 CFR 874.3300/Procode: 77 LRB
Dear Mr. Menezes:
FEB 20 1998
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
While your device has been deemed substantially equivalent to other legally marketed hearing aids, please be advised that electromagnetic interference from digital cellular telephones, as well as from other sources is increasingly becoming a concern. Typically, this interference takes the form of a buzzing sound that can range from annoying to very loud and may render a hearing aid temporarily ineffective for the wearer. Because electromagnetic interference may affect your device, you may be asked to test for electromagnetic compatibility in the future. In this interim period, we encourage you to modify your device labeling to inform practitioners and users of the potential for electromagnetic interference. Please be aware that a 510(k) submission is required for any claims that infer that your device is compatible with potential sources of electromagnetic interference, such as "compatible with digital cellular telephones", and that data supporting such claims is necessary.
{1}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
lliam Yi
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known): K 97 42 / g
Device Name: Finetone ITE Faceplate Assembly, Model M-LIN
Indications For Use: The M-LIN is a Class A linear circuit. Available options are the Low cut, High cut and Output controls.
## A. General Indications:
The indication for use of the air conductin hearing aids in this submissin is to amplify sound for individuals with imparied hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)):
| Severity: | |
|-------------|---|
| 1. Slight | ✓ |
| 2. Mild | ✓ |
| 3. Moderate | ✓ |
| 4. Severe | |
| 5. Profound | |
| Configuration: | |
|--------------------------------------------|---|
| 1. High frequency<br>Precipitously sloping | ✓ |
| 2. Gradually sloping | ✓ |
| 3. Reverse Slope | |
| 4. Flat | ✓ |
| 5. Other | |
| Other: | |
|---------------------------------|---|
| 1. Low tolerance<br>To loudness | |
| 2. High Tolerance | |
| 3. Normal Tolerance | ✓ |
B. Specific Indications (Only appropriate.):
1.
(Most psychoacoustic indications such as improved speech intelligibility in background noise, must be supported by clinical data.)
- 2. 3.
PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David G. Seysman
Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number
{3}------------------------------------------------
510(k) Number (if known): | 9742/9
Device Name: Finetone ITE Faceplate Assembly, Model M-PP
Indications For Use: The M-PP is a Class B circuit. Available options are the Low cut, High cut and ouput controls.
A. General Indications:
2.
3.
The indication for use of the air conductin hearing aids in this submissin is to amplify sound for individuals with imparied hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)):
| Severity: | |
|--------------------------------------------|---|
| 1. Slight | |
| 2. Mild | |
| 3. Moderate | |
| 4. Severe | ✓ |
| 5. Profound | |
| Configuration: | |
| 1. High frequency<br>Precipitously sloping | |
| 2. Gradually sloping | ✓ |
| 3. Reverse Slope | |
| 4. Flat | ✓ |
| 5. Other | |
| Other: | |
| 1. Low tolerance<br>To loudness | |
| 2. High Tolerance | ✓ |
| 3. Normal Tolerance | ✓ |
B. Specific Indications (Only appropriate.):
(Most psychoacoustic indications such as improved speech intelligibility in
background noise, must be supported by clinical data.)
1.PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Byrne
(Division Sign-Off)
Division Sign-Off) Division of Reproductive, Abdominal, E and Radiological Devi 510(k) Number
{4}------------------------------------------------
510(k) Number (if known): | 97 42/9
Device Name: Finetone ITE Faceplate Assembly, Model M-HLIN
Indications For Use: The M-HLIN is a Class A linear circuit utilizing a hybrid amplifier. Available options are the Low cut, High cut and Output controls.
A. General Indications:
ని.
The indication for use of the air conductin hearing aids in this submissin is to amplify sound for individuals with imparied hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)):
| Severity: | Configuration: | Other: |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|--------------------------------|
| 1. Slight | 1. High frequency Precipitously sloping | 1. Low tolerance To loudness |
| 2. Mild | 2. Gradually sloping | 2. High Tolerance |
| 3. Moderate | 3. Reverse Slope | 3. Normal Tolerance |
| 4. Severe | 4. Flat | |
| 5. Profound | 5. Other | |
| B. Specific Indications (Only appropriate.): (Most psychoacoustic indications such as improved speech intelligibility in background noise, must be supported by clinical data.) | | |
| 1. | | |
| 2. | | |
PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CØRH, Office of Device Evaluation (ODE)
Daniel L. Seymour
Division Sign Off
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K974219
{5}------------------------------------------------
510(k) Number (if known): ├ 97 7 Z / 9
Device Name: Finetone ITE Faceplate Assembly, Model M-D
Indications For Use: The M-D is a Class D linear circuit. Available options are the Low cut, High cut and Output controls.
A. General Indications:
The indication for use of the air conductin hearing aids in this submissin is to amplify sound for individuals with imparied hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)):
| Severity: | | Configuration: | | Other: | |
|-------------|--------------|-----------------------------------------|--------------|------------------------------|--------------|
| 1. Slight | | 1. High frequency Precipitously sloping | <div>✓</div> | 1. Low tolerance To loudness | |
| 2. Mild | <div>✓</div> | 2. Gradually sloping | <div>✓</div> | 2. High Tolerance | <div>✓</div> |
| 3. Moderate | <div>✓</div> | 3. Reverse Slope | | 3. Normal Tolerance | <div>✓</div> |
| 4. Severe | <div>✓</div> | 4. Flat | <div>✓</div> | | |
| 5. Profound | | 5. Other | | | |
B. Specific Indications (Only appropriate.):
(Most psychoacoustic indications such as improved speech intelligibility in background noise, must be supported by clinical data.)
| 1. | |
|----|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | the submit and the submit of the submit of the submit of the subscription of the sub- |
| 3. | . company with the country of the country of the submit and the submit and the submit of the submit of the submit of the submit of the subsequence of<br>the submit and the submit and the submit of the submit of the submit of |
PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Segerson
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K974719
{6}------------------------------------------------
510(k) Number (if known): ★タア 42/ 9
Device Name: Finetone ITE Faceplate Assembly, Model M-D/ASP.
Indications For Use: The M-D/ASP is a Automatic Signal Processing circuit utilizing Class D technology. Available options are the Asp, Low cut and High cut controls.
A. General Indications:
్రస్స
The indication for use of the air conductin hearing aids in this submissin is to amplify sound for individuals with imparied hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)):
| Severity: | | Configuration: | | Other: | |
|-------------|---|--------------------------------------------|---|---------------------------------|---|
| 1. Slight | | 1. High frequency<br>Precipitously sloping | ✓ | 1. Low tolerance<br>To loudness | ✓ |
| 2. Mild | ✓ | 2. Gradually sloping | ✓ | 2. High Tolerance | |
| 3. Moderate | ✓ | 3. Reverse Slope | | 3. Normal Tolerance | ✓ |
| 4. Severe | | 4. Flat | ✓ | | |
| 5. Profound | | 5. Other | | | |
B. Specific Indications (Only appropriate.):
(Most psychoacoustic indications such as improved speech intelligibility in background noise, must be supported by clinical data.)
1.
2.
3.
PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of ØDRH, Office of Device Evaluation (ODE)
Daniel A. Seymour
---
(Division Sign-Off)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K974219
{7}------------------------------------------------
510(k) Number (if known): K 97 4 Z 1 9
Device Name: Finetone ITE Faceplate Assembly, Model M-D/AGCI
Indications For Use: The M-D/AGCI is an input compression circuit utilizing Class D technology. Available iptions are the Low cut and High cut controls.
A. General Indications:
The indication for use of the air conductin hearing aids in this submissin is to amplify sound for individuals with imparied hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)):
| Severity: | |
|-------------|----------------------------------------------------|
| 1. Slight | <span style="text-decoration: underline;">✓</span> |
| 2. Mild | <span style="text-decoration: underline;">✓</span> |
| 3. Moderate | <span style="text-decoration: underline;">✓</span> |
| 4. Severe | |
| 5. Profound | |
| | Configuration: |
|-----------------------------------------|----------------------------------------------------|
| 1. High frequency Precipitously sloping | <span style="text-decoration: underline;">✓</span> |
| 2. Gradually sloping | <span style="text-decoration: underline;">✓</span> |
| 3. Reverse Slope | |
| 4. Flat | <span style="text-decoration: underline;">✓</span> |
| 5. Other | |
| | Other: |
|------------------------------|----------------------------------------------------|
| 1. Low tolerance To loudness | <span style="text-decoration: underline;">✓</span> |
| 2. High Tolerance | |
| 3. Normal Tolerance | |
B. Specific Indications (Only appropriate.):
(Most psychoacoustic indications such as improved speech intelligibility in background noise, must be supported by clinical data.)
- 1. 2.
3.
PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Seymore
(Division Sign-Off)
Division of Reproductive, Abdominal, EN and Radiological Devic 510(k) Number
{8}------------------------------------------------
510(k) Number (if known):
Device Name: Finetone ITE Faceplate Assembly, Model M-KAMP.
Indications For Use: The M-KAMP detects and amplifies quiet sounds. In loud situations it provides low distortion and high fedelity sound. Available options are the Threshold Kneepoint and low cut controls.
## A. General Indications:
The indication for use of the air conductin hearing aids in this submissin is to amplify sound for individuals with imparied hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)):
| Severity: | | Configuration: | | Other: | |
|-------------|---|-----------------------------------------|---|------------------------------|---|
| 1. Slight | ✓ | 1. High frequency Precipitously sloping | ✓ | 1. Low tolerance To loudness | ✓ |
| 2. Mild | ✓ | 2. Gradually sloping | ✓ | 2. High Tolerance | |
| 3. Moderate | ✓ | 3. Reverse Slope | | 3. Normal Tolerance | ✓ |
| 4. Severe | | 4. Flat | ✓ | | |
| 5. Profound | | 5. Other | | | |
B. Specific Indications (Only appropriate.):
(Most psychoacoustic indications such as improved speech intelligibility in background noise, must be supported by clinical data. } -----------
1.
2.
3.
PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|--------------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices | |
| 510(k) Number | K974219 |
{9}------------------------------------------------
510(k) Number (if known): 1997 4219
Device Name: Finetone ITE Faceplate Assembly, Model M-EQII.
Indications For Use: The M-EQII is a Wide Dynamic Range, duel channel, input compression circuit which utilizes Class D technology. Available options are the Threshold Kneepoint, Low Frequency Gain, High Frequency Gain, Crossover and Output controls.
- A. General Indications:
The indication for use of the air conduction hearing aids in this submission is to amplify sound for individuals with impaired hearing. The devices are indicated for individuals with losses in the following category(ies). (Check appropriate space(s)):
| Severity: | | Configuration: | | Other: | |
|-------------|--------------|-----------------------------------------|--------------|------------------------------|--------------|
| 1. Slight | <div> </div> | 1. High frequency Precipitously sloping | <div> </div> | 1. Low tolerance To loudness | <div> </div> |
| 2. Mild | <div> </div> | 2. Gradually sloping | <div> </div> | 2. High Tolerance | <div> </div> |
| 3. Moderate | <div> </div> | 3. Reverse Slope | <div> </div> | 3. Normal Tolerance | <div> </div> |
| 4. Severe | | 4. Flat | | | |
| 5. Profound | | 5. Other | | | |
B. Specific Indications (Only appropriate.):
(Most psychoacoustic indications such as improved speech intelligibility in background noise, must be supported by clinical data.)
- 1.
2.
- ತ.
PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Eyaluation (ODE)
(Division Sign-Off) Division of Reproductive, Abdominal, El and Radiological Devi
510(k) Number K974639
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.