95 C/1.5 CITRIC ACID HEAT PROCESSING OF POLYSULFONE DIALYZERS
K974090 · Fresenius Medical Care North America · MSE · Aug 27, 1998 · Gastroenterology, Urology
Device Facts
Record ID
K974090
Device Name
95 C/1.5 CITRIC ACID HEAT PROCESSING OF POLYSULFONE DIALYZERS
Applicant
Fresenius Medical Care North America
Product Code
MSE · Gastroenterology, Urology
Decision Date
Aug 27, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Hemoflow dialyzers are designed for acute and chronic hemodialysis. Hemoflow dialyzers are also appropriate for single or multiple use when reprocessed with 1.5% Citric Acid/95℃.
Device Story
Device is a reprocessing procedure for Fresenius polysulfone hemodialyzers; utilizes 1.5% citric acid solution and 95°C moist heat exposure for 24 hours. Procedure performed in clinical settings using convection ovens; Fresenius DRS-4 system may be used. Input: used hemodialyzers; Output: reprocessed, disinfected/sterilized hemodialyzers. Process involves thermal destruction of microorganisms; citric acid acts as non-toxic additive to enhance disinfection effectiveness. Validation confirms 18 log reduction of B. stearothermophilus spores. Healthcare providers use reprocessed dialyzers for subsequent dialysis sessions; benefits include reuse of hemodialyzers, reducing waste and costs while maintaining clinical performance equivalent to formaldehyde-based reprocessing. No difference in patient intradialytic episodes noted compared to traditional methods.
Clinical Evidence
Bench testing: thermal penetration studies in convection ovens; microbial kill effectiveness testing using B. stearothermophilus spores (18 log reduction achieved). Toxicity: ISO 10993-1 testing on dialyzers reprocessed 15+ times showed no toxicity. In vitro: membrane performance (flux) compared to formaldehyde-reprocessed dialyzers showed no difference. In vivo: clinical comparison of patients transferred from formaldehyde-reprocessed to heat-reprocessed dialyzers showed no differences in intradialytic episodes (hypotension, reactions).
Technological Characteristics
Polysulfone hemodialyzers; 1.5% citric acid (pH 2.1) solution; 95°C moist heat disinfection; convection oven; thermocouples for heat penetration monitoring. ISO 10993-1 biocompatibility standards. No software-based algorithm; manual/mechanical reprocessing procedure.
Indications for Use
Indicated for patients requiring acute or chronic hemodialysis. Appropriate for single or multiple use when reprocessed using the specified 1.5% citric acid/95°C thermal disinfection procedure.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Submission Summary (Full Text)
{0}------------------------------------------------
K974090 p. 1/3
## Summary of 510 (k) Submission and Statement of Equivalence for 95°C/1.5% Citric Acid Dialyzer Reprocessing
Hemodialyzer reuse has been practiced throughout the world by the renal community since the introduction of dialysis as a recognized replacement therapy for chronic renal disease. In the United States many forms of hemodialyzer reuse have been practiced (formaldehyde, gluteraldehyde, peracetic acid to name a few) and medical devices and procedures have been reviewed by the FDA and have been found to be substantially equivalent to medical products and practices in use prior to 1976. Fresenius Medical Care is requesting by submission of this application that the Commissioner find that the procedure and components used in the reprocessing of Fresenius polysulfone hemodialyzers with 95°C/1.5% citric acid be found substantially equivalent to other forms of hemodialyzer reuse that have been in medical practice in the US prior to 1976.
Radiant heat (moist or dry) has been used for over 100 years as a means to sterilize or disinfect utensils, solutions and containers. Scientists since the time of Pasteur noted that exposure to a specific temperature for a specific amount of time would allow for preservation of the compound or component tested. This observation of specific time/temperature exposure is the basis of current thermal sterilization of medical products and pharmaceuticals. Figure 1 in Attachment 1 shows, in graphic form, the amount of exposure time (hours) required to attain sterilization of a medical product at various temperatures (°C). In this case sterilization is defined as the amount of time at the specified temperature to attain a 6 log reduction in bacterial concentration. Table 1 shows that exposure at 121°C for 15 minutes is comparable to 105°C for 10 hours.
Also enclosed in Attachment 1 is a copy of a chapter from Principles and Methods of Sterilization titled "Thermal Destruction of Microorganisms". In this attachment relationships are presented for the relative resistance to heat for various microorganisms. As an example, vegetative microorganisms, such as Escherichia coli, are as much as 3.000.000 times more sensitive to heat than bacterial spore formers. Likewise, bacterial spore formers have different heat resistance. Bacillus stearothermophilus, the recognized indicator organism for moist heat sterilization processes, is 50-100 times more resistant to moist heat than Bacillus subtilus.
Citric acid is a natural forming, non-toxic compound that when diluted to 1.5%, weight per volume, results in a solution of approximately pH 2.1 that has no chemical effect on components used in the manufacture of Fresenius polysulfone hemodialyzers. Citric acid is commonly used as a stabilizer in food and cosmetic products as well as a bacteriostatic compound. However citric acid alone is not considered a bacteriostat or bactericide. Enclosed in Attachment 2 is a copy of test results where 1.5% citric acid solution was challenged according the USP antimicrobial effectiveness test as a bactericide. The test results indicate that a 1.5% citric acid solution does not meet the requirements of the antimicrobial effective test in that the test organisms did not exhibit an accelerated death rate different than the control, but no organisms proliferated in the solution.
Citric acid was selected as a non-toxic additive to enhance the effectiveness of low temperature disinfection procedures. The plastic components used in the manufacture of Fresenius polysulfone dialyzers are heat-labile and malleable at temperatures of 121°C so lower temperatures were required for longer periods of exposure time. The addition of
3
{1}------------------------------------------------
K974090 p. 2/3
citric acid to the dialyzer reprocessing solution (water) increased the effective kill of the indicator organism during the disinfection process.
The validation process used to qualify the 95°C heat disinfection procedure is based on the standard monograph developed by the Parenteral Drug Manufacturing Association for steam sterilization. Test dialyzers were selected after exposure to blood at least 10 reprocessing procedures. Thermocouples were positioned in dialyzers to demonstrate heat penetration variability within the dialyzer. Heat penetration studies were carried out in standard convection ovens completely filled with dialyzers to determine the "coldest" position in the oven. Various dialyzer sizes (F60 and F80) were evaluated for differences in heat penetration profiles. Studies were carried out multiple times to demonstrate reproducibility. The data collected from these studies are enclosed in Attachment 3. In summary, the temperature penetration studies demonstrated that the lower right shelf position was the coldest position within the oven, there was no difference in heat penetration studies of F60 or F80 dialyzers and that there was no "cold" position within the dialyzer (all thermocoupled positions heated at the same rate).
Based on the thermal penetration studies, 10X blood contacted reprocessed F80 dialyzers biologically seeded on the blood side with approximately 1,000,000 B. stearothermophilus spores were positioned in the "cold spot" of the oven for determination of kill effectiveness of the reprocessing procedure. Seeded dialyzers were placed in the oven at room temperature, the oven was turned on with the temperature set point at 95°C. Test dialyzers were pulled at specified hour increments, blood side drained and flushed and the contents serially plated to determine number of viable organisms. These data were used to calculate the log reduction potential of the reprocessing procedure. Three test runs were conducted. The results are enclosed in Attachment 3. Results demonstrate that, at least 18 log reduction in the test organism was achieved with the 95°C/1.5% citric acid for 24 hours (from room temperature to the test temperature) program. These results demonstrate that the heat reprocessing procedure of 95°C exposure with 1.5% citric acid results in an additional 12 log reduction of the indicator organism over standard accepted sterilization log reduction processes of 6 log. In addition to the above tests, a reprocessing run was also carried out at 90℃ for 24 hours with the same procedure. The results of this single run demonstrate at least a 12 log reduction of the test organism. Lastly, 24 dialyzers were seeded with the test organism B. stearothermophilus , heated at 95℃ for 24 hours, let set for 96 hours, blood side drained and serially plated and the dialyzer blood side filled with SCD media. No growth was observed. The data demonstrate that with worst case conditions (maximum dialyzer loading, largest dialyzer, multiple blood contact, heat from RT. to 95°C for 24 hours, oven opened to remove test samples and closed, 5°C below standard temperature) that disinfection and, in fact sterilization has been attained. In addition, 1.5% citric acid was shown to be bacteriostatic.
Toxicity testing was carried out according to the ISO 10993-1 guidelines with the exception that only one of three carcinogenicity tests was performed on dialyzer samples that were reprocessed a minimum of 15 times. The data are enclosed in Attachment 4. Results indicate no evidence of toxicity.
Attachment 5 contains the in vitro performance data for each membrane family of Fresenius polysulfone dialyzers; low (F4, F6 and F8), mid (F60M and F80M) and high flux (F60 and F80) compared to membrane performance data for 1.5% formaldehyde (HCHO) and 40℃ based on the guidelines specified in the dialyzer reuse 510 (k) document. The materials and manufacturing process used in the production of all Fresenius polysulfone dialyzers are identical. Flux rates are manipulated by accelerating the membrane precipitation rates and wall formation only. The data demonstrate that there is no difference in membrane performance due to heat reprocessing with 95°C/1.5% citric
{2}------------------------------------------------
K974090 p. 3/3
acid when compared to non-reprocessed dialyzers or dialyzers reprocessed with 1.5% HCHO and 40°C incubation.
Attachment 6 contains in vivo product performance data for all Fresenius polysulfone membrane configurations compared to in vivo product performance of dialyzers preprocessed with 1.5% HCHO with 40℃ incubation. The results show no difference in product performance due to heat reprocessing with 95°C/1.5% citric acid compared to non reprocessed hemodialyzers or dialyzers reprocessed with 1.5% HCHO with 40°C incubation.
Attachment 7 contains in vivo clinical data of comparison of patients dialyzed with hemodialyzers reprocessed with 1.5% formaldehyde and 40℃ incubation who were then transferred to dialyzers heat reprocessed with 95°C/1.5% citric acid. Patient parameters and intradialytic enisodes (hypotension, reactions, etc.) were followed. No differences were noted between the two reprocessing procedures.
Attachment 8 contains proposed clinical reprocessing procedures, convection oven operator's manual and brochure for a typical temperature recorder, and proposed manual for citric acid reprocessing with a Fresenius DRS-4 Dialyzer Reprocessing System.
Based on the enclosed information, Fresenius requests that the Commissioner find that hemodialyzer reprocessing using 1.5% citric acid/95°C for 24 hours is substantially equivalent to other reprocessing procedures in clinical use now and prior to 1976.
{3}------------------------------------------------
Rockville MD 20850
Food and Drug Administration 9200 Corporate Boulevard
Image /page/3/Picture/2 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized human figure with three wavy lines representing the head, body, and legs. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN" which is arranged in a circular fashion around the left side of the logo. The text is in uppercase letters.
AUG 27 1998
Mr. Tom Folden Director, Product Development Fresenius Medical Care 2637 Shadelands Drive Walnut Creek, CA 94598
Re: K974090
Multiple Use Labeling for Hemoflow hemodialyzers Reprocessed with 1.5% Citric Acid/95+ 2 °C Dated: May 20, 1998 Received: May 29, 1998 Regulatory Class: II and III 21 CFR §876.5820 and §876.5860 Product code: 78 MSE and 78 MSF
Dear Mr. Folden:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include i requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premaket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (2) CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrb/dsmaldsmamain.html".
Sincerely yours
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Fresenius Medical Care North America Renal Product Technologies. The logo consists of a triangular shape on the left and the company name on the right. The triangular shape is divided into four sections, each with a horizontal line. The text is stacked on top of each other, with "Fresenius Medical Care" on the top, "North America" in the middle, and "Renal Product Technologies" on the bottom.
## Indications for Use Statement
"Hemoflow dialyzers are designed for acute and chronic hemodialysis. Hemoflow dialyzers are also appropriate for single or multiple use when reprocessed with 1.5% Citric Acid/95℃."
liauth
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number 974-090
Prescription Use
(Per 21 CFR 801.109)
.
:
.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.