K972183 · Ace Medical Co. · HSB · Aug 26, 1997 · Orthopedic
Device Facts
Record ID
K972183
Device Name
DEPUY ACE A.I.M. TITANIUM TIBIAL NAIL
Applicant
Ace Medical Co.
Product Code
HSB · Orthopedic
Decision Date
Aug 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DePuy ACE A.I.M.º Titanium Tibial Nail is indicated for internal fixation of tibial fractures including transverse fractures, oblique fractures, spiral fractures, comminuted fractures, fractures, fractures with bone loss, bone transport, open fractive osteotomies, pathologic fractures, pseudoarthrosis of the tibial shaft, nonunions, malunions, metaphyseal fractures and epiphyseal fractures.
Device Story
Intramedullary fixation device for tibial fractures; titanium nail inserted into tibial canal to stabilize bone segments. Features distal bend (2° or 5° depending on diameter) and proximal dynamization slot; includes multiple cross-locking screw holes for cortical bone screws. Used by orthopedic surgeons in clinical/OR settings. Provides mechanical stability to fracture site; allows for compression or dynamization to promote healing. Benefits patient by facilitating bone alignment and stabilization during fracture recovery.
Clinical Evidence
Bench testing only. Biomechanical testing compared subject device to Synthes Tibial Nail; results indicated subject device was at least as strong or stronger than the comparator.
Technological Characteristics
Titanium 6Al-4V ELI (ASTM F-136). Diameters 8.0mm-13.0mm; lengths from 25.5cm. 8mm/9mm nails are solid; 10mm-13mm nails are cannulated. Distal bend 2° (10-12mm) or 5° (8-9mm). Proximal dynamization slot (5mm). Mechanical fixation via cross-locking screws.
Indications for Use
Indicated for internal fixation of tibial fractures in patients requiring stabilization of transverse, oblique, spiral, comminuted, segmental, or pathologic fractures; also indicated for bone transport, corrective osteotomies, pseudoarthrosis, nonunions, malunions, metaphyseal, and epiphyseal fractures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{1}------------------------------------------------
AUG 26 1997
K972183
510(k) SUMMARY
| NAME OF FIRM: | DePuy ACE Medical Company<br>2260 East El Segundo Boulevard<br>El Segundo, CA 90245 |
|-------------------------------------|-------------------------------------------------------------------------------------|
| 510(k) CONTACT PERSON: | Kathleen A. Dragovich<br>Regulatory Affairs Specialist<br>DePuy ACE Medical Company |
| TRADE NAME: | DePuy ACE A.I.M.® Titanium Tibial Nail |
| COMMON NAME: | Intramedullary Rod |
| CLASSIFICATION: | 888.3020 Rod, Fixation, Intramedullary and<br>Accessories |
| DEVICE CODE: | 87HSB |
| SUBSTANTIALLY<br>EQUIVALENT DEVICE: | DePuy ACE AIM® Titanium Tibial Nail |
## INTENDED USE:
The DePuy ACE A.I.M.º Titanium Tibial Nail is indicated for internal fixation of tibial fractures including transverse fractures, oblique fractures, spiral fractures, comminuted fractures, fractures, fractures with bone loss, bone transport, open fractive osteotomies, pathologic fractures, pseudoarthrosis of the tibial shaft, nonunions, malunions, metaphyseal fractures and epiphyseal fractures.
## DEVICE DESCRIPTION AND SUBSTANTIAL EQUIVALENCE RATIONALES:
The DePuy ACE A.I.M.® Titanium Tibial Nail is an intramedullary fixation of tibial fractures. The nail has a distal bend of 2º for the nails of 10mm, 11mm, 12mm diameters. The distal bend in the nails of 8mm and 9mm diameters is 5°. The DePuy ACE A.I.M.º Titanium Tibial Nail has diameters from 8.0mm to 13.0mm and lengths from 25.5cm. The proximal end of the 10mm, 11mm, 12mm , and 13mm nails is 13mm in diameter while the proximal end of the 8mm and 9mm nadiameter. The proximal end of the nail contains two 6mm cross locking screw holes which accept 5.5mm solid cortical bone screws and one 5mm long dynamization slot in the M-L plane. The distal end of the nail contains two 5.0mm holes in the M-L plane which accept 4.5mm solid cortical bone screws and one 5.0mm hole in the A-P plane positioned between the two transverse holes. The 8mm and 9mm nails are solid while the 10mm to 13mm nails are cannulated.
The DePuy ACE A.I.M. Titanium Tibial Nail is manufactured from Titanium 6A1-4V ELI (ASTM standard F-136).
The new DePuy ACE A.I.M.® Titanium Tibial Nail is the same basic design as the current DePuy ACE A.I.M. Titanium Tibial Nail, with the exceptions of adding a proximal dynamization slot in the M-L plane, which allows 5mm of length to dynamize across fracture site and that the two oblique proximal screw holes are moved more proximal to vield a longer working length for the nail. Also, there is an additional hole in the nail in the A-P plane. This is positioned between the current two distal transverse holes, allowing for oblique locking distally, which is more stable and allows the lagging of anterior fragments of the distal tibia. The additional holes and changes in placement are very similar to that of the Synthes Titanium Cannulated Tibial Nail and the Synthes Titanium Unreamed Tibial Nail.
Biomechanical testing was done using the Synthes product and the DePuy ACE Tibial Nail. On all tests, the DePuy Tibial Nail was either at least as strong or stronger than the Synthes Tibial Nail.
Based on this information, DePuy ACE believes that the modified DePuy ACE A.I.M. Titanium Tibial Nails are substantially equivalent to the DePuy ACE A.I.M." Titanium Tibial Nail, the Synthes Reamed Titanium Tibial Nail, and the Synthes Unreamed Titanium Tibial Nail.
ల్లో ప్ర
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Kathleen A. Dragovich ·Regulatory Affairs Specialist DePuy ACE Medical Company 2260 East El Segundo Boulevard El Segundo, California 90245-4694
AUG 26 1997
K972183 Re: DePuy ACE A.I.M.® Titanium Tibial Nail Requlatory Class: II Product Code: HSB Dated: June 6, 1997 Received: June 10, 1997
Dear Ms. Dragovich:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2 - Ms. Kathleen A. Dragovich
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known)
DePuy ACE A.I.M.® Titanium Tibial Nail Device Name:
Indication For User:
The DePuy ACE A.I.M.® Titanium Tibial Nail is indicated for internal fixation of tibial fractures including transverse fractures, oblique fractures, spiral fractures, segmental fractures, comminuted fractures, fractures with bone loss, bone transport, open fractures, corrective osteotomies, pathologic fractures, pseudoarthrosis of the tibial shaft, nonunions, malunions, metaphyseal fractures and epiphyseal fractures.
Concurrence of CDRH, Office of Device Evaluation
OR Over-The-Counter Prescription Use (Per 21 CFR 801.109) (Division Sign-Off) (Division of General Restorative Devices ി 1972 510(k) Number .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.