UNITRON MODEL SOUND F/X CIC
Device Facts
| Record ID | K970692 |
|---|---|
| Device Name | UNITRON MODEL SOUND F/X CIC |
| Applicant | Unitron Industries, Inc. |
| Product Code | ESD · Ear, Nose, Throat |
| Decision Date | May 8, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 874.3300 |
| Device Class | Class 1 |
| Attributes | Therapeutic |
Intended Use
To amplify and transmit sound via air conduction to the ear.
Device Story
Completely-In-The-Canal (CIC) hearing aid; amplifies/transmits sound via air conduction. Features two-channel wide dynamic range compression; six adjustable parameters (Low/High-channel Gain, See-Saw, Cross Over Frequency, Threshold Knee-point, power); twin average compression detectors; Class D circuitry. Powered by size 10A or 5A battery. Operated by patient; volume adjusted via screw set control. Assembled from standard industry components. Benefits patient by improving auditory perception.
Clinical Evidence
Bench testing only; compliance with ANSI Standard S3.22-1987.
Technological Characteristics
CIC form factor; Class D circuitry; two-channel wide dynamic range compression; six adjustable parameters; screw set volume control; powered by size 10A or 5A batteries. Technical specifications comply with ANSI S3.22-1987.
Indications for Use
Indicated for individuals requiring sound amplification via air conduction to the ear to address hearing impairment.
Regulatory Classification
Identification
An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.
Related Devices
- K970519 — UNITRON MODEL SOUND F/X PROGRAMMABLE ITE (SF/X PROGRAMMABLE) · Unitron Industries, Inc. · Apr 24, 1997
- K964639 — TELEX IN-THE-EAR CUSTOM HEARING AID WITH THRESHOLD COMPRESSION-II · Telex Communications, Inc. · Jan 21, 1997
- K963122 — UNITRON MODEL ICON KR · Unitron Industries, Inc. · Sep 6, 1996
- K964503 — UNITRON MODEL SOUND F/X +4 · Unitron Industries, Inc. · Nov 27, 1996
- K964511 — UNITRON MODEL SOUND F/XD +4 (SOUND F/XD +4) · Unitron Industries, Inc. · Nov 27, 1996