PERINEOMETER & VAGINAL PROBE (BIOFEEDBACK MONITOR #20 AND VAGINAL PROBE)
K970145 · Biosearch Medical Products, Inc. · HIR · Aug 11, 1997 · Obstetrics/Gynecology
Device Facts
Record ID
K970145
Device Name
PERINEOMETER & VAGINAL PROBE (BIOFEEDBACK MONITOR #20 AND VAGINAL PROBE)
Applicant
Biosearch Medical Products, Inc.
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Aug 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1425
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Perineometer (Biofeedback monitor) and vaginal probe are intended to be used to assist the patient in exercising her pelvic muscles (referred to Kegal exercises) which in the majority of cases lead to increased bladder control.
Device Story
Device consists of a vaginal probe connected to a battery-powered biofeedback monitor. Probe features a silicone shaft and inflatable silicone balloon; balloon provides gentle pressure to vaginal wall. Patient performs Kegel exercises; pelvic muscle contractions displace air within the balloon. Pressure changes are transmitted via tubing to the monitor, which converts signals into a visual display of LED lights. Used in a clinical or home setting under physician prescription. Provides real-time visual feedback of muscle activity, assisting patients in monitoring exercise progress. Benefits include enhanced pelvic floor strength and improved urinary continence.
Clinical Evidence
Bench testing only. Biocompatibility testing performed on probe components including cytotoxicity (MEM elution), hemolysis, pyrogenicity, intracutaneous toxicity, and implantation tests. All materials met USP Class VI requirements. Check valve performance verified against ANSI/HIMA MD 70.1 and British BS 5081 standards.
Technological Characteristics
Passive biofeedback system. Materials: USP Class VI silicone (probe/marker ring), medical-grade plastic (tubing/connectors). Sensing principle: pneumatic pressure displacement via inflatable balloon. Energy source: rechargeable batteries (unit disabled during charging). Connectivity: wired pneumatic connection between probe and monitor. Output: LED visual display. Sterilization: not specified.
Indications for Use
Indicated for female patients requiring pelvic muscle exercise (Kegel exercises) to improve pelvic floor strength and bladder control, typically associated with anatomical or neurological pelvic floor dysfunction.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
{0}
K970145
AUG 11 1997
# bi5search MEDICAL PRODUCTS INC.
## >510(k) SUMMARY <
## SAFETY & EFFECTIVENESS SUMMARY
## PERINEOMETER & VAGINAL PROBE
Manufactured / Sold by:
Biosearch Medical Products, Inc.
35 Industrial Parkway
P.O. Box 1700
Somerville, NJ. 08876 {USA}
Telephone: 1-908-722-5000
Fax: 1-908-722-5024
Contact Person:
Martin Dyck
*(Handwritten signature above printed name)*
Vice President of Operations,
New Product Development Coordinator
*(Handwritten signature above printed name)*
1/15/97
date submitted
Biosearch Medical Products, Inc. P.O. Box 1700 Somerville, NJ. 08876-1276 (USA)
TEL: 908-722-5000 FAX: 908-722-5024
BIO5SEARCH
{1}
K970145 3/8
# SAFETY & EFFECTIVENESS SUMMARY
## Biosearch Medical Products, Inc.
### PERINEOMETER & VAGINAL PROBE
**INTENDED USE STATEMENT:** The Perineometer (Biofeedback monitor) and vaginal probe are intended to be used to assist the patient in exercising her pelvic muscles (referred to Kegal exercises) which in the majority of cases lead to increased bladder control.
(Figure showing vaginal probe inserted)

{2}
K970145
# SAFETY & EFFECTIVENESS SUMMARY
## Biosearch Medical Products, Inc.
### PERINEOMETER & VAGINAL PROBE
## DESCRIPTION OF THE DEVICE:
The PERINEOMETER & VAGINAL PROBE, is a passive device which will allow the patient to visually monitor their Kegal exercise progress.
The **Vaginal probe** is a Single patient use only device comprised of a medical grade Silicone shaft with a silicone balloon (Inflatable to enlarge to accomodate various vaginal sizes.) and a polyurethane connector tube. The silicone balloon is mounted in such a way that when pressure is applied to it, air will be displaced through its connector tubing. This air displacement is being monitored by the biofeedback monitor which will cause the LED lights to turn on in proportion to the pressure exerted. This probe is similar to the legally marketed probe via 510(k) numbers (k902843) and (k913736) except that in this case we are inflating the balloon to provide gentle pressure to the vaginal wall. This will allow the probe to be customized for the patients individual vaginal size.
The **Perineometer (Biofeedback Monitor)** is powered by rechargeable batteries. To prevent a potential shock hazard this unit will not function while it is being recharged. The Perineometer is identical to the biofeedback monitor which had been 510(k) approved for use in fecal incontinence (k902843). Thus, in the case of this 510(k) submission, we only seek to expand its use for general Kegal exercises.
## SYSTEM DESCRIPTION:
The patient inserts the vaginal probe which is connected to the Perineometer. Once the probe is inserted the balloon is inflated until it provides gentle pressure to the vaginal wall. As the pelvic muscles expand and contract (Kegal exercises) the pressure in the probe changes and these pressure changes are converted into a visual signal on the Perineometer (biofeedback monitor). The visual signal on the Perineometer is a series of LEDs (lights) which provide the patient with a real time representation of pelvic muscle activity.

Vaginal Probe

Perineometer
2
{3}
K970145 4/8
# SAFETY & EFFECTIVENESS SUMMARY
## Biosearch Medical Products, Inc.
### PERINEOMETER & VAGINAL PROBE
**Why do pelvic muscle exercises (Kegal exercise):** Without the strength and support of the pelvic floor muscles, urinary continence is difficult to maintain. Female urinary incontinence is usually associated with anatomical, neurological pelvic floor dysfunction. Presently many Urologist/Gynecologists are recommending pelvic muscle exercises to their female patients since these exercises will serve to enhance the strength and pelvic muscle control for the patient and in many cases improve urinary continence.
**What need does this device address:** The Perineometer (Biofeedback Monitor) and Vaginal Probe makes it possible for the patient to visualize their pelvic muscle activity. This is otherwise extremely difficult for the patient to monitor, due to the anatomical location of these muscles. This procedure must be performed at regular intervals as prescribed by the physician.

{4}
K970145 5/8
# SAFETY & EFFECTIVENESS SUMMARY
## Biosearch Medical Products, Inc.
### PERINEOMETER & VAGINAL PROBE
## BIOCOMPATIBILITY DATA AND MATERIAL SPECIFICATIONS:

VAGINAL PROBE shown above is similar (except that the proximal balloon has been removed.) to the Anorectal probe which has been previously approved via the following Biosearch Medical Products 510(k)'s : k902843 (System #5 & Anorectal Probe) and k913736 (System #10 & Anorectal Probe).
* LUER LOCK CONNECTOR -> (Not inserted into a body cavity) The material used is a Medical grade plastic.
* MEDICAL GRADE TUBING -> (Not inserted into a body cavity) The material used is a Medical grade plastic.
* MOLDED TUBE ADAPTER -> (Not inserted into a body cavity) The material used is a Medical grade plastic.
* SILICONE PROBE & SILICONE MARKER RING -> The material used is USP class VI Silicone.
* LUER CONNECTOR & DUAL PORT ADAPTER -> (Not inserted into a body cavity) The material used is a Medical grade plastic.
* BALLOON INFLATION VALVE/PORT -> (Not inserted into a body cavity) The materials used are Medical grade plastics.
4
{5}
K920145 6/8
# SAFETY & EFFECTIVENESS SUMMARY
## Biosearch Medical Products, Inc.
### PERINEOMETER & VAGINAL PROBE
Cont. **BIOCOMPATIBILITY DATA AND MATERIAL SPECIFICATIONS:**

THIS IS A HAND HELD MONITOR. IT DOES NOT COME INTO CONTACT WITH MUCUS MEMBRANES OR BODY FLUIDS.
**SPECIAL NOTE**: PERINEOMETER has been previously approved via the following Biosearch Medical Products 510(k): k902843 (System #5 & Anorectal Probe).
5
{6}
K970145
# SAFETY & EFFECTIVENESS SUMMARY
## Biosearch Medical Products, Inc.
### PERINEOMETER & VAGINAL PROBE
| MATERIAL | TEST DATE | TEST PERFORMED | RESULTS |
| --- | --- | --- | --- |
| Perineometer
Model #20 | NA | Device is identical to legally marked device known as Biosearch Biofeedback monitor #5 | Previous 510(k)
k902843 |
| Anorectal Probe
Assembly | 11/11/91 | Tested entire device for Cytotoxicity - MEM Elution | Nontoxic |
| Glue used to assemble the Anorectal probe | Vendor Provided | FDA Master File | Nontoxic |
| Silicone used in Anorectal Probe | Vendor Provided | USP tests for Class VI Plastics. (Hemolysis, Pyrogenicity, Intracutaneous Toxicity, Intramuscular implant for 90 days) | Meets the requirements of USP Class VI Plastic |
| Polyurethane Tubing
Not inserted into patient. | Vendor Provided | USP tests for Class VI Plastics. (Cytotoxicity - Agarose overlay, Hemolysis, Intracutaneous Toxicity, Implantation Test) | Meets the requirements of USP Class VI Plastic |
| Luer Connectors
Not inserted into patient. | Vendor Provided | USP tests for Class VI Plastics. (Acute Systemic toxicity, Intracutaneous Toxicity, Implantation Test) | Meets the requirements of USP Class VI Plastic |
| Molded Tube Adapter
Not inserted into patient. | Vendor Provided | USP tests for Class VI. (Acute Systemic toxicity, Hemolysis-Direct & Extraction, Intracutaneous Toxicity, Muscle Implantation Test, Cytotoxicity - MEM Elution) | Meets the requirements of USP Class VI Plastic |
| Check Valve | Vendor Provided | ANSI/HIMA MD 70.1
British BS 5081,176 | Meets the requirements |
{7}
K970145 8/8
# SAFETY & EFFECTIVENESS SUMMARY
## Biosearch Medical Products, Inc.
### PERINEOMETER & VAGINAL PROBE
## EQUIVALENCE MATRIX (COMPARING USE AND TYPE)
| NAME &
510(k) No. | PROBE
(BALANCE
TYPE) | PROBE
(CIRC
TYPE) | USED FOR
REQUIREMENT | CAN BE
USED WITH
REQUIREMENT
OF OTHER
RECOGNIZER
(OPTIONAL) | USED FOR
MALE
REQUIREMENT
CONTROL | USED FOR
FEMALE
REQUIREMENT
CONTROL | PROBE
INSERVED
IN THE
RECIPIENT | PROBE
INSERVED
IN THE
VAGIN |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| BIOSEARCH
"PERINOMETER &
VAGINAL PROBE"
{This Submission} | YES | NO | YES | YES | NO | YES | NO | YES |
| BIOSEARCH
"ANORECTAL
BIOFEEDBACK
SYSTEM #5"
{k902843} | YES | NO | YES | YES | NO | NO | YES | NO |
| BIOSEARCH
"ANORECTAL
BIOFEEDBACK
SYSTEM #10"
{k913736} | YES | NO | YES | YES | NO | NO | YES | NO |
| CARDIO DESIGN
"PERITRON"
{k945611} | YES | NO | YES | NO | YES | YES | YES | YES |
| MILEX
"PERINEAL EXERCISER"
{k862410} | YES | NO | YES | YES | NO | YES | NO | YES |
| HOLLISTER
CONTIMED II"
{k891774} | YES | NO | YES | YES | YES | YES | YES | YES |
{8}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
AUG 11 1997
Mr. Martin Dyck
Vice President of Operations,
New Product Development Coordinator
Biosearch Medical Products, Inc.
P.O. Box 1700
Somerville, New Jersey 08876-1276
Re: K970145
PERINEOMETER & VAGINAL PROBE
(Biofeedback Monitor #20 and Vaginal Probe)
Dated: May 13, 1997
Received: May 15, 1997
Regulatory class: II
21 CFR §884.1425/Product code: 85 HIR
Dear Mr. Dyck:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{9}
Page ___ of ___
510(k) Number (if known): K970145
Device Name: Perineometer & Vaginal Probe (Biofeedback Monitor #20 and Vaginal Probe)
Indications For Use:
INTENDED USE STATEMENT: The Perineometer (Biofeedback monitor) and vaginal probe are intended to be used to assist the patient in exercising her pelvic muscles (referred to Kegal exercises) which may lead to increased bladder control.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Colin M Pollard
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K970145
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.