K200409 · Pelvital USA, Inc. · HIR · Jul 10, 2020 · Obstetrics/Gynecology
Device Facts
Record ID
K200409
Device Name
Pelvital System
Applicant
Pelvital USA, Inc.
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Jul 10, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1425
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K200409 · Jul 10, 2020
Pelvital System
Pelvital USA, Inc.
Published clinical literature (Nilsen et al., 2017) reporting on a prospective cohort study
The sponsor utilized a published prospective cohort study to provide additional scientific evidence supporting the safety and effectiveness of the Pelvital System for strengthening pelvic floor muscles in women with stress urinary incontinence.
Prospective cohort study; Stress urinary incontinence; Pelvic floor muscle training
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Nilsen et al. (2017); Prospective cohort study; Follow-up/Duration: 6 weeks (treatment phase); 2-year follow-up mentioned in citation
60 women with stress urinary incontinence (leakage between 5g and 50g) who had failed supervised pelvic floor muscle training; Sample Size: 60; Number of Sites: 1 (University Hospital of Northern Norway)
Not applicable for this study
Urine leakage at six weeks
Indications for Use
The Pelvital System is intended for the strengthening of the pelvic floor muscles, which has been found to help women with stress urinary incontinence.
Device Story
Pelvital System is a reusable, non-sterile vaginal device for in-home pelvic floor muscle (PFM) training. Device consists of a silicone-covered probe and a handheld control unit connected by cable. Probe contains an accelerometer, gyroscope, and an electric motor with an eccentric weight to generate oscillations. User inserts probe into vagina; performs Kegel exercises. Device provides biofeedback via control unit to confirm correct muscle contraction. Oscillations provide resistive surface for PFM conditioning. Control unit houses rechargeable battery and PCB to manage motor frequency and biofeedback display. Healthcare providers use output to monitor patient progress; patients use device to improve PFM strength and reduce stress urinary incontinence.
Clinical Evidence
Clinical study: randomized, double-blind, crossover trial (n=144) comparing PFM training with vs. without oscillations. Primary endpoint: 24-hour pad weight test at 6 weeks. Both groups showed significant improvement from baseline (p=0.0007 for treatment, p=0.009 for control). No significant difference between groups at 6 weeks (p=0.96). Safety profile: 44% of patients reported adverse events, mostly mild (Grade 1-2); one unrelated serious adverse event.
Technological Characteristics
Materials: ABS plastic housing, medical-grade silicone probe sheath. Sensing: Accelerometer and gyroscope. Energy: 3.7V rechargeable Lithium-ion Polymer battery (680-710 mAh). Dimensions: 36mm diameter, 100mm length. Connectivity: Wired probe-to-control unit. Standards: ISO 10993-1, ISO 14971, ES60601-1, IEC 60601-1-2, IEC 60601-1-11. Disinfection: Dispatch Hospital Cleaner Disinfectant Towels with Bleach.
Indications for Use
Indicated for women with stress urinary incontinence to strengthen pelvic floor muscles.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 10, 2020
Pelvital USA, Inc. % Carrie Hetrick, DDS, MSRS Principal and CEO Hetrick Medical Device Consulting, LLC 19123 West 60th Lane Golden, CO 80403
Re: K200409 Trade/Device Name: Pelvital System Regulation Number: 21 CFR§ 884.1425 Regulation Name: Perineometer Regulatory Class: II Product Code: HIR Dated: May 30, 2020 Received: June 2, 2020
Dear Carrie Hetrick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K200409
Device Name Pelvital System
Indications for Use (Describe)
The Pelvital System is intended for the strengthening of the pelvic floor muscles, which has been found to help women with stress urinary incontinence.
| Type of Use (Select <i>one</i> or <i>both</i> , as applicable) |
|----------------------------------------------------------------|
|----------------------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## Pelvital System
## K200409
#### 1. Submission Sponsor
Pelvital USA, Inc.
1000 LaSalle Avenue
SCH 435
Minneapolis, MN 55403
Phone: (612)-804-1319
Contact: Dale Wahlstrom
Title: Chief Executive Officer
#### 2. Submission Correspondent
Hetrick Medical Device Consulting, LLC
19123 West 60th Lane
Golden, CO 80403
Cell Phone: (720) 838-413
Contact: Carrie Hetrick, DDS, MSRS, Principal and CEO
Email: carrie.hetrick@gmail.com
#### 3. Date Prepared
July 9th, 2020
#### 4. Device Identification
| Trade/Proprietary Name: | Pelvital System |
|-------------------------|-------------------|
| Common/Usual Name: | Perineometer |
| Classification Name: | Perineometer |
| Regulation Number: | 21 CFR § 884.1425 |
{4}------------------------------------------------
Product Code: HIR
Device Class: Class II
Classification Panel: Obstetrics/Gynecology
#### 5. Legally Marketed Predicate Device(s)
K162689, Bioinfinity (M) Sdn. Bhd., Vibrance Pelvic Trainer
#### 6. Device Description
The Pelvital System is a repeat use, a non-sterile vaginal device intended to condition and strengthen the pelvic floor muscle (PFM) system. The model MTI-1 is designed for in-home use to enhance PFM training during normal Kegel exercises. The product consists of a probe that is inserted into the vagina and a handheld control unit that provides a biofeedback function to inform the correct muscles are being contracted during a Kegel exercise. The probe is designed to provide a resistive surface against which the user can contract the PFM, and an electric motor containing an eccentric weight that causes the probe to oscillate . The probe and the control unit are connected by a cable.
Control Unit: The control unit consists of a 3.7V Lithium-ion Polymer rechargeable battery with a capacity of 680-710 mAh and built-in safety protection. The unit also contains a PCB assembly used to control the motor speed and frequency and to provide the user with the biofeedback information. The control unit housing is made of ABS plastic (Acrylonitrile-butadiene-styrene copolymer).
Probe Unit: The probe unit contains an accelerometer and gyroscope which generate the biofeedback information. It also houses the motor and weight used to generate oscillations. The maximum diameter is 36 mm, and the maximum length is 100 mm. The design includes a winged flange that limits the maximum insertion length to 88 mm. The probe housing is cylindrical and is made of ABS plastic. The probe is the only part of the device that directly contacts the user's vaginal cavity (mucosal membrane contact, ≤24hr) and is covered entirely with a biocompatible medical-grade silicone sheath.
#### 7. Indication for Use Statement
The Pelvital System is intended for the strengthening of the pelvic floor muscles, which has been found to help women with stress urinary incontinence.
#### 8. Substantial Equivalence Discussion
The following table compares the Pelvital System to the predicate device concerning indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
{5}------------------------------------------------
| Manufacturer | Pelvital USA, Inc. | Bioinfinity (M) Sdn. Bhd. | Significant Differences |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Pelvital System | Vibrance Pelvic Trainer | |
| 510(k) Number | K200409 | K162689 | Not applicable |
| Product Code | HIR | HIR | Identical |
| Regulation Number | 21 CFR § 884.1425 | 21 CFR § 884.1425 | Identical |
| Regulation Name | Perineometer | Perineometer | Identical |
| Indications for Use | The Pelvital System is<br>intended for the<br>strengthening of the<br>pelvic floor muscles,<br>which has been<br>found to help women<br>with urinary stress<br>incontinence. | The Vibrance Pelvic<br>Trainer (VPT) is<br>intended for the<br>strengthening of the<br>pelvic floor muscles,<br>which has been found<br>to help women with<br>urinary incontinence | Same |
| Over the Counter<br>(OTC) | Yes | Yes | Identical |
| Prescription Use | Yes | No | Differs |
| Anatomical Sites | Vagina | Vagina | Identical |
| Feature | Resistive vaginal<br>exerciser | Resistive vaginal exerciser | Identical |
| Target Population | Women with urinary<br>incontinence | Women with mild<br>incontinence | Identical |
| Anatomical Site | Vagina | Vagina | Identical |
| Single Patient Device | Yes | Yes | Identical |
| Single Use or<br>Reusable | Reusable | Reusable | Identical |
| Provided Sterile | Clean, but not sterile | Clean, but not sterile | Identical |
| Biofeedback display<br>information | Yes | No | Differs. The subject<br>device displays<br>mechanical<br>transduction<br>information to the<br>user, while the<br>predicate device<br>requires the user to<br>switch out resistance<br>sheaths. |
| Device Design | A handheld<br>biofeedback device<br>and a vaginally | A handheld device<br>consisting of the Main<br>Body and three gradual | Differs. The main<br>difference between<br>these two devices is |
| Manufacturer | Pelvital USA, Inc. | Bioinfinity (M) Sdn. Bhd. | Significant Differences |
| Trade Name | Pelvital System | Vibrance Pelvic Trainer | |
| | inserted probe | Resistance Sheaths | the subject device<br>has a handheld<br>Display and Control<br>Unit and does not<br>have resistance<br>sheaths.<br>Performance testing<br>of the subject device<br>does not raise any<br>additional questions<br>for safety or efficacy<br>as performance<br>testing supports safe<br>use of the device. |
| Material | Main Body: ABS<br>Plastic, PCB Assembly,<br>rechargeable lithium<br>polymer battery (680-<br>710 mAh, 3.7 V) with<br>USB charger<br>Probe: Medical grade<br>silicone (polysiloxane)<br>probe cover, ABS<br>Plastic, PCB Assembly | Main Body: ABS Plastic,<br>PCB Assembly,<br>rechargeable lithium<br>polymer battery (140<br>mAh, 3.7 V) with USB<br>charger<br>Sheaths: Medical grade<br>silicone<br>(polydimethylsiloxane) | The main difference<br>between these two<br>devices is the<br>replacement of the<br>two low electrical<br>energy coin cell<br>batteries with a<br>rechargeable battery<br>with USB charger. |
| RoHS Compliant | Yes | Yes | Identical |
| Operating Principle | Resistive pelvic floor<br>strengthener | Resistive pelvic floor<br>strengthener | Identical |
| Resistance<br>Component | The Pelvital System<br>(MTI-1)<br>Static resistive<br>component via the<br>vaginally inserted<br>probe. | The Pelvital System<br>(VPT)<br>Electromechanical<br>spring contact offers<br>3 levels of resistance<br>sheaths with<br>increasing stiffness | Identical |
| Biocompatibility | Guidelines set forth<br>in ISO 10993 testing<br>results indicate the<br>material is<br>biocompatible,<br>nontoxic, and well<br>tolerated by mucosal<br>membranes. | Guidelines set forth in<br>ISO 10993 testing<br>results indicate the<br>material is<br>biocompatible,<br>nontoxic, and well<br>tolerated by mucosal<br>membranes. | Similar |
| Chemical Safety | Addressed by | Addressed by | Similar |
| Manufacturer | Pelvital USA, Inc. | Bioinfinity (M) Sdn. Bhd. | Significant Difference |
| Trade Name | Pelvital System | Vibrance Pelvic Trainer | |
| | biocompatibility<br>testing per ISO 10993 | biocompatibility testing<br>per ISO 10993 | |
| Instructions for Use | Manual | Manual | Similar |
| Energy Use and/or Delivered | Energy is supplied by a<br>rechargeable lithium-<br>polymer battery (680-<br>710mAh, 3.7 V) with<br>USB charging capability | Energy is supplied by a<br>rechargeable lithium-<br>polymer battery (140mAh,<br>3.7 V) with USB charging<br>capability | Similar |
| Packaging | The device is sealed<br>plastic bag and manual<br>in a cardboard box | The device is sealed plastic<br>bag and manual in a<br>cardboard box | Identical |
| Product Dimensions | 10.0 cm x 3.6 cm | 6 cm x 2 cm | Similar |
### Table 5A – Comparison of Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
#### 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of the Pelvital System and in showing substantial equivalence to the predicate devices that are subject to this 510(k) Submission, Pelvital USA, Inc. completed several tests. The Pelvital System meets all the requirements for the overall design, biocompatibility, and electrical safety testing, and performance confirm that the output meets the design inputs and specifications. The Pelvital System passed all testing stated above, as shown by the acceptable results obtained.
Testing demonstrated the Pelvital System meets all relevant standards requirements. Internal verification and validation testing confirm that the product specifications are met, which are equivalent in design and technological characteristics to the predicate device. The testing results support that the disinfection, electrical safety testing, and functional testing of the device all met or exceeded the standards for the acceptance of the device. Testing of the Pelvital System supports the claims of substantial equivalence to the predicate device. Evaluation of the performance characteristics establishes that the performance, functionality, and reliability of the Pelvital System are substantially equivalent to the predicate device.
The following testing has been performed to support substantial equivalence:
Biocompatibility - The biological safety of the Pelvital System was evaluated in accordance with ISO 10993-1:2009 and guidance document entitled Use of International Standard ISO-10993-1, "Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing within a Risk Management Process" dated June 16th, 2016. Under these, for the stated indications for use, each component of the device's biological safety was evaluated for in vitro cytotoxicity, skin sensitization, and vaginal irritation.
- . In Vitro Cytotoxicity test was performed according to ISO 10993-12 and ISO 10993- 5, with the results suggesting that the test articles did not induce a cytotoxic effect.
- . Skin Sensitization (Maximization Test) was performed according to ISO 10993-10, with the results suggesting that the device extracts did not produce skin sensitization in guinea pigs.
{8}------------------------------------------------
- . Irritation (Vaginal Irritation Study) was performed on New Zealand White Rabbits according to ISO 10993-10, with the results showing that there were no significant clinical signs and gross findings in either the control or test group, and there were no mortalities.
Electromagnetic Compatibility and Electrical Safety - The Pelvital System complies with the applicable voluntary standards, which include ES60601-1, IEC 60601-1-2, and IEC 60601-1-11. The Pelvital System complies with these voluntary standards, including IEC 60601-1-11 for Electrical Safety and systems used in the home healthcare environment.
Software - All documentation was prepared and submitted for the device per FDA guidance documents. The software was tested against the established Software Design Specifications for each of the test plans to assure the device performs as intended. The device hazard analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria of each module and the interaction of processes. The Pelvital System device passed all testing and supported the claim of substantial equivalence and safe operation.
Disinfection/Cleaning: The Pelvital System device is intended for single-patient use. Disinfection studies showed that Dispatch® Hospital Cleaner Disinfectant Towels with Bleach disposable wipes (EPA Reg. No: 56392-8) were effective in reducing the risk of worst-case vaginal test articles are considered successfully disinfected as the recovered log reduction for each test article was equal to or greater than 6 for each test replicate.
Performance Testing – Bench: Determination of the performance of the Pelvital System was carried out through Radial Compression Tests; Probe Insertion and Probe Cable Pull Tests; and, Display and Control Unit Connector Pull Tests.
Usability Testing: Several ergonomic factors were tested with users. The studies show that users find Pelvital System to be easy to use and that the ergonomic factors designed into the system reduce user error through easy to understand language. The usability testing and results were considered in the Risk Analysis. As problems were identified, both software and mechanical changes were made.
ISTA 6-FEDEX-A Transportation & Packaging: Testing demonstrated compliance with ISTA 6-FedEx A, Testing Packaged Products Weighing up to 150 Lbs.
Risk Analysis - Formal Risk Assessment of the Pelvital System was performed in accordance with ISO 14971. With respect to perceivable conditions in which the device would be subjected to a worst-case environmental or human error scenario, Pelvital USA, Inc. believes the outcomes of these risks are considered acceptable within the context of ISO 14971, and that all potential risks have been mitigated to the lowest form.
Review of Published Literature on Subject Device: The first clinical study on the Pelvital System was conducted at the University Hospital of Northern Norway. Sixty women with leakage between 5g and 50g demonstrated ability to contract pelvic floor muscles, and stress urinary incontinence (SUI) greater than
{9}------------------------------------------------
one year despite supervised pelvic floor muscle training (PFMT), were enrolled. Nilsen et al. (2017)2 included a single treatment arm where the subjects performed daily at-home PFMT for 5 minutes over six weeks while using the Pelvital System. The primary endpoint was urine leakage at six weeks. Stress tests for urine leakage were performed with a standardized volume of 3dL saline in the bladder. The women in the study who used the Pelvital System had a high cure rate for their SUI condition, comparable to surgical procedures.
This study represents valid scientific evidence from a well-controlled investigation by qualified experts to further demonstrate that there is reasonable assurance of the safety and effectiveness of the subject device under its conditions of use.
#### 10. Clinical Performance Data
A clinical study was conducted to evaluate the safety and effectiveness of the Pelvital System. The study was a randomized, double-blind, crossover trial of adult women who have been diagnosed with stress urinary incontinence. Subjects were placed into the control arm (PFMT in conjunction with no oscillations) or the treatment arm (PFMT in conjunction with oscillations). Subjects conducted their respective therapy five minutes a day over twelve weeks. After six weeks (primary endpoint), all patients in the control arm crossed over into the treatment arm.
The overall summary of the clinical study is shown:
- Study Design: The study was a randomized parallel-group design, with a one-way cross over from control to treatment at six weeks.
- Study Endpoints: The primary outcome for this study is the improvement in the severity of involuntary urine loss, determined by an improvement in 24-hour pad weight test, from baseline to week 6. Secondary outcomes include incontinence episode frequency and health-related quality of life (PGI-I, I-QOL, ICIQ-UI-SF).
- . Study Procedure: All subjects are randomized to either the treatment or control arm. Each arm conducts 5 minutes of training therapy each day and conducts bi-weekly follow-up visits. At six weeks, the control arm is crossed over to treatment, and both arms continue daily training through 12 weeks.
- . Number of Patients/Sites: 215 subjects were screened for the study. 71 failed the screen for various inclusion/exclusion criteria, and 144 were randomized into the study, 72 in each control and treatment arms. All 144 subjects were evaluated for the 6-week primary endpoint based on the intent to treat (ITT) population across four clinical sites in the United States.
<sup>1</sup> Nilsen I, Rebolledo G, Acharya G, Leivseth G. Mechanical oscillations superimposed on the pelvic floor muscles during Kegel exercises reduce urine leakage in women suffering from stress urinary incontinence: A prospective cohort study with a 2 - year follow up. Acta Obstet Gynecol Scand. 2018;97:1185–1191. https://doi.org/10.1111/aogs.13412
{10}------------------------------------------------
- Study Treatment: The primary endpoint was at six (6) weeks, with participant duration of twelve ● (12) weeks.
- Results Effectiveness: Of the 144 evaluated patients, 72 were in the treatment arm, and 72 were in the control arm. The Control arm and the Treatment arm have improved significantly over their baseline values. The Control arm utilized the Pelvital trainer without oscillations. The use of the Pelvital device without oscillations had a beneficial effect on its own as compared to baseline (p = .009), and with oscillations, the Treatment arm data compared to baseline is even more profound (p = .0007). A copy of the primary endpoint analysis is presented below:
| Visit | N | | Mean + S.D. | | |
|------------------------------------|----|----|-----------------|-----------------|-------------------------------|
| | Tx | C | Tx | C | p-value |
| Baseline | 72 | 72 | $43.9 \pm 49.7$ | $59.1 \pm 98.6$ | |
| 6 weeks | 72 | 72 | $28.7 \pm 46.1$ | $38.8 \pm 77.8$ | p=0.96 by Mann Whitney U test |
| Change | | | $17.0 \pm 40.6$ | $20.2 \pm 64.0$ | |
| Change % | | | 33% | 33% | |
| Baseline to 6 weeks: Paired t-test | | | P=0.0007 | P=0.009 | |
- Results Safety: A total of 55 patients (44%) had 89 adverse events in combination with all . treatment sessions. There was one serious adverse event (Grade 3) that was deemed unrelated to the study device. All other adverse events were graded severity levels 1 or 2.
Results of the clinical investigation support the safety, effectiveness, and indications for the use of the Pelvital System for the strengthening of the pelvic floor muscles, which has been found to help women with urinary incontinence. The clinical study confirms that the device is safe and effective as used according to the instructions for use.
#### 11. Conclusion
By definition, a device is substantially equivalent to a predicate device has the same intended use and the same technological characteristics as the previously cleared predicate device; or the device has the same intended use and different technological characteristics but can be demonstrated as substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) Submission that the differences between the Pelvital System and the predicate device do not raise any different questions regarding its safety and effectiveness. The Pelvital USA, Inc. Pelvital System, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.