K162689 · Bioinfinity (M) Sdn. Bhd. · HIR · Jun 8, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K162689
Device Name
Vibrance Pelvic Trainer
Applicant
Bioinfinity (M) Sdn. Bhd.
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Jun 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1425
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vibrance® Pelvic Trainer (VPT) is intended for the strengthening of the pelvic floor muscles, which has been found to help women with urinary incontinence.
Device Story
Intravaginal exercise device for home use; strengthens pelvic floor muscles via resistance and vibration biofeedback. Device consists of main body (ABS plastic, PCB, rechargeable battery) and three silicone resistance sheaths of varying stiffness. User inserts device into vagina; performs voluntary pelvic floor contractions. Device senses force applied by muscles; provides vibration biofeedback to confirm correct muscle engagement. Audio-guided training program included. Output allows user to verify exercise effectiveness; progressive strengthening achieved through resistance sheaths. Benefits include improved pelvic floor muscle strength and management of urinary incontinence.
Clinical Evidence
Bench testing only. Includes mechanical compression/pull-off testing, biocompatibility (ISO 10993-1:2009, cytotoxicity, sensitization, vaginal irritation), and electrical safety/EMC (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2).
Technological Characteristics
Materials: ABS plastic chassis, medical grade silicone (polydimethylsiloxane) sheaths. Power: 3.7V 140mAh rechargeable lithium polymer battery with USB charging. Dimensions: 2.4 cm max shaft diameter, 6 cm max inserted length. Sensing: Electromechanical spring contact for force sensing. Connectivity: Standalone. Sterilization: Non-sterile (clean).
Indications for Use
Indicated for women with mild urinary incontinence for the purpose of strengthening pelvic floor muscles.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 3, 2018
Bioinfinity (M) Sdn. Bhd. % Carrie Hetrick Senior Consultant, RA Emergo Group 2500 Bee Cave Road Building 1, Suite 300 Austin, Texas 78746
Re: K162689
Trade/Device Name: Vibrance® Pelvic Trainer Regulation Number: 21 CFR 884.1425 Regulation Name: Perineometer Regulatory Class: Class II Product Code: HIR Dated: May 2, 2017 Received: May 4, 2017
Dear Carrie Hetrick:
This letter corrects our substantially equivalent letter of June 8, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael T. Bailey -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162689
Device Name Vibrance® Pelvic Trainer
Indications for Use (Describe)
The Vibrance® Pelvic Trainer is intended for the strengthening of the pelvic floor muscles, which has been found to help women with urinary incontinence.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
### Vibrance® Pelvic Trainer
### K162689
#### 1. Submission Sponsor
Bioinfinity (M) Sdn. Bhd.
No. 21, Jalan 4/62A, Bandar Menjalara, Kepong
52200 Kuala Lumpur
Malaysia
Phone: +603-6270236
Fax: +603-62720093
Contact: Mr. Ng Shea Kang
Title: Chief Executive Officer
#### 2. Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78701
Cell Phone: (720) 838-413
Office Phone: (512) 327.9997
Contact: Carrie Hetrick, Senior Consultant, RA
Email: project.management@emergogroup.com
#### 3. Date Prepared
May 2, 2017
{4}------------------------------------------------
#### 4. Device Identification
| Trade/Proprietary Name: | Vibrance® Pelvic Trainer |
|-------------------------|--------------------------|
| Common/Usual Name: | Perineometer |
| Classification Name: | Perineometer |
| Regulation Number: | 21 CFR § 884.1425 |
| Product Code: | HIR (Perineometer) |
| Device Class: | Class II |
| Classification Panel: | Obstetrics/Gynecology |
#### 5. Legally Marketed Predicate Device(s)
K141893, Bioinfinity (M) Sdn. Bhd. Vibrance® Kegel Device (VKD)
The predicate device has not been subject to a design-related recall.
#### Device Description 6.
The Vibrance® Pelvic Trainer is a repeat use, non-sterile intra-vaginal exercise device intended to strengthen the pelvic floor musculature by offering resistance to an individual's voluntary contractions of these muscles. The Vibrance® Pelvic Trainer incorporates an active vibration biofeedback mechanism for the identification of the pelvic floor muscles, and the sheaths provide graduated resistance levels for the progressive strengthening of the pelvic floor muscle. The device is intended for home use.
The therapeutic effect is from the user using the device to exercise the pelvic floor muscles in accordance with the exercise program included in the instructions. The device provides biofeedback information that allows the user to verify the use of the correct muscles in the exercise routine and thus the effectiveness of the exercises undertaken. The device also includes a resistive component to the training program to strengthen the pelvic floor muscle via physiologic principles. The biofeedback mechanism works by directly sensing the force applied by the pelvic floor muscle.
The device consists of the Main Body and three Resistance Sheaths with varying degrees of stiffness. It has a PCB assembly and is a battery-powered device (140mAh, 3.7V lithium polymer battery) with USB charging capability, operating on two independent, enclosed electronic circuits for the vibration biofeedback loop and the other for audio guided training. The maximum diameter of the Vibrance® Pelvic Trainer shaft is 2.4 cm. The maximum inserted length of the shaft is 6 cm.
#### Main Body
The chassis of the device consists of an Acrylonitrile Butadiene Styrene (ABS) plastic frame with the biofeedback mechanism. It is incorporated within a tubular sheath made of medical grade silicone rubber which encloses the whole chassis. This main body and resistance sheath are the parts of the device that directly contact the user's intravaginal cavity (mucosal membrane contact, ≤24hr). The base of the tubular sheath and chassis are winged to prevent total ingestion of the vaginal
{5}------------------------------------------------
canal.
#### Resistance Sheath
Three Resistance Sheaths, each with varying levels of stiffness are included with the Vibrance® Pelvic Trainer. When a Resistance Sheath is placed over the Main Body, the challenge of the exercise is increased.
#### 7. Indication for Use Statement
The Vibrance® Pelvic Trainer (VPT) is intended for the strengthening of the pelvic floor muscles, which has been found to help women with urinary incontinence.
#### 8. Substantial Equivalence Discussion
The following table compares the Vibrance® Pelvic Trainer to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
| Manufacturer | Bioinfinity (M) Sdn. Bhd. | Bioinfinity (M) Sdn. Bhd. | Significant |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Vibrance® Pelvic Trainer | Vibrance Kegel Device | Differences |
| | Image: Vibrance Pelvic Trainer product packaging and device | Image: Vibrance Kegel Device product packaging and device | |
| 510(k) Number | K162689 | K141893 | Not applicable |
| Product Code | HIR | HIR | Identical |
| Regulation Number | 21 CFR § 884.1425 | 21 CFR § 884.1425 | Identical |
| Regulation Name | Perineometer | Perineometer | Identical |
| Indications for Use | The Vibrance® Pelvic Trainer<br>(VPT) is intended for the | The Vibrance Kegel Device (VKD)<br>is intended for the | Identical |
| Manufacturer | Bioinfinity (M) Sdn. Bhd. | Bioinfinity (M) Sdn. Bhd. | Significant |
| Trade Name | Vibrance® Pelvic Trainer | Vibrance Kegel Device | Differences |
| | strengthening of the pelvic<br>floor muscles, which has been<br>found to help women with<br>urinary incontinence | strengthening of the pelvic floor<br>muscles, which has been found<br>to help women with urinary<br>incontinence. | |
| Over the<br>Counter (OTC) | Yes | Yes | Identical |
| Anatomical Sites | Vagina | Vagina | Identical |
| Feature | Resistive vaginal exerciser | Resistive vaginal exerciser | Identical |
| Target Population | Women with mild incontinence | Women with mild incontinence | Identical |
| Anatomical Site | Vagina | Vagina | Identical |
| Single Patient Device | Yes | Yes | Identical |
| Single Use<br>or Reusable | Reusable | Reusable | Identical |
| Provided Sterile | Clean, but not sterile | Clean, but not sterile | Identical |
| Biofeedback<br>display<br>information | No | No | Identical |
| Device Design | Handheld device consisting of<br>the Main Body and three<br>gradual Resistance Sheaths | Handheld device consisting of<br>the Main Body and three<br>gradual Resistance Sheaths | Identical |
| Material | Main Body: ABS Plastic, PCB<br>Assembly, rechargeable lithium<br>polymer battery (140 mAh, 3.7<br>V) with USB charger<br>Sheaths: Medical grade silicone<br>(polydimethylsiloxane) | Main Body: ABS Plastic, PCB<br>Assembly, two low electrical<br>energy coin cell batteries<br>(CR1220)<br>Sheaths: Medical grade<br>silicone<br>(polydimethylsiloxane) | The main<br>difference<br>between these<br>two devices is<br>the<br>replacement of<br>the two low<br>electrical<br>energy coin cell<br>batteries with a<br>rechargeable<br>battery with<br>USB charger. |
| RoHs Compliant | Yes | Yes | Identical |
| Operating Principle | Resistive pelvic floor strengthener | Resistive pelvic floor strengthener | Identical |
| Resistance<br>Component | The Vibrance® Pelvic Trainer<br>(VPT) Electromechanical<br>spring contact offers 3 levels<br>of resistance sheaths with<br>increasing stiffness | The Vibrance Kegel Device<br>Electromechanical spring contact<br>offers 3 levels of resistance<br>sheaths with increasing stiffness | |
| Biocompatibility | Guidelines set forth in ISO<br>10993 testing results indicate<br>the material is biocompatible,<br>nontoxic, and well tolerated by<br>mucosal membranes. | Guidelines set forth in ISO<br>10993 testing results indicate<br>the material is biocompatible,<br>nontoxic, and well tolerated by<br>mucosal membranes. | Identical |
| Manufacturer | Bioinfinity (M) Sdn. Bhd. | Bioinfinity (M) Sdn. Bhd. | Significant |
| Trade Name | Vibrance® Pelvic Trainer | Vibrance Kegel Device | Differences |
| Chemical Safety | Addressed by biocompatibility<br>testing per ISO 10993 | Addressed by biocompatibility<br>testing per ISO 10993 | Identical |
| Instructions for Use | Manual | Manual | |
| Energy Use and/or<br>Delivered | Energy is supplied by a<br>rechargeable lithium polymer<br>battery (140mAh, 3.7 V) with USB<br>charging capability | Energy is supplied by two<br>independent replaceable 3.0 Volt<br>CR-1220 lithium<br>/manganese dioxide (Li/MnO2)<br>Button Cell batteries | Different - The<br>difference<br>between these<br>two devices is the<br>replacement of<br>the two low<br>electrical energy<br>coin cell batteries<br>with a<br>rechargeable<br>battery with USB<br>charger. |
| Packaging | Device in sealed plastic bag and<br>manual in cardboard box | Device in sealed plastic bag and<br>manual in cardboard box | Identical |
| Dimensions | 6 cm x 2 cm | 6 cm x 2 cm | Identical |
#### Table 5A – Comparison of Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
The differences in technological characteristics do not raise different questions of safety and effectiveness.
#### 9. Non-Clinical Performance Data
The following testing has been performed to support substantial equivalence:
Mechanical Testing – Compression testing was conducted on the Vibrance® Pelvic Trainer to measure the forces required to vibrate the sensor for each of the three levels of resistance sheath. Pull off test was also conducted to evaluate how much force is needed for the sheath to be removed from the main body when the device is in use but is adequate for removal of a sheath after use. A correlation measurement test was then conducted to evaluate the test results of the compression test. The mechanical testing concluded that the different components of the device would function properly and safely when exposed to loads and configurations representative of normal use.
Biocompatibility - The biological safety of the Vibrance® Pelvic Trainer was evaluated in accordance with ISO 10993-1:2009 and guidance document entitled Use of International Standard ISO-10993-1, "Biological Evaluation of Medical Devices: Part 1: Evaluation and Testing within a Risk Management Process" dated 16 June 2016. Under these, for the stated indications for use, each component of the device's biological safety was evaluated for in vitro cytotoxicity, skin sensitization, and vaginal irritation.
- In Vitro Cytotoxicity test was performed according to ISO 10993-12 and ISO 10993- 5, with the results suggesting that the test articles did not induce a cytotoxic effect.
- Skin Sensitization (Maximization Test) was performed according to ISO 10993-10, with the results suggesting that the device extracts did not produce skin sensitization in guinea pigs.
- . Irritation (Vaginal Irritation Study) was performed on New Zealand White Rabbits according to ISO 10993-10, with the results showing that there were no significant clinical signs and gross findings in either the control or test group, and there were no mortalities.
{8}------------------------------------------------
Electrical Safety and Electromagnetic Compatibility - The Vibrance® Pelvic Trainer complies with the applicable voluntary standards which include:
- . IEC 60601-1:2005 + A1:2012 Medical Electrical Equipment: Part 1: General Requirements for Basic Safety and Essential Performance.
- IEC 60601-1-11 Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment (Recognition Number 19-6)
- IEC 60601-1-2:2014 (4" edition) Medical electrical equipment - Part 1-2: General reguirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances
The Vibrance® Pelvic Trainer complies with these voluntary standards for Electrical Safety, Electromagnetic Compatibility, and for systems used in the home healthcare environment.
Software - Software information was provided in accordance with the FDA Guidance document, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005.
Risk Analysis - Formal Risk Assessment of the Vibrance® Pelvic Trainer was performed in accordance with ISO 14971. With respect to perceivable conditions in which the device would be subjected to a worst-case environmental or human error scenario, Bioinfinity believes the outcomes of these risks are considered acceptable within the context of ISO 14971, and that all potential risks have been mitigated to the lowest form.
#### 10. Performance Testing Summary
As part of demonstrating safety and effectiveness of Vibrance® Pelvic Trainer and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Bioinfinity (M) Sdn. Bhd. completed a number of tests. The Vibrance® Pelvic Trainer meets all the requirements for overall design, biocompatibility, and electrical safety testing performance confirm that the output meets the design inputs and specifications. The Vibrance® Pelvic Trainer passed all testing stated above as shown by the acceptable results obtained.
The Vibrance® Pelvic Trainer complicable voluntary standards for biocompatibility. The device passed all the testing in accordance with national and international standards.
#### 11. Statement of Substantial Equivalence
lt has been shown in this 510(k) submission that the differences between the Vibrance® Pelvic Trainer and the predicate device do not raise any different questions regarding its safety and effectiveness. The performance testing provided demonstrates that the subject device is as safe and effective as the predicate device. The Bioinfinity (M) Sdn. Bhd. Vibrance® Pelvic Trainer, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.