AXSYM B12
Device Facts
| Record ID | K965186 |
|---|---|
| Device Name | AXSYM B12 |
| Applicant | Abbott Laboratories |
| Product Code | CDD · Clinical Chemistry |
| Decision Date | Mar 24, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.1810 |
| Device Class | Class 2 |
Intended Use
AxSYM B12 is a microparticle enzyme intrinsic factor assay for the quantitative determination of vitamin B₁₂ in human serum or plasma.
Device Story
Microparticle enzyme intrinsic factor assay; quantitative measurement of vitamin B₁₂ in human serum or plasma; performed on Abbott AxSYM System; utilizes Abbott B12 calibrators and controls for accuracy/precision verification; intended for clinical laboratory use; results aid clinicians in assessing vitamin B₁₂ levels; supports diagnosis of deficiency states.
Clinical Evidence
Bench testing only. Correlation analysis performed on 337 clinical specimens comparing AxSYM B12 to IMx B12. Results: correlation coefficient 0.975, slope 1.00, y-intercept -12.371 pg/mL, standard error of y estimate 67.281. Dynamic range 60-2,000 pg/mL.
Technological Characteristics
Microparticle enzyme intrinsic factor assay; quantitative detection; compatible with human serum and plasma (tripotassium EDTA, potassium oxalate); dynamic range 60-2,000 pg/mL; automated system-based analysis.
Indications for Use
Indicated for the quantitative determination of vitamin B₁₂ in human serum or plasma to aid in the diagnosis of vitamin B₁₂ deficiency.
Regulatory Classification
Identification
A vitamin B12 test system is a device intended to measure vitamin B12 in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.
Predicate Devices
- Abbott IMx® B12
Related Devices
- K984108 — ABBOTT ARCHITECT B12 · Abbott Laboratories · Feb 3, 1999
- K051327 — ABBOTT AXSYM B12 · Abbott Laboratories · Aug 3, 2005
- K161721 — Diazyme Vitamin B12 Assay, Diazyme Vitamin B12 Calibrators Set, Diazyme Vitamin B12 Controls Set · Diazyme Laboratories · Sep 20, 2016
- K993571 — BAYER DIAGNOSTICS ACS:180 VB12, BAYER DIAGNOSTICS ADVIA CENTAUE VB12 · Bayer Corp. · Dec 16, 1999