DEPUY DUPONT PHANTOM SUTURE ANCHOR
Device Facts
| Record ID | K964521 |
|---|---|
| Device Name | DEPUY DUPONT PHANTOM SUTURE ANCHOR |
| Applicant | Depuy, Inc. |
| Product Code | MAI · Orthopedic |
| Decision Date | Feb 6, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3030 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The DePuy DuPont Phantom Suture Anchor is intended for use in the shoulder for soft tissue re-attachment and capsular repair, including rotator cuff repairs, Bankart lesion repairs and SLAP lesion repairs.
Device Story
PLLA cortical screw with flat tip and rectangular head; two parallel axial holes; pre-threaded with #2 nonabsorbable polyester suture. Used in shoulder surgery for soft tissue re-attachment and capsular repair (rotator cuff, Bankart, SLAP lesions). Available in 3.5mm or 4.5mm diameters; supplied pre-loaded on driver or separate. Implanted by orthopedic surgeons in clinical/OR settings. Provides mechanical fixation for soft tissue to bone; suture secures tissue; PLLA material resorbs over time. Benefits include secure tissue re-attachment during healing.
Clinical Evidence
Bench testing only; mechanical pull-out load testing performed in cadaver bone to compare performance against Zimmer Statak.
Technological Characteristics
Material: PLLA (poly-L-lactic acid). Design: Threaded cortical screw, flat tip, rectangular head, two parallel axial holes. Dimensions: 3.5mm or 4.5mm diameter. Suture: #2 nonabsorbable polyester. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring soft tissue re-attachment and capsular repair in the shoulder, including rotator cuff, Bankart lesion, and SLAP lesion repairs.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Zimmer Statak
- Acufex Suretac Absorbable Tack