UNITRON MODEL ICON AOHPL +4
Device Facts
| Record ID | K964504 |
|---|---|
| Device Name | UNITRON MODEL ICON AOHPL +4 |
| Applicant | Unitron Industries, Inc. |
| Product Code | ESD · Ear, Nose, Throat |
| Decision Date | Nov 27, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 874.3300 |
| Device Class | Class 1 |
| Attributes | Therapeutic |
Intended Use
To amplify and transmit sound via air conduction to the ear.
Device Story
Behind-the-ear (BTE) hearing aid; amplifies and transmits sound via air conduction. Features Class D circuitry for battery efficiency; includes power, compression limiting, active low-cut, and high-cut tone controls. Equipped with telecoil and preamplifier. Operated by user via volume control and microphone/telecoil/off switch. Powered by standard size 13 hearing aid battery. Used by patients with hearing loss to improve auditory perception.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Behind-the-ear form factor; Class D circuitry; telecoil with preamplifier; standard size 13 battery; volume control; microphone/telecoil/off switch. Technical specifications comply with ANSI Standard S 3.22-1987.
Indications for Use
Indicated for individuals requiring sound amplification via air conduction to the ear to assist with hearing impairment.
Regulatory Classification
Identification
An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.
Related Devices
- K964510 — UNITRON MODEL ICON AOHPD +4 (ICON AOHPD +4) · Unitron Industries, Inc. · Nov 27, 1996
- K964508 — UNITRON MODEL ICON AOHP +4 (ICON AOHP +4) · Unitron Industries, Inc. · Nov 27, 1996
- K964503 — UNITRON MODEL SOUND F/X +4 · Unitron Industries, Inc. · Nov 27, 1996
- K964502 — UNITRON MODEL SOUND F/XP +4 · Unitron Industries, Inc. · Nov 27, 1996
- K963122 — UNITRON MODEL ICON KR · Unitron Industries, Inc. · Sep 6, 1996