IMIG-MRI

K963953 · Imig-Mri Systems, LLC · LNH · Dec 18, 1996 · Radiology

Device Facts

Record IDK963953
Device NameIMIG-MRI
ApplicantImig-Mri Systems, LLC
Product CodeLNH · Radiology
Decision DateDec 18, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

Anatomy: head, limbs, spine, torso; Nuclei: H-1; Diagnostic uses: Imaging

Device Story

Magnetic resonance diagnostic device; 0.15 Tesla static magnetic field; generates diagnostic images of head, limbs, spine, torso. Operated by trained personnel in clinical settings. System utilizes RF power and magnetic field gradients to acquire signals; processes signals to produce anatomical images. Healthcare providers use images for diagnostic assessment. Benefits include non-invasive visualization of internal anatomy.

Clinical Evidence

Bench testing only. Performance data provided for head and body receive coils including signal-to-noise ratio (S/N > 32 head, > 31 body), uniformity (< 15% head, < 40% body), geometric distortion (< 2.5% head, < 3.5% body), slice thickness/position accuracy, and spatial resolution.

Technological Characteristics

0.15 Tesla static magnetic field; 18.4 Tesla/sec max rate of magnetic field change; 0.05 W/kg max RF power deposition; 114 dB peak acoustic noise. Includes head and body receive coils. Performance specifications: 15 cm dsv (head) and 30 cm dsv (body) volumes.

Indications for Use

Indicated for diagnostic imaging of the head, limbs, spine, and torso using H-1 nuclei.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1} DEC 18 1996 K963953 IMIG-MRI 510(k) Summary Submitter: IMIG-MRI Systems I.I.c., 300 Vesper Executive Park, Tyngsborough MA 01879 Company contact: Greg Hurst, Ph.D., Manager of Imaging Science and Applications (508) 649-8590; FAX (508) 649-8520 Regulatory Counsel: Jonathon S. Kahan, Attorney at Law Hogan &amp; Hartson 555 Thirteenth Street N.W., Washington D.C. 20004-1109 (202) 637-5794; FAX (202) 637-5910 Date of Summary Preparation: 18 September 1996 Device Name: Magnetic Resonance Diagnostic Device. Classification Number: 90LNH Device Class: Class II, under 21 CFR 892.1000 Device Trade Name: IMIG-MRI™ Intended Uses: Anatomy: head, limbs, spine, torso; Nuclei: H-1; Diagnostic uses: Imaging Predicate Device: Siemens Magnetom P8 Safety parameter data summary: Maximum static magnetic field: 0.15 Tesla Maximum rate of magnetic field change: 18.4 Tesla/sec Maximum RF power deposition: .05 W/kg Acoustic noise levels: 114 dB peak; 95 dB A-weighted RMS Performance test data summary: | (Receive coil:) | (head) | (body) | | --- | --- | --- | | Specification volume: | 15 cm dsv | 30 cm dsv | | S/N: | > 32 | > 31 | | Uniformity: | < 15 % | < 40 % | | Geometric distortion: | < 2.5 % | < 3.5 % | Slice thickness: within 10% of nominally designated value Slice position (gap): within 10% of nominally designated value Spatial resolution: nominally equivalent to pixel size General safety and effectiveness concerns: Safe and effective use of the machine is assured by associated labeling. This labeling includes: advertising brochures, Site Planning Guide, and Instructions for Use (comprised of Clinical Users Guide, User Safety Guide, User Training Guide, User Applications Guide, and User QA &amp; Maintenance Guide) Substantial equivalence: This device has the same intended use and basic technological characteristics as the Siemens Magnetom P8.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...