LM BONE ANCHOR (ORTHOPEDICS)
Device Facts
| Record ID | K963812 |
|---|---|
| Device Name | LM BONE ANCHOR (ORTHOPEDICS) |
| Applicant | Li Medical Technologies, Inc. |
| Product Code | MBI · Orthopedic |
| Decision Date | Nov 12, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
UltraFix RC Shoulder (rotator cuff repair) UltraFix Mini Hand/Wrist (ulnar or lateral collateral ligament reconstruction, radial collateral ligament, scapholunate ligament reconstruction), Foot/Ankle (hallux valgus reconstruction, midfoot reconstructions, lateral stabilization, medial stabilization, achilles tendon), Elbow (biceps tendon, ulnar collateral ligament, radial collateral ligament) UltraFix Shoulder (bankart repair, rotator cuff repair, SLAP lesion, acromio-clavicular separation, biceps tenodesis, deltoid repair, capsular shift reconstruction, capsulolabral reconstruction), Hand/Wrist (ulnar or lateral collateral ligament reconstruction, radial collateral ligament, scapholunate ligament reconstruction), Foot/Ankle (hallux valgus reconstruction, midfoot reconstructions, lateral stabilization, medial stabilization, achilles tendon), Elbow (biceps tendon, ulnar collateral ligament, radial collateral ligament), Knee (medial collateral ligament, lateral collateral ligament, patellar tendon, patellar ligament avulsions, posterior oblique ligament, joint capsule to tibia, iliotibial band tenodesis).
Device Story
Bone anchor system for soft tissue-to-bone attachment; utilizes surgical grade 316L stainless steel or 6Al-4V ELI titanium. Device features crown and center pin for suture passage. Anchoring mechanism employs cantilevered beams for bony purchase. Used by orthopedic surgeons in clinical settings for various ligament and tendon repairs across shoulder, hand, wrist, elbow, foot, ankle, and knee. Provides mechanical fixation to facilitate tissue healing.
Clinical Evidence
Bench testing only. Pre-clinical pullout strength testing performed in cadaver specimens. Results showed mean pullout strength of LM Anchor Products is substantially equivalent to Mitek predicate anchors.
Technological Characteristics
Materials: 316L stainless steel or 6Al-4V ELI titanium. Design: Cylindrical anchor with crown and center pin for suture passage. Fixation mechanism: Cantilevered beams for bony purchase.
Indications for Use
Indicated for soft tissue to bone attachment in orthopedic procedures including rotator cuff repair, ligament reconstruction (hand, wrist, elbow, knee, ankle), and tendon repair in patients requiring surgical fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Mitek Surgical Products GII QuickAnchor (K904436)
- Li Medical Bone Anchor (K951233)
- Li Medical Bone Anchor (K951472)
- Li Medical Bone Anchor (K953764)
- Li Medical Bone Anchor (K960448)
- Li Medical Bone Anchor (K960439)
- Li Medical Bone Anchor (K960825)
Related Devices
- K981755 — LM ANCHOR · Li Medical Technologies, Inc. · Jul 13, 1998
- K960439 — LM BONE ANCHOR · Li Medical Technologies, Inc. · Apr 18, 1996
- K981764 — LM ANCHOR · Li Medical Technologies, Inc. · Jul 13, 1998
- K960448 — LM ANCHOR · Li Medical Technologies, Inc. · Apr 17, 1996
- K111397 — TITAN TI SUTURE ANCHOR SYSTEM · Arthrocare Corp. · Jul 12, 2011