LM ANCHOR
Device Facts
| Record ID | K960448 |
|---|---|
| Device Name | LM ANCHOR |
| Applicant | Li Medical Technologies, Inc. |
| Product Code | MBI · Orthopedic |
| Decision Date | Apr 17, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
Shoulder (bankart repair), Hand (ulnar or lateral collateral ligament reconstruction), Wrist (scapholunate ligament reconstruction), Foot (hallux valgus reconstruction), and Ankle (midfoot reconstructions)
Device Story
LM Anchor; surgical grade 316L stainless steel bone anchor; crown and center pin design for suture passage; facilitates soft tissue-to-bone attachment. Used in orthopedic surgeries (shoulder, hand, wrist, foot, ankle). Physician-operated. Provides mechanical fixation for ligament/tendon repair. Benefits patient by securing soft tissue to bone during reconstructive procedures.
Clinical Evidence
Bench testing only. Pre-clinical cadaver specimen study compared mean pullout strength of LM Anchor to Mitek Mini Anchor, showing substantial equivalence.
Technological Characteristics
Material: 316L surgical grade stainless steel. Design: Cylindrical anchor with crown and center pin for suture passage. Fixation mechanism: Cantilevered beams. Non-powered, mechanical device.
Indications for Use
Indicated for soft tissue to bone attachment in shoulder (Bankart repair), hand (ulnar/lateral collateral ligament reconstruction), wrist (scapholunate ligament reconstruction), foot (hallux valgus reconstruction), and ankle (midfoot reconstruction) procedures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Mitek Surgical Products Mini Anchor (K921873 and K930892)
- Li Medical Bone Anchor (K951233)
Related Devices
- K981764 — LM ANCHOR · Li Medical Technologies, Inc. · Jul 13, 1998
- K960439 — LM BONE ANCHOR · Li Medical Technologies, Inc. · Apr 18, 1996
- K981755 — LM ANCHOR · Li Medical Technologies, Inc. · Jul 13, 1998
- K960825 — LM ANCHOR OR OTHER PROPRIETARY NAME · Li Medical Technologies, Inc. · Apr 30, 1996
- K963812 — LM BONE ANCHOR (ORTHOPEDICS) · Li Medical Technologies, Inc. · Nov 12, 1996